BioLAB BLPS-301 Operation Manual

Laboratory water purification system

Advertisement

Quick Links

OPERATION MANUAL
LABORATORY WATER PURIFICATION SYSTEM
BLPS-301 BLPS-302 BLPS-303
www.biolabscientific.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BLPS-301 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BioLAB BLPS-301

  • Page 1 OPERATION MANUAL LABORATORY WATER PURIFICATION SYSTEM BLPS-301 BLPS-302 BLPS-303 www.biolabscientific.com...
  • Page 2: Table Of Contents

    INDEX 1.Preface 2.Specification 3.Water Flow Chart 4.Working Environments 5.Installation 5.1 Preparation for installation 5.2 Tube and adapter’s connection 5.3 Installation steps 6. Usage Guide 6.1 Starting up 6.2 Getting corresponding pure water 6.3 Standby 6.4 Shutdown 6.5 Releasing internal air of terminal filter 6.6 Flushing UF membrane 6.7 The usage to keep high quality pure water 7.
  • Page 3: Preface

    For the water purification system’s maximum efficiency, it is suggested that the user manual should be read before installation. Any question in the installation process, please contact our technology engineers or dealers. 2.Specification BLPS-30 Model BLPS-301 BLPS-302 BLPS-304 Output -ultrapure water Up to 2 liters/minute (less output with UF cartridge) Pure water outlet...
  • Page 4: Water Flow Chart

    (0.45+0.1)μm terminal filer REMARKS The value will be influenced by temperature and feed water’s quality. 3.Water Flow Chart BLPS-301 FV Water Flow Chart 4.Working Environments - Inlet water: Ro water, distilled water, deionized water. - Inlet water pressure: 1atm - Work temperature: 5-40℃...
  • Page 5: Tube And Adapter's Connection

    5.2 Tube and adapter’s connection The adapter of the machine is high quality easy-put adapter. And material of tube is high quality’s PE Tube installation and drag diagram ATTENTION - The tube should be cut with special tube cutter for rounded cut section. And rounded cut section should be guaranteed as much as possible with other cut tools.
  • Page 6 ATTENTION Stop plug should be pulled out before the following steps. 4. Connect to water source Use 3/8” PE tube with a suitable length. One side connects with inlet water source (Ro water, distilled water, deionized water) and the other connects with the interface with blue label marked “To inlet water” at the back of machine.
  • Page 7: Usage Guide

    6. Usage Guide All data have been set in the factory. The machine will operate smoothly without any data-setting and debugging. All work state will display on the LCD. If there is abnormal state, the system will alarm automatically. If data modification is necessary, specific step is in “Microcomputer Controller”.
  • Page 8: Starting Up

    6.1 Starting up At first, make sure of that the source water is connected to the main body’s inlet. Then insert the power line into the power source and turn on the power switch, then the system begins to produce pure water. 6.2 Getting corresponding pure water Press “DP1”...
  • Page 9: The Usage To Keep High Quality Pure Water

    ATTENTION The “UF drain valve” is shut on normal condition, except for flushing UF membrane. Frequency of flushing UF membrane: At least one times every week, and at least lasting 30 seconds every times. ATTENTION Make sure that the water source and power source is not connected when the system is not in the use state for long time (for example, off...
  • Page 10: Microcomputer Controller

    7. Microcomputer Controller There are 6 buttons on the panel as shown below Specification of the buttons- Control the solenoid valve of deionized water Control the solenoid valve of ultrapure water Main menu, modify all the function and data of the system. Shift the cursor to corresponding position Adjust the data of chosen position(0-9 circle), or turn on/off...
  • Page 11: Specification Of The Main Interface

    ,then press“OK” button. 7.2 Specification of the main interface 1. Specification of system’s working status icon At the top-right corner of the main interface, system will show 4 kinds of working status-REC, Produce, FULL, No water. It respectively means: Recirculation of ultrapure water, Producing pure water, tank is full of water, no feed water or low pressure of feed water.
  • Page 12 Specification of the interface “Pre-treat: 442h”: It means that the active carbon filter’s remaining time is 442 hours (Initial value is 450 hours UV lamp: 8992h It means that the UV lamp’s remaining time is 8892 hours Initial value is 9000 hours REMARKS -UV lamp is optional.
  • Page 13 At this time, machine would stop working, and prepares forsanitization. Method of sanitization At first, open top cap of disinfection box, then put 1 piece of disinfecting tablet inside, tighten the cap again. Lastly press “MENU” button, move the cursor to icon , press “OK”...
  • Page 14: Specification Of Main Menu

    The initial code is “1000”. Inputting initial code“1000” through shift button and figure button, then press“OK” button to confirm. If code is wrong, system will warn and return to main interface. If code is right, system will switch to interface of main menu (Picture 11). Under the main menu, press“MENU”...
  • Page 15 - Specification of the interface - TDS1 It represents TDS of source water. - Max: 0800 It means when the TDS value of TDS1 exceeds 800 ppm (initial setting value), the system will warn and display the interface of feed water quality’s standard-exceeding alarm (Picture 13). TDS2 -It represents TDS of DI water.
  • Page 16 2. The purpose of feed and DI water quality’s standard-exceeding alarm The purpose of TDS1’s alarm(feed water quality’s standard-exceeding alarm):Remind that the feed water’s quality is bad, and the extra pretreatment is necessary to protect the cartridges of main body. The purpose of TDS2’s alarm(DI water quality’s standard-exceeding alarm):Remind that DI water’s quality is bad, and mixed bed resin cartridge should be replaced to protect the post cartridges.
  • Page 17 REMARKS When the resistivity of ultrapure water exceeds the standard setting value, the system will warn 1 times every 10 seconds. During this time, the solenoid valve of pure water outlet will be closed. If opening solenoid is necessary, the value of ultrapure water resistivity’s standard-exceeding must be reset. - The purpose of ultrapure water resistivity’s standard-exceeding alarm Remind that the actual value of resistivity is lower than the setting value.
  • Page 18 OFF. Otherwise, it is ON. - The purpose of UV lamp and recirculation ON/OFF - UV lamp alarm: means that life of UV lamp is over, replacement of UV lamp is necessary. - Recirculation ON/OFF: system can restrain bacteria’s increase and keep high quality’s ultrapure water. - Setting method of UV lamp and recirculation ON/OFF: - In the interface of UV lamp and recirculation ON/OFF (Picture 17), press “MENU”...
  • Page 19 UV lamp: 9000h - It means that the UV lamp’s initial setting time is 9000 hours. When its runtime exceeds 9000 hours, the system will warn and display the interface of UV lamp replacement’s reminding (Picture 20). Pre-treat runtime: clr - clear the runtime of pretreatment cartridges to zero, system will reset timer. V lamp runtime: clr - clear the runtime of UV lamp to zero, system will reset timer REMARKS - - The above 2 operation are implemented generally on the condition of replacing cartridges.
  • Page 20 Under the main menu, press “MENU” button, move the cursor to icon then press “OK” button to switch to interface of code setting (Picture 21). Specification of the interface 0000 -It means that the password is “0000” ATTENTION -Initial setting password is 1000 - Method of code setting For enhancing management of the machine, If needed, password can be reset to protect system’s parameter from modification.
  • Page 21 - Specification of the interface: - Mode: Manual - It means that the current mode of getting water is manual. Specification of manual mode - Under this mode, press the dispense button first times, then the solenoid valve of pure water is opened. Press the dispense button again, then the valve is closed.
  • Page 22: Water Quality Test

    8. Water Quality Test The system has 3 monitor sensors of water quality measuring. - The first, TDS1, monitors feed water’s quality Ro water, distilled water, deionized water Measure unit: TDS total dissolved solid, ppm - The second, TDS2, monitors deionized water’s quality high pure water Measure unit: TDS(total dissolved solid, ppm - The third, Resist, monitors ultrapure water’s quality Ultrapure water Measure unit: Resistivity MΩ.cm...
  • Page 23: Normal Trouble Diagnosis

    3. Disassembly any part of the machine or human-behavior damage; 4. Not repaired by our serviceman. Specification can be changed without any prior notice for development. Biolab Scientific Ltd. 3660 Midland Avenue, Suite 300, Toronto, Ontario M1V 0B8, Canada Email: info@biolabscientific.com | Website: www.biolabscientific.com...