Summary of Contents for Spectral Engines Liquid Scanner

  • Page 1 Liquid Scanner QUICK GUIDE QUICK GUIDE V.2.1 12/2024...
  • Page 2: Table Of Contents

    Downloadable software: Technical Specifications • SensorControl software for LED Functions Windows Maintenance • NIRONE Scanner Mobile Cleaning the Liquid Scanner 13 Change of Battery Disposal To start using Liquid Storage and Transport Scanner, please see and Warranty follow the instructions: www.
  • Page 3: Safety Information

    Intended use for users and the device must be implemented. The Liquid Scanner is a mobile material scanner Please consider these points designed for near infrared in the risk assessment spectroscopic transmission...
  • Page 4 • Use only securely closed may cause minor injuries. or locked cuvette cells to Inspect the housing and the avoid spilling onto the glass windows in the cuvette device. holder before use. • Keep potential ignition sources away from the Only use the device if all parts device.
  • Page 5: Contents Of The Case

    Contents of the case Liquid Scanner Reference cuvette Distance adapters (for 1, 2 or 5 mm cuvettes) USB-C cable Glass cuvettes (for first tests) Quick Guide Empty space for an optional White Reference Empty space for an optional Scanner Device...
  • Page 6: Main Parts

    USB-C connector power LED will start blinking LED indicator for power red. Connect USB cable to the and battery charging Liquid Scanner to recharge LED indicator for the battery. LED indicator for Bluetooth battery blinks green while Power / Bluetooth button charging.
  • Page 7: Getting Started - Pc

    Measuring the Reference adapter (in case of 10mm Signal cuvettes not necessary) Always measure the reference in Liquid scanner cuvette signal at the start of a new set holder. of measurements. Perform and save measurements through...
  • Page 8: Getting Started - Mobile App

    To switch the device ON press scanner (for example: for 5s the next to the USB DataGatherer20LS). connector at the back side of the Liquid Scanner. Pair Your Mobile App and the Device After signing in the app, you have to pair your mobile app and device.
  • Page 9: Measuring Reference Signal

    Measuring a Sample SAVE RESULTS Getting the Results Place the sample cuvette A couple seconds after into the Liquid Scanner pressing “Measure Now”, the cuvette holder. app will display the results. All saved measurements will be available in the NIRONE...
  • Page 10: Technical Specifications

    Technical Specifications Environmental Conditions SPECIFICATION VALUE Operation +10 to +40°C (non-condensing) temperature range +50 to +104°F (non-condensing) Charging +10 to +35°C (non-condensing) temperature range +50 to +95°F (non-condensing) Storage temperature -5 °C to +60 °C (non-condensing) range +23 to +140°F (non-condensing) 80 % for temperatures up to 31 °C (88 °F) Relative humidity...
  • Page 11: Specification

    SPECIFICATION VALUE Battery charging max. 320 mA current Wireless connectivity Bluetooth Low Energy BLE 4.3 Optical Specification SPECIFICATION VALUE 1.35 - 1.65 µm (S1.7) 1.55 - 1.95 µm (S2.0) standard Wavelength range 1.75 – 2.15 μm (S2.2) 2.00 - 2.45 μm (S2.5) 13 –...
  • Page 12: Led Functions

    Bluetooth® (6 dBm). Environment Conditions Intended environment with max pollution degree 2. LED Functions LED indicators Description Liquid Scanner off / Not charging / No cable / No power Liquid Scanner off/ Charging/Cable connected Liquid Scanner off/ Charged full/ Cable connected...
  • Page 13: Maintenance

    Maintenance • Use only cuvettes in good condition. Replace cuvettes with new ones Handle your Liquid Scanner if the old ones are dirty, and sample cuvettes scratched or otherwise with care. The following damaged. suggestions help you keep your device operational.
  • Page 14: Disposal

    • it is recommended to hand tight. use new gaskets. A new Connect the device to Li-Ion battery and new the USB charger and gaskets can be ordered wait until the battery from your distributor. LED indicates a fully Remove the six screws charged battery.
  • Page 15: Disclaimer

    Products will be warranty corrected. To the maximum extent Disclaimer permitted by law, neither Spectral Engines GmbH nor its suppliers will be liable for Important! any damage or losses relating Sensors, software, to the use of the Products, documentation, and other...
  • Page 16: Declarations Of Conformity

    Declarations of Conformity The Declarations of Conformity (DoC) are available at: https://support:spectralengines.com/legal-warranty Declaration of Conformity Hereby, Spectral Engines GmbH declares that the radio equipment type NIRONE is in compliance with Directives 2014/53/EU and 2011/65/EU. UKCA Declaration of Conformity Hereby, Spectral Engines GmbH declares that the radio equipment type NIRONE is in compliance with The Radio Equipment Regulations 2017 &...
  • Page 17 ISED IC compliance This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: This device may not cause interference, and This device must accept any interference, including interference that may cause undesired operation of the device.
  • Page 18 This device FCC ID: XPYNINAB1 contains ISED: 8595A-NINAB1 This documentation is subject to revision without notice. © 2024 Spectral Engines GmbH. All rights reserved.
  • Page 19 SPECTRAL ENGINES GMBH Weißkirchener Straße 2-6 61449 Steinbach Germany +49 6171 286 9760 WWW.SPECTRALENGINES.COM info@spectralengines.com...