Hu-Friedy ConFirm RR20READER Operator's Manual

20 minute incubator

Advertisement

Quick Links

ConFirm
Operator Manual
REF RR20READER
KEY
Well Blocks
Well LED
LCD Screen
Temperature Display
(Use Up/Down arrows to view)
HOME Button
NAVIGATION Up Button
NAVIGATION Down Button
ENTER Button
Barcode Scanner
Name Plate
Power Cord Inlet
USB Port
(For HuFriedyGroup Use Only)
RS-232 Communication Port
(Printer Connectivity)
RJ-45 Communication Port
(Network Connectivity)
RECOMMENDED PRINTER SPECIFICATIONS:
Thermal Line Printer
Minimum Dot Density 200 DPI x 200 DPI
32-42 Columns
1/6 inch line spacing
14K Byte Minimum Data Buffer
Interface: RS-232
20 Minute Incubator
INCUBATOR - TOP VIEW
INCUBATOR - BACK VIEW

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ConFirm RR20READER and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hu-Friedy ConFirm RR20READER

  • Page 1: Confirm ™ 20 Minute Incubator Figure Key

    ConFirm 20 Minute Incubator ™ Operator Manual REF RR20READER INCUBATOR - TOP VIEW Well Blocks Well LED LCD Screen Temperature Display (Use Up/Down arrows to view) HOME Button NAVIGATION Up Button NAVIGATION Down Button ENTER Button Barcode Scanner Name Plate Power Cord Inlet USB Port (For HuFriedyGroup Use Only)
  • Page 3: Table Of Contents

    Table of Contents F . Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 ConFirm 20 Minute Incubator Figure Key .
  • Page 4: Specifications

    C. Specifications: AC Input Voltage: 100-240 VAC +/-10% Low: 50-60 dB Audible Alarm: Medium: 70-80 dB AC Input Frequency: 50-60 Hz High: 80-100 dB Input Current: 1.0A (rms) max Weight: 3 lbs 10 oz (1.64 kg) Overvoltage: Category II 6.5 (W) x 6.0 (H) x 11 (D) inches Operating voltage: 12 VDC Dimensions:...
  • Page 5: Getting Started

    Set the Language, Date, Time and Network connection (optional) after start-up sequence . Reference section F for more information . NOTE: USB port is not used for daily operation . Typical Start-Up Sequence Warming To 59.0C Hu-Friedy Mfg. Co., LLC. <xx:xx> 1 <r>C 5 <r>C 1 <r>C 5 <r>C...
  • Page 6: Routine Operations

    E. Routine Operations Precautions: Do not use the ConFirm 20 Minute Incubator to incubate and read any biological indicator other than the ConFirm Rapid 20 ™ ™ Minute BIs . Examples of BIs that should not be used in the ConFirm ™...
  • Page 7: Options

    For a positive result, the Incubator will sound an alarm (three beeps every five minutes) and the LED adjacent to the well will turn from slow flashing green to a rapid flashing red light . The LCD screen will display a “+” next to the well number . The operator must press the ENTER button to acknowledge the result and silence the alarm .
  • Page 8 Changing the Date: Date The Date can be set during initial Incubator set up and at any time after that . Use the NAVIGATION buttons to highlight the correct month, day and year . Select <MMM> <DD> <YYYY> by pressing the ENTER button . Press ENTER to select Changing the Time:...
  • Page 9: Cleaning And Care Of The Incubator

    G. Cleaning and Care of the Incubator: Caution: Do not spray water, cleanser, or disinfectant on or into the Incubator. Caution: Do not use abrasive materials to clean or disinfect the Incubator. To avoid spills within the individual wells: • Always keep BIs in the upright position when handling to keep recovery media from spilling out of the BI . If foreign materials get into the wells of the Incubator, please follow the procedure outlined below: Step 1.
  • Page 10: Troubleshooting

    11 . Verify the temperature reading on the thermocouple thermometer . If the temperature falls out of range for incubation, contact HuFriedyGroup or any authorized dealer . Do not use failing wells . The other panel of wells may still be operational . 12 .
  • Page 11 ERROR MESSAGE AND DESCRIPTION RECOMMENDED ACTION Remove BI from Well <#>. Scan BI First. Scan BI first and then place into any empty well. If the fault returns, contact HuFriedyGroup. The BI was not scanned before inserting into well. Invalid Barcode, Scan Again. Check BI type and scan BI again.
  • Page 12: Disposal

    Made in USA N5480/1022 Hu-Friedy and ConFirm are trademarks of Hu-Friedy Mfg . Co . , LLC, its affiliates or related companies, unless otherwise noted . ® ™ All other product and company names referenced are trademarks of their respective owner .

Table of Contents