Other product or company names mentioned herein are trademarks or business names of their respective owners. There is no warranty, position or other implication of any kinds in this manual. BSWA and its employees are not liable for any direct or indirect loss of data, benefits, etc. due to any information contained or mentioned in this manual.
User manual BSWA Technology Co., Ltd. 1. Introduction 1.1 General Description XCA106 sound calibrator is used for calibrating sound level meters and other sound measurement equipment. It provides an output level of 94.0 dB at a frequency of 1 kHz. It is designed and tested according to IEC 60942-2017 standard.
User manual BSWA Technology Co., Ltd. 1.5 Specifications XCA106 Standard IEC 60942-2017, Class 2 Output level 94.0 ± 0.3dB Output frequency 1 kHz ±1.0% Total Harmonic distortion ≤3% Battery specification 9V, 6F22/6LR61 square battery Auto switch off time 100 seconds Level stabilizing time <10 seconds...
User manual BSWA Technology Co., Ltd. 2.2 Microphone Calibration XCA106 provides a nominal tone of 1kHz, which requires no weighting corrections when used with sound level meters utilizing A, B, C or Z (Flat) frequency weighting. Follow the steps below to calibrate microphones (as shown in the following pictures): Step 1 Insert the proper microphone adapter fully into the calibrator.
Before leaving the factory, the XCA106 calibrated in a controlled laboratory environment using traceable calibration equipment. It is recommended that the XCA106 calibrator should be recalibrated at least annually by BSWA or an acoustic test laboratory qualified to perform calibration.