BSWA MicW XCA106 User Manual

Sound calibrator iec 60942-2017 class 2

Advertisement

Quick Links

User Manual
XCA106 Sound Calibrator
IEC 60942-2017 Class 2
RoSH
MicW Audio,a brand of BSWA Technology Co., Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MicW XCA106 and is the answer not in the manual?

Questions and answers

Summary of Contents for BSWA MicW XCA106

  • Page 1 User Manual XCA106 Sound Calibrator IEC 60942-2017 Class 2 RoSH MicW Audio,a brand of BSWA Technology Co., Ltd.
  • Page 2 XCA106 User Manual ZKXY-III-C021-04-P0058 v1.0 Copyright © 2025 BSWA Technology Co., Ltd. The specification of product is subject to changes without notice www.MicW-audio.com Mar. 2025...
  • Page 3: Change History

    Other product or company names mentioned herein are trademarks or business names of their respective owners. There is no warranty, position or other implication of any kinds in this manual. BSWA and its employees are not liable for any direct or indirect loss of data, benefits, etc. due to any information contained or mentioned in this manual.
  • Page 4: Table Of Contents

    User manual BSWA Technology Co., Ltd. Contents Change History ........................1 Contents ...........................2 1. Introduction ..........................3 1.1 General Description ......................3 1.2 Applications ........................3 1.3 Features ...........................3 1.4 Packing List ........................3 1.5 Specifications ........................4 2. Using the XCA106 Calibrator ..................... 4 2.1 Installing the Battery .......................
  • Page 5: Introduction

    User manual BSWA Technology Co., Ltd. 1. Introduction 1.1 General Description XCA106 sound calibrator is used for calibrating sound level meters and other sound measurement equipment. It provides an output level of 94.0 dB at a frequency of 1 kHz. It is designed and tested according to IEC 60942-2017 standard.
  • Page 6: Specifications

    User manual BSWA Technology Co., Ltd. 1.5 Specifications XCA106 Standard IEC 60942-2017, Class 2 Output level 94.0 ± 0.3dB Output frequency 1 kHz ±1.0% Total Harmonic distortion ≤3% Battery specification 9V, 6F22/6LR61 square battery Auto switch off time 100 seconds Level stabilizing time <10 seconds...
  • Page 7: Microphone Calibration

    User manual BSWA Technology Co., Ltd. 2.2 Microphone Calibration XCA106 provides a nominal tone of 1kHz, which requires no weighting corrections when used with sound level meters utilizing A, B, C or Z (Flat) frequency weighting. Follow the steps below to calibrate microphones (as shown in the following pictures): Step 1 Insert the proper microphone adapter fully into the calibrator.
  • Page 8: Maintenance And Warranty

    Before leaving the factory, the XCA106 calibrated in a controlled laboratory environment using traceable calibration equipment. It is recommended that the XCA106 calibrator should be recalibrated at least annually by BSWA or an acoustic test laboratory qualified to perform calibration.
  • Page 9 MicW Audio, a brand of BSWA Technology Co., Ltd. Room 1003, North Ring Center, No.18 Yumin Road, Xicheng District, Beijing 100029, China Tel: 86-10-5128 5118 Fax: 86-10-8225 1626 E-mail: info@bswa-tech.com URL: www.bswa-tech.com...

Table of Contents