Page 2
Thank you for choosing iProven. Please read the instructions carefully in order to accurately and safely utilize this device. The BPM-617 measures blood pressure and heart rate and saves the results. It offers real-time clinically accurate readings that are simple and easy to read thanks to the Oscillometric Measuring method.
Page 3
Intended use of the BPM The BPM-617 is an upper arm digital monitor that is used for measuring blood pressure and heart rate with an arm circumference ranging from 8 ¾” - 16 ½” (22cm to 42cm). The device is intended for home use and adults only.
Page 4
Installing the batteries Open the battery cover. Insert four AA batteries as per the symbols in the battery compartment. Close the battery cover. Setting the date and time Firstly, ensure you set the time and date correctly so the device can record the results with the correct details.
Page 5
Before using your device For an accurate result and the best insight into your blood pressure pat- tern, please measure using the same arm every time. It is recommended to use the left arm which has a better blood flow. •...
Page 6
Taking a measurement Breathe slowly 5 times and press the START/STOP button to begin the measurement. Keep your arm still. First, the measured air pressure will calibrate to zero and then the cuff will inflate and start reading the blood pressure and heart rate. The symbol should indicate OK meaning the cuff is cor- rectly secured.
WHO indicator After each measurement, the indicator shows the corresponding cate- gory color on the right of the display. The colors represent the different categories of the World Health Organisation blood pressure classifica- tion as depicted in the chart. The indicator helps you quickly interpret the results of the measurement.
Page 8
Using the memory function Make sure that the date and time are set correctly. (See “Setting up the device” section) • Every measurement is automatically saved in the memory. It has a maximum capacity of 500 (250 x 2 users) measurements. To view the recorded data press the MEM button while the monitor is inactive.
Components of the BPM Cuff CUFF Air Hose PCBA Pump AIR HOSE Valve AIR CONNECTOR PLUG MEM BUTTON DC POWER SOCKET SET BUTTON START/STOP BUTTON LCD DISPLAY SYMBOL DESCRIPTION EXPLANATION Systolic blood pressur igh pressure result Diastolic blood pressure Low pressure result PULSE/min Pulse display Pulse in beats per minute...
Page 10
Battery powered mode: AC adaptor powered mode: Power supply (Please only use the recommended AC adaptor model). Display mode Measurement mode Oscillographic testing mode Rated cuff pressure: 0mmHg~299mmHg Measurement range Measurement pressure: SYS: 60mmHg~230mmHg DIA: 40mmHg~130mmHg Pulse value: (40-199)beat/minute Pressure: Accuracy 5℃-40℃within±3 mmHg Pulse value:±5%...
Page 11
Caution This device is intended for indoor, home use. * This device is not intended for public use. * This device is portable, but it is not intended for use during patient trans- port. * This device is not suitable for continuous monitoring during medical emer- gencies or operations.
Page 12
shunt, which could disrupt blood circulation and cause injuries. * Do not place the cuff on the arm on the same side of a mastectomy (es- pecially when lymph nodes have been removed). it is recommended to take measurements on the unaffected side. * Do not wrap the cuff on the same arm to which another monitoring device is applied.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Warning: Use of accessories, trans- ducers, and cables other than those specified or provided by the manufac- turer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in...
Page 14
Guidance and manufacturer’s declaration - electromagnetic emissions Emissions test Compliance RF emissions Group 1 CISPR 11 RF emissions Class [ B ] CISPR 11 Harmonic emissions Class A IEC 61000-3-2 Voltage fluctuations / flicker emissions Comply IEC 61000-3-3...
Page 15
Guidance and manufacturer’s declaration – electromagnetic Immunity IEC 60601-1-2 Immunity Test Compliance level Test level Electrostatic ±8 kV contact ±8 kV contact discharge (ESD) ±2 kV, ±4 kV, ±8 kV, ±15 kV air ±2 kV, ±4 kV, ±8 kV, ±15 kV air IEC 61000-4-2 Electrical fast ±2 kV for power supply lines...
Page 16
FCC statement This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. PROBLEM SYMPTOM CHECK THIS...
Page 17
The Warranty Period of this iProven product is 5 years from the date of purchase. A replace- ment product or product part assumes the remaining warranty of the original product purchase.
Page 18
If for any reason you are unsatisfied with our product, feel free to reach out to iProven directly. We offer a refund or will replace the product for free within 100 days from your purchase. Contact us: support@iproven.com DISTRIBUTED BY...