Page 1
SDC-01 SERIES Thank you for purchasing the StatPro Dry Cabinet. Please follow the instructions in this manual to correctly set up and use StatPro Dry Cabinet. Please retain this documentation throughout the lifetime of the equipment.
CHECK LIST OF ACCESSORIES ................1 II. SAFETY GUIDANCE ....................2 III. STATPRO DRY CABINET COMPONENTS ............2 IV. STATPRO DRY CABINET INSTALLATION AND ADJUSTMENT ......3 V. DIGITAL THERMO-HYGROMETER BATTERY REPLACEMENT ......5 VI. TROUBLESHOOTING ................... 6 VII. SPECIFICATION ....................6 I.
C. Do not operate the StatPro Dry Cabinet in a damp or wet environment. It may cause electric shock. D. Please only use 110V AC power source (alternating current) and do not share the same socket with other electrical devices (by using splitters).
9. STATPRO DRY CABINET will begin operating automatically. The power indicator light of STATPRO DRY CABINET Automatic Drying System (core unit) will light on. 10. During initial operation, it may take 24 to 48 hours for the STATPRO DRY CABINET to reach <5%RH.
Page 5
B. REMOVING CENTER STILE AND SHELF INSTALLATION 1. Open both doors at the same level. The removable center stile appears. (pic. 1) 2. On the back of the removable center stile there is a white plastic latch (referred as the Latch below) (pic.
C. MANUAL NITROGEN PURGE SYSTEM (Optional feature) ① ① Flow regulator: ② Controlling gas flow. Manual control nitrogen purge. Default setting: 0 L/min Suggest setting: 80 L/min ② Valves: ③ To close the valve, turn handle perpendicular to the pipeline. To open the valve, turn handle parallel to the pipeline.
VI. TROUBLESHOOTING A. Problem: The cabinet doors cannot be closed smoothly. Possible Cause & Solutions: 1. The cabinet is not completely level. Please adjust the 4 pieces of “Adjustable Leveling Support”. 2. The cabinet might be impinged, which causes the cabinet distortion. Please contact your local representative for further repairing.
Need help?
Do you have a question about the SDC-01 Series and is the answer not in the manual?
Questions and answers