Data Logger for Ultra-Low Temperatures, UL-2
Please read the User Manual carefully before use, and follow all operating and safety instructions!
user manual
english

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UL-2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Metria UL-2

  • Page 1 Data Logger for Ultra-Low Temperatures, UL-2 Please read the User Manual carefully before use, and follow all operating and safety instructions! user manual english...
  • Page 2: Warranty

    User Manual UL-2 Data Logger for Ultra-Low Temperatures Preface Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument. Service In order to guarantee this equipment works safely and efficiently, it must receive regular maintenance. In case of any faults, do not try to repair it yourself.
  • Page 3: Product Description

    Introduction Congratulations on your purchase of this ultra-low temperature PDF datalogger. This datalogger is designed for monitoring freezer temperature and air temperature, subject to quality control requirements. Temperature measurements are saved throughout the entire duration of the measurement period. This datalogger is equipped with a user-defined programming function. The measurement report output is implemented using a PDF file and an Excel file;...
  • Page 4: Operation

    ⑥ Battery cover: Operated by 2 CR2032 batteries. Use a coin and follow the direction indicator to open and close the cover. When installing the batteries, the anode side faces up. ⑦ START/STOP (ON/OFF) key: After installing the batteries, press the key to power on the logger. The LCD displays for 2 seconds. Press again to turn it off.
  • Page 5 Unit of Measurement (UoM) Select the unit that will be displayed on the PDF report and LCD. The options are Metric or Imperial. In Metric, the unit is Celsius for temperature. In Imperial, the unit is Fahrenheit for temperature. Password The Password function is OFF by default.
  • Page 6: Technical Data

    Step 2: Start Logging Press the “START” key for 5 seconds to begin logging. “REC” will appear and flash on the LCD to indicate that logging is activated. Measured parameters will update and alternate on the LCD every 5 seconds. If the logger is programmed with a start delay, “REC”...
  • Page 7: Error Code

    Alarm Range Air: -30.0 ºC to 70.0 ºC External: -100.0 ºC to 120.0 ºC Alarm Delay 0, 5, 30, 45, 60, 90, 120 minutes Alarm Type Single, Cumulative, Deactivated Operation Keys 3 Keys: Start/Stop, Mark, and MAX/MIN LED Indicator REC, High/Low Alarm Protection Class IP65 Directives...
  • Page 8 Nota importante para los aparatos electrónicos vendidos en España Important note for electronic devices sold in Spain Remarque importante pour les appareils électroniques vendus en Espagne Instrucciones sobre la protección del medio ambiente y la eliminación de aparatos electrónicos: Los aparatos eléctricos y electrónicos marcados con este símbolo no pueden desecharse en vertederos.
  • Page 9 www.labbox.com...