Do you have a question about the U-LH75XEAPO and is the answer not in the manual?
Questions and answers
Summary of Contents for Olympus U-LH75XEAPO
Page 1
U-LH75XEAPO 75 W XENON APO LAMP HOUSING This instruction manual is for the Olympus U-LH75XEAPO Xenon Apo Lamp Housing. To ensure safety, optimum performance and familiarize yourself fully with the use of the lamp housing, we recommend that you study this manual thoroughly before operating the system. This instruction manual pertains only to the information associated with the lamp housing.
U-LH75XEAPO CONTENTS IMPORTANT — Be sure to read this section first for safe use of the equipment. — NOMENCLATURE ASSEMBLY CENTERING THE XENON BURNER SPECIFICATIONS TROUBLESHOOTING GUIDE LAMP HOUSING INSPECTION SHEET This device complies with the requirements of both directive 2004/108/EC concerning electromagnetic com- patibility and directive 2006/95/EC concerning low voltage.
2. The xenon short-arc burner (hereinafter referred to as “xenon burner”) used in the unit should be an Olympus supplied DC-powered UXL-75XB (mfd. by Ushio Electric).
U-LH75XEAPO Warning Indication A warning stickers are attached to parts where special precaution is required when handling and using the system. Always heed the warnings. [Warning against high temperature] Xenon Apo Lamp Housing Warning indication U-LH75XEAPO [Warning against high voltage]...
Page 6
Caution If the equipment is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the equipment may also be damaged. Always operate the equipment as described in this instruction manual. The following symbols are used to set off text in this instruction manual.
ASSEMBLY Mounting the Xenon Burner (Figs. 1 & 2) The burner and lamp housing vicinity will be extremely hot after the burner is turned off and will cause injury if touched. Before replacing ² the burner, wait at least 10 minutes, or until the burner has cooled down after turning the burner off.
Page 9
U-LH75XEAPO Mounting the Lamp Housing on the Microscope Refer to the instruction manual of your microscope and be sure to mount the lamp housing horizontally and clamp securely. Do not turn on the xenon burner before installing the lamp housing...
CENTERING THE XENON BURNER # Wait for 3 to 5 minutes after turning on the burner, and start center- ing when the xenon arc has stabilized. }If a shutter is provided on the microscope side, open the shutter. ² 1. Reduce the field iris diaphragm and open the aperture iris diaphragm. 2.
Page 11
U-LH75XEAPO Precise Centering of the Mirror (Fig. 4) }The position of the mirror has been locked by the factory adjustment ³ before shipping. Perform the following adjustment immediately after the adjustment on the previous page and only when you want very precise adjustment.
U-LH75XEAPO TROUBLESHOOTING GUIDE Under certain conditions, performance of this equipment may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. Problem Cause Remedy Page a. Field of view is obscured or is not...
{If there is any ( ) mark noted, immediately stop use of the product, and contact Olympus for detailed inspections or replace the lamp housing. {If you detect an abnormality other than that listed below or with other Olympus product, also stop the use of the product and contact Olympus for detailed inspections.
Page 16
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan Wendenstrasse 14-18, 20097 Hamburg, Germany 3500 Corporate Parkway, Center Valley, Pennsylvania 18034-0610, U.S.A. 491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373 31 Gilby Road, Mount Waverley, VIC., 3149, Melbourne, Australia 5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126, U.S.A.
Need help?
Do you have a question about the U-LH75XEAPO and is the answer not in the manual?