  • Page 1 Instruction Manual Charge Generator Type 5363A 5363A_002-812e-08.17...
  • Page 2 Instruction Manual Charge Generator Type 5363A 5363A_002-812e-08.17...
  • Page 4 Foreword Information in this document is subject to change without notice. Kistler reserves the right to change or improve its products and make changes in the content without obliga- tion to notify any person or organization of such changes or improvements.
  • Page 5: Table Of Contents

    Charge Generator Type 5363A Content Introduction ..........................3 General ............................4 Disposal Instructions for Electrical and Electronic Equipment ..........4 Function............................5 Operation ............................. 6 Charge Ranges ........................6 Voltage Range 0 ... 10 V ..................... 7 Testing the Battery Charge ....................7 Change of Batteries ......................
  • Page 6: Introduction

    It will help you with the installation, maintenance, and use of this product. To the extent permitted by law Kistler does not accept any liability if this instruction manual is not followed or products other than those listed under Accessories are used.
  • Page 7: General

    Charge Generator Type 5363A 2. General Please keep this instruction manual in a safe and easily reachable spot. In case of loss please contact Kistler‘s extensive support organization. 2.1 Disposal Instructions for Electrical and Electronic Equipment Do not discard old electronic instruments in municipal trash.
  • Page 8: Function

    The function of charge amplifiers can be tested and their charge sensitivity checked with the Charge Generator Type 5363A. For an exact charge amplifier calibration, its accuracy of approx. 1 % is not adequate; the Kistler product range does, however, also contain suitable equipment for this purpose.
  • Page 9: Operation

    Charge Generator Type 5363A 4. Operation 4.1 Charge Ranges The charge is yielded at 3 of 4 BNC sockets which are labeled with the capacitor connected in series: 0,1 nF Adjustment range 0 ... 1 000 pC 1 nF Effective range 1 000 ...
  • Page 10: Voltage Range 0

    These battery types are available in local stores. Changingthe batteries requires disassembly of the device. Please contact your Kistler Service Center to send in your device for battery change. As a result of an integral semiconductor reference element, the Charge Generator does not need to be recalibrated after a battery change.
  • Page 11: Eu Declaration Of Conformity

    Kistler Instrumente AG, Winterthur Winterthur, April 2017 i.V. D. Müller Head of Electronics Products Lines 1 / 1 www.kistler.com 200.021.367 - a - YCE - Certificate - unrestricted - Released --- 18032171 Z13734B --- 23-Aug-2017 14:50 (CET) - Drd@int.kistler.com Page 8 5363A_002-812e-08.17...