Page 5
The service manual is solely provided to trained service engineers. If any problems or questions should arise, despite the information given in this manual. Please do not hesitate to call your local Skalar representative or the Skalar head office.
® Introduction of the BLUVISION ® complements the range of products for the automation of colorimetric chemistry by Skalar Analytical B.V. The BLUVISION ® is designed for laboratories that have a wide variety of sample types and matrices for analysis. The BLUVISION is solely to be used in ®...
The disc consists of 132 positions, which can all be used for (stock-) standards, reagents or QC positions. Of the 132 total positions, 24 are suitable for 50ml flasks. (Skalar cat no. 90035009, 160 pcs. natural colored and 90035008, 160 pcs. black colored) The additional 108 positions are suitable for 10ml reaction tubes.
The cuvette disc loader (2, Figure 1-1) is an automated system that loads the cuvette disc with new cuvettes (Skalar cat no. 272003600, box 576 pc.). Each cuvette consists of 10 cells, of which each cell can be used for a different test.
Option: Temperature regulated (heated) Sample and reagent pipetting (3, Figure 1-1) is done by the needle, included in the needle head (Skalar cat no. 272003505) which is connected to a dispense pump. Dilutions are prepared in the designated flasks or reaction tubes in the sample and reagent disc.
BLUVISION Analyzer Chapter 1 – Introduction - 7 - ® Step 4: Cuvette disc Figure 1-5 Cuvette disc Part Name Description Measuring site Here the absorption/transmission of the sample are measured. Cuvette This holds the cuvettes as they move round for different holder analysis stages.
A selection of 8 filters is placed to let through a specific wavelength of 340 – 900nm and filter out other wavelengths. When the reaction is finished, the light unit (Skalar article no. 272003301) determines the absorption of light by the sample at a wavelength between 340 and 900nm, set by the selection of the filter via hardware and in the software, DiscreteAccess.
BLUVISION Analyzer Chapter 1 – Introduction - 9 - ® Step 6: Cuvette waste bins Figure 1-7 Cuvette waste bin Part Name Description Waste bin Cuvettes are ejected into 1 of 2 compartments, determined by the User in the DiscreteAccess software.
Analyzer ® Display Figure 1-8 Display The BLUVISION ® display facilitates controlling the analyzer while standing next to it, as well as from a remote location (via PC). Many useful functions that are available in the software are also available within the display menu.
1.2.7.3 Tools tab - Filters Figure 1-11 Tools - Filters Part Name Description Current filter The current filter is shown Navigation The right and left arrow can be used to navigate through the filters Tools tabs Different BLUVISION elements can be selected ®...
Page 22
- 12 - Chapter 1 – Introduction BLUVISION Analyzer ® 1.2.7.4 Tools tab - Cuvettes Figure 1-12 Tools - Cuvettes Part Name Description The cuvette disc is fully loaded with cuvettes Load cuvette disc The cuvette disc is fully unloaded...
Page 23
BLUVISION Analyzer Chapter 1 – Introduction - 13 - ® 1.2.7.6 Tools tab – Needle Figure 1-14 Connection Part Name Description The needle will move up, out of the wash vessel. Needle up 1.2.7.7 Tools tab – Calibration Figure 1-15 Tools - Calibration...
- 14 - Chapter 1 – Introduction BLUVISION Analyzer ® 1.2.7.8 Connection tab Figure 1-16 Connection Part Name Description When connected to a network, the IP address of the IP Address BLUVISION ® is shown here Firmware versions of electronic boards are displayed Firmware versions here.
Never expose the unit to extreme heat, moisture or corrosive materials. • Skalar Analytical B.V. is not liable for damages arising from the failure to follow Skalar instructions for the proper use and handling of the instrument. Skalar Analytical B.V. is not liable for malfunctioning of the instrument, other damage or personal injuries arising from modifications made to the instrument without prior written approval of such modification from Skalar Management.
BLUVISION Analyzer Chapter 1 – Introduction - 17 - ® User Manual Symbols Note: Notes contain additional information about the text. Attention: Consult the instructions before starting the operation. Warning: Information about possible hazardous situations. When volatile and/or corrosive reagent is used Volatile and corrosive components may increase the risk of corrosion.
Warranty is subject to the following exception(s): Glassware breakage. Consumable products, brokerage and transport costs for the replacement parts replaced under the warranty are to the client’s account. Skalar will not be responsible for malfunctioning of the instrument if third party components are used in the instrument.
BLUVISION Analyzer Chapter 1 – Introduction - 19 - ® CUSTOMER ACCEPTANCE FORM Company Address: Post City: Country: Date: The undersigned authorized officer confirms that the performance of the instrument, as per the purchase order number as mentioned on the warranty registration form, is according to the specifications as submitted by the manufacturer.
Revision Date Edited Checked Approved 16-04-2024 Twan Velderman Digitally signed by Twan Velderman DN: cn=Twan Velderman, email=Velderman.T@skalar.com, c=US Date: 2024.04.18 10:15:24 +02'00' 12-10-2023 1.0.17 18-09-2023 V2.1 Added sleep function (section 1.2.7.1). • Changed Start and shut down procedures to make use of the sleep •...
If the light path of the cuvette is greasy/dirty, clean it with Ethanol and a clean cloth. Make sure all Ethanol is wiped off. Place the cuvettes, with the ‘Insert’ arrow towards the BLUVISION ® the tape on top, in the entry of the Cuvette disc loader.
Page 35
BLUVISION Analyzer Chapter 2 – Operating - 25 - ® 10. Repeat this for all 4 cuvette block slots > Close the Cuvette Hatch.
Figure 2-8 Lift disc cover 3. Lift the Inner disc and place it on a steady, clean surface. 4. Add bottles and/or sample tubes to the disc. 5. Place the inner disc back in the BLUVISION ® Note: The inner disc only fits in one way. (Inlets towards User) 6.
BLUVISION Analyzer Chapter 2 – Operating - 27 - ® Water supply (for analysis and reagents) Degassing water supply To avoid interference of gas during analysis, the water supply can be degassed. There are 3 options. 1. Vacuum (best) 2. Helium gas (good) 3.
‘‘Flush diluters’ till the new liquid reaches the waste, via the needle as well as via the bottom of the waste vessel (section 1.2.7.5). Empty and clean waste bin The waste bin is located behind the right front door of the BLUVISION ® Figure 2-1).
4. Right-click Ethernet > Left-click Properties. (Figure 3-2) 5. Select the top Internet protocol version (TCP/iPv4). (Figure 3-3) 6. Click Properties. 7. Select “Obtain an IP address automatically”. 8. Click OK > Close all windows > restart the PC and BLUVISION ®...
7. Select “Use the following IP address”. (Figure 3-4) 8. Fill in: a. IP address: 192.168.0.100 b. Subnet mask (auto): 255.255.255.0 c. Default gateway: 192.168.0.20 9. Click OK > Close all windows > restart the PC and BLUVISION ® Figure 3-1 Figure 3-2 Figure 3-3 Figure 3-4...
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 33 - First Log-in Note: Before an ‘Administrator’-password is set, pressing CTRL+F12 can be used to log-in. After opening DiscreteAccess, the ‘DA Login’-screen is shown: Figure 3-5 Login screen 1. Fill in ‘Administrator’ as ‘Login ID’...
- 34 - Chapter 3 – Software Introduction BLUVISION Analyzer ® User Interface description Commands Button Icon Description of function (upon clicking) Menu En-/disables selection of ‘About’, ‘Minimize’ and ‘Exit’ (1/2) Select ‘About’ to view the System and Instrument version...
Page 45
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 35 - Button Icon Description of function (upon clicking) Save Save an item Save As Save an item Print Print the shown information Edit Edit the selected item Cut selection Copy...
Page 46
- 36 - Chapter 3 – Software Introduction BLUVISION Analyzer ® Button Icon Description of function (upon clicking) Move Up Move the selection Up a Row Move Down Move the selection Down a Row Browse Browse in Windows Explorer Load...
Page 47
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 37 - Button Icon Description of function (upon clicking) Close Close the window Exit Exit the menu Exit Exit the menu...
BLUVISION Analyzer ® Home Menu Info Figure 3-6 Home Menu Part Icon Name Description Reference Connect the software to The BLUVISION ® , which 3.6.1.3 Disconnected is now disconnected Disconnect the software from the BLUVISION ® 3.6.1.3 Connected which is now connected...
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 39 - Analysis Menu Info Figure 3-7 Analysis Menu Part Button Name Description Reference Start Start analysis Barcode Scan Start Barcode Scan Stop Stop the currently running analysis Result Open the Result window to view live results...
- 40 - Chapter 3 – Software Introduction BLUVISION Analyzer ® Settings Menu Info Figure 3-8 Home Menu Part Button Name Description Reference Method Setup Open the Method Setup window Open the QC Setup window QC Setup Table Setup Open the Table Setup window...
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 41 - BV Settings Menu Info Figure 3-9 BV Settings Menu Part Button Name Description Reference Scan (system Open the Scan window admin only) Filter Wheel Open the Filter Wheel setting window...
- 42 - Chapter 3 – Software Introduction BLUVISION Analyzer ® Group and user management (Administrator only) DiscreteAccess ‘User Management’ makes it possible to create users and assign them to customized groups with specific rights and access. This feature improves the general overview and simplifies monitoring.
BLUVISION ® Analyzer Chapter 3 – Software Introduction - 43 - Create a Group 1. Select the ‘User Management’-tab. 2. Click Group Control > New Group. 3. Fill in and click Save. Delete a Group 1. Select the ‘User Management’-tab.
- 44 - Chapter 3 – Software Introduction BLUVISION Analyzer ® Application Settings ‘Application Settings’ is divided in three separate menus. (Figure 3-11) • Application Settings; Describes configuration of saving paths, scheduling settings, IP-address configuration and formatting of the run file names.
Page 55
The IP-address of the BLUVISION can be found by navigating ® to the ‘Connection’-tab on the display. (section 1.2.7.8) 3. Fill in the time in milliseconds after which the BLUVISION resumes ® operation when in ‘pause’-mode. 4. Allow Automatic (un)loading of cuvettes by (un)checking the box, respectively.
7. Click Save > confirm. Configuration Settings This menu shows the information used for communication with the SQL- server. No changes are to be made without consent of Skalar personnel. Temperature Settings To improve preservation of samples and reagent, quality of pipetting and...
BLUVISION ® CHAPTER 4: PRE ANALYSIS SOFTWARE Preface: This chapter gives a description for setup of methods and tables. Also a section is included that explains how to check hardware, using the software.
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 49 - PRE ANALYSIS SOFTWARE Method Setup A method is an established specific procedure which quantifies and/or qualifies an unknown, in this case, analyte. In this menu, the details for...
Page 60
® Note: Method settings are very critical regarding analysis and are developed by Skalar. 8. Fill in the details per method step. 9. If, for a reagent step, more of the same reagent is available, note the positions in the ‘Alternate Pos1’ and, if required, ‘Alternate Pos2’...
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 51 - 37. Select whether the Overrange Threshold is based on a specified concentration, or when the measurement is higher than the highest standard. 38. Select where to dilute the sample, in a Cuvette, Sample cup or extra cuvette.
- 52 - Chapter 4 – Pre analysis software BLUVISION Analyzer ® Edit a Method Group 1. Click Method Groups and select a method group. 2. Click Edit. Note: ‘Clear’ can be used to reset all fields. 3. Edit the group name in the ‘Group Name’-field.
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 53 - QC Setup Scheduling of quality control (QC) samples is done by setting the ‘Interval’, ‘After Calibration’ and ‘Start & End QC’ parameters. The QC measurements, set in the ‘QC Setup’ are scheduled in all analyses after the run is started.
- 54 - Chapter 4 – Pre analysis software BLUVISION Analyzer ® New Table 1. Click New. 2. Switch between ‘Method(s)’ and ‘Method Groups’ by clicking the slider. 3. Select the ‘Method(s)’ or ‘Method Groups’ for analysis by checking the box(es) in the ‘Select’-column.
Page 65
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 55 - 12. Select per sample which method should be performed, pre-dilution factors and duplo-measurements by checking the respective box(es). 13. Select whether a method has priority by clicking Priority Methods and selecting the method that has priority >...
- 56 - Chapter 4 – Pre analysis software BLUVISION Analyzer ® Include ‘Matrix spike’ (MS, recovery) To determine the Recovery the “Sample Type”: Matrix Spike (MS) is used. 1% of the sample volume (method) will be added to the sample from the spike solution.
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 57 - Import a Table The minimum requirements for importing a (partial) table into DiscreteAccess is that the use of Methods is specified (Column I and J in Figure 4-9). In the...
- 58 - Chapter 4 – Pre analysis software BLUVISION Analyzer ® Figure 4-10 Table menu during analysis Import a LIMS file for manual barcode scanner To make use of a LIMS file in DiscreteAccess, the file needs to consist of a CSV-file (comma separated value file) with the layout of the used sample table.
Page 69
BLUVISION ® Analyzer Chapter 4 – Pre analysis software - 59 - 4. Select the desired CSV-file location > click Open. 5. Navigate to the ‘Analysis’-tab. 6. Click Barcode Scan. Figure 4-13 Barcode scan 7. In the ‘Path’-section the CSV-file needs to be selected.
- 60 - Chapter 4 – Pre analysis software BLUVISION Analyzer ® Configuration of BLUVISION ® Settings (system admin only) The BLUVISION ® is equipped with many convenient options and settings that help maximize efficiency and accuracy of the analyses.
4. Click Check Position to verify. 5. Open the filter wheel on the BLUVISION and verify whether the filter is ® in the correct position > confirm on the BLUVISION ® display. 6. Click Save > OK. Figure 4-16 Filter wheel...
Chapter 4 – Pre analysis software BLUVISION Analyzer ® Tray Selection Because different racks are available for use in the BLUVISION ® , rack parameters can be set in the ‘Tray Selection’. Default values are shown below. Figure 4-17 Tray Selection The ‘Volume’...
Before analysis is started, the following should be checked and/or prepared. Refer to the designated section for additional information. • All safety instructions are read, understood and complied with. (section 1.3 - 1.9) • All steps described in the BLUVISION ® Installation Manual are executed correctly and completely.
- 66 - Chapter 5 – Analysis BLUVISION Analyzer ® Loading samples and reagent during analysis DiscreteAccess will display the status of the sample and reagent in use by the current analysis in the visualization of the Sample and Reagent disk (see Figure 5-1).
BLUVISION Analyzer Chapter 5 – Analysis - 61 - ® 1. Via the display, in the left top of the ‘Main screen’ (section 0) press the ‘add/refill’-button. 2. Select which container needs to be refilled. 3. Wait till the display shows to open the hatch > Open the hatch and refill.
- 68 - Chapter 5 – Analysis BLUVISION Analyzer ® Routine Operation Warning: Before operating the analyzer, always perform the maintenance instructions. Preparation of the analyzer 6. If the analyzer is sleeping, wake the analyzer by pressing the display. 7. Switch on the workstation.
BLUVISION Analyzer Chapter 6 – Post Analysis - 61 - ® POST ANALYSIS Manage run-files From the ‘Home’-menu click Post Analysis. Import a run Note: Dragging and dropping of run files into DiscreteAccess is not possible from network or USB-drives.
- 61 - Chapter 6 – Post Analysis BLUVISION Analyzer ® Manage run-details and results From the ‘Home’-menu click Post Analysis, double click a run-file to open. View run details Note: In Post-analysis, only part of the options are enabled.
BLUVISION Analyzer Chapter 7 - Report - 75 - ® REPORT Gathered data from analyses can be organized and displayed by designing a report format and adding data. The ‘Report Designer’ is implemented to create a uniform report structure, to which differing sets of data can be added via the ‘Report Viewer’...
Page 86
- 76 - Chapter 7 - Report BLUVISION Analyzer ® 7.1.1.1 Run information Template A table is opened, with on the left side run specific information. (1, Figure 7-2) This information can be dragged into the table, which will eventually show the values of the in ‘Report Viewer’...
BLUVISION Analyzer Chapter 7 - Report - 77 - ® Edit a Template 2. Click Open Template. 3. Select the desired Template. 4. Click Open > Edit the desired details. 5. Click: a. Save > OK or; b. Save As > Fill in a name > click OK > OK.
- 78 - Chapter 7 - Report BLUVISION Analyzer ® Report viewer options When the report is displayed in the Report Viewer, a user can select the tabs below to toggle between the methods or samples, depending on the type of report.
BLUVISION Analyzer Chapter 7 - Report - 79 - ® Export and print The final report can be exported to PDF or as a Comma Separated Value file (CSV file). Click Export to PDF, to export the selected report tab as a PDF document or Click Export all to PDF option to export all the report tabs as PDF files.
• Perform daily and weekly maintenance. • Clean the dust filter in the back cover of the analyzer. • Lubricate guideways (Skalar art. no. 90012384, section 8.5) • Update SQL and Windows. • Flush the tubing with 0.1% hypochlorite solution (section 8.2.3).
1. Place the desired rinsing chemical in R24. (Oxalic acid section 8.2.1, Hypochloride 0.1%) 2. Via the display, navigate towards the “Rinse” menu. (Figure 8-1) 3. Press Rinse Tubing From R24. (4, Figure 8-1) The BLUVISION will pickup from R24 and hold it above the rinse ® vessel.
Figure 8-2 QC setup Upgrading software and Database Patches Attention: The software should only be upgraded when recommended by Skalar Analytical B.V. The database is only to be patched when an upgrade of the software is performed or older run files are imported.
® E. Open the cuvette hatch. 6. Spray multi spray (Skalar art. No. 90012384) on the linear guideway. 7. Close the cuvette hatch. 8. Open the cuvette disk cover, incubator. 9. Remove the cuvette disk by unscrewing the 3 screws.
Page 98
- 88 - Chapter 8 - Maintenance BLUVISION ® Analyzer 6. Put the correct calibration tool in the position noted on display, take note: Figure 8-3 Calibration tools in BV A. For Sample Disk: • Use the Sample tool for "A" positions.
Page 99
BLUVISION ® Analyzer Chapter 8 - Maintenance - 89 - 17. Verify that all calibration tools are removed from the analyzer. 18. Reset the analyzer.
BLUVISION ® Analyzer Chapter 8 - Maintenance - 91 - Logbook Analyzer: Serial no.: Date of purchase: Responsible person: Date first analysis: Next Samples Date Maintenance description Performed by Remarks date cumulative...
- 92 - Chapter 8 - Maintenance BLUVISION ® Analyzer Logbook example Analyzer: BLUVISION ® Discrete Analyzer Serial no.: 17199 Date of purchase: 02-02-2017 Responsible person: Skalar Date first analysis: 05-03-2017 Performed Next Samples Date Maintenance description Remarks date cumulative 12/4 Weekly;...
- 95 - TROUBLESHOOTING Troubleshoot overview Note: (CTRL+) Click on the issue to navigate to the respective troubleshoot. Analyzer display is turned off, nothing happens. • There is no connection with the BLUVISION ® (anymore). • • X-loader error on start-up.
1. Power is not connected. Analyzer display is turned off, 2. Fuse is blown. nothing happens. Shoot 1. Check and connect all wiring to and from the BLUVISION ® 2. Replace the fuse. Trouble Cause 3. The network cable is not connected There is no connection with the (anymore).
Page 107
Replace a cuvette and verify that the sensor is orange, not blinking, but continuous orange. k. Restart the complete system. 6. Contact Skalar to upgrade the firmware and if necessary, the software. (always check compatibility and backup the database) Refer to the separate installation manual. Trouble Cause 7.
Page 108
10. Refill the sample or reagent, continue/restart the run. Perform 11 below if alternate positions or warnings in the software are desired. a. In the software, from the ‘Home’-tab, click Application Settings b. Click Application Settings c. Click the BLUVISION ® Settings-tab d. Enable ‘Level Detection’ by checking the box.
Page 109
13. Cuvette clip on new cuvette disk does not open fluently. Shoot 12. Contact Skalar. 13. Contact Skalar Trouble Cause 14. Locate the chemicals in the incubator, it may be...
Page 110
Perform another measurement when done. 16. . a. Refresh the water supply. b. Before use degas the water. (section 2.6.1) c. Place the supply vessel, insert the BLUVISION ® tubing. d. Flush diluters (via display) till all air is out of the tubing.
Page 111
BLUVISION ® Analyzer Chapter 9 - Troubleshooting - 101 - Trouble Cause 20. Flexible tubing is twisted. The Waste vessel is overflowing. 21. Waste vessel is clogged. 22. Boxer pump has stopped working. Shoot 1. Disconnect rinsing supply and remove all water.
Page 112
24. The wait time to stabilize between movements is too short. Shoot 23. Contact Skalar. 24. Contact Skalar. Trouble Cause 25. The movement is blocked. The pick and place unit stops during movement, showing a 26.
Page 113
BLUVISION ® Analyzer Chapter 9 - Troubleshooting - 103 - Trouble Cause 30. The cuvettes are loaded incorrect into the cuvette Sometimes a cuvette is loaded on storage. its side, from the cuvette loader in the cuvette disk. Shoot 30. Verify that all cuvettes are placed correctly into the cuvette storage, if so Contact Skalar.
Page 114
34. Update SQL via Windows update. 35. Number of tables is too big. Shoot 33. . a. Contact your Skalar representative for the newest software version. 34. . a. Restart the PC. b. Click Start (in Windows taskbar) c. Type: “Check for Updates”...
Page 115
Shoot 37. . a. Unload cuvette disk via the display to reset the error, see 2 in Figure 1-12. b. Contact Skalar if the problem persists. Trouble Cause 15. In-method dilutions in-cuvette using demi- When alternate diluent is used in water cannot be combined with alternate cuvette the results are incorrect.
Page 116
BLUVISION ® Analyzer Chapter 9 - Troubleshooting - 106 - Trouble Cause 18. The reagent volume needed, calculated by Table in software shows sufficient the software, does not include: Duplo/triplo reagent, but the run cannot be calibration, QCs, Extra volume in the completed because of low reagent method, Critical minimum.
BLUVISION ® Analyzer Chapter 10 - Technical Information - 109 - TECHNICAL INFORMATION Technical Specifications Frame: Stainless Steel profiles and sheets Analyzer dimensions: (W x D x H) 1210 x 790 x 560 mm Min. needed space: (W x D x H) 1210 x 650 x 560 mm...
Internal: SM3 704-1 Motor Cable 272008307 1 pc Yearly Maintenance articles Description Article No. Quantity Multispray plus lubricant 400ml for Guideways 90012384 1 pc Tubing set BLUVISION (V1) 272003800 1 set ® Tubing set BLUVISION (V2) 272003810 1 set ®...
Need help?
Do you have a question about the BLUVISION and is the answer not in the manual?
Questions and answers