Display message
&
Problem-solving guide
WHAT IT MEANS
ACTION
Full
battery
None
required.
Empty battery
: Warns when
battery is
low.
The unit can be
operated
but
recommended
to change
new batteries
tor
correct result!
Replace new batte ri es.
II display
"BAT",
the unit
can
not be
operated,
change new batteries
immediately.
DOL
LOfL
Replace new
batte
ries
immediately.
II
the sampling
volume
is tOD
low
or not
sufficiently blow.
Press
CID
button again to
repeat
the
measurement.
II
there is na breath sample blown
within
30
seconds,
the LCD
will
display
"OUT".
Press
CID
button to restart
the countdown
cyde
and
retest.
ff
the unit is
powered on
outside of the
acceptable temperature
range(0-40'C
or
32-104F) the unit
will
display
it.
Repeat the
test after the unit
has reached a
temperature
within
the operation
range.
Technical specifications
0.000
-
0.500%BAC
Indication
ol B.A.C.
0.00 - 5.00
%,
0.00 -
2.50mg/1
+/-
0.005%BAC
at O.l%BAC
at 25'C
Accuracy
+/-
0.05
%,
at 1.0
%,
at 25'C
+/-
0.025mg/1
at
0.50mg/1
at 25'C
Warm up
time
Within
18 seconds at
0.05%BAC
Response time
Within
10
seconds
at
0.05%BAC
Sensor
Fuel cell sensor
Power
supply
2 (two)
AAA size alkaline
batteries
Battery
Lile
Approx.
1000 tests
Dimensions(HxWxD)
48mm
x
115mm
x 18mm
Weight
93g induding
batteries
Calibration
Every
12
months
or
after
500 tests
Operating
temperature
0-
40'C
Manufacturer
/ Distributor
:
Need help?
Do you have a question about the DA-7100 and is the answer not in the manual?
Questions and answers