SENSIRION SGD4 Series Design-In Manual

Refrigerant detection system

Advertisement

Quick Links

Design-in Guide SGD43S-M3-Sx
For SGD43S-M3-Sx Refrigerant Detection System
The SGD4x sensor family is Sensirion's series of gas concentration sensors designed for the HVACR industry.
The sensor is fully calibrated and compensated for maximum accuracy and ease of integration. This document
aims at helping the proper design-in of a refrigerant sensor / detection system in HVACR appliances.
Contents
1
Select the right SGD43S-M3-Sx variant for your application ........................................................................................... 2
2
Sensor & Refrigerant Detection System Operation .............................................................................................................. 2
3
Mechanical Considerations ............................................................................................................................................................. 3
3.1 Appliance Integration ............................................................................................................................................................. 3
3.2 Sensor Orientation ................................................................................................................................................................... 3
3.3 Sensor Mounting ...................................................................................................................................................................... 4
3.4 Inlets .............................................................................................................................................................................................. 4
3.5 Vibration ...................................................................................................................................................................................... 4
3.6 Sunlight ........................................................................................................................................................................................ 4
4
Electrical considerations................................................................................................................................................................... 5
4.1 Connector .................................................................................................................................................................................... 5
4.2 Power supply .............................................................................................................................................................................. 5
4.3 Multicomponent System ....................................................................................................................................................... 5
5
Revision History ................................................................................................................................................................................... 6
www.sensirion.com / D1
Scan me to provide feedback
Version 1.0 - Apr 2025
1 / 7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SGD4 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for SENSIRION SGD4 Series

  • Page 1: Table Of Contents

    Design-in Guide SGD43S-M3-Sx For SGD43S-M3-Sx Refrigerant Detection System The SGD4x sensor family is Sensirion’s series of gas concentration sensors designed for the HVACR industry. The sensor is fully calibrated and compensated for maximum accuracy and ease of integration. This document aims at helping the proper design-in of a refrigerant sensor / detection system in HVACR appliances.
  • Page 2: Select The Right Sgd43S-M3-Sx Variant For Your Application

    Figure 1: Timeline of the operation of the SGD43S-M3-Sx, with a leak event. The resulting Refrigerant Detection System (RDS) states, status light and relay states are depicted on the upper part. The sensor’s refrigerant concentration reading is plotted on the bottom part. www.sensirion.com / D1 Version 1.0 – Apr 2025 2 / 7...
  • Page 3: Mechanical Considerations

    NOTICE: The sensor should not be installed with the connector facing upwards, especially in situations where water or condensate may be dripping on the sensor from above. www.sensirion.com / D1 Version 1.0 – Apr 2025 3 / 7...
  • Page 4: Sensor Mounting

    3.6 Sunlight Do not expose sensor to direct sun. Exposing SGD43S-M3-Sx to direct sunlight might introduce temperature gradients and accelerate aging. www.sensirion.com / D1 Version 1.0 – Apr 2025 4 / 7...
  • Page 5: Electrical Considerations

    Communication with each distinct sensor requires each device to have a dedicated address. The device address can be modified using the Modbus interface described in the sensor datasheet. www.sensirion.com / D1 Version 1.0 – Apr 2025 5 / 7...
  • Page 6: Revision History

    Revision History Date Version Pages Changes April 2025 Initial version www.sensirion.com / D1 Version 1.0 – Apr 2025 6 / 7...