Lambda PRECIFLOW touch Manual

Peristaltic pump
Table of Contents

Advertisement

Quick Links

LAMBDA PERISTALTIC PUMP
Revision 12/2025
MANUAL
PRECIFLOW touch
HiFLOW touch
MAXIFLOW touch
MEGAFLOW touch
Software version 5.00 or later.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PRECIFLOW touch and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lambda PRECIFLOW touch

  • Page 1 LAMBDA PERISTALTIC PUMP Revision 12/2025 MANUAL PRECIFLOW touch HiFLOW touch MAXIFLOW touch MEGAFLOW touch Software version 5.00 or later.
  • Page 2: Table Of Contents

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 TABLE OF CONTENT Technical description ........................5 Technical parameters of LAMBDA Peristaltic Pumps ............. 5 Device description ........................ 6 Safety .............................. 7 Intended use of the pump ....................7 Use restrictions ........................7 Safety for installation, operation, cleaning, maintenance and storage ......
  • Page 3 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 Start dosing .......................... 21 Stop dosing .......................... 21 Change of dosing direction ....................22 Maximum ..........................22 Set DIGITAL SPEED ......................22 Set FLOW RATE ........................23 Elapsed dosing time (Delivery time) ................. 24 7.10...
  • Page 4 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 12.2 Basic commands ......................... 41 12.3 Commands with objects ..................... 42 12.4 response object description ....................43 13 CAN bus communication (REMOTE 1) ..................45 13.1 Description of communication ..................45 13.1.1 Extended data frame ....................45 13.1.2...
  • Page 5: Technical Description

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 1. TECHNICAL DESCRIPTION 1.1 TECHNICAL PARAMETERS OF LAMBDA PERISTALTIC PUMPS PRECIFLOW HiFLOW MAXIFLOW MEGAFLOW Type Microprocessor-controlled programmable laboratory peristaltic pump Accuracy ± 5% 0–1000 rpm 0–2800 rpm 0–3200 rpm 0–3200 rpm Digital speed...
  • Page 6: Device Description

    PUMP HEAD COVER TOUCH SCREEN TUBING INPUT/OUTPUT PUMP HEAD METAL PLUNGER FOR COVER FIXATION LAMBDA REMOTE 1 CONNECTION (Smart adapters, LAMBDA MINIFOR 2 BIO) POWER SUPPLY External 12 V DC adapter PC CONNECTION USB connection LAMBDA REMOTE 2 CONNECTION (Backward compatibility and old generation functionality)
  • Page 7: Safety

    PERISTALTIC PUMPS Manual 12/2025 2 SAFETY Thank you for choosing the LAMBDA peristaltic pump - LAMBDA PRECIFLOW touch, HiFLOW touch, MAXIFLOW touch / MEGAFLOW touch. This manual from LAMBDA CZ s.r.o. offers safety information derived from laboratory experience and expertise in designing laboratory instruments.
  • Page 8: Safety For Installation, Operation, Cleaning, Maintenance And Storage

    Repairs and services can only be carried out by an authorized person who is aware of the hazard involved. LAMBDA CZ s.r.o. assumes no liability for any service or repair performed by the user, an unauthorized person, or third-party companies.
  • Page 9: Guarantee On Lambda Peristaltic Pump

    These guarantees cover proven material and manufacturing defects, provided that the instrument was used following the operational manual and advice given by LAMBDA. Making a warranty claim will not affect the duration of the warranty. Further claims are excluded.
  • Page 10: Power Supply Connection

    4 POWER SUPPLY CONNECTION 4.1 POWER SUPPLY FOR REMOTE-CONTROLLED PUMP Do not use any power supply for the LAMBDA peristaltic pump if the pump is connected to a bioreactor/fermenter LAMBDA MINIFOR or LAMBDA MINIFOR 2 BIO! 1. Plug the connector of the remote-control cable from the LAMBDA MINIFOR / LAMBDA MINIFOR 2 BIO into the corresponding socket (REMOTE) at the rear of the peristaltic pump.
  • Page 11: Loading Procedure For Tubing

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 5.1 LOADING PROCEDURE FOR TUBING Steps: 1. Remove the cover glass from the pump head: Gently press on the cover glass and rotate it counter- clockwise. Remove the cover glass in the upward direction.
  • Page 12 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 3. Test the selected speed before insertion of tubing into LAMBDA PRECIFLOW touch, HiFLOW touch or MAXIFLOW touch. For LAMBDA MEGAFLOW touch, employ the lowest speed when inserting the tubing. SPEED RANGE INDICATOR...
  • Page 13 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 4. Apply a small amount of silicone grease to the upper surface of all three plastic bearings that come into contact with the cover glass. 5. Place the cover glass back onto the pump head.
  • Page 14: Menu (User Control Logic)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 6 MENU (USER CONTROL LOGIC) The menu is only available in the STOP state when dosing is not in progress: On the main screen in stop mode,  Press the icon "MENU". ...
  • Page 15: Program Library

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 6.3 PROGRAM LIBRARY The built-in program library allows the creation of 10 programs, where each program can contain up to 100 records. Use the “UP” and “DOWN” icons to navigate horizontally and change the program. The highlighted program is considered selected for use.
  • Page 16: Units

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 6.5.2 UNITS The dosing rate unit can be selected here. The following are available: digital speed, ml/h, ml/min, l/h. In order to use the instrument in volume units (ml/h, ml/min, l/h), it is necessary to make and set a calibration constant for the tubing used.
  • Page 17: Remote 2 I/O

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025  Auto-reset when start If the option is active, each new batch run in program mode automatically starts the program from the beginning. If not active, the display shows the option “Continue or Restart”.
  • Page 18: Master Control Input

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 Setting the inpute remote signal function. This hardware signal is available on the “14 REMOTE 2 connector REMOTE 2 connector (for wiring see chapter functionality”).  Disabled The function is disabled  Stop Activation of the signal will stop the dosing ...
  • Page 19: Device Information

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 Solution names enables information about the used substance in the tubing to be shown on the display. These names can then be selected manually by clicking on them on the main screen. (For details about fluids name library, see chapter “10 Fluids name...
  • Page 20: Direct Mode (Fixed Value)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 7 DIRECT MODE (FIXED VALUE) Direct mode is a mode in which the speed is controlled by a fixed value. 7.1 HOME SCREEN DESCRIPTION The home screen is always displayed when the pump is idle. The parameter setting is done by clicking on the value.
  • Page 21: Start Dosing

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 If volume units are selected in the settings, calibration is performed (calibration constant is set), the dosing rate is automatically displayed as flow rate instead of speed. In addition, information (counter) about the dispensed volume is automatically available.
  • Page 22: Change Of Dosing Direction

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 7.5 CHANGE OF DOSING DIRECTION The arrow icon shows the currently selected direction of rotation of the pump head rotor. • Changing the flow direction is done by clicking the "DIRECTION" icon. 7.6 MAXIMUM •...
  • Page 23: Set Flow Rate

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 7.8 SET FLOW RATE Available when volume units (ml/h, ml/min or l/h) are selected. The flow rate can be changed during pumping with DIRECT flow control. To set the flow rate click on the flow rate parameter on the main screen.
  • Page 24: Elapsed Dosing Time (Delivery Time)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 7.9 ELAPSED DOSING TIME (DELIVERY TIME) The duration of elapsed dosing time is available as “Delivery time” on the main screen by clicking the right or left arrow icon. • A value is reset by clicking on the value.
  • Page 25: Calibration And Flow Rate Volume Units

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 8 CALIBRATION AND FLOW RATE VOLUME UNITS A Calibration is necessary to convert and directly set the flow rate. During calibration the amount of liquid the pump dispenses during a set time is measured. It is best to use a laboratory scale to measure the amount of liquid.
  • Page 26: Performing Calibration

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 8.1 PERFORMING CALIBRATION Go to calibration by clicking the "CALIBRATION" item in main menu. Prepare to measure the amount of liquid dispensed. (See the illustration above as an example of measurement with scale.) Steps: 1.
  • Page 27 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 3. CALIBRATION IN PROCESS Wait for calibration (dosing) to complete. After the calibration time has elapsed, the pump will automatically stop and return to the calibration screen. 4. CALIBRATION Enter the dispensed volume:...
  • Page 28: Program Mode (Pre-Defined Program)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9 PROGRAM MODE (PRE-DEFINED PROGRAM) For use in program mode it is necessary to activate the control method in the menu on pre-defined program. Program structure: Name Name of program Segment (defined by flow rate/speed, time length, type of...
  • Page 29: Dosing Screen Description

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.2 DOSING SCREEN DESCRIPTION After starting the dosing, the dosing screen will appear on the display. CURRENT SEGMENT SPEED/FLOW RATE REMAINING TIME TO NEXT SEGMENT 9.2.1 PROGRAM OVERVIEW During the program processing, it is possible to display the program's current parameters.
  • Page 30: Select Program

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.3 SELECT PROGRAM Changing or selecting the program is only possible when the pump is stopped, idle. Steps: 1. Go to the program library by clicking on the Program icon on the main screen or by clicking on Program library in the menu.
  • Page 31: Edit Name

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.4.1 EDIT NAME Click in the program name area to change it. The screen will show the current name with an alphanumeric keyboard. Confirm the change by clicking the "OK" button. PROGRAM NAME AREA 9.4.2 ADD NEW SEGMENT...
  • Page 32: Remove Segment

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.4.3 REMOVE SEGMENT 1. Select the segment you want to remove (highlighted table row) by clicking the up and down arrow icons. 2. Click the "MINUS" icon to remove a currently highlighted program segment.
  • Page 33: Action On End

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.4.4.1 ACTION ON END Choose what action shall take place to process all program segments: REPEAT PROGRAM: The program will be processed again from the first segment. The number of repetitions can be set.
  • Page 34: Calibration

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.4.4.4 CALIBRATION To use volumetric units, a conversion calibration constant must be set. The constant for the type of tubing used is obtained by calibration, see chapter Enter the calibration constant, click "SET" to confirm the new value or "BACK"...
  • Page 35: Program Pause

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 9.5.2 PROGRAM PAUSE For interruption/pause of the dosing program, click on “STOP”, on the pump display. The dosing / dosing program is interrupted. 9.5.2.1 CONTINUE OR RESTART OF THE DOSING PROGRAM AFTER A PAUSE To finish the pause, click on “START, on the pump display.
  • Page 36: Fluids Name Library

    It is used to quickly find out what preparation the pump is dispensing. The library items can be edited from a PC via the LAMBDA Device Manager software application. The number of items is limited to 32.
  • Page 37: Software Update

    1. Download the archive “pump-2-x.yy.zip” file. 2. Decompress archive containing pump software file “pump-2-x.yy.hex” 11.2 PC SOFTWARE APPLICATION The PC software application LAMBDA Device Manager (ldm) is available for pump software updates. Steps: 3. Download “ldm-x.y-x64-setup.exe” file (X, Y subsites version number) 4.
  • Page 38: Update Procedure

    1. Connect the supplied 12V DC power adapter to the pump. 2. Connect the pump to the PC using the A-B USB cable. 3. Start the Lambda Device Manager application. 4. Wait and check that the computer has detected the pump.
  • Page 39 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 7. Click the “Select file” button and browse the path to the decompressed peristaltic pump software file (pump-2-X.YY.hex). 8. Click on the “Start” button to initiate the update process. Do not disconnect the power or USB communication cable...
  • Page 40 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 Software update done 9. The progress of the process is displayed in the lower part of the window labelled "UPDATE STATUS". Wait for the process to complete. When finished, a new text window will appear: “Software update is successfully finished.”...
  • Page 41: Usb Communication

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 12 USB COMMUNICATION The peristaltic pump can be set and controlled using a USB interface via a virtual COM port. The communication is text-based and uses the JSON data transfer format. After connecting to a PC, a virtual COM port (COMx) is created automatically. Most operating systems will not require the installation of additional drivers, as the peristaltic pump is identified as a CDC-class device using the USB Descriptor.
  • Page 42: Commands With Objects

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 VALUE RESPONSE NOTES Reset configuration to SetDefaults {"ACK":1} factory defaults. Set operating mode SetOpMode 0 or 1 {"ACK":1} (RUN=1, STOP=0) GetVer Return object Version GetProcData Return object ProcData GetDeviceInfo Return object DeviceInfo Information about device...
  • Page 43: Response Object Description

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 12.4 RESPONSE OBJECT DESCRIPTION Example: {"ProcData":{"Flow":1000,"OpMode":0,"DelivTime":61128,"DelivVolume":0.6,"Directio n":1,"FluidName":"ACID","FlowUnit":0,"Calibration":200.000}} OBJECT KEYS VALUES VALUE DESCRIPTION Version String Hardware version Number Software version (major.minor) Number Serial number ProcData Flow Number Flow rate if non-rpm unit is set...
  • Page 44 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 OBJECT KEYS VALUES VALUE DESCRIPTION Current calibration constant float respect two decimal points – max. Calibration Float 999.99 FlowControl 0=Direct/1=Program Current fluid name text (max.32 FluidName String characters) 0=smooth, 1=normal (only for touch...
  • Page 45: Can Bus Communication (Remote 1)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13 CAN BUS COMMUNICATION (REMOTE 1) The peristaltic pump implements the CAN Specification 2.0B (Controller Area Network) interface. This asynchronous serial data communication protocol provides reliable communication in an electrically noisy environment. For internal purposes, the auxiliary bits in the extended identifier are used for the 1x master/n-slave communication model.
  • Page 46 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 extended data frames. The IDE bit indicates the data frame type. IDE is 1 for the extended data frame. The extended data frame Control field consists of seven bits. The first bit is the RTR. For the extended data frame, RTR is 0. The next two bits, RB1 and RB0, are reserved bits in the dominant state (logic level ‘0’).
  • Page 47: Eid Bitwise Significance

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 • CAN_PURPOSE • CAN_ROTATION It sends these commands periodically, approximately every 50 milliseconds. (For more information about commands, see Chapter 13.3) 13.1.2 EID BITWISE SIGNIFICANCE The extended data frame is composed as follows. Serial number 3932390 is used for the following examples.
  • Page 48: Message Filtering

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 EID decomposed into SID and EID for arbitration fields (master sends) EID (29 bits) – 0x083C00E6 SID (11 bits) EID (18 bits) 010 0000 1111 00 0000 0000 1110 0110 0x20F 0xE6 0x083C00E6 >> 18...
  • Page 49: Commands (Pseudo-Identifier)

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13.3 COMMANDS (PSEUDO-IDENTIFIER) The commands (pseudo-identifiers) are used to identify the type of message that the peristaltic pump is sending or that the pump is to process. Commands sent to the pump (EID bit 28 = 0 and 27 = 1) may or may not contain values. Some commands contain only the command code in the data portion of the message.
  • Page 50 LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 Extracting the serial number from the EID identifier (hexadecimal): 0x183C00E6 bitwise AND 0x3FFFFFF = 0x3C00E6 (3932390 Command Type DATA0 = 0x80 -> CAN_STATUS Command Data DATA1 = 0x03 -> Peristaltic pump Preciflow DATA2 = 0x00 -> STOP operating mode DATA3 = 0x00 ->...
  • Page 51: Read Commands

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13.3.1 READ COMMANDS The table lists the commands the pump sends and describes the information they provide. Some commands are sent asynchronously, see section (13.1). Only the meaning of the data part of the CAN message (frame) is described in the table. The whole message is described above.
  • Page 52: Write Commands

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13.3.2 WRITE COMMANDS The table contains commands that can be used to control the pump or change parameter values. Only the commands, i.e., the data part of the CAN message, are described in the table. Creating a complete CAN message/frame is necessary for writing, as described in chapter 13.1.1.
  • Page 53: Data Format

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13.3.3 DATA FORMAT • Integer - 32-bit signed integer, in little-endian format with the Least Significant byte (LSB) at lower data byte (DATA1). EXAMPLE, READ DATA0 DATA1 DATA2 DATA3 DATA4 (command code) 0x183C00E6...
  • Page 54: Heartbeat

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 • Strings (ASCII characters) Strings are represented as byte arrays (hexadecimal values of ASCII characters). The maximum length of characters is limited to 32. The end of string is identified by byte 0x00. The user application must ensure the creation of a chain based on the processing of up to 4 messages.
  • Page 55: Connector Wiring

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 13.5 CONNECTOR WIRING CAN_L 12 V DC (WHITE) (RED+BLUE) CAN_H (BLACK+ (GREEN) YELLOW) Marking Color Purpose CAN_L White CAN- CAN_H Green CAN+ Black+Yellow Ground (Device power supply) Device power supply It is used to power the entire device instead of using an external...
  • Page 56: Remote 2 Connector Functionality

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 14 REMOTE 2 CONNECTOR FUNCTIONALITY 14.1 CONNECTOR WIRING Colour Description yellow (+) input remote speed control 0-10V *) grey step signal from motor (0 and 12V) green earth, 0 V brown + 12 V (+) input remote ON/OFF;...
  • Page 57: Commands Not Containing Data

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 is the speed of rotation (3 ASCII numbers from 0 to 9; sent from the highest order digit to the lowest order digit) is the control sum in HEX format (2 ASCII signs of the type 0…9ABCDEF)
  • Page 58: List Of Accessories

    The pump will go to the local command (pump front panel is activated). 15 LIST OF ACCESSORIES • 800113 Stainless steel tubing clamp Tubing for LAMBDA PRECIFLOW touch, HiFLOW touch & MAXIFLOW touch: (Article number, material, inner diameter/outer diameter, length) • 4815-1 Silicone tubing 0.5/2.5 mm...
  • Page 59: Revision History

    LAMBDA Laboratory Instruments PERISTALTIC PUMPS Manual 12/2025 16 REVISION HISTORY Description Reorganization of chapters (direct mode / program mode) Added USB and CAN communication chapters Added chapter with connector wiring Update chapters 12.1, 12.3, 12.4 Large revision of chapter 13 Product name extension “touch”.

This manual is also suitable for:

Hiflow touchMaxiflow touchMegaflow touch