Page 3
JA – 日本語 EN - ENGLISH ICP EXPRESS MONITOR ICP EXPRESSモニタ前面パネル FRONT-PANEL A オン/オフキー A ON/OFF Key B 患者モニタゼロキー B Patient Monitor Zero Key C 患者モニタ20 mmHg較正キー JA – 日本語 EN - ENGLISH C Patient Monitor 20 mmHg D 患者モニタ100 mmHg較正キー ICP EXPRESS REAR-PANEL ICP EXPRESS背面パネル...
Page 4
ICP EXPRESS Monitor ® EN - ENGLISH FUSE REPLACEMENT JA – 日本語 ヒューズ交換 JA – 日本語 EN - ENGLISH 患者モニタ感度選択スイッチ PATIENT MONITOR SENSITIVITY SELECTION SWITCH A 5 µVの位置 A 5 µV Position B 40 µVの位置 B 40 µV Position EN - ENGLISH AC POWER CORD RETAINER JA –...
Page 5
EN – ENGLISH Codman Patient connections are electrically isolated Type CF. Do not let patient connections ® contact other conductive parts, including earth. ICP Express Monitor ® Routinely inspect all electrical plugs and connections. Do not use if damaged. CAUTION: Federal (USA) law restricts this device to sale by or on the order of a Maintain strict sterile technique when connecting the ICP EXPRESS Cable to the physician Codman Microsensor.
Page 6
EN – ENGLISH Do not submerge the tip of the Codman Microsensor or ventricular catheter vertically Front Panel controls (See Figure 1) in a cup of sterile water or saline during the zeroing process. Doing so will impose ON/OFF KEY—Turns the ICP EXPRESS operating system ON and OFF. pressure on the transducer diaphragm that is higher than atmospheric zero, resulting NOTE: When ON, the ICP EXPRESS can be powered either from an in an inaccurate zero reference value.
Page 7
EN – ENGLISH Rear Panel Connectors Operating Instructions The following instructions describe all necessary steps required to set-up and Patient Monitor Interface Connector operate the ICP EXPRESS with the Codman Microsensor ICP Transducer and an Compatible patient monitors can be connected to this receptacle external patient monitor.
Page 8
EN – ENGLISH 3 Next, the ICP EXPRESS will automatically display instructions for calibrating 3 Upon successful completion of the zeroing process, the ICP EXPRESS will display the Codman Microsensor’s three digit zero reference value. Record the displayed the patient monitor. Following the ICP EXPRESS prompt, start by pressing the key to send a 20 mm Hg or 100 mm Hg electronic reference signal zero reference value on the Codman Microsensor connector housing and on the to the patient monitor.
Page 9
Switching ICP EXPRESS Cables ICP patient monitoring system (i.e., ICP EXPRESS, ICP EXPRESS Cable, patient Integra LifeSciences recommends that the ICP EXPRESS Cable used during the monitor) is changed or substituted while patient monitoring is in progress. original transducer zeroing procedure remain with the patient at all times. However,...
Page 10
EN – ENGLISH 3 If connected to a patient monitor, the ICP EXPRESS will prompt the user to Zero Fuse Replacement and Calibrate the patient monitor as described in the Operating Instructions 1 Turn the ICP EXPRESS OFF and remove the power cord from the rear panel AC section.
Page 11
EN – ENGLISH The ICP EXPRESS will erase the previously stored zero reference value and store External Patient Monitor Interface a new zero reference value in the ICP EXPRESS Cable whenever a new Codman Output Signal Sensitivity: 5 µV/Vex/mm Hg or Microsensor is zeroed, or when the user chooses to adjust the zero reference 40 µV/Vex/mm Hg (user selected;...
Page 12
Maximum leakage current must not PRODUCT INFORMATION DISCLOSURE exceed 10 µA INTEGRA LIFESCIENCES CORPORATION (“INTEGRA”) HAS EXERCISED REASONABLE In addition check that: CARE IN THE SELECTION OF MATERIALS AND THE MANUFACTURE OF THESE PRODUCTS. INTEGRA WARRANTS THAT THESE PRODUCTS SHALL CONFORM External fuses correspond to the values prescribed.
Page 13
EN – ENGLISH WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of ICP Express Monitor 82-6637 and its accessories. Otherwise, degradation of the performance of this equipment could result. Appendix A –...
Page 14
EN – ENGLISH Table 3 Guidance and Manufacturer’s Declaration Regarding Electromagnetic Immunity – Non-life Supporting The Model 82-6637 ICP EXPRESS Monitor is intended for use in the electromagnetic environment specified below. The customer or user of the Model 82-6637 ICP EXPRESS Monitor must ensure that it is used in such an environment.
Page 15
EN – ENGLISH This page is intentionally left blank.
Page 16
EN – ENGLISH Reference Symbol Title of Symbol Meaning of Symbol Standard Number Indicates the medical device manufacturer, as Manufacturer defined in EU Directives 90/385/EEC, 93/42/ 5.1.1 EEC and 98/79/EC. Authorized representative Indicates the authorized representative in the in the European 5.1.2 European Community.