 Multifunctional Alcohol Wear Tester
MU-ABR5
 
Operation panel and diagram 
Appearance diagram 
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 ,3
mrc@mrclab.com
MRC. 2.25

Advertisement

loading
Need help?

Need help?

Do you have a question about the MU-ABR5 and is the answer not in the manual?

Questions and answers

Summary of Contents for MRC MU-ABR5

  • Page 1  Multifunctional Alcohol Wear Tester MU-ABR5   Operation panel and diagram  Appearance diagram  Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 ,3 mrc@mrclab.com MRC. 2.25...
  • Page 2 measuring rod clamping  Place the rubber in the copper sleeve. Be careful not to leave too much rubber. If the  rubber  is  too  long,  it  will  easily  swing  during  the  test,  which  will  affect  the  test  results.  Place  the  copper  sleeve  with  the  rubber  inside  the  measuring  rod  and  tighten  the  nut.  Schematic diagram of cotton cloth clamping  Replace the alcohol cotton cloth test probe, tie the test cotton cloth with a cable tie  as shown, or use a rubber band instead of a cable tie to tie it tightly.  Sample clamping diagram...
  • Page 3 Schematic diagram of wear‐resistant machine sample clamping  Tighten the test sample with a clamp as shown in the figure. The four clamps can be  adjusted  according  to  the  position  of  different  products  until  the  product  is  completely fixed.  Diagram of placement of wear test rod of wear tester  Place the test rod at the position where the wear test is required. Note that a certain  space needs to be reserved between the test rod and the red guide block, otherwise  the test weight will not be loaded on the test sample, but directly placed on the red  guide block, resulting in a zero test load.  The left screw can adjust the front and back position of the measuring rod, and the  right screw can adjust the up and down height of the measuring rod  Wear test stroke adjustment device  The black plastic nut can be adjusted to adjust the wear test stroke. Turning it  clockwise will increase the test stroke, and turning it counterclockwise will decrease  the test stroke. ...
  • Page 4 Operation Instructions  Test Preparation:  1. Adjust the speed regulator switch to the lowest setting; 2. Turn on the power switch and adjust the speed regulator to the appropriate speed; 3. According to the set items, the whole machine is in the test state; 4. Select cotton cloth or rubber head or other test media to fix in the test fixture; 5. Run the machine idling to see if there is any abnormality. After confirming that the machine is normal, conduct product testing. Operation Specifications:  1. Connect the appropriate power supply 220V according to the mark. 2. Adjust  the  speed  regulator  switch  to  the lowest  setting  to  avoid  excessive  speed and cause abnormality of the machine. 3. Turn on the power and test the machine first. If there is any abnormality, turn off the power. 4. Press the reset button to reset the counter. 5. Press and hold the > button for eight seconds to select the counter function. After pressing and holding the > button for eight seconds, the counter displays XX‐1. Press the ∧ button to set the counter function. 29‐1 is a four‐digit counter and 32‐1 is an eight‐digit counter. Set the required number of tests according to the test requirements. ...
  • Page 5 Top left1  Top left2  Top left3  Top left4  Bottom right4  Bottom right3  Bottom right2  Bottom right1  6. Place the sample to be tested on the test bench and clamp it. 7. Select the required measuring rod, fix the test fixture, and add the corresponding test weight to meet the load requirements of the test. 8. Press the start button and the whole machine will enter the test state. 9. Adjust  the  speed  knob  on  the  speed  regulator  to  make  the  machine  reach  the required test speed requirements. 10. After the whole machine tests the number of times set by the counter, it stops and is in standby state. 11. If you need to temporarily stop the machine during the test, just press the stop button. If you need to restart, just press the start button to resume the action. 12. If there is an abnormal situation during the test, please press the power switch directly to cut off the power. 13.