V 2400
EN
Installation and operating instructions
0297

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the V 2400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Durr Dental V 2400

  • Page 1 V 2400 Installation and operating instructions 0297...
  • Page 2 The current version of the installation and operating instructions is available in the Download Center: http://qr.duerrdental.com/9000-606-97 © DÜRR DENTAL SE...
  • Page 3: Table Of Contents

    Contents Contents Pipe materials ....Hose materials ....Information about electrical con- Important information nections .
  • Page 4: Important Information

    – NOTICE These installation and operating instructions Risk of extensive material/property damage apply to: V 2400 Other symbols Order number: 7137-02 These symbols are used in the document and on or in the unit: Warnings and symbols Note, e.g.
  • Page 5: Copyright Information

    Important information Safety CE labelling with the number of the noti- fied body The unit has been developed and designed in such a way that dangers are effectively ruled out Order number if used in accordance with the Intended Use. Despite this, the following residual risks can Serial number remain:...
  • Page 6: General Safety Information

    Important information The unit must not be used as a vacuum Specialist personnel ❯ cleaner. Operation Do not use chemicals containing chlorine or ❯ Unit operators must use their training and knowl- foaming chemicals. edge to ensure safe and correct handling. Operation in operating theatres of explosive ❯...
  • Page 7: Notification Requirement Of Serious Incidents

    Important information Notification requirement of serious incidents The operator/patient is required to report any serious incident that occurs in connection with the device to the manufacturer and to the com- petent authority of the Member State in which the operator and/or patient is established/resi- dent.
  • Page 8: Product Description

    V 2400, 400 V, 3~, 50/60 Hz with con- trol box ......7137-02 –...
  • Page 9: Technical Data

    Product description Technical data Electrical data 7137-02 Rated voltage 400, 3~ Mains frequency Nominal current Start-up current Motor protection Rated power Type of protection IP 21 Protection class Connections 2 x Æ 50 (DN50) Suction connection (outside) Exhaust air connection (external) Æ...
  • Page 10: Characteristic Curves

    Product description Classification Medical Device Class (MDR) Characteristic curves Fig. 1: Characteristic curves for 7137-02, operation at 50 Hz / 60 Hz, measured in accordance with ISO 10637 The pressure in the suction system is limited by the supplied auxiliary air valve (50 Hz or 60 Hz operation).
  • Page 11: Type Plate

    Product description Operation Type plate The V-suction unit is suitable for use in dry air The model identification plate is located on the suction systems. The advantage of this system is turbine housing. that the suction unit, regardless of the actual connection layout, can be installed in any availa- ble and suitable room (including upper floors or basements).
  • Page 12: Assembly

    Assembly Hose materials Assembly For waste connections and suction lines only use the following hose types: – Flexible spiral hoses made of PVC with inte- Requirements grated spiral or equivalent hoses The unit can be installed either on the same level –...
  • Page 13: System Components

    Assembly System components Current consumption of Cross-section unit [A] The system components listed below are required or recommended for various procedures > 25 and < 32 or for installation. > 32 and < 40 Condensate separator > 40 and < 50 Two condensate separators are installed on the >...
  • Page 14: Installation

    Waste water outlet building waste water system. A suitable waste Connections V 2400 water collecting container can also be used if no building waste water drainage system is available. This then needs to be emptied on a regular basis.
  • Page 15: Electrical Connections

    Assembly Electrical connections 10 Testing suction power after installation NOTICE The following procedure is recommended in Short circuit due to defective lead order to measure the maximum negative pres- Do not route wires near hot surfaces. ❯ sure of the suction system so that it can be com- pared with the performance curve: Before connecting, check that the power sup- ❯...
  • Page 16: 11 Commissioning

    Assembly 11 Commissioning NOTICE Device damage due to dirt particles Dirt particles in the hoses can make their way into the turbine of the unit if filter screen is removed prematurely. After initial installation, lift off the large ❯ suction hose for approx. 3 minutes at all workplaces one after the other.
  • Page 17: Usage

    Usage The following are required for disinfection/clean- Usage ing: ü Non-foaming disinfectant/cleaning agent that is compatible with the materials. 12 Disinfection and cleaning ü Unit care system, e.g. OroCup To pre-clean, suck up 2 litres of water with the ❯ NOTICE care system.
  • Page 18: 13 Maintenance

    Usage 13 Maintenance All maintenance work must be performed by a qualified expert or by one of our Service Techni- cians. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯ Wear protective equipment when working (e. g. impermeable gloves, protective goggles and ❯...
  • Page 19: Troubleshooting

    Troubleshooting Troubleshooting 14 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
  • Page 20 Troubleshooting Error Possible cause Remedy Suction performance too low Coarse filters in system clogged Clean coarse filters. ❯ (e.g. at separator devices) Leak in the suction pipe Check and if necessary ❯ establish leak-tightness of suction pipe and connections. Mechanical sluggishness of tur- Disassemble the unit and ❯...
  • Page 21: 15 Transporting The Unit

    Troubleshooting 15 Transporting the unit WARNING Infection due to contaminated unit Disinfect the unit before transport. ❯ Close all media connections. ❯ Wear protective equipment to avoid any risk of infection (e.g. liquid-tight protective gloves, protective goggles, face mask). Prior to disassembly, clean and disinfect the ❯...
  • Page 22: Appendix

    Appendix Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
  • Page 23 Appendix 9000-606-97/02 2305V002...