Instruction Manual Amplifier Units For RA1000 Series NEC San-ei Instruments, Ltd. Version 2 March. 2000...
Page 2
This instruction manual is to provide information that is necessary for you to safely and correctly operate the amplifier units, Omniace II RA1000 series. Please always place this instruction manual together with the amplifier units whenever you use the units, so that you can access and refer to the manual at any time.
Page 3
Reprinting or reproduction of this manual, in whole or in part, without permission is prohibited. ● NEC San-ei has made every effort to attain the completeness in contents of this manual. Please feel free to contact your dealer regarding any errors, omissions, questions or suggestions, if you find one.
Page 4
WARNING WARNING indication labels of amplifier units ●Input signal connection and permissible common mode input voltages First, confirm that the mainframe is correctly and securely grounded through the protective grounding terminal, before connecting the equipment with a device to be measured. Also, be careful so that the input voltage does not exceed the permissible range of common mode input voltage, when you connect an amplifier unit with the measurement equipment.
Page 5
WARNING ●Warning against electric shock and permissible input voltage Do not touch, by any means, metallic portions of the input section, when a high voltage input signal is being applied, to avoid a risk of electric shock. Also, it is very dangerous to apply an input voltage exceeding the range of permissible input voltages for individual amplifier units, since application of such high voltages would cause equipment failures.
Page 6
Warranty Application NEC San-ei is making every effort in maintaining a high quality control level for its products from the design to shipping phases. However, in an unlikely event of finding a symptom of failures, you should check the operational status of the equipment, the status of the electric source voltage and the connection status of various cables, before asking NEC San-ei for repair.
Page 7
Designations used in this instruction manual The following explains the meaning of designations and symbols used in this instruction manual: designations or meaning symbols If any instructions in WARNING are ignored, the ignorance could lead to one or more of the following: WANING 1.
Before Using Amplifier Units ······································································································· 1 Safety Considerations and Precautions - Warning and Caution ································································· 2 Warranty Application ··········································································· 5 Warranty Provisions ·················································································································· 5 Designations used in this instruction manual ······················································································ 6 Contents Section 1. Instructions on the use of amplifier units ........ 1-1 1.1.
Page 9
1.9. 2CH vibration & RMS amplifier units ........... 1-44 ................1.9.1. Overview 1-44 ............1.9.2. Connection with input signals 1-44 ........... 1.9.3. Setting for 2CH vibration & RMS amplifier units 1-46 ......1.9.4. Specifications of 2CH vibration & RMS amplifier units (AP11-109) 1-49 ........
Section 1. Instructions on the use of amplifier units(2CH high-resolution DC amplifier units ) 1.1 2CH high-resolution DC amplifier units 1.1.1. Overview 2CH high-resolution DC amplifier units convert input signal voltages into digital data of 16 bits with high resolution. The units have a capability of conversion time of 10 s and incorporate two(2) channels per unit, and the two channels are insulated to each other within the unit.
Section 1. Instructions on the use of amplifier units(2CH high-resolution DC amplifier units ) ・Use cables with the insulation sheath of no less than 2 kV of withstand voltages. NOTE ・Do not apply voltages exceeding the permissible common mode input voltage, since application of such voltages would lead to malfunctions or failures of equipment.
Section 1. Instructions on the use of amplifier units(2CH high-resolution DC amplifier units ) 1.1.4. Specifications of 2CH high-resolution DC amplifier units(AP11-101) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel is insulated to each other and also from cabinet.) Input coupling modes AC coupling and DC coupling 0.1, 0.2, 0.5, 1, 2 and 5 V-FS (Voltages exceeding 30V shall not be...
Section 1. Instructions on the use of amplifier units(2CH high-resolution DC amplifier units ) 1.1.5 External drawings of 2CH high-resolution DC amplifier units RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) 1.2. 2CH FFT amplifier units 1.2.1 Overview 2CH FFT amplifier units are used to FFT-transform output voltages of piezoelectric acceleration sensors built-in the amplifier or other piezoelectric acceleration sensors (used together with charge converters), or other input voltages.
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) When setting the sampling speed other than 10μs step (ex. 5μs or 11μs, etc.) or setting the NOTE analyzing speed of FFT mode faster than 40kHz, the signal waveform can not be obtained correctly. If you execute FFT in that condition, the suspected frequency component is displayed.
Page 17
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) Measurement in the Vibration sensor mode Set the oscillation sensor in this box. RA1000 AMP (95691-2076-0000)
Page 18
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) Unit setting in the Vibration sensor mode You can select and switch the unit of vibration between the following: ・m/s ・G Using remote charge converters Set the type of sensors to be connected, by selecting the one from the following: ・...
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) 1.2.4. Specifications of 2CH FFT amplifier units(AP11-102) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel is insulated to each other and also from cabinet.) AC coupling and DC coupling ※...
Section 1. Instructions on the use of amplifier units(2CH FFT amplifier units) 1.2.5. External drawings of 2CH FFT amplifier units RA1000 AMP (95691-2076-0000) 1-11...
Section 1. Instructions on the use of amplifier units(2CH high-speed DC amplifier units) 1.3. 2CH high-speed DC amplifier units 1.3.1. Overview 2CH high-speed DC amplifier units convert input analog voltages into digital signals at high speed with a sampling interval of 1s (i.e., A/D conversion) The units incorporate two(2) channels per unit and the two channels are insulated to each other within the unit.
Section 1. Instructions on the use of amplifier units(2CH high-speed DC amplifier units) ・Use cables with the insulator of no less than 2 kV of withstand voltages. NOTE ・Do not apply voltages exceeding the permissible common mode input voltage, since application of such voltages would lead to malfunctions or failures of equipment.
Section 1. Instructions on the use of amplifier units(2CH high-speed DC amplifier units) 1.3.4 Specifications of 2CH high-speed DC amplifier units(AP11-103) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel is insulated to each other and also from cabinet.) Input coupling modes AC coupling and DC coupling For voltage measurement mode:...
Section 1. Instructions on the use of amplifier units(2CH high-speed DC amplifier units) 1.3.5 External drawings of 2CH high-speed DC amplifier units 1-15 RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(2CH AC strain amplifier units) 1.4. 2CH AC strain amplifier units 1.4.1. Overview The 2CH AC strain amplifier unit is an A/D conversion unit that converts output voltages of converters of the strain gauge type or varied voltages from strain gauges connected to the input.
Section 1. Instructions on the use of amplifier units(2CH AC strain amplifier units) 1.4.3. Setting for 2CH AC strain amplifier units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment after selecting a channel you want.
Section 1. Instructions on the use of amplifier units(2CH AC strain amplifier units) 1.4.4. Synchronization between AC bridge electrical source units (RA11-109) When you use multiples of this product in parallel, you should synchronize the operation of their AC bridge electrical source units. Define one(1) of the electrical source units as a master unit for the whole system.
Section 1. Instructions on the use of amplifier units(2CH AC strain amplifier units) 1.4.5. Specifications of 2CH AC strain amplifier units(AP11-104) Number of channels 2 channels(CHs)/unit Input mode balanced input (Each channel is insulated to each other and also from cabinet in the unit.) Applied strain gauge 120...
Section 1. Instructions on the use of amplifier units(2CH AC strain amplifier units) 1.4.6. External drawings of 2CH AC strain amplifier units 1-20 RA1000 AMP (95691-2076-0000)
1.5. Event amplifier units 1.5.1. Overview Event amplifier units provide status judgment information, such as that of voltage levels (H-level and L-level) and of electric contacts (open and short). One unit can have connections of up to eight(8) inputs and all inputs have a common ground. 1.5.2.
Section 1. Instructions on the use of amplifier units(Event amplifier units) (continued from the previous page) channel numbers within logic IC cord IC clip cord test clip cord event amplifier unit brown, black brown, black brown, black red, black red, black umber, black umber umber, black...
Section 1. Instructions on the use of amplifier units(Event amplifier units) 1.5.4. Specifications of Event amplifier units(AP11-105) Number of channels 8 channels(CHs)/unit logic input (Each channel is insulated to each other; the ground of each channel is connected Input mode commonly.) You can set Volt/Contact separately for each channel input.
Section 1. Instructions on the use of amplifier units(Event amplifier units) <Logic IC probes-----accessory to event amplifier unit> RT36-163 Logic IC probes are used for connection of the event amplifier unit with electronic circuits or sequence circuits for measurement of digital signals and/or relay contact signals.
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) 1.6. 2CH TC-DC amplifier units 1.6.1. Overview 2CH TC-DC amplifier units are used to make temperature measurement by directly connecting thermocouples (R, T, J, K or W) to their input terminals.
Page 35
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) Use of TC-DC amplifier units as DC amplifiers: ・If you apply, by error, any voltages that are more than the permissible input voltage (50V in DC or in AC peak value), equipment failures would be induced due to breakdown of internal parts or other reasons.
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) 1.6.3. Setting for 2CH TC-DC amplifier units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment after selecting a channel you want. Proceed with the setting operation by following displayed instructions.
Page 37
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) When set at temperature measurement in Fahrenheit: Set the equipment after selecting a channel you want. Proceed with the setting operation by following displayed instructions. Refer to Section 2 for detailed instructions of setting procedures that are not described herein.
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) 1.6.4. Specifications of 2CH TC-DC amplifier units(AP11-106) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel in the unit is insulated to each other and also from cabinet.) Input coupling modes DC coupling Applicable thermocouples...
Section 1. Instructions on the use of amplifier units(2CH TC-DC amplifier units) 1.6.5. External drawings of 2CH TC-DC amplifier units 1-31 RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(TC-DC amplifier units) 1.7. TC-DC amplifier units 1.7.1. Overview TC-DC amplifier units are used to make temperature measurement by directly connecting thermocouples (R, T, J, K or W) to their input terminals. The units can also be used as high-sensitivity DC amplifiers. 1.7.2.
Page 41
Section 1. Instructions on the use of amplifier units(TC-DC amplifier units) Use of TC-DC amplifier units as DC amplifiers: ・If you apply, by error, any voltages that are more than the permissible input voltage (50V in DC or in AC peak value), equipment failures would be induced due to breakdown of internal parts or other reasons.
Section 1. Instructions on the use of amplifier units(TC-DC amplifier units) 1.7.3. Setting for TC-DC amplifier units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment by selecting a channel you want. Proceed with the setting operation by following displayed instructions.
Page 43
Section 1. Instructions on the use of amplifier units(TC-DC amplifier units) When set at temperature measurement in Fahrenheit: Set the equipment after selecting a channel you want. Proceed with the setting operation by following displayed instructions. Refer to Section 2 for detailed instructions of setting procedures that are not described herein.
Section 1. Instructions on the use of amplifier units(TC-DC amplifier units) 1.7.4. Specifications of TC-DC amplifier units(AP11-107) Number of channels 1 channel(CH)/unit Input mode unbalanced input (The channel in the unit is insulated from cabinet.) Input coupling mode DC coupling Applicable thermocouples R, T, J and K Use of units as thermocouple amplifiers:...
Section 1. Instructions on the use of amplifier units(F/V converter units) 1.8. F/V converter units 1.8.1. Overview F/V converters are used to convert input signal frequencies into analog voltages. ●Application of voltages of more than 100V (in DC or in AC peak value) to the input of this type of units will CAUTION lead to equipment failure.
Page 47
Section 1. Instructions on the use of amplifier units(F/V converter units) ●Permissible common mode input voltage (CMV) Use insulated BNC cables for input connection by all means, which may be attached as an optional item. Be careful to maintain the common mode input voltage (CMV) at no more than 300V (in DC or in AC peak value).
Section 1. Instructions on the use of amplifier units(F/V converter units) 1.8.3. Setting for F/V converter units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment by selecting a channel you want. Proceed with the setting operation by following displayed instructions.
Page 49
Section 1. Instructions on the use of amplifier units(F/V converter units) TIPS ※Ripple ratio and response time ripples 90%FS output signal response time input signal Ripple ratio: Ripples are defined as a type of waveform like wavelets that are contained in the output signal.
Section 1. Instructions on the use of amplifier units(F/V converter units) 1.8.4. Specifications of F/V converter units(AP11-108) Number of channels 1 channel(CH)/unit Input mode unbalanced input (The channel is insulated from cabinet.) Input coupling modes AC coupling and DC coupling Input frequency range 1 Hz - 10 kHz Triggering level...
Section 1. Instructions on the use of amplifier units(F/V converter units) 1.8.5. External drawings of F/V converter units 1-43 RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) 1.9. 2CH vibration & RMS amplifier units 1.9.1. Overview 2CH vibration & RMS amplifier units have dual capabilities of A/D-converting output voltages of piezoelectric acceleration sensors built-in the amplifier and of A/D-converting the root-mean-square value of input voltages.
Page 53
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) ・Input impedance The input impedance is approximately one(1) M. However, note that the input impedance will be lowered to approximately 15 k, when the input voltage becomes 8 V or more for the input sensitivity ranges of 0.1 - 5 V-FS(full-scale) in the DC coupling mode.
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) 1.9.3. Setting for 2CH vibration & RMS amplifier units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment by selecting a channel you want.
Page 55
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) Switching/changing of input mode: Display the Extend screen by clicking this position. Vibration, RMS output mode: Voltage measurement mode: Vibration sensor mode: 1-47 RA1000 AMP (95691-2076-0000)
Page 56
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) Unit settings in the vibration sensor mode and in the vibration sensor RMS output mode: You can set the unit of vibration at one of the following in this box: ・m/s ・G Use of the remote charge converter:...
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) 1.9.4. Specifications of 2CH vibration & RMS amplifier units(AP11-109) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel is insulated to each other and also from cabinet within the unit.) Input coupling modes AC coupling and DC coupling For voltage measurement mode and for RMS converter mode:...
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) 1.9.5. External drawings of 2CH vibration & RMS amplifier units 1-50 RA1000 AMP (95691-2076-0000)
Page 59
Section 1. Instructions on the use of amplifier units(2CH vibration & RMS amplifier units) 1-51 RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(2CH DC strain amplifier units) 1.10. 2CH DC strain amplifier units 1.10.1. Overview The 2CH DC strain amplifier unit has dual capabilities of A/D-converting voltage variations obtained from strain gauge-type converters or from strain gauges connected to the input and of converting fine voltages into 16 bits of data with high resolution. The units incorporate two(2) channels per unit and the two channels are insulated to each other within the unit.
Section 1. Instructions on the use of amplifier units(2CH DC strain amplifier units) 1.10.3. Setting for 2CH DC strain amplifier units The Detail display/screen is displayed when any one of the following indications, “ALL CH”, ”Sig.Name” or ”List”, is shown. Set the equipment by selecting a channel you want.
Page 62
Section 1. Instructions on the use of amplifier units(2CH DC strain amplifier units) For the setting of DC amplifier mode: Set the equipment by selecting a channel you want. Proceed with the setting operation by following displayed instructions. Refer to Section 2 for detailed instructions of setting procedures that are not shown in this page.
Section 1. Instructions on the use of amplifier units(2CH DC strain amplifier units) 1.10.4. Specifications of 2CH DC strain amplifier units(AP11-110) Number of channels 2 channels(CHs)/unit Input mode unbalanced input (Each channel is insulated to each other and also from cabinet.) Input coupling mode DC coupling Applicable strain gauge...
Section 1. Instructions on the use of amplifier units(2CH DC strain amplifier units) 1.10.5. External drawings of 2CH DC strain amplifier units 1-56 RA1000 AMP (95691-2076-0000)
Section 1. Instructions on the use of amplifier units(2CH Zero Suppression amplifier units) 1.11. 2CH Zero Suppression amplifier units 1.11.1. Overview The 2CH Zero Suppression Amp is an amplifier that can cancel a DC voltage element added to the input signal and amplify only fluctuating portion of the input signal.
Section 1. Instructions on the use of amplifier units(2CH Zero Suppression amplifier units) • Use a cable whose withstand voltage is at least 2 kV. NOTE • Avoid inputting voltage over the allowable common voltage since it may cause erroneous operation and malfunction.
Page 67
Section 1. Instructions on the use of amplifier units(2CH Zero Suppression amplifier units) Zero suppression voltage settings 1. Sets whether zero suppression voltage is added or not. When you use the zero suppression voltage, the switch changes as [OFF] – [ON] - [OFF] with each key pressing.
Section 1. Instructions on the use of amplifier units(2CH Zero Suppression amplifier units) 1.11.4. Specifications of 2CH Zero suppression amplifier units (AP11-111) Number of channel 2CH/unit Input type Unbalanced input (Isolation: Between channels in a unit and between a channel and case) Input coupling AC coupling and DC coupling 0.1, 0.2, 0.5, 1, 2, 5V FS (at 0.1 to 5V FS and AC coupling, ±30 V or lower)
Section 2. Common settings on amplifier Detail screens 2.1. Settings of ON/OFF/GND, All and initialization of amplifier units Produce the following settings and indications at the amp detailed screen. ON/OFF/GND: Select and set the on/off status of display and re- cording for each channel.
Page 72
Section 2. Common settings on amplifier Detail screens ※The following shows display screens when the key [List] is pressed: RA1000 AMP (95691-2076-0000)
Section 2. Common settings on amplifier Detail screens 2.2. Initial status of each type of amplifier units 1. Initial status/conditions of STD. screens: HRDC FFT HSDC ACST EV RA1000 AMP (95691-2076-0000)
Page 74
Section 2. Common settings on amplifier Detail screens TCDC TDC FV RMS DCST HRZS RA1000 AMP (95691-2076-0000)
Page 75
Section 2. Common settings on amplifier Detail screens 2. Amplifier Detail - Initial status/conditions of User Conv. screen HSDC (example) 3. Amplifier Detail - Initial status/conditions of Extend screen (example) HSDC EV( RA1100,RA1200) EV( RA1300) RA1000 AMP (95691-2076-0000)
Page 76
Section 3. Common settings for individual amplifier units (amplifier STD. screen)
Section 3. Common settings for individual amplifier units (amplifier STD. screen) 1. Input setting - amplifier screen Press the key of either [AMP] or [Tray] to display the input setting - amplifier screen. ① Press the key to set the input status of ON/OFF(/GND).
Page 78
Section 3. Common settings for individual amplifier units (amplifier STD. screen) 2. Zero (0) position The zero (0) position is defined as the position of the waveform display and recording when an input equivalent to “GND” is provided at the input terminal. Press one of the buttons to change the position in steps of 5% or 10% and/or use the jogdial to change it in steps of 1%, where the full scale...
Page 79
Section 3. Common settings for individual amplifier units (amplifier STD. screen) 4. Range setting and high sensitivity setting Range: Range setting can also be done by using the jogdial. Changing the input range to any one of 5V - 100mV ranges while the input voltage is 100V or more would lead to amplifier failures and/or providing adverse effects on the signal source.
Page 80
Section 3. Common settings for individual amplifier units (amplifier STD. screen) 6. Filter setting Filter: You can set the filter for individual amplifiers. The expression of “Filter” indicates low pass filters. Note that other types of filters than low pass filters are provided as follows: High pass filters for 2CH vibration &...
Page 81
Section 3. Common settings for individual amplifier units (amplifier STD. screen) RA1000 AMP (95691-2076-0000)
Page 83
Section 4. Extend setting 1. Entry of channel names (signal names) Enter up to four(4) arbitrary characters for each channel to set the signal name. You can see and confirm the signal name in the “ALL CH” screen. In RA1200 RA1300, signal names are printed in the waveform records as channel annotation during waveform recording.
Page 84
Section 4. Extend setting 2 Setting of waveform display colors 3 Setting of Baseline width of recording waveform Waveform color: You can select waveform display colors in the screen for individual channels from eight (8) color selections. Baseline width of recording waveform: When recording signal waveforms by RA1200, you can select and set waveform line-widths for individual channels.
Page 85
Section 4. Extend setting RA1300 Recording position, pitch of signals, amplitude, recording line thickness and display color of signals can be set. Recording position sets distance from lower edge to low waveform of No.8 signal on the screen (waveform recording). Display color of waveform changes every pressing key of each channel.
Page 86
Section 4. Extend setting 5. Switching/change of input mode Set the input mode in the Extend screen for 2CH FFT amplifiers and 2CH vibration & RMS amplifiers. For 2CH FFT amplifiers: For 2CH vibration & RMS amplifiers: RA1000 AMP (95691-2076-0000)
5. User Conv. ~ ~ Conversion of waveform amplitudes and units...
Page 89
Section 5. User Conv. ~ Conversion of waveform amplitudes and units ~ ◆ The capabilities of physical value conversion involves those of converting the raw unit of input signals into physical values or into arbitrary characters as well as the full scale, i.e., the amplitude of waveform recordings, into different ones.
Section 5. User Conv. ~ Conversion of waveform amplitudes and units ~ 5.1 How to do User Conv. Change the magnitude of the unit so that the desired portion of the input signal can be seen exactly within the full scale.
Section 5. User Conv. ~ Conversion of waveform amplitudes and units ~ 5.2 Recording/Display range Further, you can scale (enlarge) a desired portion of the output signal range that has been set with the maximum/minimum values. Three modes are available for setting the record scaling as described below. ①...
Page 92
Section 5. User Conv. ~ Conversion of waveform amplitudes and units ~ 5.2.1 How to set the Recording/Display range with maximum/minimum values Specify the desired maximum/minimum values for record scaling as shown in the display figures below. 5.2.2 How to set the Recording/Display range with sensitivity/division (Sens/div) Specify the desired sensitivity per division as well as the desired minimum value as shown in the display figures below.
Page 93
Section 5. User Conv. ~ Conversion of waveform amplitudes and units ~ RA1000 AMP (95691-2076-0000)
Section 6. Procedures for changing amplifier units...
Page 95
Section 6. Procedures for changing amplifier units Amplifier units can be changed easily, since they have plug-in mounting structure. However, mount or dismount amplifier units, only after you have turned off the power supply switch and disconnected the power supply cable from the mainframe Mounting or dismounting amplifier units while the mainframe is powered on would lead to damages to the mainframe, Omniace II.
Page 96
(1) No part, nor the whole, of this Instruction Manual may be used, copied or reproduced in any form or by any means without prior permission from NEC San-ei Corporation. (2) The content of this Instruction Manual is subject to change without prior notice.