INSTRUCTIONS FOR USE
OXIPEN
®
PULSE OXIMETER
INSTRUCTIONS FOR USE
These instructions have been prepared
with great care – should you find details,
however, which are not concurrent with
the handling of this system, please
advise us promptly so that we may
eliminate any inconsistencies as
quickly as possible.
We reserve the right to make
changes stemming from further
development in relation to the information stated in these
instructions.
All trademarks cited and illustrated in the text are trademarks of
the current owners and are recognized as being protected.
Reproduction, translation and duplication in any form – even
abridged – require the written approval of the manufacturer.
This handbook is subject to the revision service of Honeywell
Healthcare Solutions GmbH.
LIST OF ILLUSTRATIONS
FIG. 1: FRONT, BACK, AND FRONT END VIEWS ...................... 4
FIG. 2: OXIPEN® DISPLAY ................................................... 4
FIG. 3: INSTRUMENT ERROR DISPLAY ................................. 5
FIG. 4: OPENING THE BATTERY COMPARTMENT LID ............ 6
FIG. 5: CONNECTING THE SENSOR ..................................... 6
FIG. 6: DISPLAY SIGNAL ACTIVATED/DEACTIVATED ............ 7
1
GENERAL INFORMATION ......................................................... 2
2
SAFETY INFORMATION ............................................................. 2
3
INTRODUCTION ........................................................................... 3
3.1 INTENDED USE .................................................... 3
3.2 GENERAL FUNCTIONAL PRINCIPLES
AND CONDITIONS ............................................... 3
3.3 ACCURACY, MALFUNCTIONS................................. 3
4
OPERATOR CONTROLS AND SYMBOLS ............................. 4
4.1 THE INSTRUMENT ............................................... 4
4.2 THE DISPLAY ....................................................... 4
4.3 OXIPEN® SYMBOLS .............................................. 4
5
INFORMATION MESSAGING AND SIGNALS ...................... 5
5.1 DISPLAY .............................................................. 5
5.2 SIGNAL QUALITY ................................................. 5
5.3 VARIABLE PULSE TONE ....................................... 5
5.4 SIGNALS ............................................................. 5
5.5 ERROR MESSAGES .............................................. 5
6
COMMISSIONING ....................................................................... 6
6.1 UNPACKING & CHECKING THE CONSIGNMENT ....... 6
6.2 TESTS ................................................................. 6
6.3 PARTS LIST.......................................................... 6
6.4 INSERTING THE BATTERIES .................................. 6
6.5 CONNECTING THE SENSORS ................................ 6
7
OPERATION ................................................................................... 6
7.1 SWITCHING THE INSTRUMENT ON AND OFF ......... 6
7.2 BACK LIGHTING ................................................... 6
7.3 SIGNAL MUTE SWITCHING ................................... 7
7.4 POWER SAVING SWITCH-OFF ............................... 7
8
ENVITEC BY HONEYWELL SPO
SENSORS & ACCESSORIES ...................................................... 7
8.1 SELECTION AND APPLICATION OF A SENSOR. ........ 7
8.2 OVERVIEW OF ENVITEC SENSORS AND CABLES ..... 7
9
FAILURE CORRECTION AND MAINTENANCE ................... 8
9.1 CLEANING, DISINFECTION OF THE INSTRUMENT
SURFACES .......................................................... 8
9.2 CLEANING, DISINFECTION OF THE SENSORS ......... 8
9.3 MAINTENANCE AND SERVICE ............................... 8
9.4 CALIBRATION ...................................................... 8
9.5 RETURNING THE OXIPEN®. .................................. 8
9.6 DISPOSAL OF THE PRODUCT ............................... 8
10 GUARANTEE .................................................................................. 9
11 TECHNICAL DATA ........................................................................ 9
11.1 SPECIFICATIONS ................................................. 9
Issue 3
058-07-21000085
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ENVITEC OXIPEN and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Honeywell ENVITEC OXIPEN

  • Page 1 SENSORS & ACCESSORIES ............7 abridged – require the written approval of the manufacturer. This handbook is subject to the revision service of Honeywell 8.1 SELECTION AND APPLICATION OF A SENSOR..7 8.2 OVERVIEW OF ENVITEC SENSORS AND CABLES ..7 Healthcare Solutions GmbH.
  • Page 2: Safety Information

    Attention notes alert the user of necessary care for WARNING the safety and effectiveness of the OxiPen®. Only the sensors and accessory parts approved by Honeywell Healthcare Solutions GmbH may be used with OxiPen®. 2.0 SAFETY INFORMATION Sensors and accessories must be in flawless condition. The...
  • Page 3: Pulse Oximeter

    Solutions GmbH or by one of the sites authorized by EnviteC oxygenation. The absorption of dysfunctional derivatives such by Honeywell for these purposes and the instrument is as methaemoglobin or carboxyhaemoglobin is similar to that employed in accordance with the instructions for use.
  • Page 4 Date of manufacture Battery compartment, spring-loaded catch Battery compartment lid Manufacturer Instrument label Sensor socket Product number Serial number Consider disposal regulations Conformité Européenne (European Conformity) 0123 IPX2 Protection against dripping water Medical device OxiPen® Instructions for Use | sps.honeywell.com/ast...
  • Page 5 In the case of instrument failure please advise your specialist actuating the mute key and may be switched on again by dealer or Honeywell Healthcare Solutions GmbH. pressing the same key again (see 0). If the sensors have not been applied in accordance with the If the pulse tone is deactivated it will be reactivated specification OxiPen®...
  • Page 6: Operation

    Figure 5. Connecting the sensor carrier. Unpack OxiPen® and the accessories. If a part is missing or is damaged, please apply to the Honeywell Healthcare Solutions GmbH service centre or to your local Honeywell Healthcare Solutions GmbH dealer. 6 .2 Tests Prior to employing the OxiPen®...
  • Page 7 The instrument switches itself off automatically if it is not used during the first 3 minutes after being switched on (no 8 .2 Overview of EnviteC by Honeywell Sensors measurements) or if no measurement signal is detected during TABLE 8. SENSORS AND CABLES 3 minutes following the last measurement.
  • Page 8 Should problems be observed when working with the OxiPen®, which you are unable to eliminate yourself, please contact your 9 .4 Calibration specialist dealer or the service centre of Honeywell Healthcare Solutions GmbH. The instrument requires neither calibration nor maintenance.
  • Page 9: Technical Data

    Display Range accordance with our guarantee conditions. SP02 45 % to 100 % Honeywell Healthcare Solutions GmbH provides no guarantee, Pulse frequency 0 BPM to 300 BPM if the operator endangers the function of the instrument through non-observance of these instructions for use, improper...
  • Page 10 Honeywell Healthcare Solutions GmbH Alter Holzhafen 18, 23966 Wismar, Germany Phone: +49 (0)3841-360-1 Email: info@envitec.com Honeywell Sensing & Safety Technologies 830 East Arapaho Road 058-07-21000085-EN | 3 | 01/24 Richardson, TX 75081 © 2024 Honeywell International Inc. All rights reserved. www.honeywell.com...