RadPro International GmbH Table of contents Introduction ........................5 TLDcube ......................... 5 Target Audience ...................... 5 Intended Application ....................5 Not Intended Application ..................5 Technical Specification .................... 6 Safety Instructions ....................7 Basic Information and Advices ..................8 Preface ........................8 Security warning ......................
Page 3
RadPro International GmbH Filters ........................13 Reference light .......................15 TLD Materials ........................15 Installation ........................17 Prepare for Operation .....................17 Software Basics......................18 Program Start up ....................18 6.1.1 Overview ......................18 Initial setup ......................19 6.2.1 Device connection setup .................19 6.2.2 Create a heating profile ...................19 Login/Logout ......................20 6.3.1 Login .......................20...
Page 4
RadPro International GmbH 7.4.4 Measuring the actual luminescence signal ............27 7.4.5 TLStudio Data Flow ..................28 Create and edit heating profiles ................29 Create and edit Sets ....................29 Data evaluation ......................30 7.7.1 Measurement list .....................30 7.7.2 Measurement view ..................32 7.7.3 Measurement data ..................34 Messaging system ......................36 Settings .........................37 9.1.1...
TLDcube is a compact and cost efficient thermoluminescence (TLD) reader to measure, analyze and evaluate thermoluminescence materials using thermal stimulation (TL). It is developed in collaboration with Freiberg Instruments GmbH, Germany. Innovative design and software enhances easy operation and reliability of the newly developed TLD reader.
RadPro International GmbH 1.5 Technical Specification Heater Ceramic contact heater with nitrogen cooling function Base Plate Stainless steel, exchangeable Base Plate types Chip 3.2 mm x 3.2 mm, Rod 1 mm x 6 mm, Disc Ø 4.5 mm, Multi-Purpose Ø 10 mm Powder Dish Ø...
RadPro International GmbH 1.6 Safety Instructions In general handle the device with care as mentioned in this user manual. Improper usage may cause personal injury or damage to property. The TLD reader has a ceramic heating element on which the samples are placed for TL measurements. This heating can reach temperatures up to 700°C.
RadPro International GmbH 2 Basic Information and Advices 2.1 Preface It’s necessary to consider the warning notices. The defiance could cause damages to health and property. It is highly recommended that repair or service of TLDcube is carried out by a trained professional.
RadPro International GmbH 2.6 Power Supply 2.7 Mechanics Inappropriate operation can cause property damage and injuries. External tensions and vibrations should not be applied to the equipment! For a better cooling of the equipment don’t place any other hardware within 20 cm. 2.8 Hazardous Material An assignment of hazardous materials is not intended.
RadPro International GmbH 3 General Information 3.1 Basic Construction Dimensions: 276 x 148 x 164 mm Weight: Approximately 5.5 kg 3.2 Electronics Power supply Detector Motion control Microprocessor Temperature control Valve control Ethernet Powered by...
RadPro International GmbH 3.3 Sample Handling and Loading Step 1: Check if the PC or Laptop is connected to the device Step 2: Check if the Nitrogen supply is turned ON, if applicable Step 3: Turn ON the power supply Step 4: Push the Eject button in front of the device.
RadPro International GmbH 3.5 Heating Unit Thermoluminescence (TL), preheat (PH) of materials or elevated temperature measurements need a heating element. Measurements are possible during any part of a heating/cooling cycle, which may simply be a heating ramp but series of several heating steps are possible, too.
RadPro International GmbH 3.6 Photosensor • Module H10722-110 • photosensor with integrated amplifier • for standard applications in luminescence dating and dosimetry • detection unit with super bi-alkaline cathode PMT (HAMAMATSU) • sensitivity 230-700 nm • peak sensitivity 400 nm •...
RadPro International GmbH Fig: BG25 Filter Curve 3.8 Reference light This is an inbuilt unit inside the Measurement chamber of a TLDcube. This checks the functionality of the PMT unit frequently and in case of any defects in or malfunctioning of the PMT unit an error message will be displayed on the TLStudio software.
Page 16
RadPro International GmbH Types and Forms of TLD Materials: TLD Powders for dosimetry in medical physics: • Beam dosimetry • TLD Quality Assurance for postal reference dosimetry • Typical grain size 80 – 200 µm (customer specified) TLD Chips, Rods, Discs and Micro Cubes in medical physics: Diagnostic Radiology: •...
RadPro International GmbH 5 Installation Equipment is packed for shipment. Detectors are disassembled. A detail description how to pack and unpack the equipment can be found in a separate report. 1. Position the TLDcube device on a horizontal table which is qualified for 15 Kg. 2.
RadPro International GmbH 6 Software Basics TLStudio is a powerful software that works with TLDcube (TLD-reader). The software contains the complete workflow beginning with the creation of heat ramps, measure count curves and up to high level dose evaluation. 6.1 Program Start up The program can be started by clicking the desktop icon.
RadPro International GmbH 6.2 Initial setup To initially setup the program the following basic settings are mandatory: 6.2.1 Device connection setup TLStudio communicates with the device over a TCP/IP network connection. The following settings must be taken in the section “Devices” on tab “Main Control”: 1.) Log in with engineer login (user: Engineer, password: engineer) 2.) Right click in “Devices”...
RadPro International GmbH 1.) Type in a profile name 2.) Create a heat ramp in the middle. 3.) Click “save” to store the recipe on hard disc. 6.3 Login/Logout Some TLStudio functions are only accessible via special login. Possible login roles are operator (this is the standard role) and engineer. The operator login is limited to select recipes/sets and start and analyze measurements.
RadPro International GmbH 7 Measurement Each measurement is automatically saved into a folder with the selected set name. 7.1 Preliminary software set up The following steps are required before taking measurements using the software, (Note: The following steps are not required if you want to just open or analyze older files.) 1.) Select device in “Devices”...
RadPro International GmbH 7.2 Detector temperature In the device details, the current Photosensor temperature is shown. The background of this field is colored depending on the difference to the setpoint. The reader is shipped with a setpoint of 25 °C, but this point is adjustable. It should not be changed after calibration. Green means the detector temperature is in a good range.
RadPro International GmbH The measurement results (data curves) will be displayed in the analysis section of the software. 7.3.2 Measure sets with predefined elements In this mode a predefined set with elements will be measured. Therefore the Set must be created with the built in Set editor in the controlling area.
RadPro International GmbH The measurement results (data curves) will be displayed in the analysis section of the software. 7.3.3 Measure Elements based on existing Set-Calibration With this mode, it is possible to use an existing Set and calibration to measure and evaluate Elements outside this Set.
RadPro International GmbH The currently performed step is announced by the “Run state” element in the “main control” tab. In addition to this visualization, the state of the device is displayed by a lamp on the “open button” on the device itself. In idle mode the lamp will glow constant whereas a slow pulsing lamp is indicating a running measurement.
RadPro International GmbH The Gas state indicator shows if the pressure at the inlet is sufficient. It is possible that the pressure falls below the threshold during the chamber fill process, but that is not a Problem. 7.4.4 Measuring the actual luminescence signal In the actual measurement, the device will execute the assigned heating profile.
RadPro International GmbH 7.5 Create and edit heating profiles The heating profile editor tab is used to create and edit profiles. An existing profile can be opened by double clicking it in the profile list. 1.) [Create new:] Press “New”-button 2.) Create or edit heat ramp 3.) Set up regions of interest.
RadPro International GmbH 4.) Add Elements by typing a name into the “Name”-column. Press “Enter” to accept the new entry. 5.) Arrange Sets by pressing “Arrange elements” (first condition: Numbers ascending; second condition: text ascending) 6.) Select recipe to measure this Set with 7.) Click “Save”...
Page 31
RadPro International GmbH Open single measurement Open Set Save selected measurement/Set as another filename Remove selected measurement/Set from list (not from hard disc) Clear whole list (not from hard disc) Add a Folder to list (e.g. to separate measurements). Right clicking a Folder and select “Load measurement”...
RadPro International GmbH 2.) Select by check special curve checkboxes This will append the selected object to the “Measurement view” – each with another color. Example: compare heat curves 7.7.2 Measurement view The measurement view shows the selected measurements in the measurement list. Zooming can be performed by creating a mouse rectangle around the desired area.
Page 33
RadPro International GmbH Performs horizontal automatic zoom Performs vertical automatic zoom Resets scaling/zoom to default scaling Prints report to a selectable printer Prints report to XPS format Setup optional default scaling for different axis types (e.g. temperature) for better comparison between measurements Enables mouse measuring mode: 1.) Select start point per mouse click 2.) Select end point per mouse click...
RadPro International GmbH 7.7.3 Measurement data All measurement parameters are displayed in the “Measurement data” section. The regions of interest can be modified by editing the list in the lower area. With “Calculate ROIs” the counting will be performed for the new or changed regions. Powered by...
Page 35
RadPro International GmbH Example: Divide Region 1 into 2 and 3: Powered by...
RadPro International GmbH 9 Settings 9.1.1 Device settings From the context menu under the “Devices” section in “Main Control”, select “Settings…”. This menu is only available with engineer login. In the dialog box the device specific settings can be taken: 1.) “Simulation”...
RadPro International GmbH 9.1.2 Program settings The program-wide settings are available under the tab “Settings”. “Data backup path” is the standard backup path to store measurements after finishing. “Count warn limit” will warn, if the calculated counts are above the limit. “View”...
RadPro International GmbH 10 Software information For service requests or software version the tab “Info” will provide the necessary information. Powered by...