Revision Log Revisions Description Letter Date 06-25-2020 Release using Operations Manual Rev Q. Updated to Operations Manual Rev R. 03-11-2021...
Page 3
Be sure external loop is sealed & able to withstand 10 psig. Insert the male quick disconnects into the female quick disconnects. When quick disconnects are in place and the unit on, the external circulation pump will function and pump liquid at the OUTLET connection. INST-INSTALL-2050 Rev C 03/11/2021 Page 3...
Page 4
Operate the unit using only the power cord that has been supplied when purchased. Leakage current of this unit can exceed 3.5mA. Electrical power required is: 2050-1-1 115V, 60Hz, 10A 2050-1-4 230V, 50/60Hz, 5A...
Page 5
Overtime, this can cause the reservoir to overflow. To prevent severe burns, keep all extremities away from the fluid in the reservoir tank or any exterior surfaces around the tank. INST-INSTALL-2050 Rev C 03/11/2021 Page 5...
Page 6
Before draining the circulator, be sure the unit is at room temperature. To empty the circulator, remove the 1/8” MPT plug (unscrew by turning counter-clockwise) located on the lower left side of the circulator. Use distilled water, or deionized water with a resistivity between 50KΩ-CM and 1MΩ-CM. INST-INSTALL-2050 Rev C 03/11/2021 Page 6...