3-in-1 Ear, Forehead +
Touchless Thermometer
USER MANUAL
INFRARED TEMP CHECKS IN 1 SEC
This ultimate multi-use infrared thermometer is here to deliver accurate temp checks in one second.
The 3-in-1 Ear, Forehead + Touchless Thermometer has a color-coded screen and optional sound
alarm to alert you when fevers spike. Seamlessly move from in-ear temp checks when baby's
awake, to stealth forehead temp checks while baby's sleeping. In addition to tiny humans, this
thermometer works on adults, and even reads surface temperatures in touchless mode -- like the
bottle you're about to feed your feisty freeloader.
Manufactured for
Fridababy LLC, Miami, FL 33137
Warning:
Keep out of reach of children under 12 years.
Never immerse the thermometer into water or other liquids (not waterproof). For cleaning and disinfecting, please
follow the instructions in the "Care and Cleaning" section.
Never use the thermometer for purposes other than those it has been intended for. Please follow general safety
precautions when using on children.
Keep the thermometer away from direct exposure to the sun and keep it in a dust-free, dry, well-ventilated place
at a temperature between 50°F (10°C) - 104°F (40°C). Do not use the thermometer in high humidity environments.
(>95% RH)
Do not use the thermometer if there are signs of damage on the measuring sensor or on the instrument itself. If
damaged, do not attempt to repair the instrument! Please contact FridaBaby.
This thermometer consists of high-quality precision parts. Do not drop the instrument. Protect it from severe impact
and shock. Do not twist the instrument or the measuring sensor.
Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration,
changes in appetite or activity, seizure, muscle pain, shivering, sti neck, pain when urinating, etc., even in the
absence of fever.
Even in the absence of fever, those who exhibit a normal temperature may still need to receive medical attention.
People who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to
determine the severity of their illness.
Temperature elevation may signal a serious illness, especially in adults who are old, frail, have a weakened immune
system, or neonates and infants. Please seek professional advice immediately when there is a temperature elevation
and f you are taking the temperature of anyone who is:
• Over 60 years of age (Fever may be blunted or even absent in elderly patients)
• Has diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer,
chemotherapy, chronic steroid treatment, splenectomy)
• Bedridden (e.g., nursing home patient, stroke, chronic illness)
• A transplant patient (e.g., liver, heart, lung, kidney)
This thermometer is not intended for preterm babies or small-for-gestational age babies. This thermometer is not
intended to interpret hypothermic temperatures. Do not allow children to take their temperatures unattended.
Clean the thermometer probe after each use.
Do not use the thermometer on newborns or for continuous temperature monitoring purposes.
Do not take a measurement while or immediately after nursing a baby.
Patients should not drink, eat, or be physically active before/while taking the measurement.
Please Read and Save These Instructions
This product is not intended to diagnose or treat any health problem or disease and should not
be used as a substitute for the consultation and advice a physician or other medical professional.
Thermometers should only be used under the supervision of an adult.
This appliance conforms to the following standards:
ASTM E 1965-98 (Re-approved 2009)
Standard Specification for Infrared Thermometers
for Intermittent Determination of Patient Temperature
ISO80601-2-56 First Edition 2009-10-01 Medical Electrical Equipment -
Part 2-56: Particular Requirements For Basic Safety And Essential Performance
Of Clinical Thermometers For Body Temperature Measurement.
(General Plastic Surgery/General Hospital)
Contents
1 Thermometer, 2 x AAA batteries, 1 Storage Pouch, 1 Owner's Manual, 1 Quick Start Guide
Product Illustration
Ear Probe
LCD
Display
Screen
Power
/ Measure /
Mute / Unmute
Forehead Reader
(remove when
measuring
ear temperature)
Mode
/ Memory
Button
Battery
Cover
Specifications
Type: Digital Infrared Thermometer
Measure Range: Forehead + Ear: 32.0°C–42.9°C
(89.6°F–109.2°F) / Object: 0°C–100°C (32°F–199°F)
( °C / °F chosen by user)
Accuracy: Ear + Forehead mode ±0.2° C/±0.4 °F
Object mode ±1.0°C / 2.0°F (laboratory tested)
Operating mode: Direct Mode
Display: Segment LCD, Four color LED backlight
(white, green, orange, red)
Memory: Can store 40 values
Battery: AAA (1.5V DC) x 2
Battery life: Can be used up to 3,000 times
Dimension: 143 x 35 x 41mm
Weight: 80g (with battery)
FCC Statement
IMPORTANT INFORMATION REQUIRED BY THE FCC
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference; and (2) this device must accept any interference received, including interference
that may cause undesirable operation.
NOTE: This equipment has been tested and found to comply with the limits fora Class B digital device, pursuant to
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate radio frequency energy, and if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment o and on, the user is
encouraged to try to correct the interference by one or more of the following measures: 1) Reorient or relocate the
receiving antenna. 2) Increase the separation between the equipment and receiver. 3) Connect the equipment into an
outlet on a circuit di erent from that to which the receiver is connected. 4) Consult the dealer or an experienced
radio/TV technician for help.
CAUTION: Changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
Love it? Leave a review.
fridababy.com
Questions?
frida.com/help
Child Forehead Mode
(under 12yr old)
Ear Mode
Adult
Forehead
Mode
Current
Temperature
Value
Previous
Fahrenheit / Celsius
Temperature
Degrees
Value
Ambient operating range:
Temperature: 50°F - 104°F (10°C-40°C)
Humidity: 15-95%RH, non-condensing
Atmospheric pressure:70-106kPa
Storage and transportation condition:
Ambient Temperature: -20°C-55°C(-4°F-131°F)
Relative Humidity: 0-95% RH, non-condensing
Atmospheric pressure: 50kPa to 106kPa
Ingress Protection Rating: IP 22
Classification: Type BF
Model: AOJ-20A
Directions
FOREHEAD READER OFF
Thermometer automatically switches to Ear Mode.
1. Press the big Power/Measure button for 1 second to turn the thermometer on. The LCD screen will show
the last measurement taken.
2. Select your "mode" by following the directions for ear, forehead or touchless use in the tables below.
3. Press and release the Power/Measure button for 1 second again.
4. Once the beep is heard, you can read the value. "T" indicates a temperature reading.
5. The device automatically shuts o after being unused for 10 seconds.
FOR EAR USE
1) Remove the forehead reader, automatically selecting Ear mode
2) Fit the probe snuggly into the ear canal.
WARNING:
Make sure that the ear is clean, with no earwax or obstructions.
The right ear reading may di er from the reading taken at the left ear. Always take the temperature in the
same ear.
Do not force the thermometer into the ear canal or the ear canal may get injured.
Be careful when taking the temperature on a child, whose ear canal is small.
Do not insert past the point where you feel resistance.
NOTE: Children under 1 year: Pull the ear straight back. Children over 1 to adults: Pull the ear up and back. Make
sure the ear canal is straight, so that the temperature probe can receive an infrared ray from the eardrum.
FOR FOREHEAD USE
1) With the forehead reader on, press the Mode/ Memory button to choose "Adult forehead mode"
Child forehead mode
2) Touch the forehead reader to the center of the forehead.
WARNING:
If the forehead area is covered with hair, sweat, or dirt, please clean the area beforehand to improve the
reading accuracy.
Always check if the lens is clean.
Always make sure the user and the thermometer have been in the same room for at least 30 minutes
prior to the measurement.
NOTE: The forehead measurement is an indicative reading. The measured forehead temperature can fluctuate up to
1°F/0.5°C from your actual body temperature. Please be aware of the factors that influence the accuracy as
described in the "**Normal body temperature / Fever indicator" and "Warning" sections.
FOR TOUCHLESS USE
Note: This mode is ONLY for measuring the temperature of the room or an object. Not for body temp checks.
1) Press the Mode/ Memory button to choose the house icon
2) Hold the thermometer about 1-3 cm (0.39-1.18 inches) away from the object.
TEMPERATURE READING
Forehead or Ear:
Normal temperature: 89.6°F - 99.5°F (32°C - 37.5°C)
Indicator: Green light, one long beep
Low fever: 99.7°F -100.4°F (37.6°C - 38.0°C)
Indicator: Orange light, 3 short double-beeps.
High fever: 100.6°F - 109.2°F (38.1°C - 42.9°C)
Indicator: Red light, 3 short double-beeps.
It is recommended that 3-5 measurements in the same location be taken and the highest one taken as the reading
in the following situations:
Newborn infants in the first 100 days.
Children under three years of age with a compromised immune system and for whom the presence or absence
of fever is critical.
When the user is learning how to use the thermometer for the first time until they have familiarized themselves
with the instrument and obtained consistent readings.
Switch between Mute and Un-Mute: With the thermometer on, press the Measure/Mute-unmute button for 5 seconds.
F˚/ C˚ conversion: To change from Fahrenheit (F°) to Celsius (C°), Open the battery cover.
Use the toggle switch to change the °F/°C.
Memory mode: Press the Mode/ Memory button for 5 seconds to go to the memory mode. Press this
button again to check the 40 stored readings one by one. If no value has been saved, it will display "---".
NOTE: All stored readings are reset when the batteries are replaced.
Call your child's doctor right away if your child has a fever and:
Is younger than 3 months (12 weeks) and has a temperature of 100.4°F (38.0°C) or higher.
If, at any age, the fever rises above 104°F (40°C) repeatedly.
Battery Insertion and Replacement
First time use: 1) Pull battery door o and open battery chamber. 2) Insert two AAA batteries. 3) Slide battery
chamber back into place and reattach battery door.
Replacement: 1) Replace battery when appears in the lower right corner of LCD display. 2) Remove the old
batteries following the above instructions. Discard batteries lawfully. 3) Replace with two new AAA batteries.
Remove the batteries if the thermometer will not be used for more than two months.
Object
Care and Cleaning
Mode
Use an alcohol swab or cotton swab moistened with 70% alcohol to clean the thermometer casing and the
Mute /
measuring probe. After the alcohol has completely dried out, you can take a new measurement.
Un-mute
Icon
Ensure that no liquid enters the interior of the thermometer. Never use abrasive cleaning agents, thinners or
benzene for cleaning and never immerse the instrument in water or other cleaning liquids. Take care not to
Low Battery
scratch the surface of the LCD screen.
Indication
Troubleshooting / Error Codes
Performance may be degraded should one or more of the following occur:
Operation outside stated temperature and humidity range
Storage outside state temperature and humidity range
Mechanical shock
Patient temperature is below ambient temperature
Error
Problem/Cause
Er1
The ambient temperature is not in range.
Memory error or calibration is not finished. Item
ErC
will be turned o automatically after 3 seconds.
Hi
In ear or forehead mode, T>42.9°C (109.2°F)
In object mode, T>100°C (199°F)
Lo
In ear or forehead mode, T<32°C (89.6°F)
In object mode, T<0°C (32°F)
36.8
The power voltage is between 2.4V-2.7V
*
The power voltage is lower than 2.4V.
Failed to
• The battery level is too low.
power on
• Polarities of the batteries are reversed.
• The thermometer is damaged.
Reading
• The lens of the probe is dirty.
is too low
• The distance of the item and target is too far.
• You have just come from a cold environment.
Reading
You have just come from a hot environment.
is too high
**Normal Body Temperature / Fever Indicator
Body temperature can vary
from 97°F and 100.0°F and still
be considered normal. The
medically accepted "normal"
body temperature is 98.6°F.
Body temperature is commonly
lower upon waking than at any
point during the rest of an
individual's waking hours.
FOREHEAD READER ON
Thermometer defaults to selection mode.
.
.
Touchless - Room/Object:
32.0°F - 199.0°F (0°C - 100°C)
Indicator: White light, one long beep
Do not dispose of used batteries in the trash. Consult your
local recycling center for the proper way to dispose of
batteries in your area. Keep used batteries away from
children, pets and excessive heat.
Solution
One long beep, and red backlit for 3 seconds. Take a
measurement under an ambient temperature between
10°C (50.0°F) and 40°C (104°F).
One long beep, and red backlight for 3 seconds.
Contact dealer.
One long beep, and red backlight for 3 seconds.
One long beep, and red backlight for 3 seconds.
The battery icon will flicker, it means the battery is low,
but you still can use the thermometer.
The battery icon will flicker, the item will be turned o
automatically after 3 seconds. Replace with a new battery.
• Replace with a new battery.
• Ensure the batteries are in the right position.
• Contact dealer.
• Clean the lens with a cotton swab.
• Keep in contact with forehead or put probe into the ear canal.
• Stay in warm room for at least 30 min before taking a reading.
Stay in an adequately cool room for at least 30 minutes before
taking a reading.
Temperature readings will vary
A fever is defined as a body temperature that
based on the body location point of
is elevated above the normal for that person.
the reading. Oral temperature
It is important to determine what is normal for
readings between 97°F (36.1°C) and
an individual before determining if a fever is
100°F (37.7°C) are considered
present. Tracking an individual's temperature
normal. A rectal temperature reading
on a consistent basis, at the same body site
is generally 1°F (0.5°C) higher; while
and at the same time every day when the
an underarm, or axillary,
person is well or healthy will help determine
temperature will be 1°F (0.5°C) lower.
an individual's normal temperature.
or
Need help?
Do you have a question about the AOJ-20A and is the answer not in the manual?
Questions and answers