Sound Field Instrumentation ULTRASOUND POWER GENERATOR SF610 100 kHz – 10 MHz, 20 W User Manual www.gampt.de...
Page 2
GAMPT mbH. GAMPT mbH is not liable for damages occurring as a result of incorrect use, or for repairs and alterations carried out by third parties not authorized by GAMPT mbH.
Amplifier stage (ULTRASOUND AMPLIFIER) ..............11 Example of use ......................13 Technical Details ......................17 Revisions of the user manual ..................19 Please read through the whole user manual before using the device. Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Because power ultrasound is produced with the device and the connected ultrasonic probe, the ultrasonic probe must not be used on people or animals. GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
The Ultrasonic Power Generator SF610 combines both components in a compact device in combination with an efficient power supply. In addition, the SF610 has a number of safety features that can be very helpful in protecting the ultrasonic probes from overload.
PROBE RF OUT output (13) signal output of waveform generator output of ultrasound amplifier (BNC port (14) on/off switch of device for ultrasonic probe) GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Attention! Only replace the fuse with the type stated in the manual. If the wrong type of fuse is used, there is a risk of fire. Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Line 4: ultrasonic current display and power display I r m s : 0 . 0 2 A 0 . 9 2 W Fig. 5: start screen Version 1.1 GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
The signal frequency can be tapped at the TTL output (11) as a square wave signal (0-5 V). At the SIGNAL output (13), the non-amplified generator signal is supplied according to the parameters set. When the SF610 is switched on, the signal generator always wakes up with the same initial setting (see Table 1).
Page 10
P o u t : I r m s : 0 . 0 2 A 0 . 9 2 W Fig. 11: frequency (initial state) GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Page 11
4 3 . 5 V P o u t : I r m s : 0 . 0 2 A 0 . 9 2 W Fig. 17: reduce the value by turning left Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Page 12
Unit Minimum Maximum Step Size Frequency 100 000 10 000 000 Amplitude mVpp 2 000 Pulse repetition frequency 20 000 Burst length µs 65 000 GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Current continuous Display of voltage and current values The current values of transmitting voltage and current can be read in lines 3 and 4 of the LCD display (3). Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Page 14
The values of current, voltage and power are displayed flashing if the power exceeds a value of 15 W. GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Example of use In the following the procedure for operating the SF610 is described step by step using an example. The task is to operate a 2 MHz probe with sinusoidal 2 MHz bursts at a repetition frequency of 20 kHz and a burst length of 50 µs (100 oscillations per burst).
Page 16
P o u t : I r m s : 0 . 0 2 A 0 . 9 2 W Fig. 27: Burst length selected as signal variable GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Page 17
4 3 . 5 V P o u t : I r m s : 0 . 0 2 A 0 . 9 2 W Fig. 31: Signal amplitude selected as signal variable Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Page 18
Attention! Use the CURRENT and VOLTAGE rotary knobs (1) and (2) to adjust the current and voltage limits according to the connected ultrasonic probe (see also Chapter 4.2). GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Display: 4-line LCD for the display and setting of the parameter values of the signal generator and of the ultrasound generator and of the output voltage of the LASER unit. Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...
Page 20
40-80 Vpp, continually adjustable Current Limit: 100-600 mA, continually adjustable mode, frequency, burst length and pulse repetition frequency correspond to the values set at the signal generator GAMPT mbH · Hallesche Straße 99F · D-06217 Merseburg · Phone +49 3461/2786910 · www.gampt.de...
Revisions of the user manual Revisions of the user manual Revision Revision date Description 1.0.1 October 27, 2020 First edition Firmware version 1.2 (October 2020) Ultrasound Power Generator SF610 · User Manual Version 1.0.1 EN...