CWT. These warnings are identified by the following symbol: The intended use is for the measurement of AC current. PEM accepts no responsibility for any damage resulting from careless use, or failure to observe these instructions.
Page 4
Accuracy (typ.) (5% to 100% Peak Current) Calibration Nominally ±0.2% with the conductor central in the Rogowski loop. See the calibration certificate for further details. Positional Accuracy ±2% variation of accuracy with conductor position in the loop (see ‘Obtaining The Best Measurement’) Linearity ±0.05% of reading Noise (Typ)
HANDLING INSTRUCTIONS Every effort has been made to ensure that the CWT is as robust as possible. To achieve such an extremely thin Rogowski coil the CWT is necessarily delicate. To prolong the life of the coil please observe the following handling instructions. When fully inserted the ferrule will grip the coil.
The voltage ratings are appropriate for intermittent use of the CWT as a test instrument and not for continuous use in a permanent installation. For information regarding permanent installation of PEM’s Rogowski coils on high voltage equipment please consult PEM.
CWT, remove the battery cover and replace with appropriate batteries for your CWT type. Alkaline batteries that are deeply discharged are more prone to leaking. PEM strongly advise that for long-term storage, the batteries are removed from the instrument.
After switch-on the CWT may require a settling period of up to 2 minutes before providing correct current measurement. The settling period is dependent on warm-up time and the low frequency bandwidth. PEM recommend that the integrator is kept well away from any strong sources of electromagnetic interference when taking measurements. Nov 2024...
50 single screened co-axial cable type. Ω PEM does not consider the use of extension cables to be problematic from the noise viewpoint but consideration should be given to the routing of very long cables. Nov 2024...
In the event of a defect where the transducer has not been misused the CWT should be returned to PEM with all freight charges to be paid by the customer. Correction shall be in the form of repair or replacement.