Brand Name of Product
Generic Name of Product
Product Code Number(s)
Intended Use
Range of Applications for
Product
Key Specifications of Product
Shipping Conditions &
Requirements
Storage Conditions
Packaging Conditions
Shelf Life
Description of Use(s)
Preparation for Use
Document Version 3
.
Instructions for Use: RoboticArmCheck
RoboticArmCheck
TM
Test for residual hemoglobin inside the channel of a robotic instrument.
RAC-001
To test for residual blood (hemoglobin) inside a robotic instrument prior to
sterilization/disinfection process.
Robotic instruments with an internal channel that can accommodate a 1.75mm minimum
diameter swab.
MEASURING RANGE: The test kit can detect 0.1 μg of blood with a blue-green color
change either on the swab and/or in the liquid.
RAC-001 Tip size: 1.38mm
Shipping & Storage
N/A
Store RoboticArmCheck
TM
light and heat. Do not allow to freeze.
12 single use tests
Each test kit consists of:
Vial with solution A (transparent cap)
o
Vial B (green cap)
o
Swab
o
Prepackaged, chlorine-free water vial
o
Scissors
o
The reagent expires18 months from date of manufacture.
The Included prepackaged, chlorine-free water vial expires 2 years from date of
manufacture.
See package for expiration date.
Instructions for Using Product
To test for residual hemoglobin inside a robotic instrument prior to sterilization/disinfection
process.
1. Open the protective pouch of the RoboticArmCheck
2. Included are:
A. Solution vial (transparent cap)
B. Vial B (green cap)
C. Collection swab, vial of prepackaged chlorine-free water and scissors
3. Keep test away from sunlight.
.
in closed packaging in a cool place 2°C- 25°C. Keep away from
.
TM
TM
test kit.
.
Need help?
Do you have a question about the healthmark RoboticArmCheck RAC-001 and is the answer not in the manual?
Questions and answers