Ref: OM005 OCTOBER 2024 THIS IS A MEDICAL DEVICE...
Page 2
Contents General Information Case Information Spare Parts List Troubleshooting Guide Parameters that are not default EC Declaration of Conformity...
General Information Start Up Guide Abi Cots are designed to allow bereaved Connect the plug to a 240V standard UK socket • Switch the unit on at the mains and confirm that the • parents to spend a larger amount of LED light in the bottom corner of the unit is on to time with their baby in order to grieve indicate that the unit is receiving power.
General Information Switch off at socket and remove plug. • Allow 2 hours for cot to defrost. • Handling and Transport Follow maintenance procedure as below • Clean stainless steel well as below. The refrigerated cot is transported on a pallet and can •...
General Information Case Cleaning Switch off the refrigerated cot at the plug socket before commencing any cleaning operation. All the case interior and exterior panel work should be regularly cleaned with a damp cloth using mild soap solution. The use of abrasive materials on any surface should be avoided.
Page 6
Model Bond Abi Cot Features Description Refrigerated Cot Wooden Panels White Internal • • Cot Size 863mm High x 458mm Deep Brass Fittings Insulated trivet • • Cot Length 737mm Colour Maximum design 25°C / 60% RH 9010 • Ambient Conditions Temperature Range +2°C to +8°C...
Page 7
Part Number Description Colour Quantity Per Cot MA30Y42 Cot End Panel with display cutout 9010 MA30Y44 Cot End Panel no cutout 9010 MA30Y41 Cot Side Panel with hook 9010 MA30Y43 Cot Side Panel no hook 9010 EX2AC028 Embraco Condensing Unit UEMT6144U 130165 LAE Controller EX5ACA104...
Page 8
No Operation • Check power supply connected and fuse hasn’t blown • Check power cable isn’t damaged • Check RCD hasn’t tripped, this is inside the cot Cot not getting cold (call an engineer) • Refrigerant leak • Damaged compressor •...
Page 9
Controller LAE AT1-5 Parameter Sheet Refrigerant R290 Set Up Data Param Unit Default Actual Description °C/°F (1) Minimum Limit for Set Point °C/°F (1) Maximum Limit for Set Point °C/°F (1) Set Point REF / HEA °C/°F (1) OFF / ON Thermostat Differential Compressor Rest time Default Compressor run with sensor T1 failure...