  • Page 2: Revision Log

    Revision Log Revisions Description Letter Date 12-07-2017 Initial release. New cover. 04-04-2018 01-10-2019 Added FIL-124. 02-19-2020 Replaced NOZ-105 with NOZ-110, new parts pics.
  • Page 3 Air Pump Inlet Filter, (1) Fresh Air Pump Inlet Filter Gasket, (1) Silicone Lubricant Tools Required: Drill, Phillips Screwdriver Bit, Ratchet, 5/8″ Socket, Ladder, Flatblade Screwdriver Parts included in Preventative Maintenance Kit for Model 6024 are shown below. (1) HEPA Filter (1) Atomizer (1) CO2 In-Line...
  • Page 4 3. Remove top access panel by sliding towards front of chamber until (3) tabs on each side are released as shown below. Access Panel Tabs 4. Loosen (2) clamps as shown below. Note: Clamps may be colored different than shown. Clamps INST-PM-6024 Rev D 02/19/2020...
  • Page 5: Replacing Hepa Filter

    3. Install new HEPA filter into tubing assembly with small diameter HEPA filter connector facing towards front of chamber. Note: Silicone lubricant may be needed to install HEPA filter into tubing. 4. Secure HEPA filter assembly with (2) retained clamps. INST-PM-6024 Rev D 02/19/2020...
  • Page 6 3. Remove and discard old fresh air pump inlet filter gasket as shown below. Fresh Air Pump Inlet Filter Gasket 4. Install new fresh air pump filter. 5. Install new fresh air pump filter gasket. 6. Secure fresh air pump filter cover with retained screw using flatblade screwdriver. INST-PM-6024 Rev D 02/19/2020...
  • Page 7 4. Remove (4) screws securing intake diffuser using Phillips screwdriver bit as shown below. Note: Retain screws. Intake Diffuser Intake Diffuser Mounting Screws (as viewed from inside chamber under intake diffuser) 5. Remove intake diffuser by pulling towards front of chamber. INST-PM-6024 Rev D 02/19/2020...
  • Page 8 12. Align tabs on intake diffuser with drip pan. 13. Secure intake diffuser with retained (4) screws using Phillips screwdriver bit. 14. Reinstall false floor. 15. Reinstall shelf channels. 16. Reinstall shelves. 17. Plug in power cord and turn on chamber. INST-PM-6024 Rev D 02/19/2020...
  • Page 9 18. Continue to instructions for replacing additional accessory preventative maintenance items if applicable. INST-PM-6024 Rev D 02/19/2020...