Page 1
BC 28 DE Blutdruckmessgerät Gebrauchsanweisung ..........2 EN Blood pressure monitor Instructions for use ..........21 FR Tensiomètre Mode d’emploi ............. 38 ES Tensiómetro Instrucciones de uso ..........56 IT Misuratore di pressione Istruzioni per l’uso ..........74 TR Tansiyon ölçme cihazı...
Thank you for choosing a product from our range. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, weight, blood pressure, body temperature, pulse, gentle therapy, massage, beauty, baby and air. With kind regards, Your Beurer team Table of contents 1. Included in delivery ............22 7. Usage ................29 2.
1. Included in delivery Observe the instructions Check that the exterior of the cardboard delivery pack- Read the instructions before starting aging is intact and make sure that all contents are pre- work and/or operating devices or sent. Before use, ensure that there is no visible dam- machines age to the device or accessories and that all packaging Isolation of applied parts...
evaluated graphically. You are also warned of possible Permissible storage and transport Storage/Transport existing cardiac arrhythmia. temperature and humidity 4. Warning and safety notes Permissible operating temperature and Operating Notes on use humidity • In order to ensure comparable values, always measure your blood pressure at the same time of day.
Page 5
• Using the blood pressure monitor outside your home • Cardiovascular diseases may lead to incorrect meas- environment or whilst on the move (e.g. whilst travel- urements or have a detrimental effect on measurement ling in a car, ambulance or helicopter, or whilst under- accuracy.
Page 6
• Place the cuff on your wrist only. Do not place the cuff Notes on handling batteries on other parts of the body. • If your skin or eyes come into contact with battery • The blood pressure monitor can only be operated with fluid, rinse the affected areas with water and seek batteries.
5. Device description Notes on electromagnetic compatibility 1. Display • The device is suitable for use in all environments 2. Wrist cuff listed in these instructions for use, including domestic 3. Risk indicator environments. 4. Start/stop button • The use of the device may be limited in the presence 5.
6. Initial use Information on the display: 1. Systolic pressure Inserting the battery 2. Risk indicator 3. Diastolic pressure 4. Calculated pulse rate 5. Cardiac arrhythmia symbol 6. Pulse symbol 7. User memory / 8. Memory space number/ • Remove the battery compartment lid on the left side memory display for of the device.
Page 9
• Before initial use and after each time you replace the Date battery: The year flashes on the display. When inserting the batteries into the device you will be • Select the desired year using the taken to the relevant menu automatically. memory button M and confirm with •...
7. Usage If the values between the two arms are significantly dif- ferent, please consult your doctor to determine which Ensure the device is at room temperature before meas- arm should be used for the measurement. uring. The measurement can be performed on the left or right wrist.
Page 11
• Press the START/STOP button to • Now select the desired user memory by pressing the start the blood pressure monitor memory button M. If you do not select a user memory, . All displays will illuminate the measurement is stored in the most recently used briefly.
Page 12
stimulants, stress or lack of sleep. Arrhythmia can only the higher category; for the example given this would be be determined through an examination by your doctor. “High normal”. If the symbol is shown on the display after the Blood pressure Systole Diastole Action...
Page 13
Saving, accessing and deleting measured values • Press the memory button M. The results of every successful measurement are AM flashes on the display. stored together with the date and time. The old- The average value of the morning est measurement is overwritten in the event of more measurements for the last 7 days is than 60 measurements.
9. What if there are problems? • To clear the relevant user memory, you must first select a user memory. Error Possible cause Solution • Start the retrieval of the average measured values. message A flashes on the display and the average value of The systolic or Please wait one all saved measured values in this user memory is...
10. Disposal Error Possible cause Solution message Repairing and disposing of the device The blood Please wait one • Batteries do not belong in household waste. Please pressure is higher minute and repeat the dispose of empty batteries at the collection points than 300 mmHg measurement.
During the warranty period, should this product prove appropriate means. Precise instructions for checking to be incomplete or defective in functionality in accord- accuracy may be requested from the service address. ance with the following provisions, Beurer shall carry...
Page 18
Repairs or an exchange in full do not extend the war- A warranty claim shall only be considered if the buyer ranty period under any circumstances. can provide Beurer, or an authorised Beurer partner, with - a copy of the invoice/purchase receipt, and - the original product.
Need help?
Do you have a question about the BC 28 and is the answer not in the manual?
Questions and answers