Page 1
AimSafety™ PM Multi-Gas Monitor w/ Pump User Instructions IMPORTANT: Read and understand contents of this manual prior to operation. Keep these user instructions for reference.
Page 2
AimSafety PM500 Manual REV – 1.0 [34-2900-0000-2] P a g e...
AimSafety PM500 Manual TABLE OF CONTENTS Product Overview ......................6 1.1. Introduction ........................ 6 Specifications ......................... 6 2.1. LCD Display Symbols ....................8 Activation ........................9 3.1. Power On the Monitor ....................9 3.2. Powering Off the Monitor ................... 9 Pump ..........................
Page 4
AimSafety PM500 Manual Warning ➢ Do not replace or change any parts. Unauthorized replacement of parts and/or maintenance will void the warranty. ➢ Remove any debris on the surfaces of the sensor, display, LED’s, buzzer and pump hole before use.
Page 5
AimSafety PM500 Manual ➢ If a charge-generating mechanism is present, the exposed metallic part on the enclosure is capable of storing a level of electrostatic charge that could ignite IIC gases. Therefore, the user / installer shall implement precautions to prevent the build-up of electrostatic charge. This is particularly important if the equipment is being used in a CID1 or Zone 0 environment.
AimSafety PM500 Manual 1 Product Overview 1.1. Introduction The PM500 is a portable sampling type gas detector that measure up to six (6) gases simultaneously to protect workers from the disasters caused by the Oxygen deficiency, Toxic gas poisoning and gas explosions.
Page 7
AimSafety PM500 Manual Event Log: 30 EA Calibration Log: 30 EA Data Logging Bump Log: 30 EA Data Log: Two Months or longer Dimension 5.7” (L) x 3.0” (W) x 1.7” (H) (14.6 cm x 7.7 cm x 4.3 cm) Weight 17.3 oz.
AimSafety PM500 Manual 3 Activation 3.1. Power On the Monitor Press the Power/Enter button for 2 seconds and SYS boot is displayed. Once the device turns on, the version and LCD are displayed. After 10 seconds of system test, the warmup will begin.
AimSafety PM500 Manual 4 Pump 4.1. Pump Test When you change the gas tube or reconnect to the detector, test the sampling system by blocking the end of the tube. When the flow is blocked, warning alarm will be activated and the device will continue to alarm.
AimSafety PM500 Manual 5.2. Display Mode Pressing the △ or ▽ button steps the display through fourteen different display modes as shown above. Measure Value (Min) Measure Value (Max) Measure Value (STEL) Measure Value (TWA) Alarm Value( 1 Alarm Value (2...
AimSafety PM500 Manual 5.2.1. Display Mode in Detail LCD Display Description in Detail Measuring Mode (Basic Display) Display the current gas levels of the atmosphere Battery power level icon is displayed. A low gas alarm has been triggered. *In ambient air, the Oxygen level normally indicates 20.9%vol.
Page 13
AimSafety PM500 Manual LCD Display Description in Detail Display the preset low alarm levels. Display the preset high alarm levels. Display the preset STEL levels. Display the preset TWA levels. Firmware Version, Current battery voltage, Current temperature (Fahrenheit) REV – 1.0...
Page 14
AimSafety PM500 Manual LCD Display Description in Detail Date / Time Calibration Concentration Value Last Calibration Date (01.01= January 1 Remaining time to next calibration date when the calibration interval is configured. (Default: N/A) Remaining time to next bump test date when the bump interval is configured.
AimSafety PM500 Manual STEL Alarm activation: The alarm activates when the average gas levels for the last fifteen minutes exceed the STEL concentration. The audible, visual and vibration alarms stop when the device is in a safe area where the gas concentration is normal.
AimSafety PM500 Manual 6 Settings and Operation Press and hold △ + ▽ buttons for two seconds simultaneously to enter the Settings menu. In the settings menu, the Alarm Menu, Calibration Menu, Inspection (Test) Menu, and Setup Menu will display. Users can configure a setting by scrolling to the required menu and pressing the Power/Enter button in a menu.
AimSafety PM500 Manual Change alarm Clear the previous Clear the previous Change Auto value settings Min, Max value TWA & STEL value or Latch alarm mode ① ALR SET: Adjust the Low, High, STEL & TWA alarm levels ② MIN MAX CLR: Deletes the previously measured MAX and MIN levels.
Page 19
AimSafety PM500 Manual To activate a zero or span calibration, press the Power/Enter button. ① Zero Run (Zero / Fresh Air Calibration) ② SPAn Run (Span Calibration) In the Span calibration mode, you must specify which sensor(s) will be calibrated.
AimSafety PM500 Manual IMPORTANT For span calibrations, if an additional toxic or VOC (PID) sensor is included, perform the span calibration of those sensors after the other sensors are calibrated. CAUTION Before calibration, confirm that the sensor warmup is completed. The device will display normal concentrations when the warmup is complete.
Page 21
AimSafety PM500 Manual Power/Enter button to activate the test. There are function tests for the LCD, buzzer, LED’s, Backlight (Yellow), Backlight (Red), Motor and Memory. Self Test Bump Test In the BUMP RUN mode, press the Power/Enter button to enter the mode, and then select either on or off for each sensor.
AimSafety PM500 Manual 6.4. Set Menu In the “set menu”, press the Power/Enter button to enter into the mode, and then select from the four menus below by pressing the △ or ▽ button. To enter the mode, press the Power/Enter button.
AimSafety PM500 Manual 7 Log During a normal operation, data, event, calibration, and bump logs are stored in the device. The stored data can be downloaded via AimSafety PM Link with the PC program. Up to 30 log events will be saved, and once the data is full, the oldest date will be overwritten automatically and the new data will be stored.
AimSafety PM500 Manual 8 Certification √ FCC compliance This device is tested according to FCC rules Part 15 and complies with restrictions for a Class A digital device. These restrictions are designed to provide adequate protection against an industrial environment which may cause harmful interference during operation.
Page 25
AimSafety PM500 Manual ATEX CSANe 23ATEX1128X EN IEC 60079-0:2018 0080 EN 60079-1:2014 EN 60079-11:2012 II 1G II 2G Ex db ia IIC T4 Gb Ex db ia IIB T4 Gb • PM500-IR (Infrared LEL) Ex ia IIC T4 Ga Ex ia IIB T4 Ga •...
AimSafety PM500 Manual 9 Error Codes The following error codes may be displayed on the LCD if the detector is not working properly. Code Description Solution Visually inspect the pump and filter. If dirty and/or damaged, replace Err – 1 Pump Operation Error with new pump inlet and/or filters and cycle power on the monitor.
AimSafety PM500 Manual 11 Maintenance & Cleaning 11.1. Charging A universal charging adapter is included with every PM500 monitor. Only AimSafety certified adapters should be used to charge the unit. CAUTION Charge at room temperature (32°F to 104°F) Ensure the charging adapter is firmly connected to the PM500 The backlight indicates charging is taking place by switching On/Off every 1 second Do not use any alternative adapter.
Page 28
AimSafety PM500 Manual 11.5. Pump When you change or reconnect any sampling tubing, a pump test should be performed. To perform a pump test, firmly block the end of the tube and confirm that the unit goes into alarm. 11.6. Filter Ensure the dust and PTFE filters are free from debris, obstructions and discoloration.
AimSafety PM500 Manual 12 Response Factors for Combustibles and VOCs 12.1. Catalytic Bead / Pellistor LEL Sensor LEL (Lower Explosive Limit) sensors are designed to detect a wide variety of combustible gases and vapors. Different gases exhibit varying responses when detected by LEL sensors. For the most accurate readings, it is recommended that the instrument be calibrated with the target gas.
Need help?