Summary of Contents for Environmental Measurement Japan MIJ-12
Page 1
MIJ-12 Waterproof Data Logger Manual English Version Environmental Measurement Japan Environmental Measurement Japan, CO., LTD. 2-52-42 Takamidai, Fukuoka-city Higashiku, Fukuoka 811-0215, Japan TEL +81-92-608-6412 FAX +81-92-985-7844...
Page 2
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan MIJ-12 Preheat Version: How to wiring output Sig - Sig - HIOKI Sensor Cable 279-04 POWERIN 2.5V 279-04 Time POWERIN LR5042 VOLTAGE LOGGER <1M> mini 記録 SetVol PWLED 設定 停止 2秒押し <5K>...
There is a red sticker on the base. MIJ-12 Preheat-less Version: How to wiring How to wire the senser (do not require preheat) to MIJ-12 preheat-less model. Since it does not require preheating, the PR-01 board is not required. Sensor Cable ×...
Caution When using the MIJ-12 in the field, please note the following: Open the zipper as shown below and put the zip bag into the MIJ-12. There is no need to remove the silica gel from the zip bag. GOOD! 乾燥剤...
Page 5
When using the MIJ-12 in the field, please note the following: There have been cases where the lid of the MIJ-12 was closed with the cable still in place, causing water to get into the case and destroying the PR-01 and data logger.
Page 6
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan Data Logger 5041 or 5042 how to set The following settings are made at the time of shipment. ・Rec interval:10min ・Start Method:Button Operation ・Stop Method:Button Operation(one-time) ・Rec Mode:Instantaneous ・Preheat:ON/OFF from [Measurement Method] tab *Generally, we do not recommend making any changes other than to the rec interval.
Page 7
Environmental Measurement Japan Click the measurement method tab to set preheat ON/OFF. MIJ-12 Preheat-less Version: Set preheat to OFF MIJ-12 Preheat Version:Set preheat to 1 sec as below Finally, click the Submit Settings button to complete. Logging Start, Stop, and Data collect Start Logging Place the sensor.
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan The measurement screen will be displayed one second after the REC light turns to steady. This means that measurement has started normally. If the REC disappear from the display you have to try again.
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan Trouble Shooting Battery last only a few weeks Is the preheat time set too long? Although it depends on the sensor you are using, a preheat time of 1 second is generally fine. Some customers set the preheat time 10 minutes, but this unnecessarily shortens the battery life, so please change it to 1 second.
Page 10
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan Data not avilable Did you check the operation before installing it? If you are using the product for the first time, we recommend that you check its operation and perform preliminary tests before installation.
Page 11
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan ・ Try Short circuit the GND and PRH terminals of the PR-01 with tweezers, etc. Does the LED light up ? output Sig - Sig - POWERIN 279-04 POWERIN 2.5V 279-04 Time...
Page 12
MIJ-12 Waterproof Data Logger Manual Environmental Measurement Japan Output is Negative: Please check that the included LR9802 cable is plugged in the correct direction. HIOKI LR9802 cable HIOKI HIOKI HIOKI HIOKI - 5.000 5.000 DATA DATA LR5042 LR5042 VOLTAGE LOGGER...