Page 1
RJL Systems 33939 Harper Avenue Clinton Township, MI 48035 USA Voice: 800 528-4513 Fax: 586 790-0200 Web: www.rjlsystems.com Quantum VII Bioelectrical Impedance Analyzer (BIA) Operators Manual for the study of human body composition by Clinicians and Healthcare Professionals...
Table of Contents About the Quantum VII real time BIA….…………………….. 4, 5, 6 Architecture and Specifications…………… ..…………….….. Classical Mode BIA …………………...…………….….. Using the Quantum VII (Getting Started)……..……..….….. Classical BIA Screen Shots ……………….…..…...…… The Main Menu ………………........….….. 11, 12 Set WIFI SSID …...............
In Scheduled BIA mode measurements are taken at specific intervals from 2 seconds to 24 hours. Any interval above 20 seconds will cause the Quantum VII to sleep until the next sample time arrives. The instrument will then wake up for 5 seconds to take a new measurement, average them and create a new report record.
Page 5
LAN is configured. This will precisely set the clock / calendar to 1 second accuracy. Battery: The Quantum VII uses a 7.4 Volt 850 mAh Lithium Ion battery with smart protection and is CE, RoHS MSDS Certified. will last more than 6 hours of continuous operation and can be recharged ...
Page 6
An event file can have 32,768 records and must be manually reset with the CLEAR SUBJECT DATA menu entry. The only method to access any record is with the WiFi feature of the Quantum VII and its website. The Local Area Network (LAN) SSID and password will have to be set for the first time for the IP address of the website to work.
Architecture: The Quantum VII is built around a low power Arduino NANO 33 IOT (Internet of Things) with the following features: Micro-controller Low power Arm Cortex_MO+ 32 bit SAMD21 Radio Module U-box NINA-W102 including Wifi and Bluetooth Low Energy (BLE) for communications...
Before turning the analyzer on, prepare the subject (patient) as described in Performing a BIA Test page 24. Insert the subject cable into the top connector of the Quantum VII and start analyzer by pressing the middle ON button. In Classic BIA mode a splash screen will begin and take about three seconds then a green flash appears indicating there are no errors reading the calibration values.
RT max – min range measured resistance, reactance and phase angle of N number of BIA observations. Then the the Quantum VII shutdown procedure starts and turns the instrument off in 2 seconds. The file containing the Classic BIA report records is available ...
ON button and the display will change to the characteristics of the item. Touching the display window again will return the menu in most cases. When the Quantum VII is in any of its three communication modes the touch screen menu is disabled forcing the use of the ON/OFF button to turn the instrument off.
RJL Systems for product service. Calibration values are kept on record for comparison to a new calibration procedure. The first firmware release of the Quantum VII is revision • 0.050 dated 01/03/2023. The Current Ua RMS: is the actual real time 50 Khz •...
10 connection is working. The message “Hello from client” will records / second with an average appear on the Quantum VII screen and be erased when the WiFi connection. Delete Message button is pushed The Classical or Scheduled BIA records can be selected •...
Scheduled BIA Mode Sampling Time > 00:00:00 More than 2000 saved records / battery charge Start BIA logging : Press ON quickly Management: Press ON for > 2 seconds Averages : Press ON again (STD, min-max) last reading Turn completely OFF: Press ON again...
When SET SAMPLE TIME is greater than zero the scheduler is active. It is always started by press the ON button momentarily when the Quantum VII is OFF. That gives the user a chance to set up an experiment (relocate the analyzer and connecting leads) before scheduling BIA observations. It is a good idea to start the scheduler at a time that is convenient, such as the second and minute hand on a watch, to know when it will turn back on.
The only method to access any record is with the WiFi feature of the Quantum VII and its website. The Local Area Network (LAN) SSID and password will have to be set for the first time for the IP address of...
BIA measurement period. The usefulness of the Quantum VII as a research tool in studying body composition changes due to digestion, Typical BIA Measurement exercise and forced motion are many. For example, a 15...
RJL Systems Standard Manual Entries ▪ Care of the analyzer and cables ▪ Using the test resistor ▪ Subject preparation and test protocol ▪ Electrode Placement ▪ Trobleshooting Guide ▪ Standard Warranty Proper care of the analyzer and cables...
RJL analyzers are built to last. Many RJL instruments built over 20 years ago are still being used today. The instruments we design and build are subjected to our rigorous quality assurance testing procedures. Like all sensitive electronic instruments, proper care and handling will minimize the risk of damage. CLEANING &...
Test Resistors are supplied with every Analyzer - cables can be tested before BIA testing is done. RJL Systems will supply additional Test Resistors at no charge other than a shipping fee. Replacement cables can be purchased from RJL and shipped via “overnight” delivery if necessary.
Contact RJL Systems to order a new set. If either test fails, or you have questions about the results please call RJL Systems Technical Support. Note: Have your readings available or be prepared to run the test(s) again while an RJL technician is...
Preparation for subject testing (WHAT TO DISCUSS WITH SUBJECT PRIOR TO TESTING) Subjects who do not adhere to the pre-testing protocols Failure to adhere to the pre-testing protocols described on the next page could alter the way the body interacts with the electrical signal of the BIA. This could alter the body's resistance and reactance, which would result in altered body composition estimates produced by the software.
Pre-testing protocol for the subject Explain testing procedures to the subject and address any questions or concerns prior to testing. Name, gender, age, height, weight, and side of body being tested should be recorded. If exceptions to Testing Protocol are evident (i.e. subject tested sitting up due to wheelchair use) record the exception in the database record in the comment field, and test that individual in consistent manner every time.
1 Turn the Analyzer ON by pressing the (|) button. 2 The RJL Systems initial splash screen and a green flash will appear. After ten seconds the N count will start counting. The analyzer is now recording resistance, reactance and phase angle values.
Troubleshooting Guide Possible causes of STRANGE TEST RESULTS OR ERRATIC RESISTANCE/REACTANCE DISPLAY: A malfunctioning analyzer or broken/damaged subject cables (see “Using the Test Resistor to Test Analyzer/Cables). Expired electrodes. RJL electrodes have an average shelf life of approximately 15 months. Check ...
If RJL requests, the Customer will, at its own expense, deliver the equipment in appropriately protective packing to RJL Systems. RJL Systems will pay for the return shipping expense WHAT IS NOT COVERED: Accessories are not covered under this warranty.
Customer is responsible for all shipping costs – including insurance if desired - both to and from RJL Systems. If the Analyzer is a discontinued model, RJL Systems will advise if repair parts for the unit can still be obtained and at what cost After RJL Systems receives and performs a diagnostic evaluation on the unit, we will contact the customer with a repair estimate (labor and parts) before proceeding with service.
Page 31
Quantum VII Downloading Saved Data Every measurement made with the Quantum VII is saved on an internal SD card (8 GB). There are two files that contain Classical and Scheduled BIA measurements and have no size limitation. Each measurement is date and time stamped with all BIA values and characteristics.
Page 32
Creating a CSV file Select (click) the Classical BIA Rec xxx or Scheduled BIA Rec xxx menu item and your browser will fill with all saved records as text lines with a top header that labels each field. This may take a while depending on how many records are saved (more than 1000).