The information in this manual is proprietary and confidential to Whitehall Manufacturing, a division of Acorn Engineering and a member of Morris Group International. Any reproduction, dissemination or distribution of this document in whole or in part without prior written authorization of Whitehall Manufacturing is strictly prohibited. While every effort has been made to ensure the accuracy of information provided in this document, please note the information is subject to change without notice.
Page 3
Draining Instructions ...........................12 Moving your Whirlpool ........................12 Thermometer Calibration Instructions....................13 Disinfecting and Cleaning Instructions....................14 Turbine Maintenance ...........................14 Troubleshooting ............................15 Classification ...............................16 Specifications...............................17 WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 4
Indicates that user manual shall be referred to before performing any installation, operation or maintenance Alternating Current Indicates alternating current is present Manufacturer Indicates manufacturer of device Manufacture Date Indicates manufacture date of device WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 5
• If there is visible damage to the Whirlpool or packaging upon receipt, report this damage to shipping company and Whitehall. Do not use the Whirlpool if damage to it is suspected. •...
Page 6
Incident Reporting Any serious incident that has occurred in relation to the Whirlpool should be reported to Whitehall manufactur i ng. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 7
For smaller Whirlpools, the user sits comfortably next to the tub and places their extremity (hand, foot, arm or leg) in the tub. For larger Whirlpools, the user can enter the tub and sit inside. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 •...
Page 8
Hang seat on back edge of tub as WHL-90-M shown. Bottom of seat rests on WHL-105- bottom of tub. For maintenance, see the Disinfecting and Cleaning Instructions section. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 9
Seat for 610 WHS-110- mm (24 inch) wide tubs, except Lo-Boy WHL-75-M Adjustable WHL-90-M Suspension WHL-105- Seat for Lo- Boy tubs WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 10
(for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 •...
Page 11
3. Fill tub with water so that the bottom of turbine is at least 15.2 cm (6 inches) below surface of water. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 •...
Page 12
floor drain or sink to allow for draining and make sure drain valve is closed. Lock wheels. Do not move the Whirlpool with water in the tub. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 •...
Page 13
5. Re-calibration shall be performed at least every 12 months or as needed. Figure 10: Calibrating the thermometer WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 14
5. Ensure the pressure control valve is in the open position. 6. Turn on the turbine for 5 seconds and turn it off. Wait 10 minutes. WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 •...
Page 15
2. Push down and turn each of the round black fuse covers counter-clockwise on top of the turbine and set aside 3. Remove the fuses and discard WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 •...
Page 16
6. Plug in the turbine Classification Electric Shock Protection Class I Applied Parts None Ingress Protection IP24 Mode of Operation Continuous WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 17
5 - 95%, non-condensing Atmospheric Pressure 70 – 106 kPa Thermometer Range Dual scale: 5-60°C and 40-140°F Accuracy ± 1.1°C (2°F) WHITEHALL MANUFACTURING • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A. Phone 800-591-9280 • 626-968-6681 • www.whitehallmfg.com...
Page 18
Stainless steel must be properly maintained after water has been introduced into the fixture, or Whitehall's limited warranty is void. If you have any questions or require technical assistance, please call 800-743-8219.