Each pipette has been checked & calibrated at factory with procedure conforming to EN- ISO 8655-6 standards.
Depending upon use, we recommend checking of calibration every six months. However this can be adjusted to individual requirements.
The pipette is checked at maximum volume, at 50% of maximum volume and at minimum or 10% of maximum volume, whichever is higher.
(Systematic Error)
(Random Error)
An analytical balance must be used. The balance selection depends upon selected model of the pipette and sensitivity of balance reading.
Test liquid: Water, distilled or deionized, grade 3 water conforming ISO3696.
Calibration should be carried out in a draft- free room at a constant (+/- 0.5°C) temperature of water, pipette and air between 15°C to 30°C. The relative humidity must be above 50% especially with volumes under 50 µl, the air humidly should be as high as possible to reduce the effect of evaporation loss.
Special accessories for analytical balance, such as the evaporation trap are recommended for the calibration of volumes under 50 µl.
When not in use, it is recommended that your pipettor is stored in a vertical position.
Leaving the pipette on it's side can cause liquids to leak into the body of the pipette and cause corrosion.
These pipettes are available as fixed or variable, general purpose micropipettes for sampling and dispensing accurate amounts of liquid in micro volumes.
These pipettes operate on the air displacement principle and use disposable tips of any standard make. All pipettes are equipped with a built-in tip ejector.
Reference | Volume Range (µl) | Increment (µl) | Accuracy (±) % | Precision (±) % |
EASD-P02-001 | 0,1-2,5 | 0,01 | ±12,0% - 2,5% | ±6,0% - 1,6% |
EASD-P10-001 | 0,5-10 | 0,1 | ±2,5% - 1,0% | ±1,5% - 0,8% |
EASD-P20-001 | 2-20 | 0,1 | ±2,0% - 1,8% | ±2,0% - 0,4% |
EASD-P50-001 | 5-50 | 0,5 | ±3,0% - 0,9% | ±2,0% - 0,3% |
EASD-10P-001 | 10-100 | 0,5 | ±3,0% - 0,8% | ±1,5% - 0,15% |
EASD-20P-001 | 20-200 | 1,0 | ±2,0% - 0,6% | ±0,8% - 0,15% |
EASD-1KP-001 | 100-1000 | 5,0 | ±2,0% - 0,6% | ±0,7% - 0,2% |
EASD-5KP-001 | 500-5000 | 50,0 | ±2,0% - 0,5% | ±0,6% - 0,2% |
EASD-10K-001 | 1000-10000 | 100,0 | ±3,0% - 0,6% | ±0,6% - 0,2% |
These Pipette are designed and constructed for accurate and precise liquid handling. These devices are intended to dispense liquid in the volume range from 0.1uL to 10mL in combination with matching pipette tips.
These pipettes may only be operated by trained specialist staff for general laboratory purposes. All users must have read the operating manual carefully and familiarized themselves with the device mode of operation.
Note: Reverse pipetting technique is recommended for viscous solutions, solutions having tendency to foam or for dispensing very small volumes.
To maintain the best results from your pipettor, each unit should be checked every day for cleanliness. Particular attention should be paid to the tip cone(s).
This pipettor has been designed for easy in- house service. However, we also provide complete repair and calibration service. Please return your pipettor to your local distributor for repair or calibration. Before returning, please make sure that it is free from any contamination.
Check the performance of your pipettor regularly e.g every 3 months and after every in-house service or maintenance.
To clean your pipettor, use ethanol and a soft cloth or lint-free tissue. It is recommended to clean the tip cone regularly.
Note: Never disassemble the upper part of the pipette. To avoid losing or damaging fragile parts, reassemble the pipette immediately.
Trouble | Possible Reason | Correction |
Droplets left inside the tip | Unsuitable tip, non-uniform wetting of the plastic | Use new tip |
Leakage or Pipetted volume too small | Tip holder (cone) scratched or damaged | Change the tip cone |
Organic solvent as liquid | Aspirate & discard the organic solvent several times before actual pipetting by the same tip | |
Tip incorrectly attached | Attach firmly | |
Unsuitable tip | Use new tip | |
Foreign particles between tip and tip cone | Clean the tip cone | |
Insufficient amount of grease on piston & O-ring | Clean & grease O-ring and piston | |
O-ring not correctly positioned or damaged | Change the O-ring | |
Inaccuracies | Incorrect operation | Follow instructions carefully |
Calibration altered | Recalibrate according to instructions | |
Unsuitable for the particular liquid pipetting technique | Use correct pipetting technique | |
Instrument damaged | Send to repair | |
Push button jammed or moves erratically | Piston contaminated | Clean & Grease O-ring and piston |
Penetration of solvent vapors | ||
Tip ejector jammed or moves erratically | Tip cone contaminated from outside | Remove ejector collar and clean tip cone's outer surface with ethanol |
Volume setting is not properly click stopped | Click stop mechanism damaged | Send for repair |
Push button does not turn for volume setting | Use of excessive force beyond the range of pipette | Send for repair |
If help is needed, contact with your supplier/manufacturer or Labbox via:
www.labbox.com
Please, provide the customer service representative with the following information:
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the 40+ and is the answer not in the manual?
Questions and answers