Easy 40+ Manual

Calibration

Each pipette has been checked & calibrated at factory with procedure conforming to EN- ISO 8655-6 standards.
Depending upon use, we recommend checking of calibration every six months. However this can be adjusted to individual requirements.

Procedure to check calibration

The pipette is checked at maximum volume, at 50% of maximum volume and at minimum or 10% of maximum volume, whichever is higher.

  • A new tip is first pre-wetted 3-5 times and a series of ten pipetting is done at each volume.
  • Use of forward pipetting technique is recommended.
  • Calculate the inaccuracy and imprecision for all three volumes as per EN ISO 8655-6/DIN 12650 standards based on the following calculation

Conversion of weight readings to volume

  • Mean Value
  • Mean Value
  • Xi: Balance Reading
  • n: Number of Reading
  • Z: Conversion factor [example Z=1.0040 µl/mg at 25°c and 1013 hPa]

Calculation for in-accuracy

(Systematic Error)

  • A% =
  • V: Mean Value
  • Volume at which readings are taken

Calculation for Imprecision

(Random Error)

  • S =
  • CV% =
  • S: Standard Deviation
  • : Mean Value
  • n: Number of readings

Devise requirements and test conditions

An analytical balance must be used. The balance selection depends upon selected model of the pipette and sensitivity of balance reading.
Test liquid: Water, distilled or deionized, grade 3 water conforming ISO3696.

Calibration should be carried out in a draft- free room at a constant (+/- 0.5°C) temperature of water, pipette and air between 15°C to 30°C. The relative humidity must be above 50% especially with volumes under 50 µl, the air humidly should be as high as possible to reduce the effect of evaporation loss.
Special accessories for analytical balance, such as the evaporation trap are recommended for the calibration of volumes under 50 µl.

Recalibration

  1. Place the service tool intro grooves at the base of the push button as shown in the Figure 1 below. Turn it clockwise to increase & anticlockwise to decrease the volume.
  2. Repeat the 'Checking calibration' procedure.

Storage

When not in use, it is recommended that your pipettor is stored in a vertical position.
Leaving the pipette on it's side can cause liquids to leak into the body of the pipette and cause corrosion.

Product description

These pipettes are available as fixed or variable, general purpose micropipettes for sampling and dispensing accurate amounts of liquid in micro volumes.
These pipettes operate on the air displacement principle and use disposable tips of any standard make. All pipettes are equipped with a built-in tip ejector.

Product Table Comparison

Reference Volume Range (µl) Increment (µl) Accuracy (±) % Precision (±) %
EASD-P02-001 0,1-2,5 0,01 ±12,0% - 2,5% ±6,0% - 1,6%
EASD-P10-001 0,5-10 0,1 ±2,5% - 1,0% ±1,5% - 0,8%
EASD-P20-001 2-20 0,1 ±2,0% - 1,8% ±2,0% - 0,4%
EASD-P50-001 5-50 0,5 ±3,0% - 0,9% ±2,0% - 0,3%
EASD-10P-001 10-100 0,5 ±3,0% - 0,8% ±1,5% - 0,15%
EASD-20P-001 20-200 1,0 ±2,0% - 0,6% ±0,8% - 0,15%
EASD-1KP-001 100-1000 5,0 ±2,0% - 0,6% ±0,7% - 0,2%
EASD-5KP-001 500-5000 50,0 ±2,0% - 0,5% ±0,6% - 0,2%
EASD-10K-001 1000-10000 100,0 ±3,0% - 0,6% ±0,6% - 0,2%

Intended Use

These Pipette are designed and constructed for accurate and precise liquid handling. These devices are intended to dispense liquid in the volume range from 0.1uL to 10mL in combination with matching pipette tips.
These pipettes may only be operated by trained specialist staff for general laboratory purposes. All users must have read the operating manual carefully and familiarized themselves with the device mode of operation.

Pipetting Technique

  1. Forward Pipetting
    1. Press the operating button to the First stop.
    2. Dip the tip attached to the pipette into the solution to a depth of about 1cm, and slowly release the operating button. Wait for a while, then withdraw it from the liquid touching it against the edge of the reservoir to remove excess liquid adhering to the outer surface of the tip.
    3. Dispense the liquid into the receiving vessel by gently pressing the operating button to the first stop. After a second, press the operating button to the second stop. This will empty the tip completely. Remove the pipette from the vessel by sliding it up the wall of the vessel.
    4. Release the operating button to the ready position.
  2. Reverse Pipetting
    1. Press the operating button to the second stop.
    2. Dip the tip attached to the pipette into the solution to a depth of about 1cm, and slowly release the operating button. This action will fill the tip with a volume that is larger than the set volume. Wait 1-2 seconds and withdraw the tip from the liquid, touching it against the edge of the reservoir to remove excess liquid.
    3. Disperse the liquid into the receiving vessel by pressing the operating button gently and steadily to the first stop. This volume is equal to the set volume. Hold button in this position. Some liquid will remain in the tip, which should not be dispensed.
    4. The liquid remaining in the tip can be dispensed back into the original solution by pressing the button to the second stop or disposed together with the tip.
    5. Release the operating button to the ready position.

Note: Reverse pipetting technique is recommended for viscous solutions, solutions having tendency to foam or for dispensing very small volumes.

Pipetting Recommendations

  • Aspirate liquid into the pipette only when a tip is attached to its tipcone.
  • Allow liquids, tips, and pipettes to equilibrate to the ambient temperature.
  • Pre-rinsing of tip 5 times with the liquid to be dispensed is recommended. This is important especially when dispensing liquids which have a viscosity and density different from water.
  • Pre-rinse the tip several times before use when pipetting liquids at temperature different from ambient.
  • While pipetting, the pipettor should be vertically straight and tip should be dipped only a few millimeters into the liquid.
  • Always control the push button movements with the thumb for consistency.
  • Wipe the tip only if there is liquid on the outside of the tip, being careful to avoid touching the tip's orifice.
  • Don't keep the pipette in your hand while not working, to avoid transferring to body heat. Use the correct pipette tip designed for use with the pipette.
  • Select the correct pipetting technique (e.g. Reverse, forward etc.) depending on the nature of the liquid.
  • Using excessive force to turn the push button outside the range specified for it may jam the mechanism and damage the pipettor.

Maintenance

To maintain the best results from your pipettor, each unit should be checked every day for cleanliness. Particular attention should be paid to the tip cone(s).
This pipettor has been designed for easy in- house service. However, we also provide complete repair and calibration service. Please return your pipettor to your local distributor for repair or calibration. Before returning, please make sure that it is free from any contamination.
Check the performance of your pipettor regularly e.g every 3 months and after every in-house service or maintenance.

Cleaning the product

To clean your pipettor, use ethanol and a soft cloth or lint-free tissue. It is recommended to clean the tip cone regularly.

In-House Maintenance

  1. Hold down the tip ejector.
  2. Place the tooth of the opening tool between the tip ejector and the tip ejector collar to release the locking mechanism.
  3. Carefully release the tip ejector and remove the ejector collar.
  4. Place the wrench end of the opening tool over the tip cone and turn it anticlockwise.
  5. After removing the tip cone, wipe the piston, the O-ring and the tip cone with ethanol and a lint-free cloth.
    Note: Models up to 10µl have fixed O-ring location inside the tip cone, so do not remove or maintain it.
  6. Before replacing tip cone, it is recommended to grease the piston slightly using the silicone grease provided.
  7. After reassembling, use the pipettor (without liquid) several times to make sure that the grease is spread evenly.
  8. Check the pipettor calibration.

Note: Never disassemble the upper part of the pipette. To avoid losing or damaging fragile parts, reassemble the pipette immediately.

Troubleshooting

Trouble Possible Reason Correction

Droplets left inside the tip

Unsuitable tip, non-uniform wetting of the plastic Use new tip

Leakage or Pipetted volume too small

Tip holder (cone) scratched or damaged Change the tip cone
Organic solvent as liquid Aspirate & discard the organic solvent several times before actual pipetting by the same tip
Tip incorrectly attached Attach firmly
Unsuitable tip Use new tip
Foreign particles between tip and tip cone Clean the tip cone
Insufficient amount of grease on piston & O-ring Clean & grease O-ring and piston
O-ring not correctly positioned or damaged Change the O-ring

Inaccuracies

Incorrect operation Follow instructions carefully
Calibration altered Recalibrate according to instructions
Unsuitable for the particular liquid pipetting technique Use correct pipetting technique
Instrument damaged Send to repair

Push button jammed or moves erratically

Piston contaminated Clean & Grease O-ring and piston
Penetration of solvent vapors

Tip ejector jammed or moves erratically

Tip cone contaminated from outside Remove ejector collar and clean tip cone's outer surface with ethanol

Volume setting is not properly click stopped

Click stop mechanism damaged Send for repair

Push button does not turn for volume setting

Use of excessive force beyond the range of pipette Send for repair

Service

If help is needed, contact with your supplier/manufacturer or Labbox via:
www.labbox.com

Please, provide the customer service representative with the following information:

  • Serial number (on the back side)
  • Description of the problem
  • Your contact information

www.labbox.com

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Easy 40+ Manual

Advertisement

Need help?

Need help?

Do you have a question about the 40+ and is the answer not in the manual?

Questions and answers

Table of Contents