Sutron Accubar 56-0133 Operation & Maintenance Manual

Constant flow bubble gauge / recorder, dual orifice bubble gauge / recorder
Hide thumbs Also See for Accubar 56-0133:
Table of Contents

Advertisement

Quick Links

Accubar ® Constant Flow
Bubble Gauge/Recorder 56-0133
and
Dual Orifice Bubble Gauge/Recorder
56-0134
O
&
PERATIONS
M
M
AINTENANCE
ANUAL
Part No. 8800-1167
Revision – 1.83.1
May 9th, 2023

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Accubar 56-0133 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sutron Accubar 56-0133

  • Page 1 Accubar ® Constant Flow Bubble Gauge/Recorder 56-0133 Dual Orifice Bubble Gauge/Recorder 56-0134 & PERATIONS AINTENANCE ANUAL Part No. 8800-1167 Revision – 1.83.1 May 9th, 2023...
  • Page 3: Table Of Contents

    AVIGATING THE ENUS ........................36 URNING ISPLAY ..............................36 ONTRAST ..........................37 IEWING URRENT ........................38 RONT ANEL SDI-12 SENSOR OPERATION .........................41 SDI-12 R ............................41 EFERENCE Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 4 APPENDIX A – SPECIFICATIONS FOR THE BUBBLER ..............85 56-0133-XX SINGLE ORIFICE BUBBLER ....................85 56-0134-XX DUAL ORIFICE BUBBLER ....................86 WATER TEMPERATURE ........................87 APPENDIX B– SUTRON CUSTOMER SERVICE POLICY ..............89 APPENDIX C– COMMERCIAL WARRANTY ..................90 SUTRON MANUFACTURED EQUIPMENT ..................90 NON-SUTRON MANUFACTURED EQUIPMENT ................90 REPAIR AND RETURN POLICY......................90...
  • Page 5 EXTENDED WARRANTY AND ON-SITE MAINTENANCE...............90 Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 6: Introduction

    Introduction The Sutron Accubar ® Continuous Flow (CF) Bubble Gauge/Recorder (part number 56-0133-25-1) s a self- contained, precision device for measuring water levels. The device combines into a single package a pump, tank, manifold, control board, front panel (display and keypad), Accubar sensor, SDI-12 and RS232 interfaces, and enclosure for the purpose of measuring water levels using long established bubble gauge principles.
  • Page 7: Dual Orifice Bubbler

    Dual Orifice Bubbler The Sutron Accubar ® Constant Flow (CF) Dual Orifice Bubble Gauge/Recorder (part numbers 56-0134-25- 1, -2 and -3) is an enhanced version that is capable of providing redundant measurements, measuring water level in two separate bodies of water, accurately measuring water density, or providing an extended range of water level measurement.
  • Page 8: Unpacking

    Retain one shipping container in the event a factory return is necessary. Please note that if a return is required, a return material authorization (RMA) number is required. To get this RMA number, call the Sutron Customer Service Department at 703 406 2800. Pneumatic Features Single Orifice Bubbler...
  • Page 9: Pneumatic Features Dual Orifice Bubbler

    Pump Desiccant Accubar Accubar Vent Line 1 Tank Vent Valve Line 2 Bypass Line 1 Bypass Line 2 Line 1 Line 2 Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 10: Cabling

    Cabling Terminal Block Terminal Block Description Notes Earth Earth DATA SDI-12 NOT USED SDI-12 RS485 A RS485 A RS485 B RS485 B +12V Power 12V (4 amps) Power Ground Power Connections The CF Bubbler requires external +12V power to operate. The most common power source for the Bubbler is a lead-acid battery.
  • Page 11: Db9 Connector

    SDI-12 over RS-485. The Bubbler does not support connections over both the RS485 and the SDI-12 lines at the same time. MODBUS over RS485 is NOT supported. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 12: Quick Install

    Quick Install • Mount the enclosure vertically • Install the orifice line(s) in the desired location(s) • Connect power • Do a leak test • Connect orifice line(s) to bubbler and Purge the line(s) • Customize settings: Setup the orifice configuration (Dual Orifice Bubbler only) Setup the bubble rate...
  • Page 13: Setup And Operation

    Bubble Rate The bubble rate is the measure of the amount of air going down the orifice line per unit of time. The Sutron Bubbler supports two units for bubble rate: Bubbles per minute (BPM) and Standard Cubic Centimeters per Minute (SCCM).
  • Page 14: Purge

    Pump Run Time The Bubbler tracks the cumulative pump on time. That is the total amount of time that the pump has been running. It is possible to reset the pump run time. Both the single and the Dual Orifice Bubbler are equipped with a single pump. The Dual Orifice Bubbler has two tanks and operates a valve in order to select which tank is being filled by the pump.
  • Page 15: Blockage Detection

    Sutron provides a cap with each unit that can be used to cap the outlet for the leak test. When the leak test completes, the system will display a status indicating whether the unit has passed or failed the leak test along with a score.
  • Page 16: Starting The Bubbler

    A Dual Orifice Bubbler can run leak tests on each line independently. Leak test information for each line is stored separately. To start a leak test, use Diagnostic > Leak Test, or LEAK TEST via command line. Starting the Bubbler The Bubbler starts operating as soon as power is applied.
  • Page 17: Accubar Pressure Sensor

    "zero". A sensor's accuracy depends completely on the stability of the zero and span. Any error in the zero becomes an error in the final water level. Even though sensors are calibrated at the factory for zero, they all Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 18 drift and introduce error into the water level. The CF Bubbler Accubar features an autozero ability that eliminates the errors due to a drift in zero resulting in more accurate water level readings. The Bubbler will automatically autozero the Accubar at the end of a purge. The autozero will take approximately half a minute to complete.
  • Page 19 The Bubbler can help determine the correct Orifice Separation: • While viewing a live density reading on the front panel, press SET. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 20 • If you Bubbler is equipped to measure water temperature, it will ask if the lines are in pure water. If so, the Bubbler can correctly determine the density based on water temperature, and it can compute the exact Orifice Separation. •...
  • Page 21: Active Line

    How does the Bubbler know which line is in deeper water? It needs the help of the user. The field calibration offset of the shallower line should be setup to be greater than that of the deeper line. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 22 Pressure is maintained in both lines at all times. Accubar sensors are protected by a relief valve from excessive pressure. The Bubbler is will switch the Accubar out of line to prevent it from being exposed to excessive pressure. No line switch will occur if a line is blocked (which can happen only if Blockage Detection is enabled).
  • Page 23 Accubar sensor. That being said, the temperature of the second sensor can still be viewed diagnostic purposes: use Diagnostic > Accubar w/o offset , SDI-12 M6. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 24 Automeasure Automeasure refers to the Bubbler’s ability to automatically measure and log water level data. The user can determine when this will occur by changing the automeasure interval and time. Automeasure may not be turned off. Automeasure time and interval determine when the Bubbler measures and logs data. In Normal mode (see Operating Modes below), this is when the Bubbler measures and logs data: •...
  • Page 25: Dqap

    In DQAP mode, the Bubbler is continuously collecting Accubar water level samples. The samples are collected at the rate of one sample per second or slower, depending on the averaging time. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 26 Averaging Time Sampling rate 3 minutes or less One sample per second 6 minutes One sample every two seconds 9 minutes One sample every three seconds 12 minutes One sample every four seconds 15 minutes One sample every five seconds Please note that during certain operations, such as a purge, the Bubbler will not be able to collect a sample.
  • Page 27: Water Temperature

    Celsius or Fahrenheit units, as dictated by the Temp Units setting. Water Temperature Sensors Two main types of water temperature sensors can be connected to the Bubbler: Platinum RTD NTC Thermistor Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 28 After wiring the sensor, make sure to tell the Bubbler what kind of sensor it is via the Water Temp Type setting. The following options are available: • Platinum 385 • Platinum 392 • Platinum Custom • NTC Thermistor Platinum RTD If Platinum 385 or Platinum 392 are chosen as the Water Temp Type, no further setup is needed.
  • Page 29: Suspended Sediment

    A secure flash chip in the Bubbler provides a logging capacity of more than 300,000 entries. Data will not be lost if power is removed. Once the log is full, the oldest data will be overwritten. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 30 Each log entry consists of • date and time (with a second resolution) • name (e.g. Water Level) • measurement reading (optional) • measurement quality and units (optional) Here are several examples of log entries: • Water Level,10/11/2006,10:00:00,3.08,feet, • Setup Change,10/10/2006,16:22:33, •...
  • Page 31: Sd Card Interface

    Setup and SD Cards It is possible to save the current setup to an SD card. SD Card Operations > Write Bubbler Setup to Card. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 32: Setup

    A setup saved to the SD card can be transferred to a PC using an SD card reader. A setup file can be edited using a text editor on a PC (such as Notepad). Once settings are changed, the file can be saved to the SD card and sent to the Bubbler, changing the Bubbler’s setup.
  • Page 33: Bubbler Time

    10.0V, the measurement will proceed, but the result will be marked with a battery low flag. The red LED will come on whenever the battery voltage is less than required, indicating that the battery should be changed. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 34: Password

    Password You can enable password protection by configuring a password. If password protection is enabled, the user is allowed view setup and data. However, no changes to setup will be allowed until a password is entered. A password prompt will automatically appear whenever a setup change is attempted. Via front panel, go to Station Setup->Other Setup->Password.
  • Page 35: Fasttrack

    FastTrack. The Bubbler calls the pressure drop in the tube deltaP. Please consult the factory before making changes to any of the FastTrack settings. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 36: Front Panel Interface

    Front Panel Interface The Bubbler features a two line LCD interface, six buttons and two LEDs. The front panel interface can be used to setup the station, examine it’s status, view the current measurements, and view logged data. Navigating the Menus The menu has a tree structure, like directories in an operating system.
  • Page 37: Viewing Current Data

    If there are errors, the message ‘Hardware Error’ will be displayed when the front panel is turned on. In that case, you may press RIGHT for details. You may then press SET to clear the errors. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 38: Front Panel Menu Tree

    Front Panel Menu Tree Errors (only show if errors are present) Hardware error details Stage (live) and time of reading -- Press SET to calibrate Stage 2/Density/Inactive Line Press SET to calibrate (Dual Orifice)* Tank Pressure (live) Water Temperature* Suspended Sediment* Battery Voltage (live)
  • Page 39 Log Accubar Pressure Log Battery Log DQAP Samples SD Card Operations Download Log Read Setup From SD Card Write Setup to SD Card Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 40 Format SD Card Diagnostics Purge Last Purge Results Leak Test Last Leak Test Results Accubar 2 Configuration (Dual Orifice only)* Accubar Pressure Diagnostic (PSI, no field cal offset, no autozero) Tank and Restrictor Pressure Pump Run Time Accubar ID Passive Sensor Mode Flow Tube (Fast Track)* deltaP...
  • Page 41: Sdi-12 Sensor Operation

    (dual separate, dual density). If it is not configured for two readings, it will return 3 as the validity of the second reading. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 42 The standard SDI-12 commands for the standard CF Bubbler are below. The reference for Dual Orifice Bubbler is in a separate table following this one. Name Command(s) Response Example/Notes Acknowledge Active Send Identification a13 SUTRON CFBBLR1.sssssss 1.33 1 indicates a Single Orifice Bubbler sssssss is the Accubar ® serial number 1.33 is the software...
  • Page 43 (0 if there are no errors), the total number of +total reset count resets, the results of the Accubar autozero, and +Accubar autozero the restrictor autozero. +restrictor autozero Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 44 Name Command(s) Response Example/Notes Acknowledge Active Send a13 SUTRON CFBBLR2.sssssss 1.33 The serial number Identification 2 indicates a Dual Orifice Bubbler reported is that of sssssss is the Accubar ® serial number the first Accubar 1.33 is the software revision sensor.
  • Page 45 (pg 14) +tank 2 pressure (pg 14) Purge Now aM9! Atttn This command 0D0! requests a purge be +duration done now. The Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 46 Name Command(s) Response Example/Notes +pumptime response to the M +success (if not zero, see hardware error table, pg 69) command will tell how long the purge will last. The data provided is the result of the last purge: duration is how long the purge took in seconds, pump time is how long the...
  • Page 47 1 or 2 indicating water level in 0XS2+7.87+0 line 2 or density. Do not Set water level in line Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 48 Name Command(s) Response Example/Notes place a + before k. 2 to 7.87 feet d is the desired water level Note: a D0 command and in the units (see issued after XS is pg 48) indicated by u. The complete will display Bubbler will adjust the the new offset in the field calibration offset to...
  • Page 49 DD is the day of the month present), or that the time has been (01 to 31) synced to the GPS in the last 12 hours Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 50 Name Command(s) Response Example/Notes HH is the hour (military (if a GPS is present). time 0 to 23) g is a single digit that indicates the MM is the minutes presence of the GPS. 0 means no GPS SS is the seconds is present, 1 means that a GPS is present.
  • Page 51 For example, to read the current bubble rate, issue command XXS+1+5! and follow it up with a D0! command. The reply to D0 will have the bubble rate. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 52 Any settings that allow for a negative value can be set using a ‘-‘ as a delimiter, such as changing the user offset. XXS+1+24-5.5 would change the offset to –5.5 Changing the station name can use either a ‘+’ or ‘-‘ delimiter and may contain spaces: XXS+1+9+New Name+20 would change the station name to “New Name”, and the Purge Pressure to 20.
  • Page 53: Rs232 Command Line Interface

    LAST. The output by the Bubbler will look like this: Accubar Reading Water Level 7.01 feet Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 54: Downloading The Log

    Internal Temp 23.38 C Internal Sensors Restrictor Pressure 31.259 PSI Tank Pressure 31.263 PSI For a concise version, try !LAST or !MEAS; 7.01, 23.38 31.259, 31.263, Please note that if additional features are enabled, the output will have additional data: •...
  • Page 55: Auto Output

    Xpert or some such computer. E.g. “MEAS” will show Accubar reading Pressure 6.41 feet Internal Temp 23.91 C “!MEAS” will show Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 56 6.41, 23.91 NOTE: No individual setup commands are documented here. Every setup variable can be viewed by typing its name. E.g. “BUBBLE RATE” will show the current rate. Every setup variable can be changed by typing its name = new value. E.g.
  • Page 57 MEAS + Initiates, waits for, and shows the results of sensor measurements. That includes Accubar readings as well as Bubbler internal sensor. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 58 Can be followed by a setup variable name and a new value for that variable. E.g. “CHANGE STATION NAME = SUTRON” If SETUP DEFAULT is issued, it will reset the entire setup to defaults. STATUS 0 Shows system status including time, boot time, battery readings, last measurements, current internal sensor readings, and any hardware errors that may exist.
  • Page 59 Type SETUP to get a list of the whole setup. Every setup variable can be viewed by typing its name. E.g. “STATION NAME” will show the current station name. Every setup variable can be changed by typing its name = new value. E.g. “STATION NAME = SUTRON” will set the station name to “SUTRON”. AUTO OUTPUT AUTOMEASURE INTERVAL...
  • Page 60: Setup Transfer Via Hyperterminal

    LOG TEMP IN BOX MEAS NAME MODBUS BAUDRATE MODBUS DEVICE ID MODBUS ENABLE MODBUS PARITY MODBUS PROTOCOL OPERATING MODE ORIFICE CONFIG* ORIFICE SEPARATION* PURGE LENGTH PURGE PRESSURE RS232 WAKEUP SDI DATA OLD SDI ADDRESS SEPARATION UNITS STATION NAME STEIN STEIN STEIN TEMP UNITS TUBE ID...
  • Page 61 Set J4 to pins 1-2 to have the modem wake up on RTS Set J5 to pins 1-2 to route CD out the DSR pin The default settings from Sutron for the modem will work. If the settings have been changed, issue these commands to the modem: AT&F...
  • Page 62 UDP Auto Answer You can connect the Raven to the same battery powering the Bubbler; however, remember that it will increase the power consumption (both when the modem is idle and when it is connected). As a result, you will need to make sure your battery is large enough to provide the power needed by the station. - 62 -...
  • Page 63 SDI-12 over RS-485. The Bubbler does not support connections over both the RS485 and the SDI-12 lines at the same time. MODBUS over RS485 is NOT supported. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 64: Installation

    The Bubbler OUTLET pressure fitting accommodates an 3/8" OD tubing that will run down into the water. The 3/8” tubing provided by Sutron can be installed directly on the ground, buried in the ground or in a conduit. When running the tube to the water, try to ensure that the line is run with a continuous downward slope.
  • Page 65: Electrical Connections

    Leaks inside the Bubbler can be a source of inaccuracy and/or excessive pumping and use of desiccant. To check for leaks, you must cap the outlet or orifice and run the built-in leak test routine. Sutron provides a cap with each unit that can be used to cap the outlet for the leak test. When the leak test completes, the system will display a status indicating whether the unit has passed or failed the leak test along with a score.
  • Page 66: Factory Calibration

    separation between the orifices. The following table gives the density of pure water based on temperature. Remember that line 2 must be below line 1 for the system to properly measure the density. Temp Density .9999 1.000 .9997 .9991 .9982 .9970 .9957 The density of Seawater is generally between 0.1020 to 0.1030.
  • Page 67 If an estuarine site must be used, specific gravity measurements should be taken and corrections applied. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 68: Troubleshooting And Maintenance

    Troubleshooting and Maintenance Troubleshooting The following checklist will help in troubleshooting problems: Problem Possible Cause Run leak Test. Observe error codes and refer to “Hardware Pressure Error Codes” section to determine source of problem. dropping/ excessive pump operation Make sure orifice lines are fully attached and tight on the Bubbler orifice ports.
  • Page 69 Diagnostics = enabled. This will cause the system to write diagnostic information to the log. Let the system operate for at least 10 minutes and then retrieve the log and analyze it to better understand the problem. You may also e-mail the log file to Sutron’s customer service department to help understand how the system is operating.
  • Page 70: Hardware Error Codes

    Hardware Error Codes The Bubbler may record one of the following error values during its operation. The error codes may be returned in response to some SDI-12 commands, or may be shown on the front panel. Error Code Description 4001 restrictor sensor failure 4002 tank sensor failure...
  • Page 71: Maintenance

    Run a leak test if possible to verify there are no internal leaks. Download the data and events. Examine the log for any abnormal readings or events. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 72 Since the calibration data is stored in the small sensor module, you can freely swap or interchange Accubar sensors into different BUBBLERS. Sutron recommends an annual check of the Accubar calibration. - 72 -...
  • Page 73 Accubar Replacement The Accubar can be easily replaced. After acquiring a new sensor from Sutron, power down the Bubbler, open the front panel, unplug and remove the old Accubar and replace it with the new Accubar. Bringing the Benefits of Real-Time Data Collection to the World...
  • Page 74: Firmware Upgrade

    FIRMWARE Upgrade Sutron’s CF Bubbler has been designed using the most modern techniques such that at any time the system firmware may be upgraded while it is in the field preventing the need to ever return a unit to the factory for firmware upgrades.
  • Page 75 • Once the download is completed, the system will reboot. • Type the command ‘Ver’ to confirm that the upgrade was successful. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 76: Gps

    -5 hours. If a Sutron Satlink (versions 6.17 and newer) is connected to the Bubbler via SDI-12 (in addition to the Garmin GPS), note that Satlink will set the clock of the Bubbler once per day. Please ensure that Satlink’s local time offset is equal to that of the Bubbler.
  • Page 77: Gps Positioning

    Jumper in question is J19, and is located next to the RS232 connector. Place a connector on jumper J19 so that it connects 12v to the middle pin (pins 1 and 2) Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 78: Rj45 To Rs232 Connector

    A custom connector is required to get the Sutron and the Garmin GPS together. The connector bridges the RJ45 on the Garmin GPS to the RS232 on the Sutron unit. The table below provides the wiring diagram for the connector.
  • Page 79: Modbus

    Note: Care must be taken to make sure the selected protocol matches that of the client or there will be communication problems. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 80: Modbus Function Codes

    Modbus Parity Default is Even. Available choices include Even, Odd and None. Note: Care must be taken to make sure the selected parity matches that of the client or there will be communication problems. Modbus Delay before Tx Default is 10ms. This identifies the number of milliseconds to wait after asserting CTS before starting data transmission.
  • Page 81: Holding Registers

    1 value of type float starting at register 1. If the quality is also desired, change the number of values to 2 and choose ushort for the second reading type. The complete list of registers and their locations are below. Data Register Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 82 Last measured water level (MSW) Last measured water level (LSW) Quality of water level (see below) Last measured water level in PSI (does not apply field calibration offset or autozero) (MSW) Last measured water level in PSI (does not apply field calibration offset or autozero) (LSW) Quality of water level (see below)
  • Page 83 The [numbytes,data] represents an additional record of data if there is room in the message. Example: command: GL,data,02/07/2007 15:50:00,80,0 Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 84 reply: GLR,0,0,38,02/07/2007,15:51:00,VBAT,13.16,Volts,G 37,10/07/2003,15:51:10,A,10.89,5.2,-25.4 command: GL,data,02/07/2007 15:50:00,80,2 reply: GLR,0,2,37,02/07/2007,15:54:00,C,10.89,5.2,-25.4,0 The GLR response will contain as many log records as can fit into the response. The numbytes value in the GLR response does not include the comma preceding the data, and refers to the number of data bytes from the log that are being returned, not the number of packet bytes used to store the response (which would be twice the data bytes when ASCII protocol is selected).
  • Page 85: Appendix A - Specifications For The Bubbler

    10.5” x 14” x 7.5” (Compact model is 8.5” x 10.5” x 7”) Dimensions 3/8” Tube fitting Pressure Outlet Standard CFB Compact CFB Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 86: 56-0134-Xx Dual Orifice Bubbler

    Environmental Temperature -25C to +60C optional -40C to +60C (P/N: BBL-ET) Humidity 0-95% Non-condensing 56-0134-XX DUAL ORIFICE BUBBLER Electrical Power Required 8-16VDC Average Power <15 mA @12V for a 3 second measurement every 15 minutes, no purge. <30 mA @12V for 120 second measurement every 5 minutes with 30 second purge every day.
  • Page 87: Water Temperature

    Types Supported Platinum 385, Platinum 392, Thermistor, Custom Sensor Operational -40°C to +60°C Range The Following specifications only apply to specific Sutron sensors 5600-0100-100 & 5600-0030-1-050 Accuracy +0.1°C (-5°C to +45°C) Resolution 0.01°C * references to ft based on USGS conversion factor of 2.3073 ft water per PSI (pure water @ 50deg C.)
  • Page 88 Single Orifice Ordering Information Part Number Description 56-0133-25-1 Accubar Constant Flow Bubble (CFB) Gauge, 25 PSI max (17.5m/57ft) 56-0133-25-1C Accubar CFB Gauge, 25 PSI max, 6mm orifice lines 56-0133-25-1-T Accubar CFB Gauge, 25 PSI max, with water temperature capability 56-0133-25-1P Accubar CFB Gauge, 25 PSI max, without enclosure 56-0133-25-1S Compact CFB Gauge, 25 PSI max (17.5m/57ft)
  • Page 89: Appendix B- Sutron Customer Service Policy

    Of course not all problems can be fixed over the phone. Sometimes a customer needs an on-site technician to identify site related problems or troubleshoot a network. Sutron can provide these services at a reasonable cost. Call for details. If you would like to learn more about Sutron products email sales@sutron.com...
  • Page 90: Appendix C- Commercial Warranty

    The above Warranty applies only to products manufactured by Sutron. Equipment provided, but not manufactured by Sutron, is warranted and will be repaired to the extent of and according to the current terms and conditions of the respective equipment manufacturers.
  • Page 91 ............14 turning on/off ..........36 Modbus .............. 79 download log............30 modem ............... 60 DQAP ..............25 Raven ............. 61 Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164...
  • Page 92 Modem Data Old ............41 Xpert .............. 61 operation ............41 mounting ............64 reference ............41 validity ............41 sensors Accubar ............17 normal operating mode ........24 internal ............14 NTC Thermistor ..........28 restrictor ............14 tank ..............14 setting water level ..........

This manual is also suitable for:

Accubar 56-0134

Table of Contents