Summary of Contents for Lilliput Owon ADS3000 Series
Page 1
ADS3000 Series User Manual ADS3102/3104 ADS3202/3204 ADS3352/3354 ADS3502/3504 For product support, visit:www.owon.com.hk/download ※:The illustrations, interface, icons and characters in the user manual may be slightly different from the actual product. Please refer to the actual product.
Page 2
The information in this manual was correct at the time of printing. However, LILLIPUT will continue to improve products and reserves the rights to change specification at any time without notice.
General Warranty We warrant that the product will be free from defects in materials and workmanship for a period of 3 years from the date of purchase of the product by the original purchaser from our company. The warranty period for accessories such as probes is 12 months.
Table of Contents 1. General Safety Requirements ............1 2. Safety Terms And Symbols ..............3 How To Conduct A General Inspection .................5 How To Conduct Function Inspection ................... 6 3. Primary User Guide ................7 A General Knowledge Of The Structure Of The Instrument ..........8 Front Panel ........................8 Rear Panel ........................
Page 5
How To Set FFT ..................... 85 How To Set DIR(Digital Filtering) ................ 89 How To Set Pass Fail .................... 91 How To Set Counter ....................93 How To Set DVM ....................95 How To Set Decode ....................97 How To Set Others Modulation ..................107 How To Set Display System ................107 How To Save And Print ..................
Page 6
7. Appendix ..................147 Appendix A: Enclosure ......................147 Appendix B: General Care And Cleaning ................ 147...
1.General Safety Requirements General Safety Requirements Before use, please read the following safety precautions to avoid any possible bodily injury and to prevent this product or any other connected products from damage. In order to avoid any contingent danger, ensure this product is only used within the range specified.
Page 8
1.General Safety Requirements Do not operate without covers. Do not operate the instrument with covers or panels removed. Use Proper Fuse. Use only the specified type and rating fuse for this instrument. Avoid exposed circuit. Do not touch exposed junctions and components ...
2.Safety Terms And Symbols Safety Terms And Symbols Safety Terms Terms in this manual. The following terms may appear in this manual: Warning: Warning indicates the conditions or practices that could result in injury or loss of life. Caution: Caution indicates the conditions or practices that could result in damage to this product or other property.
Page 10
2.Safety Terms And Symbols To avoid body damage and prevent product and connected equipment damage, carefully read the following safety information before using the test tool. This product can only be used in the specified applications. Warning The four channels of the oscilloscope are not electrically isolated. The channels should adopt a common ground during measuring.
2.Safety Terms And Symbols Warning: To avoid fire or electrical shock, when the oscilloscope input signal connected is more than 42V peak (30Vrms) or on circuits of more than 4800VA, please take note of below items: Only use accessory insulated voltage probes and test lead. ...
2.Safety Terms And Symbols responsible for this service or our local offices. 3. Check the Complete Instrument. If it is found that there is damage to the appearance of the instrument, or the instrument can not work normally, or fails in the performance test, please get in touch with our distributor responsible for this business or our local offices.
3.Primary User Guide Primary User Guide This chapter elaborates the following topics: A general knowledge of the structure of the instrument A general knowledge of the user interface of the instrument How to conduct probe compensation How to set the probe attenuation coefficient ...
3.Primary User Guide A General Knowledge Of The Structure Of The Instrument This chapter gives a brief description and introduction to the operations and functions of the front panel of the instrument, so as to facilitate your operations of the instrument in the shortest time. The following are illustrated with four channels as an example, and for two channels, please refer to four channels.
Page 15
3.Primary User Guide set the value. Arrow keys: Move to select the parameter. 3. Home key: Return to the main homepage. 4. Touch key: Press it to disable the touch screen,the key light turns on;and press it again to enable the touch screen,the key light turns off (Note: The Touch Lock is only available in EduInstr system and does not work on the main homepage and other application interface.).
Page 16
3.Primary User Guide “Horizontal Scale” knob is to control time base scale. When pressing the knob can enable/disable zoom mode. 7. AFG control area(Optional). 8. Probe compensation: 3.3V/1kHz signal output. 9. External trigger interface. 10.Signal input port control area. 11.Microphone port. 12.USB Host Interface: When the oscilloscope is connected to an external USB device as a “master device”, the USB Host interface is used to transmit the data.
3.Primary User Guide Rear Panel Figure 3-2: Rear Panel 1. Foldable Handle. 2. Heat Emission Hole. 3. USB-C interface. 4. USB Host Interface: When the oscilloscope is connected to an external USB device as a “master device”, the USB Host interface is used to transmit the data.
3.Primary User Guide 8. LAN Interface: The network interface to connect a PC or router. 9. Foot Rest: To adjust the inclined angle of the oscilloscope. 10.Fuse. 11.Power outlet. 12.Safety lock. 13.AFG CH2 interface. 14.AFG CH1 interface. A General Knowledge Of The User Interface Of The Instrument Figure 3-3:Illustrative Drawing of Display Interfaces 1.
Page 19
3.Primary User Guide Wireless&networks setting interface. 4. U disk access identifier. 5. System set time, click the icon will switch to the Date&time setting interface. 6. Counter, DVM and others function information display bar (Note: click on the left corner corresponds to enable/disable statistic). Right swipe the information display bar to close the corresponding function.
3.Primary User Guide Oscilloscope Inspection 1. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel. Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe with rotating it to the right side. Connect the probe tip and the ground clamp to the connector of the probe compensator.
3.Primary User Guide Compensation" on P14), and connect the probe with the CH1 channel. If a probe hook tip is used, ensure that it keeps in close touch with the probe. Connect the probe tip with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector, and then push the Autoset button on the front panel.
3.Primary User Guide Caution: The default attenuation coefficient of the probe on the instrument is preset to 10X. Make sure that the set value of the attenuation switch in the probe is the same as the menu selection of the probe attenuation coefficient in the oscilloscope.
3.Primary User Guide the instrument and connect the ground terminal to the earth. How To Conduct Self-Calibration Self-calibration program is used to quickly make the oscilloscope be in the optimum condition to obtain the most accurate measurement. This program can be performed at any time. It is especially necessary when the ambient temperature reaches or exceeds 5℃.
4.Use the Android System Use the Android System Android System Homepage Window 1. Application shortcut key. If you click the instrument shortcut key , you can enter the oscilloscope interface. 2. App Drawer (Click to see all apps). 3. Task key. 4.
5.Use the Oscilloscope Use the Oscilloscope A General Knowledge Of Oscilloscope A General Knowledge Of Trigger System As shown in Figure , there are one knob and one key. The following exercises are to guide you through the use of the trigger system. Figure 5-1: Trigger control area 1.
Page 27
5.Use the Oscilloscope Figure 5-2: Vertical control area 1. Vertical Position knob to control the vertical display position of the signal. When turning the Vertical Position knob, the pointer indicating the Grounding Reference Point of the channel moves up and down following the waveform.
5.Use the Oscilloscope Turn the Vertical Scale knob to change the Vertical Scale Factor (Voltage Scale), and the scale factor of corresponding channel in the information display bar changes accordingly. 3. Press CH1、CH2、CH3、CH4 key to enable or disable the corresponding channel.
5.Use the Oscilloscope then the corresponding Horizontal Time Base in the status bar changes accordingly. Press the Horizontal Scale knob enter or exit waveform zoom mode. How To Use Touch Screen Control The touch screen can be used to control the oscilloscope by various gestures.
5.Use the Oscilloscope and down with the finger to scroll the list, as shown in the figure below. Open Main Menu: Click the icon in the lower right of the display area, the main menu window pops up, as shown below. Click each item in the main menu window to open the settings window of the corresponding function, and click the shortcut key at the top of the screen to open the corresponding function.
Page 31
5.Use the Oscilloscope Set the vertical position of the selected channel waveform (Vertical Position knob): The vertical position of the waveform can be changed by swiping your finger up and down in the blank position of the waveform display area, as shown in the figure below. ...
Page 32
5.Use the Oscilloscope Set Horizontal Position (Horizontal Position knob): The horizontal position of the waveform can be changed by swiping your finger around the waveform display area, as shown in the figure below. The control voltage gear and time base can be scaled in the following ...
5.Use the Oscilloscope In the waveform display area, double-click the screen and slide the hand up and down/left and right to zoom the control voltage scale and time base. Operate The Touch Screen In Waveform Amplification Mode Press Horizontal Scale knob to enter the waveform zoom mode, the main window is displayed at the top half of the screen and the amplified window is displayed at the bottom half of the screen.
5.Use the Oscilloscope Other Touch Screen Operations Click and drag the open menu item to move itself to the appropriate location.
Page 35
5.Use the Oscilloscope Control the horizontal or vertical cursor lines under cursor measurement, as shown in the figure below. Run/Stop: Click in the upper left of the display area to switch Run/Stop. Parameter Setting Keyboard in Menu Item: There are digital input mode and hobbing input mode.
Page 36
5.Use the Oscilloscope Hobbing input mode, as shown in the figure below. Swipe the information display bar...
5.Use the Oscilloscope Advanced User Guide for Oscilloscope In the previous chapter, basic operations of the oscilloscope, function area of the front panel and the roles of the keys and knobs are introduced for the user to determine the change of instrument settings by observing the status bar.
5.Use the Oscilloscope CH4, Math and Ref) and two knobs (Vertical Position knob and two Vertical Scale knob). Channel Settings Each channel is equipped with the independent vertical menu, and each item is set separately based on different channel. How to enable or disable the waveform (channel) (1) Press CH1, CH2, CH3, CH4 front key produces the following results: ...
Page 40
5.Use the Oscilloscope Click to open or close the waveform Inverted inversion function. Click the Numeric Display Box and turn General knob or swipe the screen up and down in the numeric select box with the Common finger to select commonly used magnification;...
Page 41
5.Use the Oscilloscope According to requirement to set offset div、Current Unit unit. Vertical position Offset Set the vertical display position of the settings waveform; due to limited screen display, the visible range is ±5div. Channel 1 is taken as an example to set the channel, and the operation steps are as follows: 1.
Page 42
5.Use the Oscilloscope channel operation menu (see “How to Set the Probe Attenuation Coefficient” in Page 15). If the probe attenuation coefficient is 1:1, the input channel ratio of the channel shall also be set to 1X, so as to avoid errors in the displayed gear factors and measured data.
Page 43
5.Use the Oscilloscope Click All. The high-frequency components contained in the measured signal can be passed through. 7. Set input impedance (1) Click the CH1 information display bar to bring up the CH1 Settings window. (2) Click Display switch in the menu to highlight it. (3) Select the oscilloscope input impedance in Input Impedance option.
5.Use the Oscilloscope See "How To Set Reference Waveform" on page 123 for the Ref key. Application of Vertical Position knob and Vertical Scale knob 1.Vertical Position knob to adjust the vertical position of corresponding channel waveform. The resolution of this control key changes with the vertical scale. 2.Vertical Scale knob to adjust the vertical resolution of corresponding channel waveform.
5.Use the Oscilloscope 2. Horizontal Scale knob: Set the horizontal scale factor for the main window or amplified window. The descriptions of horizontal system menu are shown in the table below: Menu Settings Descriptions Zoom Mode Click to open/close zoom mode. Through the navigation function, observe the situation of moving waveform.
Page 46
5.Use the Oscilloscope In normal mode, the Horizontal Position knob and the Horizontal Scale knob are used to adjust the horizontal position and the horizontal time base of the main window. In waveform amplification mode, the Horizontal Position knob and the Horizontal Scale knob are used to adjust the horizontal position and the horizontal time base of the amplified window.
5.Use the Oscilloscope (Play right) can change the play direction, and it stops playing when it reaches the leftmost or rightmost end. can close the navigation setting window. (6)Click Note: For the operations related to the touch screen,refer to “How to use touch screen control”...
5.Use the Oscilloscope number of segment acquisition, and click < > or press to move the cursor and select the digit to be set. Click , to select the length to be recorded 100k Depth on the right display box. Note: The length of the collected record is dynamic, changing with the acquire mode.
5.Use the Oscilloscope There is one knob and one keys in the Trigger Control Area. Trigger Level: Turn this knob to set the signal voltage of corresponding trigger point and press this knob to make the trigger level at the vertical midpoint of the trigger signal amplitude. Forced Trigger: Generate a trigger signal forcibly, mainly used in the “Normal”...
Page 50
5.Use the Oscilloscope Set Channel 2 as the signal source trigger signal. Set Channel 3 as the signal source trigger signal. Set Channel 4 as the signal source trigger signal. Set the external trigger input channel as the signal source trigger signal. EXT/5 Set the external trigger source divided by 5 to extend the external trigger level range.
5.Use the Oscilloscope Video Trigger Select the Video Trigger to trigger on standard video signal field or line of 525i/NTSC,625i/PAL or SECAM. Enter the video trigger and the trigger setting information is displayed in the lower part of the screen, such ,indicating that the video trigger is selected with the trigger information source of CH1 and synchronization type of line.
5.Use the Oscilloscope Auto Set to collect waveform even when no trigger condition is detected. Normal Set to collect waveform only when trigger Mode conditions are satisfied. Single Set to acquire a waveform when one trigger is detected and then stop acquisition. Pulse Trigger Pulse trigger is to determine the trigger moment based on the pulse width.
5.Use the Oscilloscope 100ns – 10s; click Numeric Input Box to input the interval to be set for restarting the trigger circuit and click the unit to confirm; or click Gear Input Box (- or +) or turn the General knob to set HoldOff the interval for restarting the trigger circuit, and click <...
5.Use the Oscilloscope Click to set the slope conditions, click time setting’s Numeric Input Box to input the slope > time to be set and click the unit to confirm; or = Time click Gear Input Box (- or +) or turn General knob <...
Page 55
5.Use the Oscilloscope The descriptions of runt trigger setting window are shown in the table below: Menu Settings Descriptions Set the trigger type of the vertical channel to Type Runt under-amplitude trigger. Set Channel 1 as the signal source trigger signal. Set Channel 2 as the signal source trigger signal.
5.Use the Oscilloscope Click Numeric Display Box and turn General knob to set the lower threshold; Lower Click 50% and set the shortcut key of trigger Threshold level in the vertical midpoint of the trigger signal amplitude. 100ns – 10s; click Numeric Input Box to input the interval to be set for restarting the trigger circuit and click the unit to confirm;...
5.Use the Oscilloscope Positive over-amplitude pulse . Polarity Negative over-amplitude pulse. Enter: Triggers when the trigger signal enters the specified trigger level range. Exit: Triggers when the trigger signal exits the specified trigger level range. Conditions Time: Specify the hold time of the input signal after entering the specified trigger level.
Page 58
5.Use the Oscilloscope falling (or rising) edge passes through the touch level is larger than the set timeout period. Enter the timeout trigger and the trigger setting information is displayed at the lower part of the screen, such as , indicating that the timeout trigger type is selected with 0.000pV of the trigger level value.
5.Use the Oscilloscope Auto Set to collect waveform even when no trigger condition is detected. Normal Set to collect waveform only when trigger Mode conditions are satisfied. Single Set to acquire a waveform when one trigger is detected and then stop acquisition. Nth Edge Trigger The oscilloscope triggers on the Nth edge that appears on the specified idle time.
5.Use the Oscilloscope Trigger on the rising edge of the input signal when voltage level meets the specified trigger level. Slope Trigger on the falling edge of the input signal when voltage level meets the specified trigger level. Set the time before starting the edge counting in the Nth edge trigger.
Page 61
5.Use the Oscilloscope Determine the trigger conditions with the logical relations.Enter the logic trigger and the trigger setting information is displayed at the lower part of the screen, such as , indicating that the logic trigger type is selected with the logic mode of AND, input mode of high level and output mode of Goes True.
Page 62
5.Use the Oscilloscope Click to select the output mode. GoseTrue: Trigger when condition turns True from False. GoesTrue GoseFalse: Trigger when condition turns False GoesFalse Output from True. True> Mode True >: Trigger when the time of true condition is True= more than the set time.
Page 63
5.Use the Oscilloscope Auto Set to collect waveform even when no trigger condition is detected. Normal Set to collect waveform only when trigger Mode conditions are satisfied. Single Set to acquire a waveform when one trigger is detected and then stop acquisition. RS232/UART Trigger RS232/UART bus is a serial data communication mode used for data transmission between computers or between the computer and the terminal.
Page 64
5.Use the Oscilloscope Select reverse polarity for data transmission. Common Click Numeric Display Box and turn General knob to set the commonly-used baud rate. Custom Click Numeric Display Box to input the baud rate to be set and click unit to confirm; or click Gear Baud Input Box (- or +) or turn the General knob to set the baud rate, and click <...
Page 65
5.Use the Oscilloscope Single Set to acquire a waveform when one trigger is detected and then stop acquisition. I2C Trigger I2C serial bus consists of two lines, SCL and SDA, with the transmission rate determined by the clock line SCL and transmission data by SDA, as shown in the figure, it can be triggered upon start, restart, stop, loss confirmation and specific equipment address or data value.
Page 66
5.Use the Oscilloscope knob to set the SDA threshold; Click 50% and set the shortcut key of trigger level in the vertical midpoint of the trigger signal amplitude. Trigger when SCL is in high level and SDA is Start from high level to low level. Trigger when another starting condition occurs Restart before the stopping condition.
5.Use the Oscilloscope Normal Set to collect waveform only when trigger conditions are satisfied. Single Set to acquire a waveform when one trigger is detected and then stop acquisition. SPI Trigger Trigger when the timeout conditions are met and the specified data is found by the oscilloscope.
5.Use the Oscilloscope the SCL signal remains idle for a specified time; Click Numeric Input Box to input the timeout period to be set and click the unit to confirm; or click Gear Input Box (- or +) or turn General knob to set the timeout period, click <...
Page 69
5.Use the Oscilloscope trigger type is selected with the trigger signal source of CH1, frame type of CAN_H and the baud rate of 1,000,000bps. The descriptions of CAN trigger setting window are as follows: Menu Settings Descriptions Type Set the bus trigger type as CAN. Set Channel 1 as the signal source trigger signal.
Page 70
5.Use the Oscilloscope Overload Standard Select format Format standard or extended. Extend Use the General knob and arrow keys on the panel to set Value the ID value required. Click Numeric Display Box and Byte select the byte length required Length for the set data, ranging from 1 Data...
5.Use the Oscilloscope LIN Trigger The LIN bus data frame format is shown in the figure below: Trigger with LIN bus based on signal Break,ID, ID/data and Data Error. The signal source and signal rate specified by LIN is required. Enter the LIN buss trigger and the trigger setting information is displayed at the lower right of the screen, such as , indicating that the LIN...
5.Use the Oscilloscope Use General knob and arrow keys on the panel to set the ID valued required. Use the General knob to specify the Byte ID/Data length of the data in bytes, ranging from 1 Length to 8. Use General knob and arrow keys on the Data panel to set the ID valued required.
Page 74
5.Use the Oscilloscope CH1 signal source being selected. It indicates that the current measuring type is added. It indicates that the current measuring type is not added. It indicates the display box of current measuring value type and measuring result display box. The currently selected measurement type is interpreted.
Page 75
5.Use the Oscilloscope Figure 5-6 Rise Time: Time that the leading edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude. Fall Time: Time that the falling edge of the first pulse in the waveform takes to fall from 90% to 10% of its amplitude.
Page 76
5.Use the Oscilloscope Figure 5-7 Vavg: The arithmetic mean over the entire waveform. Vpp: Peak-to-Peak Voltage. VRMS: The true Root Mean Square voltage over the entire waveform. Overshoot: Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms. Vmax: The maximum amplitude. The most positive peak voltage measured over the entire waveform.
Page 77
5.Use the Oscilloscope +PulseCnt : The number of positive pulses that rise above the mid reference crossing in the waveform. -PulseCnt : The number of negative pulses that fall below the mid reference crossing in the waveform. RiseCnt : The number of positive transitions from the low reference value to the high reference value in the waveform.
Page 78
5.Use the Oscilloscope Note for the following measurements: When source A in the menu is set to CH<n>, source A is CH<n>. When source B in the menu is set to CH<n>, source B is CH<n>. Delay( - ): The time difference between the rising edge of source A and the rising edge of source B at the middle of the threshold.
Page 79
5.Use the Oscilloscope Of which,PhaseA F B F is phase( - ),DelayA F B F is delay( - ), Period sourceA is source A period. Phase( - ): The phase difference between the rising edge of source A and the falling edge of source B at the middle value of the threshold is expressed in degrees.Calculation formula is: Of which,PhaseA R B F is phase( - ),DelayA R B F is delay( - ),...
Page 80
5.Use the Oscilloscope LFF: Time between Source A first falling edge and Source B last falling edge. Statistics Click Statistics in the setting window, as shown below. Click switch to open or close the statistics display window. This instrument supports statistics and displays the current value of a number of measurement results, as shown in the figure below.
Page 81
5.Use the Oscilloscope Horizontal Range: Selecting Screen indicates that the measurement range is the whole screen; selecting Cursor indicates that the measurement range is only within the cursor range. Top/Base: Set the measurement method for the top and bottom values of the amplitude.
5.Use the Oscilloscope keyboard and also can rotate the corresponding general knob to set value. When the current limit value is set to greater than or equal to the current median value, the interface prompts "exceed limit value", and the instrument automatically adjusts the lower limit value to make it lower than the median value.
5.Use the Oscilloscope to highlight it. (2) Select Full Screen to On status to open the full screen view of XY mode. How To Set Cursor Measurement Press Cursor key, click the main menu window to select Cursor option or click Cursor Shortcut at the upper part of the screen and then click the information display bar in the right of the screen, the cursor information display bar as shown in the figure below.
Page 84
5.Use the Oscilloscope Enable or disable the cursor measured Follow Line value moving with the cursor line. Selected Waveform Select the waveform channel to be Source measured by the cursor. Math Display time measurement cursor and menu; Display voltage measurement cursor and Type menu;...
Page 85
5.Use the Oscilloscope 4. Set Cursor Type Click X in the type menu to highlight it, and two yellow dotted lines X1 and X2 are displayed in the vertical direction of the screen; Click Y in the type menu to highlight it, and two yellow dotted lines Y1 and Y2 are displayed in the horizontal direction of the screen;...
Page 86
5.Use the Oscilloscope Cursor Measurement in Zoom Mode To perform a Zoom cursor measurement, perform the following steps: 1. Click Zoom shortcut at the top of screen, press Horizontal Scale knob or click the Horizontal Information Display Bar to enter the zoom mode. 2.
Page 87
5.Use the Oscilloscope Degrees in Vertical Units . 2. Press Cursor key to pop up the cursor measurement setting window. The cursor information display bar located to the right of the waveform display area displays the cursor readings. 3. In setting window,select Window as Main or FFT,can make the cursor line appear in the main waveform area or FFT waveform area.
5.Use the Oscilloscope 4. Set Cursor Type Click X in the type menu to highlight it, and two yellow dotted lines X1 and X2 are displayed in the horizontal direction of the screen; Click Y in the type menu to highlight it, and two yellow dotted lines Y1 and Y2 are displayed in the horizontal direction of the screen;...
Page 89
5.Use the Oscilloscope The descriptions of Math setting window are shown in the table below: Menu Settings Descriptions Switch Open or close the waveform mathematics. Basic Waveform calculation of simple addition, subtraction, multiplication and division for CH1, CH2, CH3 and CH4; Type Advanced Advanced waveform calculation for CH1, CH2,...
Page 90
5.Use the Oscilloscope 1. Click in the right corner of screen,then select Math ,the screen will pop up math setting window. 2. Click Switch to highlight it,and the pink waveform M will display on the screen. 3. Click Basic to highlight it. 4.
5.Use the Oscilloscope 5. Create the expression, then select Confirm in the keyboard to implement. 6. Click Display. When the switch label is highlighted on the right, it is enabled. The formula will be displayed at the lower left corner of the screen.
Page 92
5.Use the Oscilloscope frequency domain, which corresponds to the standard time domain graph of the oscilloscope. Then match these frequencies with known system frequencies, such as the system clock, oscilloscope or power supply. The FFT operation of this instrument can convert 8192 data points of the time domain waveform into frequency domain signal (the recording length of the acquisition shall be set to 512K points or above).
Page 93
5.Use the Oscilloscope FFT operating steps are as follows: 1. Click in the right corner of screen,then select FFT , the screen will pop up FFT setting window. 2. Click Switch when the switch label is highlighted on the right, it is enabled. The pink waveform M will be displayed on the screen (It is also available to click FFT shortcut softkey at the upper part of the screen).
Page 94
5.Use the Oscilloscope There are 6 FFT windows. Each one has trade-offs between frequency resolution and amplitude accuracy. What you want to measure and your source signal characteristics help you to determine which window to use. Use the following guidelines to select the best window. Type Descriptions Window...
5.Use the Oscilloscope are very close to the same value but have widely differing amplitudes (the side lobe level and shape factor are closest to the traditional Gaussian RBW). This window is also good for random signals. Notes for using FFT ...
Page 96
5.Use the Oscilloscope 1. Click DIR in the analysis module from the main menu window at the lower right of the screen. 2. In the setting window,click Switch when the switch label is highlighted on the right, it is enabled. The pink waveform M will be displayed on the screen.
5.Use the Oscilloscope to move the cursor and select the digit to be set. The voltage gear of the Math Waveform is the same as that of the current channel. Note: In slow sweep, the digital filter function is disabled. How To Set Pass Fail Click Pass Fail in the analysis module from the main menu window at the...
Page 98
5.Use the Oscilloscope Read Click to print the saved test rule. Pass/Fail: It detects whether the input signal of the channel is within the rules. If it is out of range, it is a failure; otherwise, it is a pass. It can output failure or pass signal through the built-in, configurable output port.
5.Use the Oscilloscope add it, doesn't start at 0. How To Set Counter To perform counter, follow these steps: 1. Click Counter in the analysis module in the main setting window at the bottom right of the screen The setting window will display on the screen.
Page 100
5.Use the Oscilloscope 6. Click Reset Stat., the historical data of the counter will be cleared and the statistics will be re-conducted.
5.Use the Oscilloscope How To Set DVM To perform DVM, follow these steps: 1. Click DVM in the analysis module in the main setting window at the bottom right of the screen The setting window will display on the screen. 2.
Page 102
5.Use the Oscilloscope 6. Click Reset Stat., the historical data of the DVM will be cleared and the statistics will be re-conducted. 7. Click Zoom box when the switch label is highlighted on the right, it is enabled. The zoom box will appear at the top right of the screen, and the number will match the number in the list on the right.
5.Use the Oscilloscope 8. Alarm: In the Upper Limit or Lower Limit option, click Numeric Input Box to set the upper or lower limit vale; In When Limit to set the limit condition to In Limit or Out Limit; Set the switch whether to turn on the alarm sound.
Page 104
5.Use the Oscilloscope bottom right of the screen 2. The setting window will display on the screen. Click Decode1 or Decode2 to set decode. Click Switch, when the switch label is highlighted on the right, it is enabled. 3. Select RS232/UART, I2C, SPI, CAN or LIN in Type option. Click Copy Trig, the settings for the trigger type can be copied.
Page 105
5.Use the Oscilloscope oscilloscope. (2) Adjust to the proper time base and voltage division. (3) In trigger menu, select trigger type as RS232/UART, set parameters based on the characteristics of the signal, trigger the signal correctly and obtain stable display. Refer to “RS232/UART Trigger” on P61. (4) After the signal is stabilized and triggered,click Decode in the analysis module in the main setting window at the bottom right of the screen...
Page 106
5.Use the Oscilloscope Click Numeric Display Box and turn General Common knob to set the commonly-used baud rate. Click Numeric Display Box to input the baud rate to be set and click unit to confirm; or click Gear Input Box (- or +) or turn the General knob to set the baud rate, and click <...
5.Use the Oscilloscope (4) After the signal is stabilized and triggered, click Decode in the analysis module in the main setting window at the bottom right of the screen Select the type as I2C, set parameters based on the characteristics of the signal.
Page 108
5.Use the Oscilloscope When read and write bits are not included, they With are added automatically depending on the Without function. DECIMAL Format Select the display format to decode. BINARY ASCII Click Switch, when switch label is Event highlighted on the right, it is enabled.The decode Table list will display on the screen.
Page 109
5.Use the Oscilloscope Select CH1,CH2,CH3 or CH4 as CLK. Click Numeric Display Box and turn General knob to set the lower threshold; Click 50% and set the shortcut key of trigger level in the vertical midpoint of the trigger signal amplitude.
Page 110
5.Use the Oscilloscope Select reverse polarity for data transmission. Set the clock edge to rising edge or falling edge. ClockEd The rising edge refers to acquire the data at the rising edge of the clock; the falling edge refers to acquire the data at the falling edge of the clock.
Page 111
5.Use the Oscilloscope the signal. When the parameters are set correctly, the information carried by the signal will be displayed. Note: Error Frame, Remote Frame, and Overload Frame will be identified in the "Data" column in the event table (Data Frame will not be identified). The descriptions of CAN decode setting window are as shown in the table below: Menu...
Page 112
5.Use the Oscilloscope DECIMAL Format Select the display format to decode. BINARY ASCII Switch, Click when switch label Event highlighted on the right, it is enabled.The decode Table list will display on the screen. Click Label, when the switch label is highlighted Common on the right, it is enabled.You can select Table...
5.Use the Oscilloscope Click Numeric Display Box and turn General knob to set the lower threshold; Threshold Click 50% and set the shortcut key of trigger level in the vertical midpoint of the trigger signal amplitude. Click Numeric Display Box and turn General Common knob to set the commonly-used baud rate.
Page 114
5.Use the Oscilloscope shown in the table below: Menu Settings Descriptions Only display the acquisition points. Point Type Vector filling displays the space between Vector adjacent acquisition points in the middle. Close 1Second Select the time of duration. 2Seconds Note:Currently support CH1, CH2, CH3, CH4, Persist 5Seconds FFT, XY, DIR, waveform operation models.
Page 115
5.Use the Oscilloscope Transparency Drag the slider to the right of Window Transparency item to set window brightness. The adjustable range is 0% to 100%. Afterglow: When afterglow function is used, the afterglow display of image tube can be simulated. The color of the original data retained is gradually lightened and that of new data becomes brighter.
5.Use the Oscilloscope How To Save And Print Click Save in the others module in the main setting window at the bottom right of the screen Save oscilloscope waveform and picture, set to USB internal memory by operating Save in the setting window; operate Print to print the image displayed on the oscilloscope screen.
Page 117
5.Use the Oscilloscope Save the image by editing the file name or the Name system default file name. Save Save current waveform. When the type is Set, the menu shows as following: User0 Set storage location. User9 Save the current parameter settings of the Save oscilloscope to the internal memory.
Page 118
5.Use the Oscilloscope 2、In official website install print app,steps as shown in Figute1 to Figure 5. ①Click Download to download installation package, as shown in Figure 1. Figure 1 ②Click Open to installation package, as shown in Figure 2.
Page 119
5.Use the Oscilloscope Figure 2 ③Click INSTALL,start to install print app, as shown in Figure 3. Figure 3 ④Software install produce,as shown in Figure 4.
Page 120
5.Use the Oscilloscope Figure 4 ⑤After install success, click open can open print app. As shown in Figure 5. Figure 5 3、Open the installed printing software, as shown in Figure 6. Go to the last page and check "Agree" license agreement and privacy policy. Click "I agree" to start using the software, as shown in Figure 7.
Page 121
5.Use the Oscilloscope Figure 6 Figure 7 4 、 After entering the printing software interface, click "Enable Wi-Fi to see nearby printers" to connect the printer's Wi-Fi, as shown in Figure 8. The Wi-Fi connection is successful, as shown in Figure 9.
Page 122
5.Use the Oscilloscope Figure 8 Figure 9 5、In the menu below, set the printing parameters. Click to open Inverse , the image will be printed with a white background. Click to open Time, the image will display specific printing time of the image. Select printing area: Full Screen or Waveform Area.
Page 123
5.Use the Oscilloscope the print interface, as shown in the picture below. 6、Connect the instrument to the printer,as shown in the following picture. 7、Click can set the parameter including: Copies, Paper Size, Color, Orientation, Two-sided and Pages. After setting, click the print icon to print the image, as shown below.
Page 124
5.Use the Oscilloscope Note: Image printing can only be printed over a network connection, USB connection is not valid. When printing, turn off borderless printing in More options. If WIFI and network cable are connected at the same time, the printer may not be found, do not connect at the same time.
Page 125
5.Use the Oscilloscope 3. Click Disk Management menu, and information about the USB disk will display on the right side with red mark 1 and 2. Figure 5- 9: Disk Management of computer 4. Right click 1 or 2 red mark area, choose Format. And system will pop up a warning message, click Yes.
Page 126
5.Use the Oscilloscope the warning message. Figure 5- 11: Formatting the USB disk setting 6. Formatting process. Figure 5- 12:Formatting the USB disk Check whether the USB disk is FAT32 with allocation unit size 4096 after formatting. Use Minitool Partition Wizard to format Download URL: http://www.partitionwizard.com/free-partition-manager.html Tip: There are many tools for the USB disk formatting on the market, just take...
Page 127
5.Use the Oscilloscope 1. Connect the USB disk to the computer. 2. Open the software Minitool Partition Wizard. 3. Enter the app interface and information about the USB disk will display on the right side with red mark 1 and 2. Figure 5- 13:Reload Disk 4.
Page 128
5.Use the Oscilloscope Figure 5- 15:Format setting 6. Click Apply at the top left of the menu. Then click Yes on the pop-up warning to begin formatting. Figure 5- 16:Apply setting 7. Formatting process.
5.Use the Oscilloscope Figure 5- 17:Format process 8. Format the USB disk successfully. Figure 5- 18: Format successfully How To Set Reference Waveform 100 reference waveforms can be stored in the instrument, which can be displayed with current waveform simultaneously. The stored waveform can not be adjusted after being called.
Page 130
5.Use the Oscilloscope To store CH1 channel waveform to waveform0, operate according to the following steps: 1. Open CH1 channel. 2. Click Reference in the others module in the main setting window at the bottom right of the screen 3. The setting window will display on the screen.Click Switch when the switch label is highlighted on the right, it is enabled.
5.Use the Oscilloscope When the display switch is closed, the label disappears and the reference waveform displayed on the screen will be hidden accordingly. 8. Click Label to select a common type or a Custom type. When you select a common type, you can select 31 types of labels. When you select a Custom type, you can click the input box below to enter the required labels.
5.Use the Oscilloscope screen.Click Start to perform the probe check. After completing the probe check, the check results are displayed on the screen and click Quit. If the result is undercompensation or overcompensation, please refer to “How to Implement the Probe Compensation”...
Page 133
5.Use the Oscilloscope How To Set Up a Network Discovery Service 1. Click Network in the others module in the main setting window at the bottom right of the screen 2. Click mDNS when the switch label is highlighted on the right,it is enable. LXI software can be used for network discovery services;...
Page 134
5.Use the Oscilloscope software to query the instrument. 3. Please download your own software with mDNS lookup function. Now the LXI Discovery Tool software is taken as an example to explain, and the detailed operation is as follows: (1) Open LXI app, enter the app interface,as shown in the following picture. (2) Click on the Settings in the top left corner and select "Advance Views"...
Page 135
5.Use the Oscilloscope (3) Enter "Advance Settings Views" interface, as shown below. (4) Under the“Advance Settings Views”interface, select mDNS,mDNS Service Type select“_http._tcp”, as shown below. (5) After setting, click Search, available devices can be searched, and the IP address of the device and the corresponding instrument description are displayed,as shown below.
5.Use the Oscilloscope (6) Select your own instrument, click the corresponding instrument description or click "Open Web Page" in the lower right corner to jump to the network service login interface, as shown below. Default Restore the factory settings. Click Default shortcut in the left corner of screen and click Confirm in the factory settings window, to restore the factory default state;...
5.Use the Oscilloscope About It is composed of About the Application and About the Instrument. The former is to display the latest version of the instrument; and the latter is to show the instrument Product Model, Serial Number, System Version and CheckSum.
Page 138
5.Use the Oscilloscope Synchronous output is generally used to synchronize trigger signals or measurement results with other devices or systems for more accurate and comprehensive measurement, analysis, and control. The output types of the synchronous output of this instrument are trigger output and pass/fail, which can help the user better control and analyze the trigger and validity of the measurement process.
5.Use the Oscilloscope Foundation Setting Window time: Set window display time, the value can be set Close, 5s, 10s, 15s, 20s, 25s, 30s. When the set time is up, the setting window will automatically close. Beeper: Click Beeper when the switch label is highlighted on the right, it is enabled.
Page 140
5.Use the Oscilloscope Function Item Settings Vertical Coupling DC(channel coupling remains closed) Signal open or close (channel switch Channel switch remains closed) Vertical Scale Adjust to the proper scale Channel Bandwidth Current Horizontal Position Center or two squares to left or right Horizontal Scale Adjust to the proper scale Trigger Type...
Page 141
5.Use the Oscilloscope DC level: cancel autoset, auxiliary menu set. Unknown source: cancel autoset , auxiliary menu set. Partial description of nouns: Signal period:Display 1~2 waveform periods. Multi periods:Display multiple waveform periods. Rising edge:Separate shows a rising edge of square wave. Falling edge:Separate shows a falling edge of square wave.
Page 142
5.Use the Oscilloscope Note: If the automatic waveform setting is applied, the frequency and amplitude of the measured signal shall not be less than 20Hz and 5mv, respectively. If the condition is not met, the automatic waveform setting may be invalid.
6.Technical Specifications Technical Specifications Unless otherwise stated, all technical specifications apply to digital oscilloscope with the attenuation switch of the probe set to 10X. The instrument must be operated continuously for more than thirty minutes under the specified operating temperature. ...
6.Technical Specifications Decode the I2C bus addresses (including or excluding the read/write bit), data, and ACK. Source Channel: CH1~CH4 Decode SPI bus MISO/MOSI data (4 to 32 bits). The mode supports timeout and chip select (CS). Source Channel: CH1~CH4 Decode remote frames of the CAN bus at speeds up to 1 Mb/s (ID, byte count, CRC), as well as overload frames and data frames (standard/extended ID, control field, data field, CRC, ACK).
6.Technical Specifications Command Characteristics Instruction Common Supports the standard SCPI command set support Error message Error Message Definition Support status reporting Status Reporting mechanism Support synchronization Synchronization mechanisms General Technical Specification Display Characteristic Instruction 10.1 inch Colored LCD (Liquid Crystal Display) Display Type Display 1024 (Horizontal) ×600 (Vertical) Pixels...
6.Technical Specifications Others Characteristics Instruction HDMI;USB device *1, USB Host *3 ;Trig Out(P/F); Communication Interface Type-C power supply interface ;LAN interface Power Supply 100V – 240 VACRMS, 50/60 Hz, CAT Ⅱ Power < 35W Consumption Fuse 2A,T level,250 V Touch Screen Multi-touch capacitive touch screen Description: [1]: The adapter or battery is required to support a handshake protocol of 12V,...
7.Appendix Appendix Appendix A: Enclosure (The accessories subject to final delivery.) Standard Accessories: Power Cord Quick Guide USB Cable Probe Probe Adjust Appendix B: General Care And Cleaning General Care Do not store or leave the instrument where the liquid crystal display will be exposed to direct sunlight for long periods of time.