Summary of Contents for Buchi F-360

  • Page 1 Operation Manual Recirculating Chiller F-360...
  • Page 2 CH-9230 Flawil 1 E-Mail: quality@buchi.com BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experience, especially with respect to structure, illustrations and technical details. This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for competitive purposes, nor made available to third parties.
  • Page 3: Table Of Contents

    Establishing electrical connections .................... 17 5.3.1 Connecting a mains plug ..................... 17 Installing the hoses .......................... 18 Filling and draining the coolant tank .................... 18 5.5.1 Filling the coolant tank ...................... 18 5.5.2 Draining the coolant tank ..................... 18 Operation Manual Recirculating Chiller F-360...
  • Page 4 Resetting the fuse ........................ 23 Taking out of service and disposal.................... 24 Taking out of service..........................  24 Disposal .............................  24 Returning the instrument ........................ 24 Appendix ............................ 25 10.1 Spare parts and accessories ...................... 25 10.1.1 Spare parts .......................... 25 Operation Manual Recirculating Chiller F-360...
  • Page 5: About This Document

    1.3 Connected devices In addition to these operating instructions, follow the instructions and specifications in the documentation for the connected devices. Operation Manual Recirculating Chiller F-360 5/26...
  • Page 6: Safety

    Safety-related incidents that occur while using the instrument should be reported to — the manufacturer (quality@buchi.com). BUCHI service technicians Service technicians authorized by BUCHI have attended special training courses and are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair measures. 6/26...
  • Page 7: Warning Notices In This Document

    The following warning symbols are displayed in this operation manual or on the instrument. Symbol Meaning General warning Dangerous electrical voltage Heavy item Explosive substances Flammable substances Health-harming or irritant substances Hot surface Device damage Dangerous gases Corrosive Cuts by sharp edges Operation Manual Recirculating Chiller F-360 7/26...
  • Page 8: Residual Risks

    Always wear appropriate personal protective equipment such as safety goggles, protective clothing and gloves. Make sure that the personal protective equipment meets the requirements of the safety data sheets for all chemicals used. 8/26 Operation Manual Recirculating Chiller F-360...
  • Page 9: Modifications

    Unauthorized modifications can effect safety and lead to accidents. Use only genuine BUCHI accessories, spare parts and consumables. Carry out technical changes only with prior written approval from BUCHI. Only allow changes to be made by BUCHI service technicians. BUCHI accepts no liability for damage, faults and malfunctions resulting from unauthorized modifications.
  • Page 10: Product Description

    3.1.1 Process overview Fig. 1: Process overview Circulation pump Heat exchanger Heat exchange container Compressor Thermostatic expansion valve In-line valve Filter dryer 10/26 Operation Manual Recirculating Chiller F-360...
  • Page 11: Configuration

    3.2.1 Front view Fig. 2: Front view Display (see Chapter 3.2.2 "Dis- Air in for ventilation play", page 12) Navigation control Castor wheels Coolant level indicator (see Chap- Air out for ventilation ter 3.2.3 "Fill level indicator", page 12) On-/Off masterswitch Operation Manual Recirculating Chiller F-360 11/26...
  • Page 12: Display

    Indicates connection to the BUCHI vacuum controller Actual temperature Options activated if control knob is pressed Set temperature locked Set temperature 3.2.3 Fill level indicator Max fill level mark Min fill level mark Current fill level indication 12/26 Operation Manual Recirculating Chiller F-360...
  • Page 13: Rear View

    Product description | 3 3.2.4 Rear view Fig. 3: Rear view COM port Filling connection for coolant Power cable Slots for air flow out Main coolant flow IN Castor wheels Main coolant flow OUT Drain of coolant Operation Manual Recirculating Chiller F-360 13/26...
  • Page 14: Type Plate

    Accessories are delivered as per the purchase order, order confirmation, and delivery note. 3.4 Technical data 3.4.1 Recirculating Chiller F-360 Dimensions (W x D x H) 650 x 850 x1300 mm Weight 220 kg Cooling capacity at 15 °C* 6000 W 14/26 Operation Manual Recirculating Chiller F-360...
  • Page 15: Ambient Conditions

    3.5.2 Fill level sensor The instrument has an integrated fill level sensor. The fill level sensor checks whether the instrument is filled with the minimum amount of coolant. See also Chapter 3.2.3 "Fill level indicator", page 12. Operation Manual Recirculating Chiller F-360 15/26...
  • Page 16: Transport And Storage

    Empty the coolant tank before storing the instrument. After storage, check the instrument and all tubing for damage and replace if necessary. 4.3 Moving the instrument Release the castor breaks. Move the instrument to the designated place. Lock the castor breaks. 16/26 Operation Manual Recirculating Chiller F-360...
  • Page 17: Installation

    Have the installation carried out by an electrician or a person with similar expert knowledge. Carry out the installation according to the instructions. See Guide for electrical installation 5.3.1 Connecting a mains plug Precondition: The instrument is delivered with a mains cable connected to the instrument. Operation Manual Recirculating Chiller F-360 17/26...
  • Page 18: Installing The Hoses

    Place a suitable receptacle below the drain tap (see location of the tap in chapter Chapter 3.2 "Configuration", page 11). Open the drain tap and allow coolant to run into the receptacle. Wait until the the cooland drained. Close the drain tap. 18/26 Operation Manual Recirculating Chiller F-360...
  • Page 19 BÜCHI Labortechnik AG Installation | 5 Connect other devices to the instrument. See Chapter 5.4 "Installing the hoses", page 18. Operation Manual Recirculating Chiller F-360 19/26...
  • Page 20: Operation

    The padlock symbol appears on the display. ð To cancel the lock function, press and hold the navigation control. The padlock symbol disappears. ð 6.4 Shutting down the instrument Switch the On/Off master switch to Off. 20/26 Operation Manual Recirculating Chiller F-360...
  • Page 21: Cleaning And Servicing

    Users may only carry out the servicing and cleaning operations described in this section. Any servicing and repair work which involves opening up the casing may only be carried out by BUCHI service technicians. 7.1 Regular maintenance work Action Additional information...
  • Page 22: Help With Faults

    Allow it to cool down. Clean the air intake (see Chapter 7.2 "Cleaning Temperature error the housing", page 21). Switch the instrument on again. If the problem cannot be rectified contact BUCHI Customer Service. Coolant tank empty Switch off the instrument. or level too low, Allow it to cool down.
  • Page 23: Resetting The Fuse

    Disconnect the instrument from the power supply. Open the housing with the provided tool. Reset the fuse marked with F1 by pushing it up. The fuse is reset and in its upper position. ð Operation Manual Recirculating Chiller F-360 23/26...
  • Page 24: Taking Out Of Service And Disposal

    When disposing, observe the disposal regulations of the materials used. Materials used see Chapter 3.4 "Technical data", page 14. 9.3 Returning the instrument Before returning the instrument, contact the BÜCHI Labortechnik AG Service Department. https://www.buchi.com/contact 24/26 Operation Manual Recirculating Chiller F-360...
  • Page 25: Appendix

    Appendix | 10 10 Appendix 10.1 Spare parts and accessories Use only genuine BUCHI consumables and spare parts in order to ensure correct, safe and reliable operation of the system. NOTE Any modifications of spare parts or assemblies are only allowed with the prior written permission of BUCHI.
  • Page 26 11594550 | A en We are represented by more than 100 distribution partners worldwide. Find your local representative at: www.buchi.com...