Table of Contents

Advertisement

Quick Links

16-Port Manifold
User Manual
Including A0311, A0311-S, A0310
Picarro Inc.
3105 Patrick Henry Drive
Santa Clara, CA 95054, USA
Phone: +1 408 962 3944
www.picarro.com
Document Number 40-0038 Revision G

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A0311 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for PICARRO A0311

  • Page 1 16-Port Manifold User Manual Including A0311, A0311-S, A0310 Picarro Inc. 3105 Patrick Henry Drive Santa Clara, CA 95054, USA Phone: +1 408 962 3944 www.picarro.com Document Number 40-0038 Revision G...
  • Page 2 DISCLAIMER AND RESERVATION OF RIGHTS Picarro has prepared this manual solely for the information and use by its customers as a guide for the selection, installation, operation, and maintenance of the products described.
  • Page 3: Contact Information

    Contact Information General Technical Support: Email: support@picarro.com Phone: + 1 408 962 3991 European Technical Support: Email: support@picarro.com Phone: +31 85 888 1650 Customer Service: Email: orders@picarro.com Phone: + 1 408 962 3992 A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 4: Table Of Contents

    General Safety ..................... 17 Unpacking ......................18 Shipping, Handling and Storage ..............18 Inspect the Shipping Box ................18 Unpack the A0311 or A0310 Shipping Boxes ..........18 Unpack the A0311-S Shipping Box .............. 20 Recirculation Pump ..................21 Installation ......................22 Installation Safety ..................
  • Page 5 Table of Contents Using the Channel Selector Hand-Pad – A0311 .......... 45 Using the Channel Selector Hand-Pad – A0311-S ........45 Calibration, Flow Pathways ................47 Calibration ....................47 Single-port Operation ................... 49 Recirculation with the A0310 ................50 Reducing Position-to-Position Carryover ............. 50 Sweeping Out Headspace ................
  • Page 6 List of Figures List of Figures Figure 1: A0311 Front and Back Panels (Two A0311 Used for A0310) ......11 Figure 2: A0311-S Front and Back Panels ..............12 Figure 3: A0311 or A0310 Accessory Kit ............... 19 Figure 4: Splitter Cable ....................19 Figure 5: A0311-S Accessory Kit ..................
  • Page 7 List of Figures Figure 28: Hand-Pad Channel Selector – A0311-S ............46 Figure 29: Simplified Calibration Check Setup for A0311 ..........47 Figure 30: Showing Valve Sequencer Window ............. 48 Figure 31: Starting Valve Sequencer from Action Menu..........49 Figure 32: Setup for Eliminating Position-to-Position Carryover ........
  • Page 8 List of Tables List of Tables Table 1: A0311 and A0310 Specifications ..............13 Table 2: A0311-S Specifications ................... 14 Table 3: Acronyms, Formulas, Units, and Symbols ............15 Table 4: Warning/Information Icon Types ..............16 Table 5: Valve Sequencer Validation Steps ..............48 Table 6: Valve Sequencer Setup for Headspace Characterization .......
  • Page 9: Introduction

    These Picarro products comply with all required CE specifications. Intended Use The A0311 16-Port Distribution Manifold samples up to 16 gas sources in any order specified by the user. During operation, the selected line is routed through the valve into the analyzer. The 15 lines that are not selected terminate in the valve.
  • Page 10 • Unselected gas streams terminate at the valve. • Broadly compatible with all Picarro analyzers with the exception of those with known surface and chemical compatibility issues (such as the G2103, G2108, G2204, G2205, G2307, and PI2114). A0311-S 16-Port Distribution Manifold •...
  • Page 11: Figure 1: A0311 Front And Back Panels (Two A0311 Used For A0310)

    1/8” Manifold Manifold Inlets Inlets Hand Pad Connector On/Off Switch, 1/8 inch Swagelok Fuse, and Power to Analyzer Receptacle DB9 to Analyzer COM2 Figure 1: A0311 Front and Back Panels (Two A0311 Used for A0310) A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 12: Figure 2: A0311-S Front And Back Panels

    Hand Pad Connector 3/8 inch Swagelok to Vacuum Pump 1/4” Manifold 1/4 inch Inlets Swagelok to Analyzer DB9 to Analyzer On/Off Switch, COM2 Fuse, and Power Receptacle Figure 2: A0311-S Front and Back Panels P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 13: A0311 And A0310 Specifications

    Introduction A0311 and A0310 Specifications Table 1: A0311 and A0310 Specifications Parameter Specification Number of Gas Inlets Sample Handling Materials Stainless Steel Temperature Range Storage: 0 °C to 70 °C (32 °F to 158 °F) Operation: 0 °C to 35 °C (32 °F to 95 °F) Ambient Humidity Range <...
  • Page 14: A0311-S Specifications

    ± 10% of nominal voltage Fluctuation Fuse Ratings 1.5 A at 250 VAC, slow blow 5 mm x 20 mm (0.196 in. x 0.78 in.) Software Controlled by Picarro Analyzer Software Valve Sequencer Interface. Liquid Ingress Protection None P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 15: Acronyms

    Bold Italic text identifies section reference links. • Bold text is for actions to take (such as clicking on a UI button), caution and warning statements, and text you should type or select in screens. A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 16: Safety

    2. Safety Warning Symbols Icon notes and warnings are used in Picarro manuals. The purpose of these icons is to provide a visual convention to alert you of important information. They indicate dangers to either the operator or to the analyzer, and other important information.
  • Page 17: General Safety

    CE label. This CE label is located on the rear of the instrument. Using a Picarro manifold in a manner not specified by Picarro may result in damage to the unit and render it unsafe to operate.
  • Page 18: Unpacking

    Do not stack boxes more than five high. Inspect the Shipping Box Picarro products are inspected and tested before leaving the factory. The shipping boxes provide proven safety from most dropping, crushing or spiking events. If the equipment arrives damaged, photograph the damage and contact Picarro (email pictures if possible) for consultation on best course of action.
  • Page 19: Figure 3: A0311 Or A0310 Accessory Kit

    NOTE Figure 3 shows the accessories kit minus the power cable. Figure 3: A0311 or A0310 Accessory Kit Users who purchase an A0310 dual manifold kit will also receive a splitter cable (Figure 4) for the analyzer that allows the user to control both manifolds directly from the same communications port on the analyzer.
  • Page 20: Unpack The A0311-S Shipping Box

    (See warning in section 4.1, Installation Safety). CAUTION https://www.swagelok.com/en/catalog/Product/Detail?part=SS-400-P The A0311-S contains a universal power supply that accepts a power source range from 100-240 VAC, 50/60 Hz, 100 VA max. NOTE Figure 5 shows the accessories kit minus the power cable.
  • Page 21: Recirculation Pump

    While most Picarro instruments will run with one A2000 pump on the downstream side of the instrument (exception for this manual: A0310 configurations), the A0311-S kit comes with a second A2000 pump which should be connected to the A0311-S itself, following guidance in section 4.4, Connection Procedure A0311- S.
  • Page 22: Installation

    The instruments must be supported by a shelf or additional rails attached to the rack. If the A0311-S is mounted separately, it can be mounted using a front rack WARNING mount kit alone without additional support.
  • Page 23: Confirm Manifold Receives Line Power

    PERSONAL INJURY: When using the A0311-S with hazardous gases, cap all unused ports. In the event of a failure of the A0311-S vacuum pump, or if the user fails to turn on the pump, gas flowing from active ports could flow out of the unused/uncapped positions.
  • Page 24 16 unconnected and uncapped. Connect the A0311 power cable to line power. Turn the manifold ON using the On/Off switch on the A0311 back panel. With A0311 powered on, start the pump and then analyzer. This lets the analyzer recognize the newly installed hardware.
  • Page 25: Figure 8: A0311 Setup Schematic

    Installation Figure 8: A0311 Setup Schematic A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 26: Figure 9: A0311 Or A0310 Back Panel Connections

    Manifold Inlets Hand Pad Connector Figure 10: Hand Pad Connector Port on A0311 Front Panel Next, follow the steps outlined in section 5, Initialize the Sequencer (Windows) or section 6, Initialize the Sequencer (Linux), as appropriate for your operating system.
  • Page 27: Connection Procedure A0311-S

    Installation Connection Procedure A0311-S A schematic for A0311-S connections and gas flows is shown below. Figure 11: A0311-S Setup Schematic A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 28: Figure 12 A0311-S Manifold Rear Panel Connections

    See Figure 12 for the manifold back panel connectors. With A0311-S and analyzer turned off, attach the RS-232 cable from the A0311-S COM MPV port (the DB9 connector on the back panel) to the analyzer serial port (COM2 port). Remove any/all caps from the lines.
  • Page 29: Connection Procedure A0310

    Installation Connection Procedure A0310 A schematic for A0310 connections and gas flows is shown below. A Picarro A0702 Recirculation Pump is used with this setup. Refer to the A0702 User Manual (Picarro PN 40050) for detailed instructions on setup and use of this pump.
  • Page 30: Figure 15: A0310 Recirculation Setup, Front View

    To connect a Picarro analyzer to the A0310 recirculation manifold kit: With A0311 and analyzer turned off, stack the two A0311s on top of each other, on top of the analyzer (Figure 15). The upper manifold is the Sample Out Selector Manifold, while the bottom is the Sample Return Selector Manifold.
  • Page 31: Figure 16: A0310 Recirculation Setup, Rear View

    A0311 inlet positions 1 through 16 unconnected and uncapped. Connect the recirculation pump to the vacuum line on the instrument following the steps below. Customers may also want to consult the A0702 manual (Picarro PN 40050) for further guidance. It is very A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 32: Figure 17: Tightening Vacuum Adapter Fitting

    Using the 11/16” wrench, tighten 1/4” to 1/2” turn. When connecting the vacuum line 3/8” nut to the vacuum port on the Picarro analyzer, ensure that the adaptor union sits straight when tightened in place. See Figure 18 (left and middle panes) for a reference for a correct fitting alignment and see Figure 18 (right pane) for an example of a poor alignment.
  • Page 33: Figure 18: Vcr To Analyzer Adapters - Ensure Proper Alignment And

    Always use a new gasket when reconnecting VCR fittings. A used gasket can be identified by a circular depression on the surface: CAUTION Vacuum VCR Female Fitting New 1/4” Ni/Ag Gasket Figure 19: Inserting New Gasket into VCR Female Fitting A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 34: Figure 20: A0311 Back Panel Connections

    Users who purchased their A0310 kit prior to mid-2022 will need to source an adaptor to connect the 1/4” fitting on the adaptor to the 1/8” fitting on the manifold, as Picarro did not provide this piece until recently. The relevant part numbers for this adapter (a single 1/8”...
  • Page 35 “1”. If the hand pad doesn’t iluminate, or if the valve doesn’t turn, refer to section 12, Troubleshooting. 17. With both A0311 powered on, but only the bottom manifold connected to the instrument COM port, start the pump and then analyzer. This lets the analyzer recognize the newly installed hardware.
  • Page 36: Initialize The Sequencer (Windows)

    5. Initialize the Sequencer (Windows) Verify Analyzer Recognized Valve Automatically With the instrument measuring, a manifold connected and turned on, navigate to the Picarro host software. Click on the Tools menu, and select Show/Hide Valve Sequencer (Figure 22). This should open the valve sequencer interface in the background.
  • Page 37: Figure 23: External Valve Sequencer Window

    Initialize the Sequencer (Windows) A0311 and A0311-S users may now proceed to section 7, External Valve Sequencer Window. Shaded green means manifold is successfully connected and communicating with the analyzer Figure 23: External Valve Sequencer Window A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 38: Initialize The Sequencer (Linux)

    Confirm the valve is connected into COM2 on the back of the analyzer, and then restart the analyzer. The Distribution Manifold should now be initialized. A0311 and A0311-S users may now proceed to section 7, External Valve Sequencer Window. P/N 40-0038 Rev. G...
  • Page 39: External Valve Sequencer Window

    The External Valve Sequencer window serves a triple purpose. It can be used for the rotary (also known as multiport or multiposition) valve covered in this document; for controlling up to six valves in front ends like the Picarro A0314 Small Sample Introduction Module (SSIM) for leak testing; or for controlling additional 12V valves, typically 3-way solenoids for switching between gas sources.
  • Page 40: File Menu

    Load a Valve Sequence File Click here to load saved files. For Windows OS, the sequence files are in: C:\Picarro\G2000\InstrConfig\ValveSequencer\name of sequence file For Linux OS, the sequence files are in: /home/picaro/SI2000/InstrConfig/ValveSequencer/name of sequence file Save a Valve Sequence File...
  • Page 41: External Valve Sequencer Window Features

    5. Use the Rot. Valve Code column to define the steps of a run. 6. Sets the total number of steps available. Figure 26: External Valve Sequencer Field Descriptions A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 42 Picarro analyzers. This cable comes with six Molex connectors labeled V1-V6, and with equivalent connectors and pins for the mating fittings on solenoid valve cables. Picarro is able to quote and provide 3-way solenoid valves with either 1/4” fittings (S3112) or 1/8”...
  • Page 43: Set Up A Sequence Using The Valve Sequencer Window

    A variety of sequences can be created and saved using the External Valve Sequencer window. This section explains how to set up a sequence using the Sequencer window. To set up a sampling sequence: On the Picarro analyzer GUI, go to: • Tools •...
  • Page 44: Load And/Or Run A Saved Sequence

    Enter the step duration in the Duration (min) field, where the duration of the step is in minutes. If duration values are set to less than 0.1 minutes, they may not be carried out correctly. Save the valve sequence as: C:\Picarro\G2000\InstrConfig\ValveSequencer\Name of the sequence file (Windows) /home/picaro/SI2000/InstrConfig/ValveSequencer/name of the sequence file (Linux) Start sequence by hitting Apply, and then Start Sequence from the Action menu.
  • Page 45: Using The Channel Selector Hand-Pad - A0311

    Pressing the STEP and HOME buttons at the same time moves the actuator to the previous position. Figure 27: Hand-Pad Channel Selector – A0311 or A0310 Using the Channel Selector Hand-Pad – A0311-S The channel selector hand-pad shown in Figure 28 overrides the GUI set valve sequence.
  • Page 46: Figure 28: Hand-Pad Channel Selector - A0311-S

    External Valve Sequencer Window Figure 28: Hand-Pad Channel Selector – A0311-S P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 47: Calibration, Flow Pathways

    A0311. On the A0311, a single position is selected at a time (Figure 8), and all other positions dead-end in the valve. In the A0311-S (Figure 11), a single position is selected to go to the instrument, and is pulled in by the instrument pump, while all other 15 positions are flushed continuously by a second pump connected to the back of the A0311-S manifold box.
  • Page 48: Figure 30: Showing Valve Sequencer Window

    2, and that the Remaining Time (min) entry begins counting down from 10 minutes. If the sequencer does not start, consult section 12, Troubleshooting in this manual, especially regarding com ports. P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 49: Single-Port Operation

    Ensure the analyzer is off. With A0311 connected and turned on from the rear pannel, use the hand pad to send the valve to Position 1. Start the pump and then the instrument.
  • Page 50: Recirculation With The A0310

    Because the details of these configurations are complex, Picarro cannot guarantee that a configuration will work for a given use case and encourages users to explore any additional system elements that may be needed.
  • Page 51: Figure 32: Setup For Eliminating Position-To-Position Carryover

    Step 6: Duration: 10 minutes Rot Valve Code: 5 Desc: Run third sample … Etc., until the user reaches Valve Position 16. Step n: Duration: 100 minutes Rot Valve Code: 1 Desc: Return to ambient A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 52 Consult your data sheet, and limit runs without headspace purging to ensure headspace gas build-up does not lead to data quality issues through ringdown timeouts or interference. If sweepout is desired, CAUTION consider the approach recommended in the next section. P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 53: Sweeping Out Headspace

    The suggestions in this section have not been tested by Picarro, but broadly they could be put to use if an appropriate 4-way valve with low pressure drop is chosen.
  • Page 54: Figure 33: Notional Configuration For Sweep-Out Of Chamber Headspace

    Recirculation with the A0310 Figure 33: Notional Configuration for Sweep-Out of Chamber Headspace P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 55: Additional Best Practices

    Providing leak-tight fittings for sample jars or other containers. Picarro provides two possible fittings in the A0702 fittings kit that adapt VCR 1/4” to 1/8” Male NPT that can be used for single container purposes. However, for Ball Jars, we typically recommend slightly different bulkhead fittings.
  • Page 56: Interpreting Results Of Recirculation Work

    Headspace work is inevitably challenging from a mass conservation perspective. Using a syringe to removing headspace aliquots at set time intervals may reduce carryover concerns but decreases the headspace pressure (potentially changing system dynamics), which can lead to inward leaks, and eventually challenges P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 57: Figure 36: A0310 Valve Sequence Showing Equilibration Periods

    A to some later time period B. There are no discrete sample values by position provided by the A0310 configuration, unlike with some Picarro peripherals controlled by coordinators (e.g. SSIM A0314, Water Autosampler A0325). Data produced with the A0310 is saved to the UserLog data files, found under C:/Picarro/UserData/datalog_user/YYYY/MM/DD.
  • Page 58: Pharmaceutical Use (Pi2114 + A0311-S)

    10. Pharmaceutical Use (PI2114 + A0311-S) Pharmaceutical customers wishing to multiplex vapor-phase hydrogen peroxide (VHP, or VPHP) sampling will want to use the A0311-S manifold, as it contains compatible sample handling materials, and keeps all lines constantly refreshed to reduce position-to-position memory effects. Though this manifold is an excellent choice for this application, there are some important caveats about its use, which we outline here.
  • Page 59: Maintenance

    Put a new fuse (1.5 A/250 VAC, SB 5 x 20 mm) in the fuse holder. Push the fuse holder into the rear panel. You will hear a click when it is fully inserted. Release fuse holder with flathead Fuse Holder screwdriver Figure 38: Replacing the Fuse A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 60: Cleaning

    Maintenance Cleaning Clean the outside of the A0311 or A0311-S with a clean dry cloth. Only certified service technicians should access or clean the inside of the A0311 unless otherwise instructed. P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 61: Troubleshooting

    Builds after mid-2022 contain standoffs designed to prevent this from occurring. If your hand pad will not power up, you will need to remove the lid of the A0311 or A0311-S box to reconnect the power supply cable.
  • Page 62: Pressure Unlocked" Errors

    Reconnected Figure 40: Inside Manifold Box – Disconnected and Reconnected Power Cable 12.3 “Pressure Unlocked” errors If the Picarro data viewer software shows “Pressure unlocked” errors, check the following: That the position(s) of the manifold(s), is as expected, and/or consistent if the A0310 is being used.
  • Page 63: Sampling Port Does Not Change

    12.4 Sampling Port Does Not Change The sampling port doesn't change after commanding the software to do so. I have directed the A0311/A0311-S to sample from another port, but I'm still measuring what I have connected to the previous port. Why doesn't the valve change...
  • Page 64: Dilution Using Purge Gas On A0310

    Position 3 with the empty sample jar). The decrease in the stable concentration value as a fraction of the previous value will give the dilution factor. E.g., 1000→900→810→729 would suggest a consistent 10% dilution effect. P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 65: Appendix A - Setting Up Contained Pump Exhaust Flow

    A.2 Tools and Parts Required • Long flathead screwdriver (6” x 5/16” recommended) • 9/16” open end wrench • Swagelok ISO parallel thread adapter 1/4"-1/8” SS-400-1-2RS (Picarro PN 22928) • Swagelok gasket SS-2-RS-2V (Picarro PN 22929) • 1/4" tubing and stainless-steel ferrule set •...
  • Page 66: Figure 42: Pump Noise Dampener Removal

    9/16" wrench, tighten the nut approximately seven flats (420 degrees). A guide to this process can be found on the Picarro Video Gallery here: https://mktg.picarro.com/acton/media/39674/picarro-video-gallery P/N 40-0038 Rev. G...
  • Page 67: Figure 43: Pump Exhaust Line Adapter Fittings

    Ventilation exhaust tubing should be 1/4" or larger to avoid back pressure build- up in the analyzer pump. If the vent line is more than a few feet long, it must be sized up to 3/8” or 1/2”. NOTE A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 68: Appendix B - Steps If Analyzer Can't Recognize Valve

    If the analyzer does not automatically recognize the multi position valve, follow the steps below. Run the Stop CRDS Software Executable If your analyzer uses the Windows Operating System, go to the Picarro Utilities folder on the Desktop, and run the Stop CRDS Software and Serial/Socket Port Manager executable.
  • Page 69: Figure 46: Setup Tool Selection

    Confirm that the Mode dropdown states the Picarro mode being used. Click Apply to save the changes, and then Exit. 10. Restart the software from the Start Instrument icon on the desktop.
  • Page 70: Figure 47: Picarro Analyzer Setup Tool - Port Manager

    Check Mode from then Exit this dropdown list Figure 47: Picarro Analyzer Setup Tool – Port Manager Display Valve Sequencer GUI Though it is possible the instrument will recognize the MP valve(s) when the user restarts the software, it is often necessary to restart the computer first.
  • Page 71: Figure 48: Show/Hide Sequencer Under Tools Drop Down Menu

    Steps if Analyzer Can’t Recognize Valve Figure 48: Show/Hide Sequencer under Tools Drop Down Menu Figure 49: External Valve Sequencer Window A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...
  • Page 72: Appendix C - Third Party Fittings And Parts

    Pre to Mid 2022 SS-200-NFSET SS-200-NFSET See section 4.5, Connection Procedure A0310 Sample Jar 1/4” A0310 Legris Grainger bulkhead push to See section 9, PN 3116 56 00 PN 1PFC6 connect fitting Recirculation with the A0310 P/N 40-0038 Rev. G A0311/A0311-S/A0310 UM...
  • Page 73 See section 9.2, PN 1943712VDC (Humphrey website Sweeping Out provides distributor Headspace search) Various – Determined 3-Way Solenoid S3112 Valve 1/4” fittings by user Various – Determined 3-Way Solenoid S3136 Valve 1/8” fittings by user A0311/A0311-S/A0310 UM P/N 40-0038 Rev. G...

This manual is also suitable for:

A0311-sA0310

Table of Contents