Cubro Aeria 8 Pro User Manual

Cubro Aeria 8 Pro User Manual

Mattress replacement

Advertisement

Quick Links

User guide
Aeria 8 Pro and Aeria 8 Pro Bariatric
mattress replacement
Code: 2160, 2165
Contents
2
2
3
4
6
7
8
10
14
UG_2160-2165_Aeria 8 Pro_0819
M A X .
220kg
2160
M A X .
250kg
2165
16
18
19
20
21
21
22
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Aeria 8 Pro and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Cubro Aeria 8 Pro

  • Page 1: Table Of Contents

    User guide Aeria 8 Pro and Aeria 8 Pro Bariatric mattress replacement Code: 2160, 2165 M A X . 220kg 2160 M A X . 250kg 2165 Contents Definitions Troubleshooting System overview Technical specifications: Aeria 8 Pro System features Technical specifications:...
  • Page 2: Definitions

    WARNING! Risk of electric shock. WARNING! Explosion hazard. Important information. System overview Introducing the Aeria 8 Pro and 8 Pro Bariatric mattress replacement systems 1 Dynamic mattress replacement 2 Digital control unit with transport function Straps 360° zip...
  • Page 3: System Features

    • The Aeria 8 Pro and 8 Pro Bariatric are indicated for the prevention and treatment of skin breakdown and pressure ulcers in patients at high to very high risk • Constructed from transverse cells 21 x 220mm high that cyclically inflate and deflate in an alternating pattern, providing gentle and dynamic support •...
  • Page 4: Intended Use

    • the prevention and treatment of skin breakdown and pressure ulcers in patients at high to very high risk Contraindications Patient conditions for which the application of pressure relief therapy on the Aeria 8 Pro and 8 Pro Bariatric mattress replacement systems is contraindicated include: • unstable spinal cord injury •...
  • Page 5 The Aeria 8 Pro and 8 Pro Bariatric mattress replacements can be fitted to most standard hospital or single bed bases The Aeria 8 Pro and 8 Pro Bariatric digital control unit can be fitted to the foot or head board of most hospital or home care beds...
  • Page 6: Safety Precautions

    • Avoid blocking the air intakes of the control unit, located at the rear of the unit • Only the control unit and mattress combination as indicated by Cubro should be used, otherwise the correct function of the product cannot be guaranteed •...
  • Page 7: Protection Against Hazards

    Power cable • The system should never be operated with a worn or damaged power cable Should the power cable be found to be worn or damaged, contact Cubro for a replacement Interference • Significant risks of reciprocal interference may be posed by the presence of the system...
  • Page 8: System Set Up

    Carefully unpack the system and locate all items shown in the picture below Inspect each item for any damage that may have occurred during shipping Any damage or missing components should be reported to Cubro as soon as possible Control unit...
  • Page 9 9 Once the mattress is fully inflated, bedding WARNING! Do not store the can be replaced Sheets should be loose system in direct sunlight. and not fitted so they do not interfere with cell alternation sales@cubro.co.nz 0800 656 527 cubro.co.nz UG_2160-2165_Aeria 8 Pro_0819...
  • Page 10: Operation

    Operation Control unit layout 1. Power button Press the Power button for at least two seconds to turn on the control unit 2. System OK LED Once the system is ready to operate the green LED is turned on and remains activated This LED will only turn off if there is a failure, to indicate that the system is no longer ready to operate...
  • Page 11 WARNING! Wait at least 12 minutes between any pressure adjustment or patient assessment as the system may take a full cycle to adjust to the new pressure setting. sales@cubro.co.nz 0800 656 527 cubro.co.nz UG_2160-2165_Aeria 8 Pro_0819...
  • Page 12 Operation (cont.) Static Mode The system will automatically In Static Mode, all mattress cells remain fully start in Static Mode, and revert to inflated, thereby terminating the alternation Alternating Mode once operational cycle and any therapeutic patient benefits pressure is reached. This mode should be used to create a firm base for stable patient handling and transport The system will operate in Static Mode...
  • Page 13 WARNING! If the problem persists, Alarm Meaning contact Cubro for further advice about indicator repair. Do not try to open the control unit. Opening the unit could cause Initial failure personal injury or equipment damage.
  • Page 14: Care And Maintenance

    Care and maintenance WARNING! The mattress is not protected against excessive amounts of water. Disconnect power supply before cleaning. Failure to do so could result in equipment damage or electric shock. WARNING! Switch off and disconnect the control unit from mains power supply before cleaning.
  • Page 15 WARNING! It is recommended that fuse only (T1AL/250V).Preventive the air filter be replaced each year. inspections and calibration are Replacement air filters are available not required. from Cubro. sales@cubro.co.nz 0800 656 527 cubro.co.nz UG_2160-2165_Aeria 8 Pro_0819...
  • Page 16: Troubleshooting

    Troubleshooting If the problem cannot be resolved, please contact Cubro for further support Problem Cause Solution Control unit The control 1 Check the control unit is connected to mains power does not unit may not be outlet with the correct voltage operate;...
  • Page 17 2 Check effective system performance by conducting ‘bottoming out’ too low for the a ‘bottoming out’ test as described on page 11 while lying flat patient’s weight on the mattress sales@cubro.co.nz 0800 656 527 cubro.co.nz UG_2160-2165_Aeria 8 Pro_0819...
  • Page 18: Technical Specifications: Aeria 8 Pro

    Technical specifications: Aeria 8 Pro Classification Mattress replacement Equipment not suitable for use in the Dimensions 2000 x 880 x 220mm presence of a flammable anaesthetic mixture (inflated) with air, oxygen or nitrous oxide Weight 8 5kg Non-continuous Type BF Equipment...
  • Page 19: Technical Specifications: Aeria 8 Pro Bariatric

    Technical specifications: Aeria 8 Pro Bariatric Classification Mattress replacement Equipment not suitable for use in the Dimensions 2000 x 1070 x 220mm presence of a flammable anaesthetic mixture (inflated) with air, oxygen or nitrous oxide Weight 10kg Non-continuous Type BF Equipment...
  • Page 20: Safety Labels

    Initial failure necessary. User guide Alarm functions EXAMPLE ONLY New Approach Directives) Caution symbol Aeria 8 Pro, 8Pro XL, XXL & 10 Pro User Guide Operating instructions Type BF Applied Part Pressure arrow buttons Important information Serial number Class II Equipment Electrical hazard Symbol for ”Authoristed Representative in...
  • Page 21: Servicing

    149 Taurikura Drive, Tauriko, Tauranga, NZ website for further information on the correct disposal For additional copies of this user guide or for a different format, please contact Cubro. WARNING! Do not make any changes or modifications to this product without consultation from Cubro.
  • Page 22: Limited Warranty

    4 CUBRO LTD has no obligation to the purchaser and is not obligated to honour all or any part of this warranty unless the following procedure is followed by the purchaser: a Before making a warranty claim, the purchaser should ensure that the product is defective by following standard ‘trouble shooting’...
  • Page 23 © Cubro Limited 2019 ALL RIGHTS RESERVED: No part of this publication shall be adapted, modified, reproduced, copied, or transmitted in any form or by any means including written electronic, mechanical, reprographic, photocopying, or recording means Furthermore, this publication shall not be stored in whole,...
  • Page 24 0800 656 527 Get in touch sales@cubro.co.nz cubro.co.nz UG_2160-2165_Aeria 8 Pro_0819...

This manual is also suitable for:

Aeria 8 pro bariatric21602165