CPAC SteriSURE 3100 Manual

Dry heat sterilizers with 60 minute sterilization cycle times

Advertisement

Quick Links

Dry Heat Sterilizers with
60 Minute Sterilization Cycle Times
MODEL
3100
MODEL
2100
2364 Leicester Road, P.O. Box 175, Leicester, NY 14481
Phone (585) 382-3223 ● Fax (585) 382-9481 ●
www.cpac.com
December 2, 2016

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SteriSURE 3100 and is the answer not in the manual?

Questions and answers

Summary of Contents for CPAC SteriSURE 3100

  • Page 1 Dry Heat Sterilizers with 60 Minute Sterilization Cycle Times MODEL 3100 MODEL 2100 2364 Leicester Road, P.O. Box 175, Leicester, NY 14481 Phone (585) 382-3223 ● Fax (585) 382-9481 ● www.cpac.com December 2, 2016...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS RECOMMENDATIONS – IMPORTANT SAFEGUARDS ACCESSORIES MATERIALS INTEGRITY CAUTIONS SAFETY NOTES CONCERNING TEMPERATURE OPERATING INSTRUCTIONS CLEANING AND MAINTENANCE REPLACEMENT FUSES SIMULATED USE TESTING ERROR CODES BIOLOGICAL TESTING 9-10 UNIT ELECTRICAL RATINGS USB PORT ENVIRONMENTAL CONDITIONS WARRANTY...
  • Page 3: Recommendations - Important Safeguards

    RECOMMENDATIONS Read the entire instruction manual before installation or operation of the SteriSURE Dry Heat Transfer Sterilizer. It will help you to understand the operation of the system, how various sub assemblies work together and the operating sequence of the controls. WARNING: NEVER ATTEMPT TO PERFORM ANY ELECTRICAL TROUBLESHOOTING ADJUSTMENT(S) OR SERVICE(S) UNLESS YOU ARE A QUALIFIED ELECTRICIAN, ELECTRONICS TECHNICIAN OR FACTORY TRAINED...
  • Page 4: Accessories

    SteriSURE STERILIZERS The technology used in the SteriSURE sterilizer utilizes the latest advancements in dry heat sterilization. We are confident you will find it a valuable and cost saving addition to your practice. The SteriSURE sterilizer* is intended for indoor use in hospitals and dental, orthodontic and health care facilities.
  • Page 5: Operating Instructions

    After room temperature instruments are placed in the sterilizer, the temperature may drop a few degrees depending on the size of the load. The cycle will not begin until 320°F (160°C) has been reestablished. If the temperature drops below 310°F (154°C) during a sterilization cycle, the timer will stop and an error code will display.
  • Page 6: Maximum Loading

    In order to ensure sterility, instruments intended to be stored must be sealed in SteriSURE nylon pouches prior to sterilization. SteriSURE Part No. 400636, NYLON SELF SEAL POUCHES 2" X 10" SteriSURE Part No. 400651, NYLON SELF SEAL POUCHES 3” X 10” SteriSURE Part No.
  • Page 7: Cleaning And Maintenance

    All of the electronic and mechanical components used in the sterilizer’s construction are long life, heavy-duty parts that require no maintenance. An authorized service representative or CPAC should be contacted if any of these parts need to be replaced.
  • Page 8: Replacement Fuses

    The inner door gasket should be replaced if it shows sign of wear or if it comes loose. Order Gasket Kit #211001 for Model 2100 and #311001 for Model 3100. Contact your authorized service representative or call CPAC at (585) 382-3223. REPLACEMENT FUSES Models 2100 &...
  • Page 9: Error Codes

    If one of the following error codes appears, push and release the ON (WARM- UP)/STANDBY button to reset the unit. This will reestablish the error detect logic and will eliminate false error codes that may occur. If the error code persists, call your authorized service representative or CPAC at (585) 382-3223. Process Failures:...
  • Page 10 Recommended and Required Equipment For monitoring dry heat, biological indicators containing Bacillus atrophaeus should be used along with the recommended test load. Spore test strips, as well as test services are widely available through universities and commercial services.  Chemical indicators supplied by SteriSURE are recommended, part. no. 400635 ...
  • Page 11: Unit Electrical Ratings

    UNIT ELECTRICAL RATINGS MODEL 2100 115 VAC, 50/60 Hz, 10 Amps MODEL 3100 115 VAC, 50/60 Hz, 10 Amps MODEL 2100-220V 230 VAC, 50/60 Hz, 5 Amps MODEL 3100-220V 230 VAC, 50/60 Hz, 5 Amps USB PORT The Steri-SURE Sterilizer is capable of and will need to be updated if power is storing sterilization cycle data on a USB lost to the sterilizer.
  • Page 12: Environmental Conditions

    All units must have a Return Material Authorization code (RMA) visible on the returned item. RMA’s can be obtained by calling CPAC at (585) 382-3223. EVERY EFFORT HAS BEEN MADE TO ENSURE THE COMPLETE ACCURACY OF THE CONTENTS OF THIS MANUAL.

This manual is also suitable for:

Sterisure 2100