TABLE OF CONTENTS RECOMMENDATIONS – IMPORTANT SAFEGUARDS ACCESSORIES MATERIALS INTEGRITY CAUTIONS SAFETY NOTES CONCERNING TEMPERATURE OPERATING INSTRUCTIONS CLEANING AND MAINTENANCE REPLACEMENT FUSES SIMULATED USE TESTING ERROR CODES BIOLOGICAL TESTING 9-10 UNIT ELECTRICAL RATINGS USB PORT ENVIRONMENTAL CONDITIONS WARRANTY...
RECOMMENDATIONS Read the entire instruction manual before installation or operation of the SteriSURE Dry Heat Transfer Sterilizer. It will help you to understand the operation of the system, how various sub assemblies work together and the operating sequence of the controls. WARNING: NEVER ATTEMPT TO PERFORM ANY ELECTRICAL TROUBLESHOOTING ADJUSTMENT(S) OR SERVICE(S) UNLESS YOU ARE A QUALIFIED ELECTRICIAN, ELECTRONICS TECHNICIAN OR FACTORY TRAINED...
SteriSURE STERILIZERS The technology used in the SteriSURE sterilizer utilizes the latest advancements in dry heat sterilization. We are confident you will find it a valuable and cost saving addition to your practice. The SteriSURE sterilizer* is intended for indoor use in hospitals and dental, orthodontic and health care facilities.
After room temperature instruments are placed in the sterilizer, the temperature may drop a few degrees depending on the size of the load. The cycle will not begin until 320°F (160°C) has been reestablished. If the temperature drops below 310°F (154°C) during a sterilization cycle, the timer will stop and an error code will display.
In order to ensure sterility, instruments intended to be stored must be sealed in SteriSURE nylon pouches prior to sterilization. SteriSURE Part No. 400636, NYLON SELF SEAL POUCHES 2" X 10" SteriSURE Part No. 400651, NYLON SELF SEAL POUCHES 3” X 10” SteriSURE Part No.
All of the electronic and mechanical components used in the sterilizer’s construction are long life, heavy-duty parts that require no maintenance. An authorized service representative or CPAC should be contacted if any of these parts need to be replaced.
The inner door gasket should be replaced if it shows sign of wear or if it comes loose. Order Gasket Kit #211001 for Model 2100 and #311001 for Model 3100. Contact your authorized service representative or call CPAC at (585) 382-3223. REPLACEMENT FUSES Models 2100 &...
If one of the following error codes appears, push and release the ON (WARM- UP)/STANDBY button to reset the unit. This will reestablish the error detect logic and will eliminate false error codes that may occur. If the error code persists, call your authorized service representative or CPAC at (585) 382-3223. Process Failures:...
Page 10
Recommended and Required Equipment For monitoring dry heat, biological indicators containing Bacillus atrophaeus should be used along with the recommended test load. Spore test strips, as well as test services are widely available through universities and commercial services. Chemical indicators supplied by SteriSURE are recommended, part. no. 400635 ...
UNIT ELECTRICAL RATINGS MODEL 2100 115 VAC, 50/60 Hz, 10 Amps MODEL 3100 115 VAC, 50/60 Hz, 10 Amps MODEL 2100-220V 230 VAC, 50/60 Hz, 5 Amps MODEL 3100-220V 230 VAC, 50/60 Hz, 5 Amps USB PORT The Steri-SURE Sterilizer is capable of and will need to be updated if power is storing sterilization cycle data on a USB lost to the sterilizer.
All units must have a Return Material Authorization code (RMA) visible on the returned item. RMA’s can be obtained by calling CPAC at (585) 382-3223. EVERY EFFORT HAS BEEN MADE TO ENSURE THE COMPLETE ACCURACY OF THE CONTENTS OF THIS MANUAL.