Human Loopamp 972000 Instructions For Use Manual

Mtbc detection kit

Advertisement

Available languages

Available languages

Quick Links

I nstruct i ons for Use
380236-D
English
L oopamp MTBC Detect i on K i t
I NTENDED USE
The L oopamp MTBC Detection Kit is a qualitative in vitro
diagnostic test to detect Mycobacterium tuberculosis complex (MTBC)
DNA extracted from the sputum in patients with any symptom
indicative of MTBC infection. The kit aids in the diagnosis of MTBC
infection and is intended to be used in professional laboratories and
hospitals by adequately trained personnel. The result can be
interpreted either by automated turbidimeter or visually under UV
irradiation.
TEST PR I NC I P L ES
Th i s product i s based on the nuc l e i c ac i d amp li f i cat i on method,
L AMP ( L oop-med i ated I sotherma l Amp li f i cat i on ), deve l oped by E i ken
Chem i ca l Co., L td.
The characteristics of the L AMP method are as follows: (1) on l y one
enzyme i s requ i red, and the amp li f i cat i on react i on proceeds under
1),2)
i sothermal cond i t i ons;
(2) i t has extreme l y h i gh spec i f i c i ty because of
the use of four pr i mers recogn i z i ng s i x d i st i nct reg i ons on the target; (3)
i t has a h i gh amp li f i cat i on eff i c i ency and can produce a high
concentration of amplified product in a short time, which makes visual or
automated detection possible.
3),4)
The pr i mers prov i ded w i th th i s product are des i gned i n the DNA
gyrase subun i t B (gyrB) and I nsert i on sequence I S6110 ( I S) reg i on of
the MTBC genome DNA, which has been conf i rmed by the a li gnment
ana l ys i s of the se l ected base sequences of MTBC and nontubercu l ous
mycobacter i a to have a we ll -conserved base sequence i n MTBC.
The DNA from untreated sputum or NA L C-NaOH-treated sputum i s
extracted us i ng the L oopamp PURE DNA Extract i on K i t (sold
separately). Then, the DNA so l ut i on i s d i spensed i nto a react i on tube.
The
strand
d i sp l acement
tr i phosphates (dATP, dCTP, dGTP, and dTTP ) , ca l ce i n, reaction buffers,
and MTBC-spec i f i c pr i mers are stored i n the dr i ed form i n the cap of the
react i on tube. This dr i ed L AMP reagent (MTBC detect i on reagent
(dMTB)) is dissolved when the DNA solution is added. The react i on tube
i s then i ncubated at 67.0 C, and the DNA i s amp li f i ed through cata l ys i s
by the strand d i sp l acement DNA po l ymerase per the L AMP react i on.
The detect i on of amp li f i ed products i s based on the turb i d i ty
measurement of a by-product, magnes i um pyrophosphate ( wh i te
prec i p i tate).
3)
Alternatively, a v i sua l judgment under UV irradiation can
be used i nstead of turb i d i ty measurement. Before react i on, ca l ce i n i n
the reagent i s i n the quenched state because of manganese i ons bound
to it; however, once the L AMP react i on i s started, pyrophosphate i ons
are generated and b i nd out the manganese i ons, and ca l ce i n becomes
4)
f l uorescent.
KIT CONTENTS
Reagents are stab l e unt il the date on the l abe l, assum i ng the
conta i ner rema i ns unopened at a storage temperature of 230 C. The
reagents are also confirmed stable after opening the container when
adhered to these procedural directions.
MTBC detect i on reagent ······························ 2 x 48 tubes
The fo ll ow i ng reagents i n the dr i ed form are conta i ned
i n each react i on tube.
Bst DNA po l ymerase
Deoxynuc l eot i de tr i phosphates
Magnes i um su l fate
Ca l ce i n
Manganese ch l or i de
Pr i mers
b)
Pos i t i ve contro l MTB (PC MTB)
Negat i ve contro l MTB (NC MTB) ····················· 3 x 0.5 m L
30 µL dropper ········································ 1 x 18 droppers
a) Bst DNA po l ymerase der i ved from Bac ill us stearothermoph il us i s
a strand d i sp l acement DNA po l ymerase that l acks 53
exonuc l ease act i v i ty.
972000
DNA
po l ymerase,
deoxynuc l eot i de
a)
c)
···················· 1 x 0.4 m L
SSP available in EUDAMED
b) Pr i mers des i gned i n the gyrB and I S reg i on of the MTBC genome
DNA, pur i f i ed from synthes i zed o li gonuc l eot i des by HP L C.
c) The PC MTB conta i ns a product resu l t i ng from the i n v i tro
amp li f i cat i on of a temp l ate genome DNA of Mycobacter i um
tubercu l os i s H37Rv (GenBank No. NC_000962 ) or i g i n.
Abbrev i at i ons of the fo ll ow i ng reagents and L ot No., as well as
manufacturer (EKN), are pr i nted on the conta i ners as shown be l ow:
Reagents
Pos i t i ve contro l MTB
Negat i ve contro l MTB
*Metrological Traceability Information
The Positive Control is derived from the amplification product of
genomic DNA originating from Mycobacterium tuberculosis H37Rv
(GenBank No. NC_000962). The specific region of Mycobacterium
tuberculosis H37Rv, which is the target of interest, serves as the
template for the amplification of the PC MTB DNA fragment. This
fragment is quantitated through photospectrometric analysis, and its
DNA concentration of PC MTB is adjusted at 1,000 copies/μL.
WARN I NGS AND PRECAUT I ONS
(1) For i n v i tro d i agnost i c use on l y.
(2) Th i s product i s des i gned on l y for c li n i ca l d i agnos i s of MTBC from
sputum samp l es of human or i g i n. Do not use it for other purposes.
(3) When us i ng th i s product, a l ways fo ll ow th ese Instructions for Use.
(4) Do not freeze the reagents.
(5) Do not use any exp i red reagent.
(6) Do not m i x d i fferent l ots.
(7) Do not rep l en i sh any reagent.
(8) The p erformance of the Loopamp MTBC Detection Kit depends
on the operator prof i c i ency and adherence to procedura l
d i rect i ons. Test i ng shou l d be done by adequate l y tra i ned
personne l strictly according to the instructions provided.
(9) Remove the requ i red number of react i on tubes from the packag i ng
before using them and re-sea l the a l um i n i um pouch i mmed i ate l y.
(10) Do not remove the des i ccant from the a l um i n i um pouch. H i gh-
l eve l hum i d i ty may deter i orate the dr i ed L AMP reagent i n the
react i on tubes.
(11) Exposure to heat, humidity, and li ght m i ght deter i orate the dMTB.
Thus, remove on l y the requ i red number of react i on tubes ( sum of
samp l es and contro l s ) and re-seal the aluminium pouch
immediately.
(12) Read the instruction manual and ensure that the required equipment
(turbidimeter or incubator) is available before commencing the
procedure.
(13) Sputum samp l es pose a potent i a l r i sk for i nfect i on. Take a ll
necessary prevent i ve measures to avo i d b i ohazard.
(14) The PC MTB and the NC MTB conta i n a sma ll amount of sod i um
az i de as a preservat i ve. As sod i um az i de i s c l ass i f i ed as tox i c,
avo i d any contact w i th eyes, mouth, or sk i n.
(15) I n case of acc i denta l contact of any reagent w i th eyes, mouth, or
sk i n, i mmed i ate l y r i nse the affected s i te w i th running water and
seek medical advice.
(16) Do not d il ute or add the PC MTB to the samp l es. Instead, use the
PC MTB on l y as descr i bed i n th i s package i nsert to avo i d DNA
contam i nat i on.
(17) Store the PC MTB and any pos i t i ve sputum samp l es separate l y
from the other k i t reagents.
(18) The cap of each react i on tube conta i ns the dMTB i n the dr i ed form.
Do not touch the i ns i de of the cap.
(19) Before us i ng the react i on tubes, check carefu ll y i f they have any
cracks or scratches. Damaged tubes m i ght g i ve fa l se resu l ts and
l ead to DNA contam i nat i on of the i ncubator and work area.
( 20 ) Do not expose react i on tubes to UV li ght before the end of the
L AMP react i on. Pro l onged exposure to UV li ght m i ght damage the
tubes and l ead to fa l se resu l ts.
( 21 ) When UV light i s used for v i sua l f l uorescence judgment, do not
stare d i rect l y at it. As UV li ght i s harmfu l to the eyes, even
watch i ng for a short per i od can i rr i tate eyes and cause symptoms
s i m il ar to conjunct i v i t i s. Instead, use a g l ass screen or wear
1
L abe lli ng on the tube Code on the cap
PC MTB
PC
L ot No., EKN
MTB
NC MTB
NC
L ot No., EKN
MTB
5)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Loopamp 972000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Human Loopamp 972000

  • Page 1 Do not use it for other purposes.
  • Page 2: Measurement Procedure

    corresponds to approximately 30 µL. protective goggles/glasses or a UV-blocking eye mask whenever (3) D i spense 30 µL of the negative control solution i nto a react i on tube l ook i ng d i rect l y at UV light. us i ng the L oopamp...
  • Page 3: Procedura L Notes

    as i nva li d, and the test shou l d be repeated. PROCEDURA L NOTES After conf i rm i ng that the run i s va li d, eva l uate samp l es as fo ll ows: ...
  • Page 4 bov i s, and Mycobacterium africanum was conf i rmed. L AMP 6. I nformat i on about a Ca li brator Pos i t i ve Negat i ve The performance test for th i s product used the p l asm i d DNA Untreated Pos i t i ve conta i n i ng the gyrB and I S reg i on of the genome DNA of...
  • Page 5 In vitro diagnostic Manufacturer Temperature limitation medical device Authorized Contains sufficient Representative in Batch code for <n> tests the European Community Importer HUMAN Gesellschaft für Biochemica und Diagnostica mbH Max-Planck-Ring 21, 65205 Wiesbaden, Germany https://www.eiken.co.jp/en *Date of Revision: 2023-10-31...
  • Page 6: Notice D'utilisation

    Notice d'utilisation RCSP disponible sur EUDAMED 380236-D 972000 Compte-gouttes 30 µl ····················· 1 x 18 compte-gouttes Français a) La Bst DNA polymerase dérivée de Bacillus stearothermophilus est un ADN polymérase à déplacement de brin qui manque Loopamp MTBC Detection Kit d'activité...
  • Page 7: Élimination Des Déchets

    Remarque : un contrôle négatif doit être utilisé à chaque fois. ne présentent pas de fissures ou de rayures. Des tubes (3)Contrôle positif MTB(PC MTB) endommagés peuvent donner de faux résultats et entraîner une Agiter (ou centrifuger) le tube pour recueillir le contenu au fond contamination de l'incubateur et de la zone de travail par l'ADN.
  • Page 8: Interprétation Des Résultats

    dans le contrôle négatif, la réaction d'amplification se déroule PC MTB. Distribuer les échantillons et la solution de contrôle correctement (Fig. 1). En revanche, si une autre situation se négatif et fermer tous les tubes de réaction avant de distribuer le produit, la réaction d'amplification peut se dérouler de manière PC MTB.
  • Page 9: Caractéristiques De Performances

    turbidité mesurée par le turbidimètre en temps réel HumaTurb C+A Comme le montrent les tableaux ci-dessous, le taux de n'est pas lié à la concentration d'ADN matrice. concordance global était favorable : LAMP pour expectorations non traitées PCR : 91,5 % (291/318 sujets) ;...
  • Page 10: Informations Pour Commander

    Retourner les tubes de réaction pour recueillir la solution Importateur dans le bouchon. Patienter 2 minutes. Patienter 2 minutes avec les tubes de réaction à HUMAN Gesellschaft für Biochemica und Diagnostica mbH Retourner les tubes de réaction cinq fois pour mélanger le contenu, puis les centrifuger à nouveau.
  • Page 11: Instrucciones De Uso

    Instrucciones de uso SSP disponible en EUDAMED 380236-D 972000 a) La ADN polimerasa de Bst obtenida a partir de Bacillus Español stearothermophilus es una ADN polimerasa con actividad de desplazamiento de la hebra que carece de actividad exonucleasa Loopamp MTBC Detection Kit en sentido 53.
  • Page 12 (19) Antes de utilizar los tubos de reacción, compruebe detenidamente denominará «solución de control negativo») siguiendo las si presentan grietas o arañazos. Los tubos dañados podrían dar instrucciones. Nota: Siempre se debe incluir un control negativo. resultados falsos y provocar la contaminación con ADN de la (3)Control positivo de MTB(PC MTB)...
  • Page 13: Interpretación De Los Resultados

    Si la turbidez aumenta en el control positivo, pero no en el control (12) Golpetee o centrifugue el tubo de PC MTB antes de utilizarlo para negativo, la reacción de amplificación está teniendo lugar que el contenido se acumule en el fondo. Cierre el tubo correctamente (Fig.
  • Page 14: Información Para Pedidos

    En cuanto a los fármacos, nuestros estudios internos han revelado producida por la sangre). Por lo tanto, consideramos que los dos que la determinación no se ve afectada por la presencia de isoniazida tipos de evaluaciones son idénticas. Las tablas a continuación (100 µg/ml), etambutol (20 µg/ml), rifampicina (100 µg/ml), pirazinamida muestran los resultados obtenidos mediante detección de la (500 µg/ml), kanamicina (20 µg/ml), y estreptomicina (500 µg/ml).
  • Page 15 Amplificación (Evite la formación de burbujas) Importador Coloque los tubos de reacción en el bloque de reacción HUMAN Gesellschaft für Biochemica del turbidímetro. und Diagnostica mbH Max-Planck-Ring 21, 65205 Wiesbaden (Alemania) Como se indica en las instrucciones de uso del turbidímetro, inicie la reacción y, a continuación,...
  • Page 16 Gebrauchsanweisung SSP in EUDAMED verfügbar 380236-D 972000 Negativkontrolle MTB (NC MTB) ······················· 3 x 0,5 ml Deutsch 30 µl Tropfer ··········································· 1 x 18 Tropfer a) Die aus Geobacillus stearothermophilus stammende Bst DNA- Loopamp MTBC Detection Kit Polymerase ist eine strangversetzende DNA-Polymerase ohne 53-Exonukleaseaktivität.
  • Page 17 (18) Die Verschlusskappe jedes Reaktionsgefäßes enthält das dMTB Loopamp PURE DNA Extraction enthalten ist. in getrockneter Form. Verschlussinnenseite nicht berühren. Verarbeitung der NC MTB die Gebrauchsanweisung beachten (im (19) Reaktionsgefäße vor Verwendung sorgfältig auf Risse oder Folgenden wird extrahierte Kratzer prüfen. Beschädigte Gefäße können falsche Ergebnisse „Negativkontrolllösung“...
  • Page 18: Interpretation Der Ergebnisse

    (3) Reaktionsgefäße einsetzen und Messung starten. Jegliche Kontamination anderer Proben mit der PC MTB (4) Turbidimeter-Anzeige auf einen Trübungsanstieg der Positiv- und vermeiden. Proben Negativkontrolllösung Negativkontrollen prüfen. Steigt die Trübung in der Positivkontrolle dispensieren und vor der Dispension der PC MTB alle Negativkontrolle jedoch nicht,...
  • Page 19 INTERFERENZEN LAMP für unbehandeltes Sputum vs. TRC: 94,3 % (230/244); LAMP Interne Studien zeigten keine Beeinflussung der Messung durch das für vorbehandeltes Sputum vs. TRC: 93,0 % (227/244). Außerdem Vorhandensein von freiem Bilirubin (91,0 mg/dl), konjugiertem Bilirubin stimmten die Auswertungen der Echtzeit-Trübungsmessung und (101,0 mg/dl), Chylus (Formazin-Trübung: 7300) und hämolisiertem des optischen Fluoreszenznachweises vollständig überein, mit Hämoglobin (2475 mg/dl).
  • Page 20 In Serviceanleitung Katalognummer Verfallsdatum nachschlagen Medizinprodukt für Hersteller Temperaturbegrenzung In-vitro-Diagnostik Bevollmächtigter Inhalt ausreichend Vertreter in der Chargennummer für <n> Tests Europäischen Gemeinschaft Importeur HUMAN Gesellschaft für Biochemica und Diagnostica mbH Max-Planck-Ring 21, 65205 Wiesbaden, Germany https://www.eiken.co.jp/en *Datum der Revision: 2023-10-31...

Table of Contents