.........................13 • mV / ORP parameter ..........................15 mV / ORP parameter setup ......................15 • High impedance testing ........................17 Instrument configuration menu .......................17 Warranty ............................19 • Warranty duration and limitations ...................19 Disposal ............................19 UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 4
The keyboard layout and the innovative high- definition colour LCD display make the instrument extremely practical, quick to operate and easy to use. The XS PX Vio instrument performs the signal simulation of a conventional pH, ORP or ISE electrode in the field with: pH: -2 ...
The proper functioning of the device and the safety of the operator are only guaranteed if all normal laboratory safety regulations are observed and if all the specific safety measures described in this manual are followed. UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 6
The owner of the instrument must inform users about the safe use of the instrument in their workplace and the management of potential risks and provide the required protective equipment. When using chemicals or solvents, follow the manufacturer's safety data sheets. UM Simulatore PX Vio EN rev.0 17.09.2024...
Environmental operating conditions 0 ... +60 °C Maximum permissible humidity <95 % non-condensing System dimensions 185 * 85 * 45 mm System weight 400 g Reference standards EMC 2014/30/EU RoHS 2011/65/EU EN 61326-1 EN 61010-1 UM Simulatore PX Vio EN rev.0 17.09.2024...
Contact your local distributor to be updated on accessories and sales codes. • installation The device leaves the factory ready for use by the user. • • Batteries are included. UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 9
In the setup menu where value entry is required, press and hold to quickly edit. IMPORTANT: • When sleep mode is active (after two minutes of inactivity) press any key to reactivate the display brightness. • Only at this point do the buttons regain their function. UM Simulatore PX Vio EN rev.0 17.09.2024...
In simulation mode, the active measurement parameter is indicated in the top left-hand of the display (e.g. Above the parameter value, the parameter settings are shown. EXAMPLE: simulated parameter pH / AUTOMATIC mode (default values) / USA buffer group UM Simulatore PX Vio EN rev.0 17.09.2024...
NO TEXT The instrument is in simulation mode. Setup menu button to enter SETUP mode, choose the desired setting • In measurement mode, press the by pressing the arrow buttons and confirming with UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 12
The XS PX Vio simulator can simulate the known pH values of the most widely used calibration buffer solutions. If the default values are not what the user needs, up to 5 can be entered via the programming menu.
Page 13
Pressing the arrow buttons will will change the simulated value scroll through the default by 0,01. Press and hold to scroll values of the selected buffer by a value of 0,1 and then 1 UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 14
The icon in the top left of the display shows the parameter in use and the text above the value shows the parameter settings. EXAMPLE: pH parameter AUTOMATIC mode (default values) / buffer group UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 15
/ ORP parameter The PX Vio simulator can simulate the most common mV values to perform electrochemical measurements. If the default values are not what the user needs, up to 5 can be entered via the programming menu. In CONTINUOUS mode, using the arrow buttons, the user can move quickly over the entire mV range from -1999 to +1999 mV.
INDOOR P3.3 BRIGHTNESS LOW - NORMAL - HIGH NORMAL P3.4 SLEEP MODE OFF - 2 MIN - 5 MIN 2 MIN P3.5 RESET SETTINGS YES - NO P3.6 AUTO OFF YES - NO UM Simulatore PX Vio EN rev.0 17.09.2024...
Page 18
NO: the instrument always remains switched on even if it is not being used. • Note: The correct and systematic use of P3.2 / P3.3 / P3.4 / P3.6 will significantly extend battery life. UM Simulatore PX Vio EN rev.0 17.09.2024...
Disposal This appliance is subject to regulations for electronic devices. Dispose of the appliance in accordance with existing local regulations. UM Simulatore PX Vio EN rev.0 17.09.2024...