Connect VuLink Telemetry or Data Logger
Use VuLink to network and transmit remote data from instruments. Add and
configure each instrument with the VuSitu mobile app. You can also connect
instruments to a 3rd party data logger.
2
Connect Up-Hole Cable
If you are using a vented cable splitter, use a vented up-hole cable.
3
Connect Cable Splitter
Always deploy with the female connector oriented down. Improper orientation can
result in data drop-outs or physical damage.
4
Connect Instrument to Female Connector (Vented)
Connect an instrument to the female down-hole connector. If you are using a
vented cable splitter, only the female connector is vented. Do not attach VuLink or
additional splitters to the female connector.
5
Connect Additional Cables to Male Connector
Connect additional cables, instruments, and splitters to the male down-hole
connector. The male side of the splitter is not vented.
6
Connect Additional Instruments
Network up to 8 instruments together. VuLink automatically sets a unique address
when each instrument is added in VuSitu. Connect one new instrument at a time and
add it to VuLink in VuSitu before connecting the next instrument.
If you are using a 3rd-party data logger, set a unique address for each instrument
before connecting the full cable network. You can configure instrument addresses
from the Instrument Settings menu in VuSitu.
Rugged Cable Splitter
Quick Start Guide
2024-10-07 | 0085842

Advertisement

loading
Need help?

Need help?

Do you have a question about the VuLink and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for In-situ VuLink

  • Page 1 Instrument Settings menu in VuSitu. 2024-10-07 | 0085842 Information subject to change without notice. In-Situ, In-Situ logo, “Water Simplified”, VuLink, RDO, RuggedCable, TROLL, BaroTROLL, HydroVu, VuSitu, Baro Merge, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.
  • Page 2 500 mm (19.69 in.) total length 42.7 mm (1.68 in.) maximum width 18.5 mm (0.73) Dimensions connector diameter Cable extension Black polyurethane, 400 mm (15.7 in.) long Conductors 6 conductors, 24 AWG, polypropylene insulation Operating temperature -25° to 55° C (-13 to 131° F) Warranty One year from date of purchase www.in-situ.com...