Tourmaline icp pressure bar, 1000 psi, 5 mv/psi, 0.2 microsecond rise time, rohs compliant, in-line amplifier (40 pages)
Summary of Contents for Amphenol PCB Piezotronics 421B3 Series
Page 1
Model 421B30 Differential charge amplifier for Dynamic pressure sensors to be used with PCB 176 sensors Installation and Operating Manual For assistance with the operation of this product, contact the PCB Piezotronics, Inc. Toll-free: 716-684-0001 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Repair and Maintenance Returning Equipment If factory repair is required, our representatives will PCB guarantees Total Customer Satisfaction through its provide you with a Return Material Authorization (RMA) “Lifetime Warranty Plus” on all Platinum Stock Products number, which we use to reference any information you sold by PCB and through its limited warranties on all other have already provided and expedite the repair process.
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
Table of Contents Introduction ....................................3 Description ....................................3 Installation ....................................3 Wiring and Operation .................................. 4 Calculation of Gain ..................................5 Structural Diagram ..................................6 Ordering Information ................................... 7 Special Considerations ................................. 7 Use with High Temperature Sensors ............................7 ESD Warning Information ................................
Introduction The 421B3X-Series Differential Charge Amplifiers are designed to convert the high-impedance output of a differential, charge-mode piezoelectric sensor to a low-impedance voltage output. These amplifiers may be used with either quartz or ceramic, differential charge-mode piezoelectric sensors. Description The amplifier operates on a 22 to 28 VDC power supply. The amplifier employs a high gain amplifier to perform the impedance transformation.
Wiring and Operation Input Signal: • Differential Piezoelectric Sensor: Connect the sensor’s positive charge output to the positive input (IN1+) and the sensor’s negative charge output to negative input (IN2-) using low-noise cable. • Non-Differential Piezoelectric Sensor: Connect the sensor output to the negative input (IN2-) and short the positive input (IN1+) and common (Ⱶ) using low-noise cable.
Note: The value of the resistor must be ≤250 Ω and the input voltage (the node labeled POWER in the diagram below) must be ≥22VDC. If a larger resistor value is used, this will cause the unit to not work properly and/or will limit the maximum output current. Calculation of Gain An amplifier is selected by specifying an input sensitivity and an output sensitivity.
Example: An amplifier ordered with 10 pC/g input sensitivity and a 100 mV/g output sensitivity would have a gain of 100 mV/g / 10 pC/g = 10 mV/pC. Structural Diagram The amplifier structural diagram is shown below: The amplifier converts the charge-based signal from a piezoelectric transducer into a current and voltage signal.
Ordering Information To configure a 421B3X-Series amplifier, construct a full model number according to the below scheme. INPUT SENSITIVITY OUTPUT SENSITIVITY FILTERS High- MODEL Value Units Value Units Low-Pass Pass 0001-1000 G/B/P 0001-5000 G/I/M/P/B X5-10 002-200 421B3X 0010-2000 mV/g 002=200 Hz 0001-0200 X5=0.5 1000-5000 mV/in/s...
ESD Warning Information Warning 1 – ESD sensitivity The 421B3X Series Differential Charge Amplifier should be opened by PCB factory-qualified personnel only. This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid injury. Warning 2 –...
Warranty A. Total Customer Satisfaction. PCB guarantees Total Customer Satisfaction through its “Lifetime Warranty Plus” on all Platinum Stock Products sold by PCB and through its Limited Warranties on all other PCB Stock, Standard and Special products. B. Platinum Stock Products – Lifetime Warranty Plus. Under PCB’s Lifetime Warranty Plus, if any PCB Platinum Stock Product ever fails, PCB will repair, replace or exchange the product at no charge.
Return Procedure To expedite returned instrumentation, contact a factory PCB Customer Service Representative for a RETURN MATERIAL AUTHORIZATION (RMA) NUMBER. When calling, please have information such as model, serial number and description of the problem available. When preparing the package for return, it is helpful to provide a written description of the symptoms and problems experienced with the equipment.
Page 16
Model Number Revision: NR DIFFERENTIAL INPUT IN-LINE CHARGE AMPLIFIER 421B30 ECN #: 49931 OPTION AL VERSION S Performance EN GLISH Optional versions have identical specifications and accessories as listed for the standard model except Sensitivity(± 5 %)(Voltage Output) 20.6 mV/pC 20.6 mV/pC [1][2] where noted below.
Page 17
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. RELEASED TO DRAFTING 49931 2.76 [70.0] 2.17 [55.1] 1.77 [45.0] ...
Need help?
Do you have a question about the PCB Piezotronics 421B3 Series and is the answer not in the manual?
Questions and answers