Do you have a question about the CFV-40A and is the answer not in the manual?
Questions and answers
Summary of Contents for KYOWA CFV-40A
Page 1
The product should not be used in any manners other than as described in this Instruction Manual. This product is a copyrighted work of KYOWA ELECTRONIC INSTRUMENTS CO., LTD. All rights reserved. Copying or reproduction of this product, whether in part or whole, is strictly prohibited without permission.
This Instruction Manual is related to the use of CFV-40A/CFV-40A-1, F/V converter card (hereinafter referred to as the F/V card). Although different input connectors are used for the CFV-40A and CFV-40A-1, specifications, etc. are used in common between the two.
PRIOR TO USE For safe use of the F/V card, be sure to read the following precautions. KYOWA ELECTRONIC INSTRUMENTS CO., LTD. assumes no liability for any damages resulting from the user’s failure to comply with the safety precautions. SAFETY SYMBOLS For safety operation of the F/V card, the following “WARNING”...
NOTATIONS USED IN THE INSTRUCTION MANUAL Informational Notes Certain notations are used as necessary to attract your attention to information that requires special care when handing the F/V card, and to information provided for reference purposes. Examples of Notations NOTE Essential precautions required when handling the F/V card.
1. OUTLINE OF THE PRODUCT 1-1 OUTLINE • The F/V card is a converter card designed to convert data of up to 20 kHz frequency into voltage. • Capable of measuring 4 channel data with 1 card. • The F/V card can supply power to sensors. (Approx. 12 VDC with Max. 50 mA) •...
1-4 PIN ASSIGNMENT AND CONNECTION OF CONNECTORS [pin assignment] 1) Input Connector CFV-40A CFV-40A-1 Pin assignment of input connectors (TAJIMI ELECTRONICS made) (LEMO made) (Common to both connectors) Signal Name Sensor power supply (+12 V) Input COM Sensor power supply (Input COM)
Page 11
Monitor output ○: Can use sensor power supply. ×: Cannot use sensor power supply. 1) Connecting CFV-40A <Accessories for CFV-40A> (1) When connecting only signal with a BNC cable, connect the accessory input conversion adaptor FV-1A. How to connect is described in the following figure.
Page 12
<Opitionals for CFV-40A> (2) When connecting sensors that require power supply with a TAJIMI ELECTRONICS made connector, connect the optional input cable U-12. How to connect is described in the following figure. To connect,hold this portion and push it in until it locks.
Page 13
D C 3) Optionals in common to both CFV-40A and CFV-40A-1 (1) When connecting an output cable to a monitor connector, connect the optional output cable U-64. How to connect is described in the following figure. A BNC connector is used for the output cable.
4) After replacing the conditioner card, power ON the EDX-2000A/B and check the system menu screen. If the Slot No. Card Type “Card Type” CFV-40A is displayed on the “Slot No.” CFV-40A column, it means that the conditioner card is properly replaced.
CFV-40A-1 is displayed on the “Card Type” column. (The screen at the right shows that ‘1’ CFV-40A is installed in the slot No. l.) Here, select the cursor to item 2) <Condition> on the system menu screen and press the ENTER key.
Page 16
Condition Setting Screen Select item 2) <Condition> from the system menu to display the screen at the right. Here, select item 1) <Channel Cond> on the <Condition> main menu and press the ENTER key. MEMO If no conditioner card is installed, CH column and items that are not available for setting conditions are indicated with “*.”...
Page 17
Mode Select item 2) <Mode> from the <Channel Cond> menu to display the screen at the right with “MEAS Mode” setting window. Here, select the cursor to ZC/S, ZC/N, or TTL according to amplitude or noise of the predicted input signal and then, press the ENTER key.
Page 18
Range Select item 3) <Range> from the <Channel Cond> menu to display the screen at the right with “Range” selecting window. Here, select any one of the following frequencies according to the predicted input signal. 50 Hz, 100 Hz, 500 Hz, 1 kHz, 2 kHz, 5 kHz, 10 kHz, 20 kHz, and OFF High Pass Filter Select item 4) <Hi Pass>...
Page 19
CAL ON/OFF Select item 5) <CAL ON/OFF> from the <Channl Cond> menu to display the screen at the right with “CAL ON/OFF” setting window. Here, to conduct CAL on the “Measurement” screen, select the cursor to “ON” and if not, to “OFF”...
Page 20
To conduct the CAL on the measurement screen (see the right figure), select the cursor to 4) CAL ON/OFF and and press the ENTER key. The following operation is repeated every time the ENTER key is pressed. Monitoring CAL:ON(+) (CAL OFF) CAL:ON(-) CALrange CAL:ON(+) CAL:ON(-) 100%...
3. SPECIFICATIONS 3-1 HARDWARE SPECIFICATIONS Name ..........CFV-40A / CFV-40A-1, F/V Converter Card Compatible Sensors ......AC signal output sensor Number of Measuring Channels ..4 Input Signal ........AC (Zero Cross), TTL level (including open-collector signal) Comparator Level ......Switching Zero cross (0 V)/TTL level Input Voltage Range ......•...
Page 22
One F/V card can be installed ・EDX-100A-2 Two F/V cards can be installed ・EDX-100A-4, EDX-200A-4H Two F/V cards can be installed ・EDX-5000A-64/80 Two F/V cards can be installed Input connector ........CFV-40A TAJIMI ELECTRONICS made connector type CFV-40A-1 LEMO made connector type...
4. OPTIONALS Input Cable (U-12) <Dedicated for CFV-40A> TAJIMI ELECTRONICS made input connector is necessary for connecting sensors that require power supply. (Length: Approx. 1.5 m) For connection and wiring, see “1-4 PIN ASSIGNMENT AND CONNECTION OF CONNECTORS.” white black...