DAT3692/2 – Hemoflow 400 Trouble Shooting Guide Copy No: Effective date: 16/02/2022 Hemoflow 400: Troubleshooting Guide This guide is designed to be used to resolve Hemoflow 400 issues on session Contents 1. Correct set up and calibration of Hemoflow 400 2. Common Hemoflow 400 related issues 2.1 Overweight Collection (COO)
Dropping the Hemoflow 400 directly onto its tray will cause damage to the load cell by pushing the barb connectors upwards and directly affect the weighing function Calibration of a Hemoflow 400 is not a weight check, the user is telling the machine what 500g equals ...
2.1- Overweight Collection (COO) Why might this occur on session? The Hemoflow 400 is a sensitive medical device. Its ability to weigh collection packs accurately can be affected by a number of things such as: Incorrect equipment set up...
Ensure the donation chair is set up on an even surface and is steady on the floor, check the chair table is also steady If a tilt alarm sounds, move the Hemoflow 400 to sit on an even surface such as the edge of a table or desk ...
2.7- Time and date/calibration in the middle of session The time displayed on a Hemoflow 400 must not be used to record any times on session paperwork If the time is out by a small margin (for example after the clocks change) this does not require adjusting and the device can continue to be used ...
DAT3692/2 – Hemoflow 400 Trouble Shooting Guide Copy No: Effective date: 16/02/2022 3.0- For information – Diagram of Hemoflow 400 parts Pinch Motor Assembly Barb Connectors Pinch Motor Covers Pinch Cam Pinch Motor 4.0 - Other Training and Guidance Material...
Need help?
Do you have a question about the Hemoflow 400 and is the answer not in the manual?
Questions and answers
what is the acceptable range when performing the daily quality controlof the hemoflow 400sx