AUTOMATIC Summary Introduction ............................... 8 Preface ............................... 8 Use in accordance with provisions ....................8 Guarantee ............................8 Exclusion of liability ........................... 8 Indications ............................8 Counter indications ........................... 8 Version information ........................... 9 Symbols in the manual ........................9 Manufacturer contacts ........................
Page 5
AUTOMATIC 5.2.3 Inserting and removing the batteries ..................26 5.2.4 Recharge procedure ........................ 27 Defibrillation PADs ........................... 28 5.3.1 Defibrillation PADs for Adults SAV-C0846 ................28 5.3.2 PADs for Children SAV-C0016 ....................28 5.3.3 Positioning of defibrillation PADs .................... 29 Memory Card and USB ........................
Page 6
AUTOMATIC 10.3 Devices ............................. 49 10.4 Geo Saver ............................50 10.5 Saver Log ............................51 10.6 Geoloc Log ............................52 10.7 Streaming............................53 10.8 List of saved ECG Streaming ......................54 10.9 Tracks ............................... 55 Maintenance ............................56 11.1 After each use ..........................56 11.2 Ordinary maintenance ........................
Page 7
AUTOMATIC 13.3 Recommended separation distance between portable and mobile RF communication equipment and Geo Saver device ..........................69 Simbology ............................70 Certifications ............................71 15.1 EC Certificate ........................... 71 15.2 IMQ Brand ............................73 Geo Saver Series Defibrillator Warranty ..................... 74 Product registration ..........................
Introduction 1.1 Preface Thank you for having chosen the defibrillator of A.M.I Italia S.r.l. model Geo Saver. So that you can correctly use the device it is necessary, before usage, to carefully read this user manual. The User Manual of Geo Saver contains the instructions for its use in compliance with its function and purpose. For a function free of error and to achieve the right benefits, it is fundamental to respect the prescriptions indicated in this user manual, to guarantee the safety of the patient, of the rescuer and of any third parties.
AUTOMATIC 1.7 Version information This user manual has a version number. The version number changes every time the manual is updated for changes made to the function of the device or to the device itself. The contents of this user manual are subject to change without notice.
AUTOMATIC Safety instructions For a correct use of the Geo Saver defibrillator, users must be aware of the safety factors listed below. Please read them carefully. The Geo Saver defibrillator, individually and in connection with its standard and optional (original) accessories, complies with the safety regulations currently in force and is in compliance with the provisions of the directives on medical products.
AUTOMATIC In a domestic environment, keep the defibrillator out of the reach of children and pets. Do not apply the defibrillation electrodes directly on an implanted pacemaker to avoid any errors in the interpretation of the device and to avoid damage to the pacemaker through the defibrillation impulse.
AUTOMATIC Description of the device 3.1 Device Information Geo Saver Automatic is called PAD or Public Access Defibrillator. Geo Saver Automatic is a FULLY AUTOMATIC external defibrillator meaning that if an arrhythmia that requires a shock is detected the device will automatically deliver the defibrillation shock. Designed to automatically detect and analyze the victim's heart rhythm, it is capable of delivering one or more defibrillation shocks if ventricular fibrillation or ventricular tachycardia (monomorphic or polymorphic with beat>...
Page 13
AUTOMATIC Compared to a classic defibrillator, the main differences are: Remote control Telemetry Geo-location Teleservice Management of AEDs and AEDs’ users Updates through remote connection Remote control Thanks to the AMISAVERCLOUD it is possible to control the device quickly and easily. By accessing the section dedicated to each Geo Saver it is possible to configure each aspect by sending simple commands and knowing the last ones sent.
AUTOMATIC 3.2 Service activation procedure In order for the Geo Saver series defibrillators to fully provide the features for which they were designed, it is necessary to follow the service activation procedure. Check that the SIM card (*) supplied is correctly inserted; Make sure you are in an area covered by a GSM / GPRS signal;...
AUTOMATIC Description of device details 4.1 General structure of the device Image 1 Description Compartment for PADS connector or ECG cable Live call button Microphone for environmental recordings Status mini display Carrying handle Keyboard with buttons and light icons IrDA port (service only, present only in the models with TFT display) Loudspeaker Live call microphone...
AUTOMATIC 4.2 Keys, icons and indicators Image 3 Function Function “Patient Type” Indicator Indicates the type of mode in use: Product logo Adult if you insert Adult PADs Indicates the model of the device Child if you insert pediatric PADs "Place PADs"...
AUTOMATIC 4.4 Standard and optional accessories of the device The Geo Saver defibrillator comes with the following standard accessories: Code Image Quantity Description Geo Saver 200J SGA-B0990 1 Unit (Version 200J or 360J) Geo Saver 360J SGA-B0991 Adult Pads SAV-C0846 1 Unit Li-SOCl SAV-C1032...
Page 18
AUTOMATIC The following are the optional Geo Saver accessories that can be purchased separately: Code Image Quantity Description 1 Unit SAV-C1033 ACC Rechargeable Li-Ion battery SAV-C1035 1 Unit Charger SAV-C1037 1 Unit GS40A15-P1J Power supply N.01 Charger SAV-C1034 1Unit N.01 GS40A15-P1J Power supply (Contains 3 units) N.01 Power supply cable SAV-C0016...
AUTOMATIC Parts and accessories of the Geo Saver 5.1 AED Batteries The Geo Saver defibrillator can work with two different types of batteries: (SAV-C1032) Non-rechargeable Li-SOCl battery (SAV-C1033) Rechargeable Li-Ion battery 5.1.1 Non-rechargeable Li-SOCl battery (SAV-C1032) The non-rechargeable battery with Li-SOCl technology (SAV-C1032) is supplied fully charged and ready for use.
5.1.3 Suggestions for a proper maintenance of rechargeable battery SAV-C1033 A.M.I Italia recommend that batteries SAV-C1033 left in a “storage stage” to be fully recharged at least every 4 months from the receipt of the goods and to be recharged regularly every 4 months when attached to the device "ready to use"...
AUTOMATIC 5.1.4 Inserting and removing the batteries To be able to operate the Geo Saver the insertion of a battery is required. Below are detailed instructions for correctly installing the batteries (rechargeable or non-rechargeable) in the Geo Saver . Image 6 Image 7 Image 8 ...
AUTOMATIC 5.1.5 Recharging station for rechargeable batteries The charging station (SAV-C1034) allows you to recharge rechargeable batteries with Li-Ion technology ACC model (SAV-C1033) of the Geo Saver. The charging station consists of the following parts: Charger (SAV-C1035) image (11) ...
AUTOMATIC 5.1.7 Recharge procedure Place the charger on a perfectly horizontal shelf and firmly attached to the floor Connect the power supply (SAV-C1037) to the charger and then to the power outlet The LED on the charger will flash green, indicating that it is ready to charge Insert the battery to be charged into the battery charger as shown in image (15) Image 15 The recharging station allows you to recharge exclusively original ACC rechargeable Li ion batteries (SAV-C1033) of...
AUTOMATIC 5.2 Geoloc Batteries The Geoloc Geo Locator can work with two different types of batteries: (SAV-C1038) Non-rechargeable Li-SOCl battery (SAV-C1039) ACC Rechargeable Li-ion battery 5.2.1 Non-rechargeable Li-SOCl battery (SAV-C1038) The non-rechargeable battery with Li-SOCl technology (SAV-C1038) is supplied fully charged and ready for use. The Li-SOCl non-rechargeable battery has been designed to have a long battery life and no maintenance whatsoever.
AUTOMATIC 5.2.2 Rechargeable Li ion battery (SAV-C1039) The rechargeable battery with Li-ion technology (SAV-C1039) of the Geoloc is recommended for those who intensively use the unique features of the Geo Saver- Geoloc -AMISAVERCLOUD system (Streaming, Speakerphone calls, tracking). Being rechargeable, it allows operators to reduce management costs and guarantees a greater number of connections.
AUTOMATIC 5.2.3 Inserting and removing the batteries In order to function, the Geoloc module requires the insertion of a battery. Below are the detailed instructions for correctly installing the batteries (rechargeable or non-rechargeable) for the Geoloc module inside the Geo Saver. Image 18 Image 19 Image 20...
AUTOMATIC 5.2.4 Recharge procedure Remove the rechargeable battery (SAV-C1039) from the Geo Saver Connect the battery charger (SAV-C1040) to the power outlet Hook the connector of the battery charger / power supply into the battery as shown in image (23) Image 23 The recharging station allows recharging exclusively original ACC rechargeable Li ion batteries (SAV-C1039) of A.M.I.
AUTOMATIC 5.3 Defibrillation PADs The Geo Saver allows the use of two different defibrillation PADs depending on the patient to be treated: Defibrillation PADs for Adults model SAV-C0846 Defibrillation PADs for Children model SAV-C0016 5.3.1 Defibrillation PADs for Adults SAV-C0846 The SAV-C0846 defibrillation PADs are pre-gelled disposable types.
AUTOMATIC 5.3.3 Positioning of defibrillation PADs The correct placement of the PADs is essential for an efficient analysis of the patient's heart rhythm and for the consequent delivery of the shock (if necessary). Always refer to the instructions given both on the packaging of the PADs and directly on each individual pad. The PADs of the Geo Saver are polarized type;...
AUTOMATIC 5.4 Memory Card and USB The Geo Saver allows the recording of data on the internal memory as well as on the external memory card. Supported memory cards are SD/SDHC cards with capacities up to 8GB Image 24 To install a Memory Card in the Geo Saver follow this procedure: A.
AUTOMATIC Auto test The Geo Saver has been designed to be a totally safe device, always ready for use and able to automatically and constantly verify correct operation, minimizing maintenance operations. The Geo Saver performs different types of self-tests: Activation: Every time a battery is inserted in the device ...
AUTOMATIC 6.2 ACTIVATION Test The Geo Saver performs functional tests only if the battery is installed. Each time a battery is inserted, the device will perform a diagnostic ACTIVATION test. During this test the device performs a complete control (firmware/hardware), which involves a consumption of the battery equal to a shock, therefore it is advisable, once performed, not to remove the battery from the device.
AUTOMATIC 6.4 POWER ON Test The Geo Saver performs self-diagnostic tests each time it is turned on, either from the power button or by inserting the battery. Following to the battery insertion a diagnostic test of ACTIVATION is performed. This self-test involves a fair amount of energy consumption.
AUTOMATIC Information button The Geo Saver is equipped with an “ ” button, through which the operator can view various useful information about the device in use on the LCD display. This button can only be used when the device is switched on and is automatically disabled in the event of an ongoing emergency.
AUTOMATIC Defibrillation If you need to help a person with sudden cardiac arrest, remember to implement the sequence of actions recommended by ERC and AHA 2017. The ERC has established a rescue protocol to be respected during the resuscitation of a person suffering from sudden cardiac arrest.
AUTOMATIC 8.2 Preparation of the patient In order to correctly position the defibrillation PADs on the chest, it is necessary to carry out the following preliminary operations: Image 28 Remove clothing from the patient's chest If the patient's chest has thick hair it might be necessary to shave Pads positioning points.
AUTOMATIC 8.4 Cardiac rhythm analysis The Geo Saver automatically analyzes the patient's heart rhythm. This phase is highlighted by the following commands: Voice commands Display Luminous Icons Keyboard Icon do not touch Do not touch the patient the patient on with fixed light Cardiac rhythm analysis in Icon type of pads...
AUTOMATIC 8.5 Shockable rhythm If a Ventricular Fibrillation or Tachycardia is detected, the Geo Saver will inform the operator using the following commands: Voice commands Display Icons/ Luminous buttons Do not touch the Shock recommended patient icon on fixed light Keep distance, charging Keep distance, the shock will be automatically delivered in 5 (five)
AUTOMATIC 8.6 Change of rhythm The Geo Saver analyzes the patient's heart rhythm continuously, during the resuscitation phase. If the device after recommending the shock detects a change in the heart rate of the patient who no longer needs a defibrillation, it will carry out the automatic disarmament.
Page 40
AUTOMATIC The following table shows the main operations to be performed and the relative visual-text-vocal commands provided by Geo Saver Type of command Instruction Operations to be performed (Geo Saver) Geo Saver A. Make sure the patient is lying on a flat surface B.
AUTOMATIC Recording, viewing and archiving the data The Geo Saver defibrillator is able to record and store both the SERVICE data of the device and the complete data of the rescue operations carried out. Data recording and archiving is done automatically (cannot be deactivated by the user) both on the internal memory of the device and on the memory card when installed.
AUTOMATIC Web Management AMISAVERCLOUD is the platform through which a user can check the status of its Geo Saver. To access the functions of AMISAVERCLOUD there is a mechanism of authentication with password. When a Geo Saver is purchased, a new user is generated in AMISAVERCLOUD by referring to an e-mail provided by the customer.
AUTOMATIC 10.1 Main features 10.1.1 Changing the language AMISAVERCLOUD is a multilingual cloud service. You can select one of the languages available through the specific icon on the top right. 10.1.2 User profile and password The current user has the possibility to view his profile and modify the password via the links located on the menu.
AUTOMATIC 10.1.3 Device management Each user can be associated with one or more Geo Savers. These are listed under the "DEVICES" menu item and displayed on a map that briefly summarizes their status. It is possible to view the details of each device by following the link in the table (or inside the map) for each of them. When viewing a Geo Saver detail, the menu is expanded with the following items: ...
AUTOMATIC 10.1.4 ECG Streaming To activate the ECG transmission, switch on the Geo Saver and then enable ECG transmission. While the ECG is streaming from the Geo Saver, a notification is displayed on the cloud and the "STREAMING" section becomes accessible as a "GEO SAVER" sub-level of the menu. The device in the ECG check phase is highlighted in the Geo Saver table and a link to the section appears on the same page.
AUTOMATIC 10.1.5 Auto tracking Among the command that can be sent to the Geo Saver through the cloud there’s enable/disable Auto tracking. This functionality can be found in the GEOSAVER page –GEOLOC section. When this function is active the Geo Saver, after having detected a movement, will start sending its positions to the cloud.
AUTOMATIC 10.1.7 Firmware update If AMI Italia makes new firmware available, the user is notified with an icon on the device table. If a device has a less recent version of the firmware than the last one released on the line corresponding to the Geo Saver in question, an icon will be displayed, by clicking that you are taken to the device page.
AUTOMATIC 10.2 Dashboard After logging in to the cloud you are redirected to the Dashboard. In this page there is a map with all Geo Savers devices -linked to the current users- located through a coloured icon. The colour of the icon is representative of the state of the Geo Saver.
AUTOMATIC 10.3 Devices In the section dedicated to the Devices, all the devices to which the current user has access are shown in a paged table. These devices can be filtered by: Membership user Range of serial numbers of the Saver ...
AUTOMATIC 10.4 Geo Saver The section dedicated to the specific Geo Saver can be divided into three parts At the top enabling / disabling of alarms On the left everything related to the Saver On the right everything related to the Geoloc (concealed) In the area relating to the Saver we can distinguish: Collapsible panel for sending a command Collapsible panel with the last commands sent and relative completion status...
AUTOMATIC 10.5 Saver Log The section dedicated to logs received from the Saver presents a list of logs that can be expanded by clicking on them with the mouse. On the left side of the page there is a box where it is possible to filter the list of logs arrived by arrival date. An additional filter can be added selecting only the logs with errors and / or warnings.
AUTOMATIC 10.6 Geoloc Log The section dedicated to logs received from the Geoloc presents a list of logs received which can be expanded by clicking on them with the mouse. On the left side of the page there is a box where it is possible to filter the list of logs arrived by arrival date. An additional filter can be added by selecting only the logs with errors.
AUTOMATIC 10.7 Streaming During the streaming phase of a Geo Saver the link to this section will become visible, presenting: A graph showing the streaming in real time Some relevant data on the current stream The GPS position on a map During the streaming, a Growl notification is also presented with the link to this section and a red panel on the Geo Saver page.
AUTOMATIC 10.8 List of saved ECG Streaming The section dedicated to the list of ECG streams for the specific Geo Saver is made visible only in the presence of at least one file and presents: A graph to playback the streaming ...
AUTOMATIC 10.9 Tracks The section dedicated to the saved tracks for the specific Geo Saver is made visible only in the presence of at least one file and presents: A map A filter to select the set of dates of interest ...
AUTOMATIC Maintenance The Geo Saver defibrillator was designed to make maintenance operations as simple and autonomous as possible. In fact, thanks to the control tests carried out in total autonomy by the device, it is not necessary to perform any extraordinary maintenance, but only routine maintenance which consists of a frequent visual check of the LED and the control display, together with a visual inspection of the relative accessories.
AUTOMATIC 11.3 Cleaning The structure of the Geo Saver defibrillator, including the connection port of the defibrillation electrodes, can be sanitized using a soft cloth dampened with one of the cleaning solutions listed below: Isopropyl alcohol (70% solution) Soap water Bleach (30 ml per liter of water) Detergents containing ammonia Detergents containing glutaraldehyde...
AUTOMATIC 11.5 Troubleshooting Guide The following table lists the symptoms, the possible causes and the possible corrective actions of the problems that may arise. For more information on the implementation of corrective actions, refer to the other sections of the user's manual. If the failure of the unit persists, request assistance.
AUTOMATIC Technical specifications The technical specifications of the Geo Saver defibrillator, its parts and accessories are shown below. 12.1 Physical characteristics Category Nominal specifications Dimensions 29,5 x 23,0 x 11,5 cm With battery Li-SOCl 2,54 Kg + Adult Pad (2,63 Kg) (SAV-C1032) Weight With battery Li-Ion...
AUTOMATIC 12.4 Alarms table Priority Cause Visual signal HIGH Device ready to deliver the shock LED Shock Icon flashing HIGH Low battery (<1% capacity) Control LED flashing 12.5 Controls and indicators Category Nominal specifications ON / OFF button (device switching on and off) Buttons “i”...
AUTOMATIC 12.7 Defibrillator Category Nominal specifications Waveform Biphasic Truncated Exponential (BTE) The waveform parameters are automatically adjusted according to the patient's impedance. In the graph on the left t represents the duration of phase 1 (ms), t represents the duration of phase 2 (ms), t is the delay between phases, U indicates the peak voltage, t...
AUTOMATIC 12.9 Patient analysis system Category Nominal specifications Determines the patient's impedance and evaluates the ECG rhythm and signal quality to Function determine whether or not the shock delivery is appropriate. 20 - 200 Ω Impedance range ≥4 seconds (with new fully charged battery) respecting IEC/EN 60601-2-4 ECG analysis time 97% Respects the guidelines IEC/EN 60601-2-4 2002(3) sources AHADB, MITDB and EDB Sensibility...
AUTOMATIC 12.12 Rechargeable battery Category Nominal specifications REF (Model) SAV-C1033 Type Li ion (lithium ions) Rechargeable 21,6 VDC - 2100 mAh Voltage Capacity Standard 200J 250 continuous shocks Power 360J 160 continuous shocks ECG analysis 21 continuous hours ≤ 2,5 hours with charging station SAV-C1035 Charging time Shelf Life 2 years or 300 charge/shock cycles (the one that occurs first)
AUTOMATIC 12.16 Timing of Shock cycles Charging time performance in accordance with 60601-2-4 (201.101) Specific Result The maximum time between the beginning of the ECG rhythm analysis and the completion of < 30 seconds the charge at maximum energy The maximum time from turn on to completion of the charge at maximum energy <...
AUTOMATIC Compliance with electromagnetic emission standards The following paragraphs will specify the compliance with electromagnetic emission standards: Guidelines and manufacturer's declaration - Electromagnetic emissions Guidelines and manufacturer's declaration - Electromagnetic immunity Recommended distances between portable and mobile radiofrequency communication equipment and the AED 13.1 Guidelines and manufacturer's declaration - Electromagnetic emissions The Geo Saver was designed to be used in electromagnetic environments with features listed below.
Page 68
AUTOMATIC Test level Immunity test Compliance level Electromagnetic environment - Guidelines IEC/EN 60601-1 Supply frequency Power frequency magnetic fields must be at levels no (magnetic field) higher than those of stations located in typical heavy 50/60 Hz 3 A/m 80 A/m industrial applications, power plants and control rooms of high voltage substations.
AUTOMATIC 13.3 Recommended separation distance between portable and mobile RF communication equipment and Geo Saver device The Geo Saver must be used in an electromagnetic environment in which radiated RF interference is controlled. The customer or the operator of the Geo Saver can help prevent electromagnetic interference by maintaining the minimum distances recommended below, between the portable and mobile RF communications equipment (transmitters) and the Geo Saver, based on the maximum output power of the devices of communication.
AUTOMATIC Simbology ILCOR Universal Symbols for AED IMQ brand High Voltage Electrical Hazard CE mark with identification number General Notices: Refer to the consultation IP56 Degree of protection of the appliance against of accompanying documents before using dust and water (including battery) the appliance Type BF, Defibrillation-proof Equipment Serial Number...
AUTOMATIC Geo Saver Series Defibrillator Warranty 1 Warranty Restriction This warranty does not cover, as it does not constitute cases of original A.M.I. Italia Srl guarantees the original purchasers that its Geo Saver non-conformity, the normal wear and tear of components subject to decay series defibrillators and related accessories and batteries are free from any during use such as buttons, LEDs and battery contacts.
AUTOMATIC Product registration In order to guarantee a correct and rapid traceability of the product sold, we ask you to complete the form below and send it by fax or registered letter to A.M.I. Italia S.r.l.