TABLE OF CONTENTS INTRODUCTION ....................................- 1 - Necessary information for the patient and those around them ........- 1 - Advice for use ..........................- 1 - Indications ............................- 2 - Counterindications ......................... - 2 - Precautions ............................- 2 - Warning..............................
WINNCARE FRANCE INTRODUCTION ECESSARY INFORMATION FOR THE PATIENT AND THOSE AROUND THEM This product is a support to help prevent and treat bedsores. Why have you been prescribed this support? Your state of health impairs your mobility and puts you at risk of bedsores. What is a bedsore? A bedsore is a more or less skin deep lesion caused by putting excessive and prolonged pressure on tissues between the body and a load bearing surface.
WINNCARE FRANCE NDICATIONS According to the CNEDiMTS (the French National Committee for Evaluation of Medical Devices and Health Technologies) report of 22 December 2009 and in accordance with expert medical opinions: Prevention and aid in the treatment of bedsore(s) in stages 1 to 4 (according to medical opinion) for patients up during the day, in bed for more than 15 hours and presenting an "average"...
WINNCARE FRANCE • In order to comply with risk assessment under ISO standard 14971, an assessment was carried out on all "hospital bed - therapeutic mattresses (Axtair) - accessories". The risk assessment shows that there may be a risk of bodily trapping if the bedridden person presents mental confusion and/or agitation.
WINNCARE FRANCE ACKAGE ONTENTS USHION • 1 cushion • 1 AXENSOR® system placed in the cushion cover • 1 user manual PERATING PRINCIPLE • Alternating pressure helps to avoid prolonged vascular compression which is likely to lead to tissue hypoxia. •...
WINNCARE FRANCE Connect the rapid inflation pump to the CPR. Keep the CPR and the pump on the same line to avoid jamming the CPR with the weight of the pump. Improper use may lead to reduced airtightness of the CPR. Press down on the rapid inflation button on the pump until the mattress is inflated (around 1 minute) - Note: The level of inflation is not important as this will be adjusted automatically by the compressor.
Page 9
WINNCARE FRANCE NOTES: • Make sure the size of the cushion is compatible with the chair seat. Where applicable, check the height adjustment of the chair armrests and footrests after having seated the user. • To take into account the parameters for the new support, the compressor must be restarted by disconnecting and reconnecting the electric cable.
WINNCARE FRANCE Indicates the compressor is on Flickers to indicate inflation of the mattress After it has gone off, the patient can be placed on it When this light is lit, it indicates that the keypad is locked: it is not possible to change mode.
WINNCARE FRANCE : CPR (M MERGENCY CARDIOPULMONARY RESUSCITATION ATTRESSES In case of heart stoppage: Turn the end of the CPR valve to the "OPEN" position. The mattress deflates and the back of the thorax will be flat on the base of the bed in less than 15 seconds in order to allow External Cardiac Massage to be carried out.
WINNCARE FRANCE EMOVING ATTRESSES Unplug the compressor from the mains Open the CPR valve to deflate the mattress. Total removal of air from the mattress can be facilitated by use of the inflation/deflation pump connected to the CPR. Position the compressor in the centre of the mattress. Roll the mattress up, starting at the feet Strap up the mattress.
Page 13
WINNCARE FRANCE WARNING: • Keep the device away from sources of humidity, direct sunlight and constant heat sources. • Prevent the risk of shocks or damage by sharp objects. NOTE: • The MANUFACTURER shall make available, on request, the circuit diagrams, lists of components, descriptions, calibration procedures and any other useful information to UPKEEP and MAINTENANCE STAFF to carry out authorised repairs in accordance with the contract which ties the requesting party to the manufacturer, ASKLE.
WINNCARE FRANCE AXENSOR® CONNECT: Details on the operation of the compressor can be transmitted on a PC via a Bluetooth connection. Visit the site www.winncare.fr , AXENSOR® CONNECT menu to download the software and find out how it works. When you stop using the software, check that the compressor starts up again in one of the three operating modes (DYNAMIC, STATIC or CARE).
WINNCARE FRANCE ATTRESSES AND USHIONS COVERS On Mattresses, two zips located inside the cover at the foot and head of the mattress allow the elements to be quickly separated. The cells and the AXENSOR® system remain attached and must be manually disinfected: use a dishcloth doused in a CE certified surface detergent or detergent/disinfection solution at the concentration recommended for use by the manufacturer.
WINNCARE FRANCE IAGRAM OF ADVISED MAINTENANCE AND DISINFECTION Receipt of Full Device Pre-Treatment of surface (Decontamination) Spread over surface by µfibres Alcohol-based solution [Recommended: hygiene mask, gloves, apron] Removal by the waste Cover changed? industry Replacement cover identical Physical cleanliness to the original protection Unacceptable Bio-cleaning surface maintenance...
WINNCARE FRANCE ALARMS TROUBLESHOOTING The following table shows all the device alarms, what triggers them, their severity as well as possible troubleshooting. Signal Severity Trigger Possible troubleshooting Condition Time perio Flickering light 1- Plug the compressor back in alarm 2 - Check that state of the Electrical fault: and presence of electrical grid...
WINNCARE FRANCE • Sound and visual alarms are designed to be visible and audible from 2 metres away from the compressor. • In order to make sure the over-pressure and leakage alarms are working on cushions connected to compressors having a batch number lower than 1431, we recommend updating the embarked software: for more information on this update procedure, contact Asklé.
WINNCARE FRANCE Flickering static and Pneumatic problem dynamic light Key to need for Meaning check: Signal Recommended date of revision past Timer flickering (operating time higher than 17,500 hours since quickly last revision) Timer flickering Revision scheduled within 2 months slowly Timer fixed Revision scheduled in over 2 months...
Page 20
In case of removal of this label by the user or an unauthorised third party, the integrity of the compressor can no longer be proven; ASKLE SANTE may refuse to carry out the repair under guarantee, and is authorised to cancel the latter.
WINNCARE FRANCE TECHNICAL FEATURES MATTRESSES SUPPORT FEATURES VAXT6/MA/AUTO VAXT6/MA/MAX Validated user weight – – 165 kg 180 kg Weight of support 8 Kg 10 kg Size of support 195 x 87 x 17 cm 195 x 87 x 20.5 cm Therapeutic air cells Number: 18 Number: 18...
Page 22
WINNCARE FRANCE COMPRESSOR FEATURES Operating modes - Alternating: 1 cell out of 2 - Static low pressure - Care: renewable, lasting 30 minutes Weight of compressor 2.7 Kg Size of compressor 22 x 25 x 11.5 cm Pressure adjustment Automatic Cycle times Between 17 and 20 mins Pump flow (indicative)
The Axtair AUTOMORPHO AXENSOR® requires special precautions in terms of electromagnetic compatibility and must be installed and commissioned in accordance with the EMC information provided on demand by ASKLE SANTE, and which are available on its website. Portable and mobile radio frequency communications equipment can affect the Axtair AUTOMORPHO AXENSOR®.