D M 3 S LT R AV I E W E R V I C E A N U A L Wireless Setup Overview ..........................3-1 Unpacking and Inspection ....................3-1 Before starting........................3-1 Wireless Adapter Setup.......................3-2 Wireless Adapter Hardware Installation ................3-5 DM3 Network Communication Setup .................3-7 Troubleshooting Network Connection Issues ...............
Page 5
D M 3 S LT R AV I E W E R V I C E A N U A L Spacelabs SpO2 Messages .................... 7-4 Troubleshooting Flow Chart: Temperature ..............7-6 Service Tools Appendix A — Symbols www.spacelabshealthcare.com I I I...
Page 6
D M 3 S LT R AV I E W E R V I C E A N U A L www.spacelabshealthcare.com...
D M 3 S LT R AV I E W E R V I C E A N U A L Introduction Overview Spacelabs Healthcare products are designed and manufactured under good manufacturing practices and in compliance with all applicable regulatory requirements.
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N • Telephone support • Loaner equipment • Factory service For information on any of these services, contact Spacelabs Healthcare Technical Support at (800) 522-7025.
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N Warnings, Cautions, and Notes The formats below show warnings, cautions, and notes throughout this manual. Be sure to read all warnings, cautions, and notes included in each section of this manual.
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N Electrical Specifications Designated Class II, Type BF (SpO Type CF) by applicable electrical safety standards.
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N Battery • BATTERY HANDLING. The Ultraview DM3 monitor contains a lithium ion coin cell battery and transport battery pack.
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N Operating time eight hours (minimum) Charging time three hours to 90% (four hours to full charge) Replacement Recommended replacement interval: two years Safety Tips:...
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N EN 1060-3:1997 EN 1060-4:2004 EN ISO 9919:2005 ANSI/AAMI SP10:2002; SP10/A1:2003; and SP10/A2:2006 Display Type Active matrix widescreen LCD...
Page 14
D M 3 S LT R AV I E W E R V I C E A N U A L N T R O D U C T I O N 1 - 8 www.spacelabshealthcare.com...
D M 3 S LT R AV I E W E R V I C E A N U A L Setup Menus Overview This chapter describes how to use the Ultraview DM3 setup menus to review and change monitor settings. Ultraview DM3 monitors ship with factory defaults, but you can change them to meet the needs of patients or to comply with protocol for your facility.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-1 Monitor Setup window in SPOT mode Figure 2-2 Monitor Setup window in MON mode Viewing Setup Categories From the Monitor Setup window, you can access functions and submenus in three categories.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Clinical Settings To gain access to the Clinical Setup submenus from the Monitor Setup window Touch Clinical.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Alerts (for SPOT mode only) To adjust settings for alerts from the Clinical Setup window Touch the Alerts button.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S To enable or disable alerts for a setting Touch On or Off. For High Limit To set the high limit for the setting, above which an alert is generated, touch the arrow keys.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S System To set system-wide parameters from the Clinical Setup window Touch the System button. Figure 2-6 System Setup window These settings are available on the System Setup window.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Table 2-2 System Setup Settings Setting Description Save Monitor Defaults Saves the current configuration settings and uses them as the power-on default user settings, instead of the factory-default settings.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-7 NIBP Initial Inflation window These settings are available on the NIBP Initial Inflation window. Table 2-3 NIBP Initial Inflation Default Settings Setting Defaults...
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-8 NIBP Protocol 1 Setup window (MON mode) To change NIBP protocols for each time interval To set the number of readings required, touch the arrow keys.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-9 Configuration window Alarms (for MON mode only) To adjust settings for alarms from the Clinical Setup window Touch the Alarms button.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-11 Alarms Setup window, second window These settings are available on the second page of the Alarms Setup window. Table 2-4 Alarms Setup Settings Setting Description...
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-12 Print Setup window These settings are available on the Print Setup window. Table 2-5 Print Setup Settings Setting Description...
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S If all modifiers except Patient ID are set to Disabled, when Save is touched to store a measurement record, the Save Spot Check Data - Supplemental window is bypassed.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-14 Biomedical Setup window These settings are available on the Biomedical Setup window. Table 2-6 Biomedical Setup Settings Setting Description...
Page 29
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Figure 2-15 Communications Setup window These settings are available from the Communications Setup window. The settings are used to establish connectivity with ICS G2 (version 4.02.02 or newer) for transmission of vital sign measurements.
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Table 2-7 Communications Settings Setting Description IP address of your DNS server. (This button is disabled if DHCP is On.) DHCP Touch On if the device receives its network addressing information from DHCP.
Page 31
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S Table 2-8 Service Setup Settings Setting Description Simulated data mode Touch On or Off to enable or disable the display of simulated data to perform diagnostic procedures on the monitor.
Page 32
D M 3 S LT R AV I E W E R V I C E A N U A L E T U P E N U S 2 - 1 8 www.spacelabshealthcare.com...
D M 3 S LT R AV I E W E R V I C E A N U A L Wireless Setup Overview The Wireless Adapter Kit (040-1644-00) allows the Ultraview DM3 monitor to wirelessly transmit patient data across the hospital network for viewing in ICS G2 Clinical Access Trends, or in a hospital EMR via ICS HL7 interface.
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Table 3-1 Kit Contents: 040-1644-00 (Continued) Description Quantity Part Number MOUNT, BRACKET, WLAN ADAPTER, 010-1970-00, 91330 407-0915-00...
Page 35
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Figure 3-1 Web GUI login screen The Setup Wizard is then shown. Selected Network Adapter mode. Figure 3-2 Operating mode selection screen The Network Adapter page shows.
Page 36
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Figure 3-3 Network adapter configuration page A wireless connection login window shows. Enter the proper security/ authentication information and touch Connect.
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Wireless Adapter Hardware Installation Using the two M3X6 screws, attach the bottom half of the wireless housing to the mounting bracket.
Page 38
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P With the status LEDs facing up, insert the Wireless Adapter into the top half of the wireless housing.
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P DM3 Network Communication Setup Before the DM3 can transmit data to the ICS G2 server, the DM3 Communication settings must be configured.
Page 40
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P • The Encryption setting is disabled by default. Encryption is only available for DM3 monitors that are authenticated on a domain controller that is running Microsoft Windows Server 2003.
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Troubleshooting Network Connection Issues If DM3 connection to the ICS server is not successfully achieved, consult the following information for possible assistance in resolving the problem.
Page 42
D M 3 S LT R AV I E W E R V I C E A N U A L I R E L E S S E T U P Table 3-2 Troubleshooting (Continued) Symptom Possible Causes Things to Try The network status icon shows the Incorrect software version.
D M 3 S LT R AV I E W E R V I C E A N U A L Routine Maintenance Overview The Ultraview DM3 monitor does not require routine service other than cleaning, battery maintenance, and service activity required by your facility. Only qualified service personnel perform periodic inspections of the monitor.
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Recommended Maintenance Procedures Table 4-1 shows the recommended maintenance procedures for the Ultraview DM3 monitor and accessories.
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Cleaning • Use only recommended cleaning solutions, or you can void the warranty from the manufacturer.
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Monitors, Cables, and Printers To clean the exterior of monitors, cables, and printers Prepare the cleaning solution according to the instructions from the manufacturer.
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Touchscreen Clean the touchscreen with a soft cloth moistened with either 70% isopropyl alcohol solution or soapy water.
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Figure 4-2 91330 Accessories and Part Numbers Item Accessory Part Number...
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Table 4-2 Replacement Kit Part Number Battery Replacement 019-0547-xx...
Page 50
D M 3 S LT R AV I E W E R V I C E A N U A L O U T I N E A I N T E N A N C E Figure 4-4 Recorder with paper correctly installed 4 - 8 www.spacelabshealthcare.com...
D M 3 S LT R AV I E W E R V I C E A N U A L Functional Verification Overview This chapter describes the procedure to verify that the Ultraview DM3 monitor functions correctly. A qualified service technician checks the monitor for acceptable performance in 12-month interval.
D M 3 S LT R AV I E W E R V I C E A N U A L U N C T I O N A L E R I F I C AT I O N The equipment listed is optional: •...
Page 53
D M 3 S LT R AV I E W E R V I C E A N U A L U N C T I O N A L E R I F I C AT I O N Set up the Fluke 2XLFE simulator as follows: Power ON the simulator.
D M 3 S LT R AV I E W E R V I C E A N U A L U N C T I O N A L E R I F I C AT I O N Table 5-1 Saturation levels for SpO accuracy test Simulator Setting...
D M 3 S LT R AV I E W E R V I C E A N U A L U N C T I O N A L E R I F I C AT I O N NIBP Calibration/Leakage Test To check NIBP calibration and leakage Connect the NIBP test fixture to the NIBP port on the monitor, as shown in...
Page 58
D M 3 S LT R AV I E W E R V I C E A N U A L U N C T I O N A L E R I F I C AT I O N Note: Tubing is sold by the foot.
D M 3 S LT R AV I E W E R V I C E A N U A L Software Installation Overview This chapter tells how to install new software versions if needed. Preinstallation Before you update the monitor software Power ON the monitor.
D M 3 S LT R AV I E W E R V I C E A N U A L O F T W A R E N S TA L L AT I O N Figure 6-1 Configuration window Verify that the S/N (serial number) on the screen matches the label on the back of the monitor.
D M 3 S LT R AV I E W E R V I C E A N U A L O F T W A R E N S TA L L AT I O N Figure 6-2 Location of USB connector Power the monitor ON.
Page 64
D M 3 S LT R AV I E W E R V I C E A N U A L O F T W A R E N S TA L L AT I O N Verify that the S/N (serial number) on the screen matches the label on the back of the monitor.
D M 3 S LT R AV I E W E R V I C E A N U A L Troubleshooting Overview This chapter provides helpful information about troubleshooting the Ultraview DM3 monitor. Troubleshooting Table Table 7-1 helps you solve problems that you may find when you use the Ultraview DM3 monitor.
Page 66
D M 3 S LT R AV I E W E R V I C E A N U A L R O U B L E S H O O T I N G Table 7-1 Troubleshooting Symptom Possible Causes Things to Try Blood pressure measurement Hypertensive patient.
D M 3 S LT R AV I E W E R V I C E A N U A L R O U B L E S H O O T I N G Table 7-1 Troubleshooting Symptom Possible Causes Things to Try Network connection is not working DM3 is not connecting to the...
D M 3 S LT R AV I E W E R V I C E A N U A L R O U B L E S H O O T I N G • Replace the SpO sensor and adapter. •...
Page 69
D M 3 S LT R AV I E W E R V I C E A N U A L R O U B L E S H O O T I N G • Failed or compromised TruLink or Nellcor/Novametrix-compatible sensors or adapters •...
Page 70
D M 3 S LT R AV I E W E R V I C E A N U A L R O U B L E S H O O T I N G Troubleshooting Flow Chart: Temperature START Take a temperature reading using proper measurement...
Page 71
D M 3 S LT R AV I E W E R V I C E A N U A L Service Tools Table 8-1 Service Tools Description Part Number Touchscreen Calibration Tool, USB 011-0224-00 Battery replacement kit 019-0547-00 www.spacelabshealthcare.com 8 - 1...
Page 72
D M 3 S LT R AV I E W E R V I C E A N U A L E R V I C E O O L S 8 - 2 www.spacelabshealthcare.com...
Page 73
D M 3 S LT R AV I E W E R V I C E A N U A L Appendix A — Symbols The following list of international and safety symbols describes all symbols used on Spacelabs Healthcare products. No one product contains every symbol. Note: Graphic elements of certain keys and symbols may vary between product lines.
Page 74
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S RECORD Key Dynamic Network Access (DNA) Key SPECIAL FUNCTIONS Key NORMAL SCREEN Key SAVE Key...
Page 75
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Service Message PREVIOUS MENU Key HOME Key Arrows On Direction...
Page 76
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S STANDBY Key Power ON/OFF Key Keyboard Connection Mouse Connection PAUSE or INTERRUPT...
Page 77
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Alarm Audio ON Alarm Audio OFF Alarm Audio Paused Low Priority Alarm...
Page 78
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Activate Recorder for Graphics Reset START (NIBP) Key Power Indicator LED...
Page 79
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Monitor Setup Select Program Options Set Initial Conditions Menu Access Special Function Menu Return Unit to Monitor Mode...
Page 80
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Enlarge, Zoom Input/Output PCMCIA Card Touchscreen, External Universal Serial Bus...
Page 81
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Battery Replace only with the appropriate battery. Battery Status Battery Replace only with the appropriate battery.
Page 82
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Alternating Current Both Direct and Alternating Current Functional Earth Ground Fuse Equipotentiality Terminal...
Page 83
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S IEC 60601-1 Class II equipment, double-isolated. The unit displaying this symbol does not require a grounded outlet.
Page 84
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S NIBP Cuff, Single Hose NIBP Cuff, Dual Hose NIBP Cuff, Surface Applied to Patient THIS SIDE TO PATIENT...
Page 85
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S NIBP Cuff, Neonatal 5 Size (8 to 15 cm) NEONATAL 5 NIBP Cuff, Thigh Size (38-50 cm) THIGH...
Page 86
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Environmental Shipping/Storage Altitude Limitations 12,200 m Environmental Shipping/Storage Temperature Limitations Fragile...
Page 87
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Do Not Reuse; Single Use Only Reusable Drip-Proof IPX1...
Page 88
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Does not contain hazardous substances — Europe Does not contain hazardous substances —...
Page 89
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S ABBREVIATIONS USED AS SYMBOLS ARE SHOWN BELOW. 1 - 32 Access Codes 1 Through 32 Amperes...
Page 90
D M 3 S A — S LT R AV I E W E R V I C E A N U A L P P E N D I X Y M B O L S Nitrous Oxide NIBP Noninvasive Blood Pressure nibp Oxygen Availability...