Dario® Blood Glucose Monitoring Systems User Guide Applicable to meters models: 1021-04 Dario with Lightning connector (in US - Dario® LC) 1021-06 Dario with USB-C connector (in US -Dario® USB-C)
Page 2
8 Hatokhen St. North Industrial Park Caesarea 3088900, Israel www.mydario.com Dario® is a trademark of LabStyle Innovation Ltd . European Patent application number 11780305.6 United States: Product is covered by one or more patents and/or patent applications – the list is detailed at the Dario website at www.mydario.com.
YSTEM ............11 AFETY EFINITIONS ........12 RACTICES SED IN UIDE .............15 ENERAL RECAUTIONS ..........16 DDITIONAL NFORMATION CHAPTER 2 DARIO SYSTEM OVERVIEW AND FEATURES ..17 ........17 ARIO YSTEM AT A LANCE ............18 ARIO OMPONENTS OMPATIBLE LATFORMS WITH THE ARIO LOOD LUCOSE ............20...
Page 4
......38 ERFORMING A ONTROL OLUTION CHAPTER 4 TESTING YOUR BLOOD GLUCOSE LEVEL ..44 .......44 LOOD LUCOSE ESTING RECAUTIONS ...........45 TRIP RECAUTIONS ..47 ACTORS NTERFERE WITH LOOD LUCOSE ESTING ..........48 NTERPRETING THE ESULTS .............49 EASUREMENT ANGE ....50 ACTORS EAD TO NACCURATE ESULTS NSERTING A ANCET INTO THE...
Page 5
............79 ROBLEM OLVING “P ” EST STRIP INSERTED BUT LEASE INSERT A NEW STRIP KEEPS ..........79 APPEARING ON THE SCREEN .........80 OU CANNOT PIERCE SKIN WITH LANCET . 80 HE LANCET DOES NOT FIT CORRECTLY INTO THE LANCING DEVICE .
Glucose Monitoring for a Mobile Age Congratulations on your choice of the Dario System. The Dario System was created to make life easier for people with diabetes by providing the right tools, information, and support in your pocket. It is a smart...
User Guide Scope The user guide is covering the use of the Dario LC and USB-C Blood Glucose Monitoring System with the following meter models: Dario LC (model 1021-04) for use with selected iPhone mobile devices with Lightning connector Dario USB-C (model 1021-06) for use with...
The Dario with Lightning connector (in US - Dario® LC) and Dario with USB-C connector (in US - Dario® USB-C) Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Dario with Lightning connector (in US - Dario®...
• Testing glucose from sites other than the fingertip. • Testing patients who are critically ill, in shock, or dehydrated. Certain health conditions, such as serious illness and dehydration, can cause false test results in glucose meters. You should not use this system to test your blood glucose if you are critically ill.
Page 10
All parts of the kit are considered bio-hazardous and can potentially transmit infectious diseases even after you have performed disinfection. This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures.
These instructions provide the information that you need for safe use and handling of your Dario. Incorrect use of this system may lead to an inaccurate glucose reading or damage to your device.
Practices Used in This User Guide The following sections describe the formats and symbols used throughout this user guide, on the Dario System packaging and inserts, and on actual system parts. Procedure Instructions • Procedure instructions are clearly identified and presented as numbered steps .
Page 13
Italics System response that occurs as a result of pressing a button or performing a task. Bold Italics Emphasize important information. Symbol Description Manufacturer Dario System LabStyle Under Directive Innovation Ltd. 0344 IVDD 98/79/EC Beijing Sterile Lancets Ruicheng Under Directive...
Page 14
Symbol Description Symbol Description LabStyle Innovation Dario logo Ltd. logo Consult instructions Do not use if package for use is damaged In Vitro Diagnostic...
General Precautions This section provides information about potential hazards that may harm your Dario System or cause it to not function properly . Inspect the Dario System prior to use. Avoid using any part of the system with broken seals or expired dates.
The blood glucose reading displayed on your smart mobile device by Dario System should not be the only information you use to make decisions about your health. Always consult your health care provider to interpret and understand the measurement results.
Dario Components The Dario System components are shown in the image below and described in the following table.
Page 19
The Dario Housing (also referred to as Housing the device housing or simply “housing”) is the main part of the Dario all-in-one device. It holds all the parts in place and consists of the lancing device, the meter slot, and the housing for the test strips cartridge.
Compatible Platforms with the Dario Blood Glucose Monitoring System The Dario LC and Dario USB-C Glucose Meters use smart phone technology to view and store unlimited numbers of glucose test results. You can also track other diabetes information and view graphs and...
You can even share your diabetes management information with your health care provider. The Dario LC System is compatible for use on iOS smart mobile devices. The Dario LC Glucose Meter currently works with the following Apple® iPhone® devices running iOS 12,13, 14 and iPhone...
Page 22
• Dario with Lightning connector (in US - Dario® LC) and Dario with USB-C connector (in US - Dario® USB-C) Blood Glucose Monitoring System Quick Guide • Dario with Lightning connector (in US - Dario® LC) and Dario with USB-C connector (in US - Dario®...
Assembling the Dario System for the First Time To set up the Dario System for the first time, you must complete the following procedures: • Insert a test strip cartridge, as described in Inserting the Test Strip Cartridge...
Downloading and Installing the Dario App on page • Set up the Dario App on your smart mobile device using the Dario App instructions or the instructions provided on the Dario website at: USA: www.mydario.com Canada: www.mydario.ca...
Ensure you have properly inserted the cartridge, as described in steps 3 and 4. Inserting a Lancet To insert a lancet : 1. Remove the orange lancet device cover from the Dario Housing, as shown in the images below.
Page 26
• If the lancet does not set into place or the lancet holder does not hold the lancet properly, insert another sterile lancet • If the problem persists, you may need to replace your Dario Housing. Contact Customer Support, for contact information please see page 5.
Setting Lancet Depth (Optional) You can adjust the puncture depth of the lancet by moving the lever between the numbers 1 and 4 (printed on the dial). 1 is the shallowest depth and 4 is the deepest. It is recommended that you start with the lancet depth set to 1 and adjust the level upwards if necessary .
Google Play. The Dario App downloads and installs just like any other app on your smart mobile device. The Dario App is free of charge and is customized for your smart mobile device. Downloading the Dario App from the Apple App Store or Google Play requires data connectivity.
Page 29
For accurate use of the Dario System with the app, the time and date must be correctly set on your smart mobile device. For instructions on setting the time and date, refer to the documentation for your device. To download the Dario App: 1.
Dario icon . The icon appears on your smart mobile device’s Home screen. 2. Select Sign Up to register a Dario user. Dario App registration requires email and Password to create your account. 3. Continue by following the instructions on the screen to register.
The First Entry screen appears, as shown in the image below. You can edit your email address and password and set the Dario App language and region by tapping the Menu icon and selecting SETTINGS > ACCOUNT INFORMATION Setting Your Personal and Diabetes Profiles To set your personal profile: 1.
Page 32
4. Press DONE To set your diabetes profile: 1. Tap the Menu icon in the upper left corner and select SETTINGS > DIABETES PROFILE, and under SETTINGS > SET MY GOALS 2. Tap the pencil icon in the upper right corner to edit the information.
Compare the results on the meter with the values displayed on the test strip cartridge. The Dario Control Solution package can be ordered separately as indicated in page 23. The Dario Control Solution package comes with two bottles of control solution (Level 1 (M), Level 2 (H)).
Control Solution Test Principle The Dario Control Solution contains a known amount of glucose that reacts with chemicals on the Dario Test Strip. Following the same procedure, you would use to check your blood, you will apply a drop of control solution to a test strip.
The control solution is not for use as a cleaning solution. Chemical Composition Dario Control Solution is an aqueous solution that contains d-glucose as the reactive ingredient in the approximate concentration. The remaining chemical components of the Dario Control Solution are: •...
Page 36
When not in use, store the meter in the Dario Housing. Do not leave your meter connected to your smart mobile device with the screen in ON mode for more than two (2) minutes, as the meter can get warm which can affect the accuracy of the reading.
Removing a Test Strip from the Cartridge You should have a test strip cartridge already inserted in your Dario Housing. If this is not the case, refer to Inserting the Test Strip Cartridge on page To remove a test strip from the cartridge: 1.
• Do not touch the yellow testing site when handling the test strip. • Do not use a damaged test strip. Performing a Control Solution Test To perform a control test: 1. Tap the Dario icon on your smart mobile device to launch the Dario App.
Page 39
2. Insert the meter into the port of your smart mobile device, as shown in the image below. A “Connecting...” message appears in the Dario App and then the screen displays the following instruction: INSERT A NEW STRIP...
Page 40
Verify the test strip direction, as shown in the image below. The Dario App screen instructs you to place a drop of blood on the test strip, as shown in the image below. 5. Shake the control solution container and apply a...
Page 41
(6) seconds and is automatically saved to memory . You can delete this control test from memory, if desired. Alternatively, the Dario System includes a function for specific control solution tests. Control solution tested values must be tested under the "Control solution test"...
Page 42
Applying too much or too little control solution to the test strip may cause out-of- range results . Do not use your Dario System to test your blood if your control solution result falls outside of the acceptable target values.
Page 43
test" function: Menu icon , selecting HELP&INFO > CONTROL SOLUTION TEST if not done the results will be included in the blood glucose averaging functions of the device. 10. Place the meter back into the housing by holding the meter at an angle with the part that connects to the smart mobile device facing away from the ridged orange panel.
. Do not put any fluids into the test strip port . Once you start the Dario App, follow the prompts on the screen . Blood Glucose Testing Precautions...
. listed in Test Strip Precautions Test Strip Precautions • Only use Dario Test Strips. Using other test strips with the Dario Glucose Meter can produce inaccurate results. • Inspect each test strip cartridge before opening it for the first time. If you see any damage to the cartridge lid, or if anything prevents the lid from closing properly, do not use the test strips.
Page 46
• Store unused test strips in their original cartridge. Do not remove test strips from the test strip cartridge and put them into another container. • Close the cartridge tightly right after you remove a test strip. This helps to keep the test strips dry. •...
• Do not bend, cut, or alter the test strips. • Do not get dirt, food, or other substances on the test strip. • It is recommended that you use the test strip immediately after removing it from the cartridge. It is not recommended that you use a test strip that has been outside of the cartridge for more than three (3) minutes.
. If the test result is inconsistent with how you feel, check your Dario System by running a control solution check, as described in Performing a Control Solution Test on page 38.
Measurement Range Your meter measures blood glucose results in the range of 20 mg/dL to 600 mg/dL (1.1 - 33.3 mmol/L). Results outside of this range should be reviewed according to the following information . • If your test result is lower than 20 mg/dL (1.1 mmol/L), “LOW! Measure with a new strip, if recurs contact your physician immediately”...
Incorrect testing technique or improper storage or handling of the Dario System may also lead to inaccurate test results which are inconsistent with how you feel. The following conditions could lead to incorrect results: •...
Be careful not to touch or bend the lancet. Step Two: Opening the Dario App Before connecting the Dario Glucose Meter into your smart mobile device, open the Dario App by tapping the Dario icon on your smart mobile device Home screen.
Step Three (relevant to the US market only): Inserting the Smart Mobile Device into the Disposable Cover Before starting the testing procedure, you should cover your phone with a disposable cover. To avoid the risk of transmitting bloodborne pathogens, it is recommended that you use a disposable sleeve to cover your smart mobile device while taking the blood glucose test.
A “Connecting...” message appears in the Dario App, and then the screen displays the following instruction: INSERT A NEW STRIP.
Page 54
• Unwashed hands could affect your test results. • Alcohol can affect test results. • When not in use, store the meter in the Dario Housing. Do not leave your meter connected to your smart mobile device with the screen in ON mode for more than two (2) minutes, as it can get warm and can affect the accuracy of the test.
Page 55
Removing a Test Strip from the Cartridge You should have a test strip cartridge already inserted in your Dario Housing. If this is not the case, refer to Inserting the Test Strip Cartridge on page 24.
Page 56
6. Hold the test strip with the gold contacts facing upward. 7. Without bending the strip, insert it into the test strip port of the Dario Glucose Meter as far as it will go. Verify the test strip direction, as shown in the...
Page 57
The Dario App instructs you to place a drop of blood on the test strip, as shown in the image below. Performing a Glucose Test If you receive a phone call while the glucose measurement is in progress, Dario will stop the test and the results will not be saved.
Page 58
your thumb to prick your finger, as shown in the images below. 3. Apply a drop of blood to the tip of the yellow window of the test strip, as shown in the images below. Capillary action will pull the blood into the strip.
Page 59
The glucose test result is automatically saved to memory. If a result does not display, you can view it in your Logbook in the Dario App. In case an error message appears, you may need to start the measurement process again.
When the test strip is removed from the glucose meter, the following message appears in the Dario App: “Please insert new test strip.” Step Six: Returning the Meter to the Dario Housing To return the meter to the Dario Housing: 1. Disconnect the meter from your smart mobile device.
(> 810 mg/dL), you may get inaccurate results with this system. This system should not be used when undergoing xylose absorption tests. Packed cell volume: Dario BGMS Test Strip suitable range is 20~60%.
Cleaning and Disinfection The outer surfaces should be cleaned when you notice that your Dario is visibly dirty. It is advised to clean and disinfect your Dario System once a week to avoid transmitting bloodborne pathogens.
• In the USA - use a Super Sani-Cloth • In Canada – use Germidice These wipes have been proven safe for use with the Dario meter, housing, orange lancet cover, and white cartridge cover. Super Sani-Cloth and Germidice are available online through amazon.com and officedepot.com.
Page 64
• It is very important to let the liquid from the wipe sit on the outer surface for at least two (2) minutes. Be extremely careful to avoid getting liquid onto the test strip port. The following images demonstrate how to wipes to clean and disinfect your Dario System.
Page 65
Wipe off any remaining residues with a clean, soft, and dry paper cloth . Your Dario System has an expected life of three (3) years. It should withstand the cleaning and disinfection recommended (156 cycles) in this user guide.
Page 66
Please contact your health care provider for questions outside of the customer support hours. The cleaning and disinfection performance testing on the Dario System was done using the Super Sani- Cloth (in Canada use Germidice ) with EPA registration number 9480-4 (Active ingredient: Quaternary ammonium chlorides and isopropanol).
Page 67
• Spray any cleaning solution directly onto the glucose meter or the lancing device. • Submerge any part of the Dario System in liquid. • Attempt to clean any internal parts. • Attempt to dry any parts using a blow dryer.
The blood glucose result is displayed on Dario App screen. By touching a drop of blood to the tip of Dario Blood Glucose Test Strip, the strip’s reaction chamber automatically draws the blood into the strip through capillary action.
6. Do not transfer test strips from one cartridge to another. The Dario Test Strip is structured as follows: Yellow Window Apply a drop of blood here. • Before Applying Blood –...
• Dario Blood Glucose Test strips should be used with Dario/Dario LC/Dario USB-C meters only. • Insert a test strip into the test strip port in the glucose meter with gold-colored bars facing up and toward the meter. • A test strip is sensitive to humidity. Therefore, you...
• Consult with your healthcare professional for the blood glucose range that is appropriate for you. Test Result Range The Dario, Dario LC and Dario USB-C meters read blood glucose results between 20-600 mg/dL (1.1- 33.3 mmol/L). Unexpected Results High or low blood glucose results can indicate potentially serious medical conditions.
• Repeat the test with a new strip. • Run a control solution test with Dario Glucose Control Solution. • If the control solution test result is within the...
Control solution check ensures that you are doing a test correctly and that your system is working properly. Make sure you use the proper control solution for the test strips you have: Dario® Blood Glucose Test Strips require Dario® Glucose Control Solution.
Test control solution at 64-86°F (18-30°C) environment. Control Solution Procedure To conduct a control solution test, refer to Performing a Control Solution Test on page 38. Understanding Your Control Solution Test • If your glucose control solution test result is within the acceptable range, you can feel confident that your test strips and meter are working correctly.
Glucose Monitoring Devices, Oct 2003, Vol. 5, No. 5: 749 -768. CHAPTER 7 TROUBLESHOOTING This chapter provides information about all the symbols and error messages that may display in the Dario App. To ensure safe and accurate use of Dario,...
Please contact your health care provider for questions outside of the customer support hours. An error message can appear in the Dario App as a message within the results area, as shown in the images below. Troubleshooting The following table lists some typical conditions that may occur with the Dario System.
Page 77
Message Content Possible Cause / Recommended Action The strip is used, Perform a new measurement please insert a with a new test strip. new strip Temperature out The meter was left in an of range. Move to extreme temperature an area between environment.
Page 78
If this message recurs, contact your health care provider Immediately. If your measurement results are inconsistent with the way you feel, contact customer support. Dario device Communication failure. failure (0x00). If Try disconnecting and the problem reconnecting the meter.
If another device is connected to the Lightning or USB-C port, disconnect it and connect the Dario Glucose Meter. You can still use the Dario App if another device is connected. If you received this message while the Dario Glucose Meter is connected, disconnect and reconnect the meter.
The lancet does not fit correctly into the lancing device. The lancet may be faulty or you may be using a lancet that is not approved for use with the Dario System. Replace the lancet with a different Dario Lancet. If the problem persists, you may need to replace the Dario Housing.
Replace the lancet with a different Dario Lancet. If the problem persists, you may need to replace the Dario Housing. Contact Customer Support, for contact information please see page 92. Lancing device does not operate correctly. The lancing device spring or handle may be faulty.
My smart mobile device battery died. If the battery of your smart mobile device is out of power, charge your device or use a different device. Download the Dario App onto that device and login to your account.
The glucose meter detects the current and converts it into a blood glucose reading. Compatibility Dario is compatible for use on selected smart mobile devices. For a complete list of compatible devices, please refer to: • USA: www.mydario.com/support/getting-started /...
Page 84
Over-the-Counter Use - Guidance for Industry and Food and Drug Administration Staff” - October 2016. The Dario® LC study showed the following results for all subjects’ samples: ”96.3% within ±15% and 99.4% within ±20% of the medical laboratory values at the entire glucose concentrations range”.
Page 85
(STD) 3mg/dL (0.15 mmol/dl) at the below 100mg/dL (5.55mmol/L), for coefficient of variation (CV) 4% at the above 100mg/dL (5.55mmol/L). The precision studies using control solution (Intermediate Precision) and blood (Repeatability Precision) are shown below. Intermediate Precision: Dario® LC BGMS 30-50 51-110 111-150 151- 250 251-400...
Coefficient 3.3% 3.3% 3.8 % 2.8 % 3.8 % of Variation *The Dario® USB-C meter is equivalent to the Dario® LC and can be used with Android SMDs. Technical Specifications Meter Specifications Result Range 20-600 mg/dL (1.1-33.3 mmol/L) Calibration Plasma-equivalent...
Do not store near solvents and ionizing radiation sources. Compliance with Electrical and Safety Standards The Dario Blood Glucose Meter has been tested for • Electrostatic discharge as specified in IEC 61000-4-2. • Immunity to radio frequency interference at the frequency range and test levels specified in IEC 61000-4-3.
Dario glucose meter device does not connect to the mobile device wirelessly. Users are required to unlock their smart mobile device prior accessing the Dario App to take a glucose measure test, as such they are assured of ongoing authorization access.
Page 89
Dario App and cloud servers. Users data is stored and transmitted through the Dario App and/or web portal to and from the remote server in encrypted form.
(1) year from the date of the original purchase. If during this time the Dario does not work properly because of a defect in materials or workmanship, LabStyle Innovation Ltd. agrees to replace or repair, free of charge, any and all parts proven to be defective and subject to warranty.
This warranty does not apply to the performance of a Dario System that has been damaged by accident or has been altered, misused, tampered with, or abused in any way. This warranty only applies to the original purchaser of the meter and/or its agents.
Contact Customer Support You may contact customer support with any questions or issues about Dario and its accessories using one of the following methods: • By phone: • USA: toll-free Monday to Friday, 9:00AM – 5:00PM EST at 1-800-895-5921 or www.mydario.com...
Need help?
Do you have a question about the LC and is the answer not in the manual?
Questions and answers