Froilabo Cryovita CS200 Operating And Maintenance Instructions Manual

For use with voyager & origin
Table of Contents

Advertisement

Quick Links

OPERATING AND MAINTENANCE INSTRUCTIONS
CS200 Controller for use with Voyager & Origin Series
• DO NOT ATTEMPT TO USE OR MAINTAIN ANY LIQUID NITROGEN FREEZER UNTIL YOU
READ AND UNDERSTAND THESE INSTRUCTIONS.
DO NOT PERMIT UNTRAINED PERSONS TO USE OR MAINTAIN THIS UNIT.
• IF YOU DO NOT FULLY UNDERSTAND THESE INSTRUCTIONS, CONTACT YOUR
SUPPLIER FOR FURTHER INFORMATION.
CAUTION - SAFETY FIRST!
CS200 Control Systems

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Cryovita CS200 and is the answer not in the manual?

Questions and answers

Summary of Contents for Froilabo Cryovita CS200

  • Page 1 CS200 Control Systems OPERATING AND MAINTENANCE INSTRUCTIONS CS200 Controller for use with Voyager & Origin Series CAUTION - SAFETY FIRST! • DO NOT ATTEMPT TO USE OR MAINTAIN ANY LIQUID NITROGEN FREEZER UNTIL YOU READ AND UNDERSTAND THESE INSTRUCTIONS. • DO NOT PERMIT UNTRAINED PERSONS TO USE OR MAINTAIN THIS UNIT.
  • Page 2: Table Of Contents

    Table of Contents SAFETY PRECAUTIONS ..................5 Liquid Nitrogen (LN ) ........................5 Extreme Cold - Cautionary Statement ..................5 Keep Equipment Area Well Ventilated..................5 Liquid Nitrogen System ....................... 5 Electrical ............................5 GENERAL INFORMATION .................. 6 CS200 Series Control System Specifications ................7 DELIVERY AND RETURNS..................
  • Page 3 Table of Contents Test Alarms ..........................37 Logging ............................37 Display Brightness ........................38 Making Adjustments to the CS200 Series Control System Sensor Assembly ...... 38 Removing/Installing the Solenoid Valve .................. 39 Removing/Installing the Controller 10K/24K Units .............. 39 Removing/Installing the Thermocouple ................... 40 Removing/Installing the Sensor Probes ...................
  • Page 4: Safety Precautions

    Safety Precautions Liquid Nitrogen Nitrogen is an inert, colorless, odorless, and tasteless gas making up four-fifths of the air you breathe – and can be very dangerous. Air is roughly one-fifth WARNING: oxygen. Liquid nitrogen is at a temperature of -196°C (-320°F) under normal The following safety atmospheric pressure.
  • Page 5: General Information

    CS200 SERIES Control Systems are designed to work with Voyager & Origin Series vessels from Froilabo. The features are designed to provide a safe environment for samples while at the same time tracking all relevant information associated with the freezer.
  • Page 7: Cs200 Series Control System Specifications

    General Information CS200 SERIES Control System Specifications Specifications Control Type Level Control & Temperature Control Level Measurement Sensor Type 8-Thermistor Fixed (standard) Range 8 inch range (8-Thermistor) Redundancy Multiple discrete points Temperature Measurement Sensor Type Type T Thermocouple Accuracy or 1.5% of reading Resolution Number of channels C, F, K, R...
  • Page 8 User Interface Display Type Vacuum Fluorescent Display (VFD) Buttons Level, Temperature and Alarm Information “At a Glance” status LED status wheel Filling 1 LED Menu Access 1 LED Language English and German Control Tests Power Up Self-Test Control system check Thermistor Status Battery voltage Control voltages...
  • Page 9 General Information CS200 SERIES Control System Specifications (cont’d) Buttons Power Turns power on/off Fill/Defog Open Valve Stop Close Valve Menu Access Menu Mute Silence audible Enter Save a setting or select a menu choice Back/Exit Leave a setting unchanged or back out of menu. Up arrow Scroll the menu system or increase a value Down arrow...
  • Page 10: Unpacking And Inspection

    (5) days from receipt of goods. If damage or loss is apparent, please notify the shipping agent immediately. In all cases, Froilabo should be notified so we can assist if needed in filing damage claims.
  • Page 11: Installation

    Installation Getting Unit into Service Your cryostorage system comes with complete instructions for how you should remove the unit from the crate and put it into service. Read both this manual and your cryostorage system’s manual before beginning any installation. Make sure to follow any required procedures and safety guidelines when you are connecting your Liquid Nitrogen source.
  • Page 12: Power Supply Connection

    Installation power supplies and electrical accessories for use with the CS200 SERIES WARNING: Control System and they are not listed or referenced in this manual, it is the This equipment responsibility of that third-party supplier to determine compliance with the is intended for standards and tests in the EN Compliance tables found on pages 48–51.
  • Page 13: Initial Fill

    Control Components CS200 Series Control System Components The CS200 SERIES Control System for the Froilabo Origin and Voyager series freezers consist of the following components: Main Control Module VFD (Vacuum Florescent Display) Module...
  • Page 14: Cs200 Controller Display Guide

    Operation CS200 Controller Display Guide Figure 2.0 CS200 Controller Display Operational Theory The CS200 SERIES Control System automatically maintains the Liquid Nitrogen (LN level and monitors temperature in the cryostorage freezer. Operational conditions are monitored, and any alarm is triggered if necessary. Operations data is stored in High Level Alarm memory on the control board.
  • Page 15 Operation The LN level is indicated on the display and is determined by the number of thermistors submerged in LN according to the table below: Table 1.0 8-Thermistor Sensor (G=GAS, L=LIQUID) 8-THERMISTOR SENSOR OFFSET DISPLAY NOTE - Offset value for Origin 94K only = 3.0 in (126 mm). Note: Please see Alarm Conditions description of...
  • Page 16 Operation Alarm Problem Detected The lid has been opened for a period which is longer than the designated Lid Open Too Long Alarm alarm setting. Valve Stuck Open Alarm The solenoid valve is stuck open. Unauthorized Access Warning The lid has been opened and an incorrect identification has been entered. Usage Warning The consumption of LN has increased and should be checked.
  • Page 17: Operating Parameters

    100/240 VAC (50/60 Hz). UL approval for the system as a whole is not required since the control operates on low voltage. If your power source differs, or is subject to disruption or line surges due to other equipment on line, consult your Froilabo representative. Remote Alarm If an error condition occurs after a user defined period of time, a remote alarm can be initiated.
  • Page 18: Temperature Monitoring

    Operation keypad to define new start and stop levels, and these levels may be set independently to vary the liquid level deference between fills. Prior to the initial fill of the freezer, a determination should be made whether vapor phase or liquid phase storage will be utilized. All units are supplied with an eight-thermistor assembly and a freeze-guard sensor unless otherwise specified.
  • Page 19 Operation Table 3.0 Annual Maintenance Schedule CS200 SERIES Control Examine for exposure to moisture, wear and tear, connector problems, and damage to the faceplate or buttons. In addition, periodic firmware updates may be important Harness Assembly Examine for damage to the cable and damage to connectors, Battery Examine connection cable and connector for damage.
  • Page 20: Controller Operation

    Controller Operation This section of the operating manual is for Froilabo approved equipment that uses the CS200 SERIES Control System. Introduction The CS200 Control System, temperature and LN level controller is designed for easy operation and reliable uninterrupted service. This controller will maintain the selected liquid level range of LN in your freezer as well as providing audible and visual alarms for any non-conforming conditions that may occur.
  • Page 21 Controller Operation Features & Settings Some control settings can be Enter and Exit the menu system by pressing: changed through the menu system. When changing settings, single Move down through the menu system by pressing: button presses will increment/ decrement a value one step at a time.
  • Page 22 Controller Operation View Only The View Only menu selection allows the user to view settings but will not allow changes to any settings. This restriction is designated with a lock symbol. Refer to Figure 8.1 Menu System - View Only on page 27 to navigate the “View Only”...
  • Page 23 Controller Operation LEVEL • Thermistor Status: Gives the user the status for the 8-thermistors on the level sensor assembly, the freezeguard sensor and the lid switch. • Sensor Positions: Allows the user to set the positions for low level alarm, start fill, stop fill, high level alarm and the sensor offset.
  • Page 24 Next Fill: Displays the next scheduled fill. • Maintenance Reminder: Allows the user or service company to acknowledge a reminder that preventive maintenance should be performed on the freezer. Contact Froilabo Customer Service for instructions how to reset the reminders. •...
  • Page 25 Supplier: User-defined. • Distributor: User-defined. • Froilabo: +33 (0) 4 78 04 75 75 / commercial@froilabo.com • Pacer Digital: info@pacerdigital.com / +1 317-849-7887 SELECT LANGUAGE: Allows the user to select the desired language. SYSTEM INFORMATION: Displays the control serial number, the firmware version, the freezer ID and any options that are enabled.
  • Page 26: Temperature

    Controller Operation Temperature The temperature in the Cryostorage freezer is measured at the location of the thermocouple. The installation location is different depending on the model of the freezer but usually the thermocouple is positioned level with the top storage box. This may mean that temperatures displayed may be slightly warmer than the temperature experienced by the samples or product stored in the freezer.
  • Page 27 Controller Operation Menu System MENU VIEW ONLY TEMPERATURE THERMOCOUPLE SELECT T/C 1 On/Off T/C 2 On/Off TEST TEMPERATURE SYSTEM CHECK THERMOCOUPLES T/C 1 OK/OPEN/Off T/C 2 OK/OPEN/Off TEST TEMP ALARMS Off - High Temp - Low Temp CONTROL BY TEMPERATURE TEMP CONTROL ON/OFF On/Off TEMP CONTROL RANGE...
  • Page 28 SECURITY SETTINGS PASSWORD POWER PASSWORD CHANGE REMOTE PASSWORD CONTACT SUPPORT CONTACT INFO LN2 SUPPLIER CONTACT INFO DISTRIBUTOR CONTACT INFO Froilabo PACER CONTACT INFO DIGITAL ENGLISH- GERMAN SELECT LANGUAGE SERIAL # 12345 SYSTEM INFORMATION Options VERSION 1. 03 FREEZER ID FRZ01...
  • Page 29 Controller Operation Factory Defaults Thermocouple #1 Thermocouple #2 Control by Temperature High Temperature Alarm #1 -100°C High Temperature Alarm #2 -100°C Low Temperature Alarm #1 Low Temperature Alarm #2 Temperature Units Celsius Sensor Position Freezer dependent Valve Open Duration Sensor Type 8-thermistor sensor (standard) Freezeguard feature Sensor Error Mute...
  • Page 30: Interconnection Block Diagram

    Controller Operation Troubleshooting If the Froilabo cryostorage freezer with the CS200 SERIES Control System installed experiences problems or appears that it is not operating at optimum efficiency, please contact your distributor for assistance. The CS200 SERIES Control System has incorporated state of the art diagnostic tools to assist in the identification and correction of any issues that may arise.
  • Page 31: Wiring Diagram

    Controller Operation Wiring Diagram CS200 Harness Wiring Diagram 4 wires position 1 P1P1 Pod 1 4 wires position 2 P1P2 10 wires Pod 2 position 1 P2P1 10 wires position 2 P2P2 Blue P1P1 Battery P1P1 Yellow + White/Yellow P2P1 Remote Alarm Orange + White/Orange P2P1...
  • Page 32: Installation & Setup

    Controller Operation Installation & Set-up The CS200 CONTROL SYSTEM consists of the following components: • Main Control • VFD Display Module • Wiring Harness Assembly • Power Supply • Thermocouple Assembly • Lid Switch Assembly • Sensor Assembly • Cryogenic Solenoid Valve •...
  • Page 33 Controller Operation Connect the level sensor assembly into the pigtail with the round connector terminating at the control end of the harness assembly. The connector is a round locking connector that is keyed so it can only be plugged in one way. Connector to Main Control Connector for Level Sensor Figure 12.0 Harness Assembly...
  • Page 34: External Connector Ratings

    Controller Operation Cryowire Dual Communication Ports Remote Alarm Connector Connect the Remote Alarm plug into the panel at the end of the wiring harness assembly. It is keyed so that it can only be plugged in one way. Figure 14.0 Battery Power Cryowire...
  • Page 35: Led Status Wheel Flash Patterns

    Controller Operation LED Status Wheel Flash Patterns Condition LED Flash Property Normal Clockwise pattern, 1 LED per second General Error Clockwise pattern, rapid rotation, flash Table 7.0 LED Status Wheel Flash Patterns Temperature Thermocouple Select The chamber temperature is monitored with 1 or 2 Type T thermocouples. The thermocouple is placed in the chamber to monitor temperature level at the top of the inventory system.
  • Page 36: Test Temperature System

    Controller Operation Access the menu by pressing Highlight CHANGE SETTINGS and press Highlight TEMPERATURE and press Highlight CALIBRATE and press Highlight CALIBRATE THERMOCOUPLE 1. The display will read Dip the thermocouple into ice water. Submerge the thermocouple into an ice water bath. The bar graph will fluctuate. When a good stable reading in obtained, the bar graph will decrease to 1 or 2 bars.
  • Page 37: Alarms And Error Conditions

    Controller Operation has occurred. This occurs because the control cannot differentiate between a very high resistance (when a thermistor is in LN ) and an infinite resistance (when an open circuit appears in the level sensing circuitry). Alarms and Error Conditions The CS200 SERIES Control System control monitors many different conditions in the freezer and has a full complement of alarms associated with these different conditions.
  • Page 38: Display Brightness

    The sensor assembly is preset at the factory for vapor phase storage. If adjustments need to be made, the following procedure will simplify the process. The CS200 SERIES Control System control installed on the Froilabo cryostorage units operates with specially designed software to match the design characteristics of your freezer. Refer to the chart below to see the versions and their difference.
  • Page 39: Removing/Installing The Solenoid Valve

    Controller Operation Removing/Installing the Solenoid Valve The LN and power must be turned off before beginning work on the solenoid valve. For Origin Electrical Units Disconnect the solenoid valve lead connection from the back of controller board. To remove the solenoid valve, loosen the compression fitting that connects the plumbing tubing to the fill tube.
  • Page 40 Controller Operation Figure 16.0 Controller Removal CS200 Control Systems...
  • Page 41: Removing/Installing The Thermocouple

    Controller Operation Removing/Installing the Thermocouple Remove the controller using the procedures outlined for your particular refrigerator model in Removing the Controller section. Disconnect only the thermocouple lead connection from the controller board. Gently pull the thermocouple from the sensor tube. To install a thermocouple, feed the thermocouple lead into the sensor tube to the elevation in the storage chamber you want to monitor.
  • Page 43 Controller Operation To make adjustments to a sensor assembly in a refrigerator filled with LN , the following procedure can be used: • Measure the LN liquid level in the refrigerator. • Take this measured level and subtract the offset to determine how many sensors should be in liquid. •...
  • Page 44: Replacement Parts

    Replacement Parts Plumbing Assembly 13 14 Figure 18.1 Origin 20K/80K/94K Figure 18.0 Origin 38K/40K Plumbing Assembly Parts Item Part No. Description 6913-9077 Valve, Relief, 1/4" MPT, 100 psig 45151590 Connector, 3/8" ODT x 1/4" FNPT LA20-9C90 Tube, Relief Valve – Origin 20K LA40-9C90 Tube, Relief Valve –Origin 38K/40K LA80-9C90...
  • Page 45 Replacement Parts Controller Assembly Figure 18.3 Origin Controller Assembly Parts Controller Assembly Parts Item Part No. Description LA40-6C68 Final Plumbing Assembly Origin 20K CS LA40-8C69 Final Plumbing Assembly Origin 38K CS LA40-8C69 Final Plumbing Assembly Origin 40K CS LA40-6C68 Final Plumbing Assembly Origin 80K CS LA40-6C68 Final Plumbing Assembly Origin 94K CS 5140-1216...
  • Page 46: Controller Assembly

    Replacement Parts Controller Assembly 8, 9, 10 11, 12, Connect (ITEM 7) (2x) and (ITEM 8) Sensor Tube Controller Assembly Parts Item Part No. Description R10K-8C68 Plumbing Assembly, CS200 Controller 5140-1279 Solenoid Valve Harness, Voyager-CS-Series, 36” 5140-1229 12 VDC Transformer No Plug LA80-9C51 Transformer Cover, Origin CS 5140-1285...
  • Page 47: Troubleshooting

    Troubleshooting Symptoms The key to troubleshooting your CS200 SERIES Control System and your cryostorage system is to determine which component in the system is the source of the problem. Determine if the problem is occurring in any of the following subsystems: Supply Vessel, Transfer Line, Power Source, Temperature, Level Sensing, Security, Lid Switch, Solenoid Valve, Control Display, Alarm System, Communications.
  • Page 48: Indicates Low Ln Supply

    Troubleshooting Press FILL button to re-open fill solenoid valve. • • Continue to lower the sensor until the STOP FILL thermistor is immersed in the LN . The fill should stop after a slight delay. • Simulate an over fill by lowering the HIGH LEVEL Alarm thermistor into the LN HIGH LEVEL alarm should sound within 10 seconds.
  • Page 49: Fill Solenoid Cycles On And Off

    Troubleshooting Make sure the thermocouple is clean and dry before and after each bath. • • Reposition the thermocouple at the desired location to monitor or control temperature. Fill Solenoid Cycles On and Off CS200 SERIES Control System features a timer function whereby the valve open duration is limited, allowed to shut for a short warm-up and then opened again.
  • Page 50: Push Buttons Will Not Respond

    Troubleshooting Push Buttons Will Not Respond Moisture may have migrated into the sealed faceplate. Call your distributor for a Material Return Authorization. Refer to the Service and Maintenance section of this manual to obtain information regarding equipment returns. Liquid Level Readout is Incorrect Liquid level is determined by a 8-thermistor assembly inside the protective sensor tube.
  • Page 51: Service And Maintenance History Log

    Date: NOTE: All cryostorage systems must be cleaned and decontaminated prior to return to Froilabo for repair or maintenance and must be accompanied by a written statement to this effect. Any cryostorage system received without this statement will be returned to the sender, freight collect. Contact Customer Service by telephone: +33 (0) 4 78 04 75 75 Email: commercial@froilabo.com...
  • Page 52: Appendix

    Appendix Certifications & Listings This product complies with the following standards and directives: CB Scheme Report and Certificate • • EN 60101-1 • IEC 61010-1 UL 61010-1 • • CAN/CSAC22.2#61010-1 • 93/42/EEC Medical Device Directive for the European Union • IEC 60601-1-2, 2007 Edition 3.0 (EMC Directive) RoHS Directive •...
  • Page 53 Appendix TABLE 11: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The CS CONTROL SYSTEM is intended for use in the electromagnetic environment specified below. The customer or the end user of the CS CONTROL SYSTEM should assure that it is used in such an environment.
  • Page 54 Appendix TABLE 12: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The CS CONTROL SYSTEM is intended for use in the electromagnetic environment specified below. The customer or the end user of the CS CONTROL SYSTEM should assure that it is used in such an environment.
  • Page 55 Appendix TABLE 13 Recommended separation distances between portable and mobile RF communications equipment and the CS CONTROL SYSTEM. The CS CONTROL SYSTEM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CS CONTROL SYSTEM can help prevent electromagnetic interference by maintaining a minimum distance between the portable and mobile RF communications equipment (transmitters) and the CS...
  • Page 56: Declaration Of Conformity

    Appendix EC Declaration of Conformity The undersigned representing the manufacturer: Pacer Digital Systems, Inc. Attn: Kevin Oeff 8658 Castle Park Drive, Suite 103 Indianapolis, IN 46256, USA Herewith declared that the Product: LN Level Control for Cryostorage System Model/Type ref.: CS CONTROL SYSTEM is in conformity with the Essential requirements of the following EC Directives when subject to correct installation, maintenance and use conforming to its(their)intended purpose, to the applicable regulations and standards, to our operation and...
  • Page 57: Warranty

    Froilabo warrants that each of its electronic controllers and displays will be free from defects in material and workmanship, in the normal service for which the product was manufactured, for a period of two (2) years from the original date of purchase. Froilabo at its option will either repair or replace any item covered under this warranty.