The User Manual is based on general product applications. Specific clinical assessments should be based on individual client physical and functional assessments by a healthcare professional. Manufactured By: Air Kinetic Technologies Corp. Distributed By: HME Home Health Ltd. (www.HMEBC.com, Marketingtickets@hmebc.com) IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE OPERATING THIS DEVICE NOTE - CAUTION AND WARNING STATEMENTS: NOTE –...
Page 3
HANDED OVER TO AN APPLICABLE COLLECTION POINT FOR THE RECYCLING OF ELECTRICAL AND ELECTRONIC EQUIPMENT. UL CERTIFICATION LOGO (COMPLIANCE WITH IEC60601-1) WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH IEC60601-1. HME Home Health Ltd. HME Signature Series Dove Mattress 3 | P a g e...
9. Disposal of Air Mattress ............................15 10. Troubleshooting ..............................16 11. Technical Data ..............................17 11.1 Product Specification ........................... 17 11.2 EMC Information (120V) ..........................18 HME Home Health Ltd. HME Signature Series Dove Mattress 4 | P a g e...
• The HME Signature Series Swan Low Air Loss Mattress System may be used in a variety of settings including, but not limited to individual home care settings and long-term care of whom suffers from pressure ulcers, or pain management as prescribed by the physician.
“click” sound will secure the connection. Connect the Fowler’s Position Detector cable to the pump Step 7 unit by pushing in to secure the connection.. HME Home Health Ltd. HME Signature Series Dove Mattress 6 | P a g e...
Page 7
5 minutes for the mattress to be inflated. Step 7 Disconnect the hose from the Control Unit and put on the hose connector Transport Cap to keep the mattress inflated. HME Home Health Ltd. HME Signature Series Dove Mattress 7 | P a g e...
Pulsate cycle time can be selected from 1-20 minutes by pressing the Cycle button. Operate or Standby Press this button to start operating or go into standby. HME Home Health Ltd. HME Signature Series Dove Mattress 8 | P a g e...
Page 9
NOTE: the control unit will enter Self Calibration mode and the airflow will be suspended temporarily. Once the pump finishes Self Calibration mode it will automatically return and continue the previous mode. HME Home Health Ltd. HME Signature Series Dove Mattress 9 | P a g e...
Note: A power failure alarm will be triggered if the power is switched off in operating mode (refer to Audiovisual Alarm). Press the power switch to restart the system or press Alarm Mute to turn off the system (refer to Alarm Mute). HME Home Health Ltd. HME Signature Series Dove Mattress 10 | P a g e...
Control Unit or the Power Switch is pressed (intentionally or unintentionally), the “Power Failure” indicator will light up along with the buzzer and will last 20 minutes. HME Home Health Ltd. HME Signature Series Dove Mattress 11 | P a g e...
Low Pressure Low Pressure Overtime Alarm Pressure Service REPEAT High Pressure High Pressure Overtime Alarm Service (Cycle 4 sec.) High Temperature High Ambient Temperature Alarm HME Home Health Ltd. HME Signature Series Dove Mattress 12 | P a g e...
Page 13
NONE Lift-Up Low Lift-Up Angle >20°(±15°) sec.) Service REPEAT Service Indicator (Cycle 4.5 Calibration NONE Calibration Uncompleted sec.) Uncompleted Calibration NONE REPEAT Calibration Completed Completed HME Home Health Ltd. HME Signature Series Dove Mattress 13 | P a g e...
40°C. Exceeding the temperature can cause significant damage to the mattress cover. CAUTION: Do not use phenolic-based products for cleaning. CAUTION: After cleaning, dry the mattress without direct exposure to sunlight. HME Home Health Ltd. HME Signature Series Dove Mattress 14 | P a g e...
Check and replace the air filter regularly if the environment is dirty. • 9. Disposal of Air Mattress When the air mattress is no longer useable, the mattress and the Control Unit may be discarded. • HME Home Health Ltd. HME Signature Series Dove Mattress 15 | P a g e...
Weight and Comfort Level Reference Table) and wait for a few minutes for better comfort. • Follow the procedures “The low-pressure light is constantly flashing, and the alarm has sounded” for inspection. HME Home Health Ltd. HME Signature Series Dove Mattress 16 | P a g e...
-- 25 °C to 70 °C; 10 % to 90 % RH ≦3000m ENVIRONMENT PRESSURE 70 kPa-101.3kPa ENVIRONMENT HORIZONTAL LEVEL WATER AND DUST PROTECTION CLASSIFICATION IP21 HME Home Health Ltd. HME Signature Series Dove Mattress 17 | P a g e...
Harmonic emissions Not applicable the public low-voltage power supply network that supplies buildings used for domestic IEC 61000-3-2 Not applicable purposes. HME Home Health Ltd. HME Signature Series Dove Mattress 18 | P a g e...
Page 19
NOTE UT is the a.c. mains voltage prior to application of the test level. HME Home Health Ltd. HME Signature Series Dove Mattress 19 | P a g e...
Page 20
At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by NOTE 2: absorption and reflection from structures, objects, and people. HME Home Health Ltd. HME Signature Series Dove Mattress 20 | P a g e...
Page 21
380 –390 TETRA 400 modulation b) 18 Hz FM c) 430 – GMRS 460, ±5 kHz FRS 460 deviation 704 – LTE Band Pulse modulation b) HME Home Health Ltd. HME Signature Series Dove Mattress 21 | P a g e...
Page 22
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while does not represent actual modulation, it would be worst case. HME Home Health Ltd. HME Signature Series Dove Mattress 22 | P a g e...
Page 23
No. 3, Gongye 5th Rd., Pingzhen Dist., Taoyuan City 324, Taiwan TEL: +886-3-4194567 www.aktc.com.tw Importer: HME Home Health Ltd. 130 – 4011 Viking Way Richmond, BC V6V 2K9 Canada HME Home Health Ltd. HME Signature Series Dove Mattress 23 | P a g e...
Need help?
Do you have a question about the Signature Series and is the answer not in the manual?